K Number
K040115

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-01-30

(10 days)

Product Code
Regulation Number
862.1495
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Piccolo Magnesium Test System (presently contained on the Renal Function Panel Plus Reagent Disc) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of magnesium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia.

Device Description

The Piccolo Renal Function Panel Plus Reagent Disc (which contains the Piccolo Magnesium Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.

AI/ML Overview

The provided document describes the safety and effectiveness of the Piccolo® Magnesium Test System. The device is intended for the in vitro quantitative determination of magnesium in heparinized whole blood, heparinized plasma, or serum.

Here's an analysis of the acceptance criteria and the study performance:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" with pass/fail thresholds for each metric. Instead, it presents performance data (linearity, precision) and draws a conclusion of substantial equivalence to a predicate device. For the purpose of this analysis, we will infer the "acceptance criteria" through comparison to the predicate device's characteristics and general expectations for clinical laboratory devices.

Performance MetricPredicate Device (Vitros 950 Magnesium Slides)Piccolo Magnesium Test System Reported PerformanceImplied Acceptance Criteria (Based on Substantial Equivalence)Met?
LinearityNot explicitly stated (implied to be acceptable)Slope: 0.992, Intercept: -0.05, Corr. Coefficient: 0.999Correlation Coefficient ≥ 0.99 (or comparable to predicate)Yes
Sensitivity0.2 mg/dL0.1 mg/dLEqual to or better than predicate (≤ 0.2 mg/dL)Yes
Assay Range0.2 - 10.0 mg/dL0.1 - 8.0 mg/dLComparable to predicate's clinical relevance RangeYes*
Within-Run Precision (CV)Not explicitly statedLevel 1: 1.7%, Level 2: 1.0%CVs typically < 5% or comparable to predicateYes
Total Precision (CV)Not explicitly statedLevel 1: 3.4%, Level 2: 2.6%CVs typically < 5% or comparable to predicateYes
Sample Type ComparabilityHeparinized plasma, serum, urineHeparinized venous whole blood, serumDemonstrated comparability for declared sample typesYes

*Note on Assay Range: While the Piccolo's upper range (8.0 mg/dL) is slightly lower than the predicate (10.0 mg/dL), the lower range is better (0.1 mg/dL vs 0.2 mg/dL). The document concludes that these values support substantial equivalence, implying the clinical utility of the 0.1 - 8.0 mg/dL range is acceptable.

2. Sample Size Used for the Test Set and Data Provenance:

  • Linearity Study: Sample sizes are not explicitly stated, but are generally multi-point dilutions across the assay range.
  • Precision Study:
    • Within-Run and Total Precision: "n=80" for each level (Level 1 and Level 2) for Magnesium. This 'n' likely refers to the number of replicates performed to achieve the precision data.
  • Sample Type Comparison: Sample sizes are not explicitly stated.
  • Data Provenance: The document does not specify the country of origin of the data. The studies appear to be prospective as they were specifically performed to evaluate the performance of the Piccolo Magnesium Test System.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not provided in the document. For in vitro diagnostic devices like the Piccolo Magnesium Test System, "ground truth" is typically established by reference methods or comparison to established, legally marketed devices. The document references the Vitros 950 Chemistry System as the predicate device, implying its performance serves as a comparative "ground truth" or benchmark.

4. Adjudication Method for the Test Set:

This information is not applicable and therefore not provided. Adjudication methods (e.g., 2+1) are typically used in clinical trials where human interpretation or expert consensus is required to establish a diagnosis, especially in imaging studies. For a quantitative chemical analyzer, the "truth" is derived from the analytical performance (e.g., linearity, precision, accuracy against a reference method or validated predicate).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. This type of study is specifically designed for evaluating the impact of AI or new technologies on human reader performance, often in diagnostic imaging. The Piccolo Magnesium Test System is a standalone analytical device, not an AI-assisted interpretation tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance study was done. The entire document focuses on the analytical performance of the Piccolo Magnesium Test System as a standalone device (algorithm only, in the sense of the instrument's processing logic), without human-in-the-loop performance as it relates to diagnostic interpretation. The studies on linearity, precision, and sample type comparability are all standalone performance assessments.

7. The Type of Ground Truth Used:

The primary "ground truth" used for comparison appears to be the performance characteristics of a legally marketed predicate device (Vitros 950 Chemistry System - Magnesium Slides) and established analytical chemistry principles. While not explicitly stated, the linearity and precision studies would have used control materials or patient samples with known or well-characterized magnesium concentrations, which act as the ground truth for those measurements. The "gold standard" for magnesium measurement itself is a well-defined chemical assay.

8. The Sample Size for the Training Set:

This information is not applicable and therefore not provided. The Piccolo Magnesium Test System is a traditional in vitro diagnostic device, not an AI/ML device that requires a distinct "training set" in the computational learning sense. The device's factory calibration and lot-specific data, referenced in the calibration method, suggest internal validation and calibration processes, but not a "training set" like that used for algorithms.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable for the reasons stated above (not an AI/ML device). The "ground truth" for its calibration would be based on certified reference materials and established metrological traceability for magnesium concentrations.

{0}------------------------------------------------

JAN 5 0 2004

onle Road, Union City, CA 94587 510 . 675-6500 Fax 510 . 441-6150

ABAXIS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K040115

1. Applicant Information:

Date Prepared:January 15, 2004
Name:Abaxis, Inc.
Address:3240 Whipple RoadUnion City, CA 94587
Contact Person:Dennis M. Bleile, PhD
Phone Number:(510) 675-6515

2. Device Information:

Fax Number:

ClassificationClass I
Trade Name:Piccolo® Magnesium Test System

(510) 441-6150

862.1495 Classification Name: Magnesium Test system

Identification of legally marketed device to which the submitter claims 3. equivalence:

The following table identifies the legally marketed device to which Abaxis claims equivalence:

Predicate Device
Predicate DeviceManufacturer510(k)NumberDate of SEDetermination
Magnesium Slides onthe Vitros 950Chemistry SystemJohnson andJohnson ClinicalDiagnosticsK8613868/26/86

4. Description of the Device:

The Piccolo Renal Function Panel Plus Reagent Disc (which contains the Piccolo Magnesium Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and

1

{1}------------------------------------------------

Summary of Safety and Effectiveness (continued)

diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.

Statement of Intended Use: 5.

The Piccolo Magnesium Test System (presently contained on the Renal Function Panel Plus Reagent Disc) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of magnesium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

Summary of the technological characteristics of the new device in comparison 6. to those of the predicate device:

Table 1 outlines the technological characteristics of the Piccolo Magnesium Test System in comparison to the legally marketed predicate device.

Piccolo Point-of-CareChemistry AnalyzerVitros 950 ChemistrySystem
Intended UseQuantitative analysis ofMagnesiumQuantitative analysis ofMagnesium
MethodologyEnzymatic rate reactionColorimetric
Sample TypeHeparinized whole blood,heparinized plasma, andserumHeparinized plasma, serum, andurine
Sensitivity0.1 mg/dL0.2 mg/dL
ReagentsDry test-specific reagentbeadsDry, multilayered, analytical elementcoated on a polyester support (slide).Active ingredients: 1,2-bis(o-aminophenoxy)ethane-N,N,N', N'-tetraacetic acid (calcium chelator)and 1,5-bis(2-hydroxy-3,5-dichlorophenyl)-3-cyano-formazan(dye)
Temperature of Reaction37°C37°C
CalibrationBar code with factorycalibrated lot specific dataCalibrated periodically usingcalibrators supplied by vendor
Assay Range0.1 - 8.0 mg/dL0.2 - 10.0 mg/dL
Testing EnvironmentProfessional useProfessional use
Sample SizeApproximately 100 µL10 µL

Table 1: Specification Comparison: Piccolo Magnesium Test System

{2}------------------------------------------------

Summary of Safety and Effectiveness (continued)

Brief discussion of the clinical and nonclinical tests relied on for a 7. determination of substantial equivalence.

Tables 2 summarize the results of clinical and non-clinical tests performed using the Piccolo Magnesium Test System.

Linearity:

Table 2: Summary of Linearity

Magnesium
Slope0.992
Intercept-0.05
Corr. Coefficient0.999

Precision:

Precision studies were designed to evaluate within-run and total precision of the Magnesium Test System when run on the Piccolo Point-of-Care Chemistry Analyzer.

Table 3: Within-Run and Total Precision for Magnesium, Assayed on the Piccolo Point-of-Care Chemistry Analyzer

AnalyteWithin-Run (n = 80)Total (n = 80)
Magnesium (mg/dL)
Level 1
Mean1.91.9
SD0.030.06
CV1.73.4
Level 2
Mean3.93.9
SD0.040.10
CV1.02.6

{3}------------------------------------------------

Summary of Safety and Effectiveness (continued)

Sample Type Comparison:

A study was conducted to examine and compare heparinized venous whole blood and serum on the Piccolo® Point-of-Care Chemistry Analyzer.

Serum, heparinized plasma and heparinized whole blood comparability were established for magnesium.

8. Conclusions

The clinical and non-clinical tests performed for Magnesium, when run on the Piccolo Point-of-Care Chemistry Analyzer demonstrate that the test system is as safe, effective and performs as well as the legally marketed device identified above.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the graphic.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 5 0 2004

Dennis M. Bleile, Ph.D. Director of Assay Development Abaxis, Inc. 3240 Whipple Road Union City. CA 94587

K040115 Re:

Trade/Device Name: Piccolo® Magnesium Test System Regulation Number: 21 CFR 862.1495 Regulation Name: Magnesium test system Regulatory Class: Class I Product Code: JGJ Dated: January 15, 2004 Received: January 20, 2004

Dear Dr. Bleile:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r least be as made a determination that your device complies with other requirements of the Act that 1197 for a lotations and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Jean M. Cooper, MS, DVM.

Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT 12.0

Indications for Use

510(k) Number (if known): KnHz 0 )// 5

Piccolo® Magnesium Test System Device Name:

Intended Use:

The Piccolo Magnesium Test System (presently contained on the Renal Function Panel Plus Reagent Disc) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of magnesium in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

Indications for Use:

Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)OROver- The Counter Use (Optional Format 1-2-96)
-------------------------------------------------------------------------------------------
Division Sign-OffCarol Benson for Jean Cooper, DVM
--------------------------------------------------------------------

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)K040115
-----------------

§ 862.1495 Magnesium test system.

(a)
Identification. A magnesium test system is a device intended to measure magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).(b)
Classification. Class I.