K Number
K040328

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-11-24

(288 days)

Product Code
Regulation Number
876.5540
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

THE FREE FLOW™ DIALYSIS CATHETER IS INDICATED FOR USE IN ATTAINING LONG-TERM ACCESS FOR HEMODIALYSIS AND APHERESIS. IT MAY BE INSERTED PERCUTANEOUSLY AND PRIMARILY PLACED IN THE INTERNAL JUGULAR VEIN OF AN ADULT PATIENT. ALTERNATE INSERTION SITES INCLUDE SUBCLAVIAN VEIN AS REQUIRED. THE CURVED FREE-FLOW™ CATHETER IS INTENDED FOR INTERNAL JUGULAR VEIN INSERTION. THIS CATHETER IS INDICATED FOR A DURATION NOT TO EXCEED (12) MONTHS.

Device Description

The Free Flow™ Long-Term Coaxial Catheter is manufactured by Medcomp. The catheter is polyurethane (Carbothane) that provides two dedicated (arterial/venous) access lumens. Each lumen is connected through an extension line with female luer connectors. The transition between lumen and extension is housed within & molded hub. To prevent recirculation, the venous end of the lumen is approximately 1. J" longer than the arterial. The catheter is capable of providing consistent flows up to 450 ml/min at a pressure of less than 200mm Hg. The catheter lumen is composed of a soft, thermo sensitive, Product Features: polyurethane material that is rigid upon insertion and once it reaches body temperature it becomes soft to reduce vessel trauma. The catheter assembly contains a pre-loaded stylet for ease of insertion. The catheter hub is molded from soft, pliable polyurethane to increase patient comfort. The dialysis extensions are color coded with a red luer and a clamp for the arterial lumen, a blue luer and a clamp for the venous lumen for easy identification. Physical Dimensions: The 15.5 French catheter will be available in a straight lumen design in 24, 28 and 32cm.

AI/ML Overview

The provided text describes a 510(k) submission for the Free Flow™ Catheter, a hemodialysis device, and focuses on demonstrating its substantial equivalence to a predicate device through in vitro performance testing. It does not involve AI/ML, clinical studies, or human image interpretation, thus many of the requested categories are not applicable.

Here's an attempt to answer based on the provided text, indicating "Not Applicable" where the information is not present or relevant to this type of device submission:

Acceptance Criteria and Device Performance

The acceptance criteria for the Free Flow™ Catheter are implicitly derived from the performance of the legally marketed predicate device (Medcomp Ash Split-Cath® II Catheter, K040328). The study's goal was to demonstrate "substantial equivalence" to this predicate device by ensuring the Free Flow™ Catheter's in vitro performance was comparable.

Acceptance Criteria/Performance MetricReported Device Performance
Force @ BreakMet (demonstrated substantial equivalence to predicate)
Air/Liquid LeakMet (demonstrated substantial equivalence to predicate)
RecirculationMet (demonstrated substantial equivalence to predicate)
Flow vs. PressureMet (demonstrated substantial equivalence to predicate)
Cuff ShearMet (demonstrated substantial equivalence to predicate)
Gravity FlowMet (demonstrated substantial equivalence to predicate)
Flow Rate (up to)450 ml/min at < 200mm Hg. (This is a design specification, implied to be met).

Study Details (with AI/ML context in mind, noting non-applicability)

  1. Sample size used for the test set and the data provenance: Not Applicable. This was an in vitro performance study comparing a new device to a predicate, not an AI/ML model test set. The "test set" would refer to the physical catheters and the conditions under which they were tested in a lab. The data provenance is in vitro testing conducted by Medcomp.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not Applicable. Ground truth as typically defined for AI/ML (e.g., expert consensus on medical images) does not apply here. The "ground truth" for this device is its physical and functional performance as measured in in vitro tests against established engineering and performance benchmarks, and in comparison to a predicate device.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not Applicable. This is not an AI/ML study involving human adjudication of results.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not Applicable. No MRMC study was performed. This is a medical device (hemodialysis catheter) approval, not an AI/ML algorithm.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not Applicable. No AI/ML algorithm was involved.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for this submission is the in vitro performance specifications and the established performance of the predicate device (Medcomp Ash Split-Cath® II Catheter, K040328), against which the Free Flow™ Catheter was compared to demonstrate substantial equivalence across various physical and hydrodynamic parameters.

  7. The sample size for the training set: Not Applicable. There is no AI/ML model or "training set" in this context.

  8. How the ground truth for the training set was established: Not Applicable. There is no AI/ML model or "training set" in this context.

{0}------------------------------------------------

510(k) SUMM

Submitter Information: A.

Submitter:NOV 2 4 2004MEDCOMP®1499 Delp DriveHarleysville, PA 19438(215) 256-4201 Telephone(215) 256-9191 Fax
Contact:Rosanna SeveriniCompliance Group Leader
Date Prepared:October 15, 2004
Trade Name:Free Flow™ Catheter
Common Name:Hemodialysis Catheter, Implanted
Classification:78 MSD
C.F.R. Section:876.5540
Predicate Device:K040328 Medcomp Ash Split-Cath(®) IICatheter

Device Description: D.

в.

்.

The Free Flow™ Long-Term Coaxial Catheter is manufactured by Medcomp. The catheter is polyurethane (Carbothane) that provides two dedicated (arterial/venous) access lumens. Each lumen is connected through an extension line with female luer connectors. The transition between lumen and extension is housed within & molded hub. To prevent recirculation, the venous end of the lumen is approximately 1. J" longer than the arterial. The catheter is capable of providing consistent flows up to 450 ml/min at a pressure of less than 200mm Hg.

The catheter lumen is composed of a soft, thermo sensitive, Product Features: polyurethane material that is rigid upon insertion and once it reaches body temperature it becomes soft to reduce vessel trauma. The catheter assembly contains a pre-loaded stylet for ease of insertion.

The catheter hub is molded from soft, pliable polyurethane to increase patient comfort.

The dialysis extensions are color coded with a red luer and a clamp for the arterial lumen, a blue luer and a clamp for the venous lumen for easy identification.

Physical Dimensions: The 15.5 French catheter will be available in a straight lumen design in 24, 28 and 32cm.

E. Intended Use:

The Free Flow™ dialysis catheter is indicated or use in attaining long-term access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily claced in the internal jugular vein of an adult patient. Alternate insertion sites include subclavian vein as required. This catheter is indicated for a duration not to exceed (12) months.

{1}------------------------------------------------

Koyagi. R.

F. Comparison to Predicate Device:

The technological characteristics of the Free Flow™ Long-Term Coaxial Catheter are substantially equivalent to the predicate devices in terms of intended use, insertion method, materials, performance, labeling, manufacturing process, and method of sterilization.

The modifications include:

  • Addition of pre-loaded stylet for ease of insertion .
  • No side holes .
  • Lumen design change .

G. Performance Data:

In vitro performance data for the Free Flow™ Long-Term Coaxial Catheter, including force @ break, air/liquid leak, recirculation, flow vs. pressure, cuff shear and gravity flow demonstrate that this device is substantially equivalent to the legally marketed predicate device.

Clinical studies were not deemed necessary since in vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to legally marketed predicate device.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2 4 2004

Ms. Rosanna Severini Compliance Group Leader Medical Components, Inc. 1499 Delp Drive HARLEYSVILLE PA 19438

Re: K040328

Trade/Device Name: Free Flow™ Long-Term Coaxial Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: 78 MSD Dated: November 15, 2004 Received: November 18, 2004

Dear Ms. Severini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase vour device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Rosanna Severini

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K040328

Device Name: FREE FLOW™ LONG-TERM COAXIAL CATHETER

Indications for use:

THE FREE FLOW™ DIALYSIS CATHETER IS INDICATED FOR USE IN ATTAINING LONG-TERM ACCESS FOR HEMODIALYSIS AND APHERESIS.

IT MAY BE INSERTED PERCUTANEOUSLY AND PRIMARILY PLACED IN THE INTERNAL JUGULAR VEIN OF AN ADULT PATIENT. ALTERNATE INSERTION SITES INCLUDE SUBCLAVIAN VEIN AS REQUIRED.

THE CURVED FREE-FLOW™ CATHETER IS INTENDED FOR INTERNAL JUGULAR VEIN INSERTION.

THIS CATHETER IS INDICATED FOR A DURATION NOT TO EXCEED (12) MONTHS.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter

Nancy C. Brogdon
(Division Sign-Off)

(Optional Format 1-2-96)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K040328

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.