K Number
K040475

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-05-21

(87 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for the MicroMax™ Suture Anchor include use in soft tissue reattachment procedures in the shoulder, wrist, elbow and knee. Specific indications are as follows:

Shoulder Indications:

  • Bankart repair
  • SLAP lesion repair
  • Acromioclavicular (ACL) separation
  • Rotator cuff repair
  • Capsule repair or capsuloabral reconstruction
  • Biceps tenodesis
  • Deltoid repair

Wrist Indications:

  • Scapholunate ligament reconstruction

Elbow Indications:

  • Tennis elbow repair
  • Ulnar or radial collateral ligament reconstruction
  • Biceps tendon reattachment
  • Medial and lateral repairs

Knee Indications (Extra-capsular repair):

  • Medial collateral ligament repair
  • Lateral collateral ligament repair
  • Posterior oblique ligament repair
  • Joint capsule closure
  • Iliotibial band tenodesis reconstruction
  • Patellar ligament/tendon repair
  • Vastus medialis obliquus (VMO) muscle advancement

The device is pre-loaded with suture for use at the discretion of the physician.

Device Description

This device is a resorbable suture anchor designed to attach soft tissue to bone. The resorbable fixation anchor is comprised of L-Lactide / Glycolide material. The poly L-Lactic/polyglycolic acid copolymer degrades by hydrolysis into L-lactic and glycolic acids. These hydrolytic products are then further degraded into carbon dioxide and water via the cellular Krebs cycle. The material and rarer degraged into carron anonial and o the predicate device, LactoSorb® L15 Soft Tissue Screw and Washer, K012572.

The suture anchor measures 3.0mm in diameter and 9.0mm in length. It is a two-piece assembly, consisting of a body and a head portion. The body portion engages the bone, which is enhanced by means of seven circumferential ribs. The ribs measure 3.25mm in diameter. In addition, the body has two wings, that when deployed will flange outward into the bone. Upon deployment, the diameter at the flanged wings is 4mm.

The head portion provides a means to drive the anchor in as well as to attach suture to the anchor. The anchor has an eyelet through the head portion to allow suture to pass through which will be used to fasten the tissue to bone. The device will be pre-loaded with any legally marketed, size 2 suture.

Instrumentation is provided for proper use and placement of the device.

AI/ML Overview

Here's an analysis of the provided text regarding the MicroMax™ Suture Anchor, focusing on acceptance criteria and supporting studies:

It is important to note that the provided document is a 510(k) summary for a medical device seeking substantial equivalence to a legally marketed predicate device. This type of submission relies on demonstrating that the new device is as safe and effective as a device already on the market, rather than proving efficacy from scratch with extensive clinical trials.

Acceptance Criteria and Reported Device Performance

The document states: "Non-Clinical Testing: Mechanical testing was performed to establish substantial equivalence to the predicate devices." Implied acceptance criteria would revolve around these mechanical test results demonstrating equivalence.

Given the information, the specific numerical acceptance criteria (e.g., minimum tensile strength, degradation rate, etc.) and the precise reported performance values for the MicroMax™ Suture Anchor are not explicitly detailed in this summary. The summary broadly indicates that the characteristics are "similar to predicate devices."

Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criterion (Implied)MicroMax™ Suture Anchor Performance (Implied)
Material composition and degradation propertiesSimilar to LactoSorb® L15 (L-Lactide / Glycolide material, degrades by hydrolysis)
Device dimensions (diameter, length, rib diameter, flanged diameter)Matches specified dimensions (3.0mm dia, 9.0mm length, 3.25mm ribs, 4mm flanged)
Mechanical properties (e.g., strength, security of fixation)Passed mechanical testing to establish substantial equivalence to predicate device. (Specific values not provided)
Suture compatibilityPre-loaded with legally marketed, size 2 suture.

Study Details:

  1. Sample Size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified for mechanical testing.
    • Data Provenance: Not specified, but generally, mechanical testing data for a 510(k) submission is generated in a laboratory setting, typically in the country where the manufacturer is located (USA in this case). It would be considered prospective data collection for the purpose of the submission.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. Mechanical testing does not typically involve human experts in establishing "ground truth" in the same way clinical or image-based studies do. The "ground truth" for mechanical testing is derived from established engineering standards and test methodologies.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are relevant for clinical studies or studies involving subjective human assessment. Mechanical testing results are objective measurements.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device for surgical fixation, not an AI-based diagnostic or imaging device. Therefore, MRMC studies and AI assistance are not relevant to this submission.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical mechanical testing, the "ground truth" is defined by engineering standards and established test methods that evaluate properties like material degradation, dimensional conformity, and mechanical strength (e.g., pull-out strength, fatigue life, etc., although not explicitly listed in this summary). The comparison is against the performance of the legally marketed predicate device.
  7. The sample size for the training set:

    • Not Applicable. This device did not use a "training set" in the context of machine learning or AI.
  8. How the ground truth for the training set was established:

    • Not Applicable. No training set was used.

Summary of what is present and absent:

  • The document clearly states that non-clinical mechanical testing was performed.
  • The purpose of this testing was to establish substantial equivalence to a predicate device (LactoSorb® Soft Tissue Screw and Washer, K012572).
  • The material (L-Lactide / Glycolide) and general design are explicitly stated to be similar to the predicate.
  • No clinical testing was used for this submission to establish substantial equivalence.
  • Detailed numerical results of the mechanical tests or the specific acceptance thresholds for those tests are not provided in this summary. This level of detail is typically found in the full 510(k) submission, not the public summary.
  • This submission method (510(k)) focuses on equivalence to a previously cleared device, rather than new efficacy claims from a de novo study.

{0}------------------------------------------------

MAY 2 1 2004

BIOMET

K040475-
page 1 of 2

510(k) Summary

Biomet Manufacturing Corp. Applicant/Sponsor: Contact Person: Kacy Arnold

Requlatory Specialist

MicroMax® Suture Anchor Proprietary Name:

Common Name: Suture Anchor

Fastener, Fixation, Biodegradable soft tissue, 888.3045 Classification Name:

Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

LactoSorb® Soft Tissue Screw and Washer (K012572)

Device Description:

.

This device is a resorbable suture anchor designed to attach soft tissue to bone. The resorbable fixation anchor is comprised of L-Lactide / Glycolide material. The poly L-Lactic/polyglycolic acid copolymer degrades by hydrolysis into L-lactic and glycolic acids. These hydrolytic products are then further degraded into carbon dioxide and water via the cellular Krebs cycle. The material and rarer degraged into carron anonial and o the predicate device, LactoSorb® L15 Soft Tissue Screw and Washer, K012572.

The suture anchor measures 3.0mm in diameter and 9.0mm in length. It is a two-piece assembly, consisting of a body and a head portion. The body portion engages the bone, which is enhanced by means of seven circumferential ribs. The ribs measure 3.25mm in diameter. In addition, the body has two wings, that when deployed will flange outward into the bone. Upon deployment, the diameter at the flanged wings is 4mm.

The head portion provides a means to drive the anchor in as well as to attach suture to the anchor. The anchor has an eyelet through the head portion to allow suture to pass through which will be used to fasten the tissue to bone. The device will be pre-loaded with any legally marketed, size 2 suture.

Instrumentation is provided for proper use and placement of the device.

MAILING ADDRESS 120. Box 587 Warsaw, IN 46581 0587

.

SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582

()[+][()[)[+](()+[(+)[(+(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+)(+) 574.267.6639

FAX "" 1. 36 . " . 86 . " . 8 . "

E-MAIL biomer@bionnet.com

{1}------------------------------------------------

K040475-
page 2 of 2

Intended Use:

ed Use:
The MicroMax™ Suture Anchor is intended for use in soft tissue reattachment procedures in the The Fileronax - Specific indications are as follows:

Shoulder Indications:

  • Bankart repair .
  • SLAP lesion repair �
  • Acromioclavicular (ACL) separation .
  • Rotator cuff repair .
  • Capsule repair or capsuloabral reconstruction .
  • Biceps tenodesis, deltoid repair �

Wrist Indications:

  • Scapholunate ligament reconstruction .

Elbow Indications:

  • Tennis elbow repair .
  • Ulnar or radial collateral ligament reconstruction .
  • Biceps tendon reattachment .
  • Medial and lateral repairs �

Knee Indications (Extra-capsular repair):

  • Medial collateral ligament repair .
  • Lateral collateral ligament repair .
  • Posterior oblique ligament repair .
  • Joint capsule closure .
  • Iliotibial band tenodesis reconstruction .
  • Patellar ligament/tendon repair .
  • Vastus medialis obliquus (VMO) muscle advancement .

The device is pre-loaded with suture for use at the discretion of the physician.

Summary of Technologies: The MicroMax™ Suture Anchor technological characteristics (material and design) are similar to predicate devices.

Non-Clinical Testing: Mechanical testing was performed to establish substantial equivalence to the predicate devices.

Clinical Testing: Clinical testing was not used to establish substantial equivalence to predicate devices.

All trademarks are property of Biomet, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three wavy lines or streams, which is a common element in the department's branding.

Public Health Service

MAY 2 1 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Kacy Arnold, RN, MBA Regulatory Specialist Biomet Manufacturing Corporation P.O. Box 587 Warsaw, Indiana 46581

Re: K040475

Trade/Device Name: MicroMax™ Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: February 23, 2004 Received: February 24, 2004

Dear Ms. Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enerebate) to the enactment date of the Medical Device Amendments, or to conniser that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmotion (110. (110.) that the device, subject to the general controls provisions of the Act. The I ou may, dierere, misions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your device is elaboried (toval controls. Existing major regulations affecting your device can may be subject to Salental Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Kacy Arnold, RN, MBA

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter with and you to begin mailioning of substantial equivalence of your device to a legally premarket notincation: The PDF intelling of your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific darno 12-7 301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. IMIsoranuming of reference to premaisonsibilities under the Act may be obtained from the Other general miorination on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

L. Mark A. Milliman

Celia M. Witten, Ph.D., M.D. Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

K040475

Device Name:

MicroMax™ Suture Anchor

Indications For Use:

Indications for the MicroMax™ Suture Anchor include use in soft tissue reattachment procedures
in the MicroMax™ Suture Anchor inclinations and fallows: Indications for the Microriax - Sucare Andrews
in the shoulder, wrist, elbow and knee. Specific indications are as follows:
in the shoulder, wrist, elbow and knee. Specifi

Shoulder Indications.

  • Bankart repair ●
  • SLAP lesion repair ●
  • Acromioclavicular (ACL) separation .
  • Rotator cuff repair .
  • Capsule repair or capsuloabral reconstruction ●
  • Biceps tenodesis ◆
  • Deltoid repair .

Wrist Indications:

  • Scapholunate ligament reconstruction ●

Elbow Indications:

  • Tennis elbow repair .
  • Ulnar or radial collateral ligament reconstruction .
  • Biceps tendon reattachment �
  • Medial and lateral repairs .

Knee Indications (Extra-capsular repair):

  • Medial collateral ligament repair .
  • Lateral collateral ligament repair .
  • Posterior oblique ligament repair .
  • Joint capsule closure �
  • Iliotibial band tenodesis reconstruction .
  • Patellar ligament/tendon repair .
  • Vastus medialis obliquus (VMO) muscle advancement �

The device is pre-loaded with suture for use at the discretion of the physician.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use no
(21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Mark A. Mullean
Division Sign-Off

Division of General, Restorati and Neurological Devices

510(k) Number

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.