K Number
K040529

Validate with FDA (Live)

Device Name
COCOON HYGIENIST
Manufacturer
Date Cleared
2004-08-19

(171 days)

Product Code
Regulation Number
872.6640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cocoon® Hygienist is a dental operative unit intended to supply utilities to and serve as a base for dental tools and accessories for use by dental professionals.

Device Description

The Cocoon Hygienist is a dental operative unit that provides the dentist with primary requirements for dental operations. The delivery system is designed to be used as an interface device to connect the dental operating hand instruments to the appropriate supply utilities such as air, water suction, drain, and electricity. The unit is supplied with controls that allow the dentist. dental hygienist, or operator to set the water flow to the hand pieces. The water flow of the syringe, directly connected to main water supply, depends on the input water pressure. The output air pressure is preset in the unit.

AI/ML Overview

The provided text is a 510(k) summary for the Satelec Cocoon Hygienist, a dental operative unit. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria through clinical trials or performance testing with quantitative metrics as might be found for a novel or more complex medical device.

Based on the provided document, here's what can be extracted and what cannot be determined:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Substantial Equivalence to Predicate Device"The technical characteristics of the Cocoon Hygienist are almost identical to those of the Boyd Industries Inc., Dental Operative Units (K020833)."
Functionality (supply utilities, serve as base for dental tools)"The Cocoon Hygienist is a dental operative unit that provides the dentist with primary requirements for dental operations." "The intended use... is to supply utilities to and serve as a base for dental tools and accessories for the dental professional."
Safety and Effectiveness"Differences that exist between the systems relating to technical specifications, materials, physical appearance, and control systems are minor and do not affect the relative safety or effectiveness of the Cocoon Hygienist."
Given the nature of a 510(k) for a Class I dental operative unit, specific quantitative acceptance criteria (e.g., minimum flow rates, pressure ranges, durability cycles) and detailed study results demonstrating performance against these criteria are not typically included in the summary. The "acceptance criteria" here are primarily met by demonstrating substantial equivalence to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The 510(k) summary for a device like this focuses on design comparison and functional equivalence to a predicate, not on a clinical or performance study with a test set of data points or patients.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. This typically applies to studies involving human interpretation (e.g., image analysis, diagnosis), which is not relevant for a dental operative unit's 510(k) submission.

4. Adjudication Method for the Test Set

This information is not provided, as there is no mention of a test set requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Reader Improvement

This information is not provided. MRMC studies are relevant for diagnostic devices where human readers interpret results, which is not the case for this dental operative unit.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

This information is not provided. This device is a mechanical/electrical unit, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the substantial equivalence claim is effectively the performance and safety profile of the predicate device (Boyd Industries Inc., Boyd Delivery Units K020833). The manufacturer asserts that the technical and functional characteristics of the Cocoon Hygienist are "almost identical" and that minor differences do not affect safety or effectiveness, thus relying on the established history of the predicate. No specific pathology, outcomes data, or expert consensus was used as "ground truth" for this particular submission.

8. The Sample Size for the Training Set

This information is not applicable/provided. There is no mention of a "training set" in the context of device development or validation for this type of medical device submission. Training sets are typically associated with machine learning or AI models.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided for the same reasons as point 8.

{0}------------------------------------------------

K040529

Pascal Dupevron Regulatory Affairs

8. SMDA Summary of Safety and Effectiveness - "510(k) Summary"

  • A. Submitter Information
    SATELEC Z.1. du Phare, BP 216 17, Avenue Gustave Eiffel 33708 Merignac Cedex FRANCE

Telephone: 011 33 5 5634 0607

Contact Person:

Date Prepared: February 25, 2004

  • B. Device Identification
Common/Usual Name:Delivery System
Proprietary Name:Cocoon Hygienist

C. Identification of Predicate Devices

The Cocoon Hygienist is substantially equivalent to the predicate device by Boyd Industries, Inc., Boyd Delivery Units (K020833) previously cleared by FDA and currently marketed.

D. Device Description

The Cocoon Hygienist is a dental operative unit that provides the dentist with primary requirements for dental operations. The delivery system is designed to be used as an interface device to connect the dental operating hand instruments to the appropriate supply utilities such as air, water suction, drain, and electricity. The unit is supplied with controls that allow the dentist. dental hygienist, or operator to set the water flow to the hand pieces. The water flow of the syringe, directly connected to main water supply, depends on the input water pressure. The output air pressure is preset in the unit.

The intended use of the Satelec Cocoon Hygienist is to supply utilities to and serve as a base for dental tools and accessories for the dental professional.

E. Substantial Equivalence:

The technical characteristics of the Cocoon Hygienist are almost identical to those of the Boyd Industries Inc., Dental Operative Units (K020833). Differences that exist between the systems relating to technical specifications, materials, physical appearance, and control systems are minor and do not affect the relative safety or effectiveness of the Cocoon Hygienist.

: 行 :

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 9 2004

Satelec C/O Dr. Jean-Luc Boulnois President Interactive Consulting, Incorporated 70 Walnut Street Wellesley, Massachusetts 02481

Re: K040529

Trade/Device Name: Cocoon® Hygienist Regulation Number: 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: August 10, 2004 Received: August 11, 2004

Dear Dr. Boulnois:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Boswell

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

Ko40529

Device Name: Cocoon® Hygienist

Indications for Use:

The Cocoon® Hygienist is a dental operative unit intended to supply utilities to and serve as a base for dental tools and accessories for use by dental professionals.

Please refer to the attached listing for a complete description.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Romano

n of Anesthesiology, General Hospital. Infection Control Dental D

510(k) Number: J(040526

Prescription Use X (Per 21 CFR 801.109)

OR

Over-The-Counter Use

§ 872.6640 Dental operative unit and accessories.

(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.