K Number
K040367

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-04-07

(54 days)

Product Code
Regulation Number
868.5665
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Electromed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.

Device Description

The proposed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport. The primary components of the MedPulse® System include an Air-Pulse Generator, an Inflatable Vest and Vest/Generator Connector Hose. The Air-Pulse Generator produces oscillating pressurized air-pulses that are delivered to the Inflatable Vest by the Vest/Generator Hose. The air-pulses produced by the Generator cause the Vest to rapidly inflate and deflate against the external chest wall of a patient to promote airway clearance by creating high-frequency chest wall oscillation (HFCWO) resulting in mobilization of bronchial secretions.

AI/ML Overview

The Electromed MedPulse® Respiratory Vest System Model 2000ez was deemed substantially equivalent to a predicate device (MedPulse® Respiratory Vest System Model 2000, 510(K) number: K982889). Therefore, the acceptance criteria and study design focused on demonstrating that changes made to the new model did not negatively impact its performance or safety compared to the predicate.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Maintain identical indications and intended use as the predicate device.The MedPulse® Respiratory Vest System Model 2000ez has identical indications and intended use: "to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport."
Maintain identical technological characteristics and principles of operation.The Model 2000ez continues to use the same method of generating pressurized air-pulses (dual acting pneumatic diaphragm technology). Users can still adjust air pulse frequency and pressure amplitude. The Inflatable Vest and Vest/Generator Connector Hose remain unchanged. The only changes are to the Air-Pulse Generator's motor (servo brush to brushless) and frequency control (analog to digital), both of which were described as maintaining or improving performance without changing the fundamental operating principle.
No change in output performance compared to the predicate device.Functional performance tests were conducted and explicitly stated to have "demonstrated no change in output performance." While specific quantitative metrics are not provided, the claim is that the output performance remains equivalent.
No new safety issues to result from the proposed changes.Electrical safety tests were performed and explicitly stated to have "demonstrated no ... safety issues to result from the changes proposed."
Enhanced reliability and longer lifespan of the motor (brushless motor change).The change to a brushless motor was intended to make the motor "more reliable and have a longer life span." While this is an stated intent, the performance testing section focuses on demonstrating no change in output performance or safety, rather than providing specific data on enhanced lifespan/reliability.
Enhanced robustness and reliability of frequency control (digital control change).The change to digital frequency control was considered to make the motor "more robust and reliable." Similar to the motor lifespan, the performance testing focused on maintaining output performance and safety, not on quantifying the enhancement in robustness or reliability from this specific change. The digital control also presented the opportunity to "remember" user settings, improving ease and reliability of use, though this is a feature enhancement rather than a core performance metric for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for the Test Set: Not explicitly stated in the provided text. The document mentions "Functional performance and electrical safety tests were performed," but does not specify the number of units tested or the duration of testing.
  • Data Provenance: The study was conducted by Electromed, Inc., (New Prague, MN, USA) for submission to the FDA. Given it is a premarket notification for a new version of an existing device, the data is prospective testing of the new device and comparative testing against the predicate.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This type of information is not applicable to this submission. The submission is for a medical device (Respiratory Vest System) that does not involve diagnostic image interpretation or complex clinical judgment that would require expert consensus for a test set ground truth. The "ground truth" here is the established performance and safety characteristics of the predicate device, against which the modified device is being compared.

4. Adjudication Method for the Test Set:

  • This information is not applicable as the evaluation involved functional performance and electrical safety testing, not subjective interpretation requiring adjudication among experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human readers interpret results, and the AI's impact on reader performance is being evaluated. This device is a therapeutic physical therapy device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Yes, in essence, a standalone performance study was done. The "Functional performance and electrical safety tests" conducted on the MedPulse® Respiratory Vest System Model 2000ez represent a standalone evaluation of the device's operational characteristics (e.g., air pulse generation, frequency, pressure stability, electrical integrity) independent of human operation. The intent was to show that the device itself, with its modifications, performs no worse than the predicate.

7. Type of Ground Truth Used:

  • The "ground truth" for this substantial equivalence claim is the established performance and safety characteristics of the predicate device (MedPulse® Respiratory Vest System Model 2000, K982889). The study's goal was to demonstrate that the new device's performance and safety metrics were equivalent to or better than this established baseline. This would involve objective measurements against design specifications and relevant safety standards.

8. Sample Size for the Training Set:

  • This information is not applicable. This is a hardware medical device, not an AI or algorithm-driven system that requires a "training set" in the machine learning sense. The device's design is based on engineering principles and previous device models, not on learning from a dataset.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not applicable for the reasons stated above (not an AI/ML device).

{0}------------------------------------------------

APR - 7 2004

Image /page/0/Picture/1 description: The image shows a series of alphanumeric characters, likely a code or identifier. The characters are "K0403657" and are oriented diagonally in the image. The text appears to be handwritten or printed with a slightly uneven quality.

ELECTROMED, INC.

502 Sixth Avenue N.W. New Prague, MN 56071 Phone: 952-758-9299 · Fax: 952-758-1941

February 7, 2004

SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990.

Submitter:Chet SievertRegulatory AffairsElectromed, Inc.502 Sixth Avenue N.W.New Prague, MN 56071Phone: 952.758.9299Fax: 952.758.1941
Contact person:Chet Sievert
Name of Device:MedPulse® Respiratory Vest SystemModel 2000ez
Classification:Powered Percussor, Class II
Predicate Device:MedPulse® Respiratory Vest SystemModel 2000510(K) number: K982889

Description of Device:

The proposed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport. The primary components of the MedPulse® System include an Air-Pulse Generator, an Inflatable Vest and Vest/Generator Connector Hose. The Air-Pulse Generator produces oscillating pressurized air-pulses that are delivered to the Inflatable Vest by the Vest/Generator Hose. The air-pulses produced by the Generator cause the Vest to rapidly inflate and deflate against the external chest wall of a patient to promote airway clearance by creating high-frequency chest wall oscillation (HFCWO) resulting in mobilization of bronchial secretions.

{1}------------------------------------------------

Intended Use:

The intended use of the Electromed MedPulse® Respiratory Vest System Model 2000ez is to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.

Comparison of Technological Characteristics:

The proposed MedPulse® Respiratory Vest System Model 2000ez has the same technological characteristics and principles of operation as Electromed's previously cleared Medpulse® Respiratory Vest System Model 2000 (K982889). The proposed MedPulse® System will continue to use the same method of generating pressurized air-pulses as the Predicate, i.e., by dual acting pneumatic diaphragm technology. The user of the Proposed System will continue to be able to adjust air pulse frequency and pressure amplitude as the Predicate System. The Inflatable Vest and Vest/Generator Connector Hose components of the Modified MedPulse® System will remain unchanged from the Predicate System.

The reason for this submission is due to two component changes to the Air-Pulse Generator's motor. The first proposed change is to switch from a servo brush motor to a brushless motor. The motor's output performance will remain the same. A change to a brushless technology is intended to make the motor more reliable and have a longer life span. The second proposed change is to change the motor's frequency control from an analog component (i.e., potentiometer) to a digital control component. The change from analog to digital frequency control is considered to make the Air-Pulse Generator's motor more robust and reliable. The motor's digital frequency control also presents the opportunity to use the same component to control the Timer feature on the Control Panel that regulates the duration of treatment. This change allows the System to "remember" the user's previous settings for ease and more reliable use.

Performance Testing:

The proposed MedPulse® Respiratory Vest System Model 2000ez was tested and compared to the predicate Medpulse® Respiratory Vest System Model 2000 (K982889). Functional performance and electrical safety tests were performed that demonstrated no change in output performance or safety issues to result from the changes proposed.

Substantial Equivalence:

The proposed MedPulse® Respiratory Vest System Model 2000ez has the identical indications and intended use, technological characteristics and principles of operation as the Predicate device. Functional performance and electrical safety testing have demonstrated no changes in safety or effectiveness. Thus, the proposed MedPulse® Respiratory Vest System Model 2000ez is substantially equivalent to the proposed Medpulse® Respiratory Vest System Model 2000 (K982889).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 7 2004

Mr. Chet Sievert Regulatory Consultant Electromed, Incorporated 502 Sixth Avenue N.W. New Prague, MN 56071

Re: K040367

Trade Name: Medpulse Respiratory Vest System, Model 2000ez Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: II Product Code: BYI Dated: March 16, 2004 Received: March 19, 2004

Dear Mr. Sievert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDAravay, publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Chet Sievert

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

for

Sincerely yours,

stt Wuttersheue

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number: K040367

Device Name: MEDPULSE RESPIRATORY VEST SYSTEM, MODEL 2000ez

Indications For Use:

The Electromed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

JtW.h

sthesiology, General Hospital,

510(k) Number: K040367

Page 1 of 1

§ 868.5665 Powered percussor.

(a)
Identification. A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.(b)
Classification. Class II (performance standards).