K Number
K040247

Validate with FDA (Live)

Date Cleared
2004-03-03

(29 days)

Product Code
Regulation Number
876.5540
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use for temporary cannulation for vascular access for extracorporeal blood treatment. The device is intended for single use only and is for temporary catheterization of less than 30 days. The device has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries when withdrawing and discarding the needle after treatment.

Device Description

Needle, Fistula

AI/ML Overview

The provided text is a 510(k) clearance letter from the FDA for a medical device called the "Plume Safety AVF Needle." This document grants market clearance based on substantial equivalence to a predicate device, rather than providing details about an acceptance criteria study as typically conducted for novel AI/software devices.

Therefore, most of the information requested in your prompt regarding acceptance criteria, study design, expert involvement, and performance metrics is not available in this type of regulatory document.

Here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Not Available. This document does not describe specific performance acceptance criteria or report performance data. It is a regulatory clearance letter based on substantial equivalence.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Available. No test set or data provenance details are provided as this is not a study report.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Available. No experts or ground truth establishment are mentioned as this is not a study report.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Available. No adjudication method is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This device is a physical medical needle, not an AI/software device. Therefore, an MRMC study or AI assistance is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Available. Ground truth is not relevant or mentioned for this type of device clearance.

8. The sample size for the training set

  • Not Available. No training set is relevant or mentioned.

9. How the ground truth for the training set was established

  • Not Available. No training set or ground truth establishment is relevant or mentioned.

Information that is available (though not directly addressing your specific questions about study criteria):

  • Device Name: Plume Safety AVF Needle
  • Indications For Use: "Use for temporary cannulation for vascular access for extracorporeal blood treatment. The device is intended for single use only and is for temporary catheterization of less than 30 days. The device has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries when withdrawing and discarding the needle after treatment."
  • Regulatory Classification: Class II
  • Predicate Device Equivalence: The clearance is based on the device being "substantially equivalent" to legally marketed predicate devices. This means the FDA found that the new device is as safe and effective as a legally marketed device and does not raise different questions of safety and effectiveness. The specifics of the predicate device are not included in this extract, but would have been part of the 510(k) submission.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stylized wings.

MAR - 3 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Swee Cheau, Chong Manager of Regulatory Affairs & Quality Assurance JMS North America Corporation 22320 Foothill Blvd. Suite 350 HAYWARD CA 94541

Re: K040247

Trade/Device Name: Plume Safety AVF Needle Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Product Code: 78 FIE Regulation Number: 21 CFR $880.5200 Regulation Name: Intravascular catheter Product Code: 78 FOZ Regulatory Class: II Dated: January 30, 2004 Received: February 3, 2004

Dear Ms. Chong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be rubject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (scctions 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your intial equivalence of your device to a legally premarket notification. The FDA miding or basicing of active and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r the regulation If you desire specific advice tor your device on one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the pronocourance the regulation entitled, "Misbranding
Office of Compliance at (301) 594-4639. Also, please at their Other separal Other of Compliance at (501) 594-4057. Taxo, production of South Pingion of Small by reference to premarket nouncation (21 cc Act may be obtained from the Division of Small
information on your responsibilities under the Act may be objacts from the (800) 63 information on your responsibilities unec the Asstance at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-friend/demag Manufacturers, International and Collsamor i Lists and Collection of Schoolsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices ()ffice of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{2}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(K) Number:

K040247

Device Name:

Needle, Fistula

Indications For Use:

Use for temporary cannulation for vascular access for extracorporeal blood treatment. The device is intended for single use only and is for temporary catheterization of less than 30 days. The device has an anti-stick feature integrated as part of the Needle Set which aids in prevention of needle-stick injuries when withdrawing and discarding the needle after treatment.

(Please Do Not WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseOROver-The-Counter Use
(Per 21 CFR Section 801.109)

Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices

510(k) NumberK040247
------------------------

10

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.