K Number
K040702

Validate with FDA (Live)

Date Cleared
2004-04-27

(41 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80L YZ, and meets all requirements of ASTM Standard D5250-00E4.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Physical and Dimensions Testing: Based on ASTM-D-5250-00E4, Inspection Level S-2, AQL 4.0Meets requirements
Pinhole Test (Water Fill Test): FDA 1000 ml. Water Fill Test based on ASTM D-5151-99, AQL 2.5, Inspection Level IMeets requirements
Primary Skin Irritation:No primary skin irritant reactions
Skin Sensitization (Allergic Contact Dermatitis):No sensitization reactions
Residual Powder Test (for "powder-free" claim): Based on ASTM D6124-01 for Starch, no more than 2 mg powder per gloveMeets "powder-free" claims (contains no more than 2 mg powder per glove)

2. Sample Size Used for the Test Set and Data Provenance

The document specifies "samplings of AQL 2.5, Inspection Level I" and "Inspection Level S-2, AQL 4.0" for the physical and pinhole tests. However, it does not explicitly state the exact numerical sample sizes used for these tests or for the biocompatibility tests (skin irritation and sensitization). The data provenance is not mentioned, but it can be inferred that the testing was conducted by or for Grand Work Plastics Products Co., Ltd., likely in China or through contracted labs. The data is presented as retrospective in nature, reflecting completed tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This device is a patient examination glove, and the "ground truth" for its performance is established through standardized laboratory and biocompatibility tests, not expert consensus on medical images or diagnoses.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods are typically associated with subjective assessments or diagnostic interpretations by experts, which are not relevant for the type of testing performed on patient examination gloves.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation, which is not the case for patient examination gloves.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

Yes, a standalone study was performed. The device's performance was evaluated through various non-clinical tests (physical, dimensional, pinhole, biocompatibility, residual powder) without human intervention in its function as a diagnostic aid. The study assesses the inherent properties and protective capabilities of the glove itself.

7. Type of Ground Truth Used

The ground truth used is based on:

  • Standardized ASTM specifications: ASTM-D-5250-00E4 for physical and dimensional properties, ASTM D-5151-99 for pinhole, and ASTM D6124-01 for residual powder. These standards define objective, measurable criteria.
  • Biocompatibility testing results: Ensuring no primary skin irritation or sensitization reactions, based on established toxicological principles.

8. Sample Size for the Training Set

Not applicable. This device does not involve machine learning or AI, so there is no concept of a "training set." The testing involved manufacturing samples directly.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: _ KO40702

1. Submitter's Identification:

Mr. Shuangzu Ma Grand Work Plastics Products Co., Ltd. Donggao Industrial Zone Zanhuang, Hebei Province P.R. China

Date Summary Prepared: March 01, 2004

2. Name of the Device:

Grand Work Plastics Products Co., Ltd. Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free

3. Predicate Device Information:

Shijiazhuang Great Eagle Plastic Products Co., Ltd. Synthetic Powder Free (Yellow) Vinyl Patient Examination Gloves (K992861)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80L YZ, and meets all requirements of ASTM Standard D5250-00E4.

5. Intended Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

{1}------------------------------------------------

6. Comparison to Predicate Devices:

Grand Work Plastics Products Co., Ltd. Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free, is substantially equivalent in safety and effectiveness to the Shijiazhuang Great Eagle Plastic Products Co., Ltd. Synthetic Powder Free (Yellow) Vinyl Patient Examination Gloves.

7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows:

The standards used for Grand Work Plastics Products Co., Ltd. glove production are based on ASTM-D-5250-00E4. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test based on ASTM D-5151-99 was also conducted with samplings of AQL 2.5, Inspection Level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

A Residual Powder Test that based on ASTM D6124-01 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free' claims (contain no more than 2 mg powder per glove).

8. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

9. Conclusions:

Grand Work Plastics Co., Ltd. Synthetic (Yellow) Vinyl Patient Examination gloves--- Powder Free conform fully to ASTM-D-5250-00€4 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the HHS logo, which consists of a stylized caduceus with three snakes intertwined around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the logo. The seal is in black and white.

Public Health Service

APR 2 7 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Grand Work Plastic Products Company Limited Mr. James Chu Official Correspondents Gloveco, Incorporated 590 West Central Avenue, # D Brea, California 92821

Re: K040702

Trade/Device Name: Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: March 12, 2004 Received: March 17, 2004

Dear Mr. Chu:

We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announccments concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Chu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosurc

{4}------------------------------------------------

Attachment A

Page 1 of 1

KO40702 510(k) NUMBER (IF KNOWN) : Grand Work Plastics Products Co., Ltd. DEVICE NAME: INDICATIONS FOR USE: Synthetic (Yellow) Vinyl Patient Examination Gloves-Powder Free

A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kei Mulby

(Division Sign-Off) Division of Anesthesiologi Infection Control, Denta

510(k) Number:

Page 1 of

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.