AMEDITECH IMMU TEST DRUG SCREEN PANEL
K040092 · Ameditech, Inc. · DKZ · Mar 8, 2004 · Clinical Toxicology
Device Facts
| Record ID | K040092 |
| Device Name | AMEDITECH IMMU TEST DRUG SCREEN PANEL |
| Applicant | Ameditech, Inc. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | Mar 8, 2004 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Intended Use
The Ameditech ImmuTest Drug Screen Panel is a one-step panel immunoassay for the qualitative detection of amphetamine, cocaine metabolite (benzovlecgonine), methamphetamine, opiates, phencyclidine, and THC in human urine. The cutoff concentrations for this panel test are amphetamine at 1000 ng/ml, cocaine metabolite at 300 ng/ml, methamphetamine at 500 ng/ml, opiates at 2000 ng/ml, phencyclidine at 25 ng/ml, and THC at 50 ng/ml. This test kit is used to obtain a visual, qualitative results and is intended for professional use.
Device Story
Device modification involving placement of six previously cleared, unmodified devices into a single plastic test strip holder. Modification does not alter fundamental scientific technology or intended use. Device intended for clinical use; operated by healthcare professionals. Output remains consistent with original cleared device performance.
Clinical Evidence
No clinical data provided; device relies on analytical performance characteristics typical of lateral flow immunoassay drug screening tests.
Technological Characteristics
Modification consists of a plastic test strip holder containing six unmodified, previously cleared devices. Fundamental scientific technology remains unchanged. No new materials or energy sources introduced.
Indications for Use
Indicated for professional use for qualitative detection of amphetamine, cocaine metabolite, methamphetamine, opiates, phencyclidine, and THC in human urine at specified cutoff concentrations.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
- K233062 — BioSieveTM Multi-Drug Urine Test Panel; BioSieveTM Multi-Drug Urine Test Panel Rx · Vivachek Biotech (Hangzhou) Co., Ltd. · Nov 2, 2023
- K031759 — ACON SPECTRUM MULTI-DRUG MULTI-LINE DRUG SCREEN TEST CARD AND ACON SPECTRUM MULTI-DRUG MULTI-LINE DRUG SCREEN TEST CARD · ACON Laboratories, Inc. · Aug 4, 2003
- K091612 — UCP RAPID DRUG SCREENING BUPRENORPHINE, AMP 300, MAMP 500, COC 150 TEST; · Ucp Biosciences, Inc. · Feb 24, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the snakes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR - 8 2004
John Wu. Ph.D. Director of Quality Assurance Ameditech, Inc. 10340 Camino Santa Fe - Suite F San Diego, CA 92121
Re: k040092
> Trade/Device Name: Ameditech ImmuTest Drug Screen Panel Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ; DIO; DJC; DJG; LCM; DKE Dated: January 5, 2004 Received: January 16, 2004
Dear Dr. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Fedcral Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
## Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, MS, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Attachment C
## Indications for Use Statement
Page of
K040092 510(k) Number (if known):
Ameditech ImmuTest Drug Screen Panel Device Name:
Indications For Use:
The Ameditech ImmuTest Drug Screen Panel is a one-step panel immunoassay for the qualitative detection of amphetamine, cocaine metabolite (benzovlecgonine), methamphetamine, opiates, phencyclidine, and THC in human urine. The cutoff concentrations for this panel test are amphetamine at 1000 ng/ml, cocaine metabolite at 300 ng/ml, methamphetamine at 500 ng/ml, opiates at 2000 ng/ml, phencyclidine at 25 ng/ml, and THC at 50 ng/ml.
This test kit is used to obtain a visual, qualitative results and is intended for professional use.
Carol Benson
Division Sign-Off
Office of In Vitro Diagnos Device Evaluation and Sa
510(k) k040092
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)