K Number
K040097

Validate with FDA (Live)

Date Cleared
2004-02-13

(24 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Medline Strider MIDI 3 scooter is to provide mobility to persons that have difficulty walking for prolonged periods of time. They are intended to increase the mobility of persons who are both physically and mentally capable of assessing and reacting to any driving situation.

The Medline Strider MIDI 3 scooter is constructed for use both indoors and outdoors. They are for use indoors, on pavement, sidewalks, pedestrian zones and areas free from motor traffic. They must not be driven on public roads with the exception of crossing or when no pavements are available.

Device Description

Not Found

AI/ML Overview

I am sorry, but the provided text is a 510(k) clearance letter for a medical device (Medline Strider MIDI 3), not a study report. It does not contain any information regarding acceptance criteria, device performance metrics, sample sizes, expert qualifications, or ground truth establishment.

Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the given input.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 3 2004

Ms. Andrea Haferkamp Corporate Director of Regulatory Affairs Medline Industries, Inc. One Medline Place Mundelein, Illinois 60060-4486

Re: K040097

Trade/Device Name: Medline Strider MIDI 3 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: II Product Code: INI Dated: January 16, 2004 Received: January 20, 2004

Dear Ms. Haferkamp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Page 2 - Ms. Andrea Haferkamp

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours

Mark McMillan

Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): K 04:0097

Medline Strider MIDI 3 Device Name:

Indications For Use:

The intended use of the Medline Strider MIDI 3 scooter is to provide mobility to persons that have difficulty walking for prolonged periods of time. They are intended to increase the mobility of persons who are both physically and mentally capable of assessing and reacting to any driving situation.

The Medline Strider MIDI 3 scooter is constructed for use both indoors and outdoors. They are for use indoors, on pavement, sidewalks, pedestrian zones and areas free from motor traffic. They must not be driven on public roads with the exception of crossing or when no pavements are available.

Prescription Use (Parl 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Office of Device Evaluation (ODE)

Mark A. Mulkern

Division of General, Restorative, and Neurological Devices

Page 1 of __

510(k) Number K040097

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).