K Number
K033989

Validate with FDA (Live)

Date Cleared
2004-01-22

(30 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The deflectable catheter system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

Device Description

The Medtronic SelectSite Model C304 Deflectable Catheter System is designed to facilitate the introduction of transvenous devices within the chambers and coronary vasculature of the heart and the introduction of leads in the left heart via the coronary sinus.

The Model C304-S59 deflectable catheter has a single lumen for passage of transvenous devices with a maximum outer diameter of 5.6 French (1.9 mm) and a usable length of 30 cm.

The Model C304-L69 deflectable catheter has a single lumen for passage of transvenous devices with a maximum outer diameter of 5.6 French (1.9 mm) and a usable length of 40 cm.

The deflectable catheter features a deflecting distal section controlled by the deflectable catheter handle for placement of transvenous devices within the chambers of the heart and for coronary sinus cannulation. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.

AI/ML Overview

The provided text describes a 510(k) summary for the Medtronic SelectSite™ Model C304 Deflectable Catheter System. This document focuses on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria or reporting on a clinical effectiveness study with specific performance metrics for the device itself.

Therefore, much of the requested information regarding acceptance criteria, performance metrics, sample sizes for test and training sets, expert qualifications, adjudication methods, and ground truth types is not available in the provided text, as these are typically part of a clinical trial or performance study, which this summary does not detail beyond "In Vitro testing".

Here's what can be extracted based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implied)Reported Device Performance
All specified design and performance requirements (for in vitro tests)"The SelectSite Model C304 Deflectable Catheter System met all specified design and performance requirements."
Biocompatibility consistent with ISO-10993"All materials were found to be biocompatible and in compliance to ISO 10993-1."
Sterilization validation (100% Ethylene Oxide)"The SelectSite Model C304 Deflectable Catheter System will be sterilized using a validated 100% Ethylene Oxide (EtO) sterilization process."
Substantial equivalence to predicate device(s)"Medtronic, Inc. considers the SelectSite Model C304-S59 and C304-L69 Deflectable Catheter Systems to be substantially equivalent to legally marketed predicate device(s)."

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The study mentioned is "In Vitro testing".
  • Data Provenance: Not specified, but "In Vitro testing" implies laboratory-based testing, not human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not applicable, as this was an "In Vitro testing" study.
  • Qualifications of Experts: Not applicable.

4. Adjudication method for the test set:

  • Adjudication Method: Not applicable, as this was an "In Vitro testing" study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No. This is a medical device (catheter) and the study described is "In Vitro testing", not an AI diagnostic tool involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance Study: Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used:

  • Type of Ground Truth: For the "In Vitro testing", the ground truth would be based on engineering specifications, physical measurements, and adherence to established standards (e.g., ISO-10993 for biocompatibility) to verify the device's design and performance.

8. The sample size for the training set:

  • Sample Size: Not applicable. This document describes "In Vitro testing" for a physical device, not an AI model that requires a training set.

9. How the ground truth for the training set was established:

  • Ground Truth Establishment: Not applicable.

{0}------------------------------------------------

K033989 p1/2

JAN 2 2 2004

Special 510(k) 510(k) Summary of Substantial Equivalence

SelectSite™ Model C304 KXXXXXXX

510(K) SUMMARY OF SUBSTANTIAL EQUIVALENCE

Date Prepared:22 December 2003
Submitter:Medtronic Inc.7000 Central Avenue NEMinneapolis, MN 55432
Contact:Kristy K. Mollner, MS, RACSr. Regulatory Affairs Specialist
Telephone:(763) 514-4189
Fax:(763) 514-6424
Email:kristina.mollner@medtronic.com
Proprietary NameMedtronic® SelectSite™ Model C304-S59 DeflectableCatheter System
Medtronic® SelectSite™ Model C304-L69 DeflectableCatheter System
Common Name:Catheter, Percutaneous
Device Classification:Class II, 21 CFR § 870.1250
Product Code:74 DQY

Device Description

The Medtronic SelectSite Model C304 Deflectable Catheter System is designed to facilitate the introduction of transvenous devices within the chambers and coronary vasculature of the heart and the introduction of leads in the left heart via the coronary sinus.

The Model C304-S59 deflectable catheter has a single lumen for passage of transvenous devices with a maximum outer diameter of 5.6 French (1.9 mm) and a usable length of 30 cm.

The Model C304-L69 deflectable catheter has a single lumen for passage of transvenous devices with a maximum outer diameter of 5.6 French (1.9 mm) and a usable length of 40 cm.

Image /page/0/Picture/10 description: The image shows the Medtronic logo with the word "Confidential" to the right of it. The Medtronic logo consists of a stylized image of a person inside of a circle to the left of the word "Medtronic". The word "Medtronic" is in a bold, sans-serif font. The word "Confidential" is in a smaller, sans-serif font.

{1}------------------------------------------------

Special 510(k) 510(k) Summary of Substantial Equivalence SelectSite™ Model C304 KXXXXXXX

The deflectable catheter features a deflecting distal section controlled by the deflectable catheter handle for placement of transvenous devices within the chambers of the heart and for coronary sinus cannulation. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.

Indications for Use

The deflectable catheter system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

Substantially Equivalent Devices

Predicate DeviceManufacturer510(k) Number
Medtronic 10600 Deflectable Catheter Delivery SystemMedtronicK013517
Medtronic Attain Model 6226DEF Deflectable Catheter SystemMedtronicK032312
Introducer Valve Model 6228VALCook Vascular IncorporatedK010128
Universal Slitter Model 6228SLTMedtronicClass 1 device exempt from pre-market notification

Summary of Studies

In Vitro testing was performed to support substantial equivalence to the predicate device(s). The SclectSite Model C304 Deflectable Catheter System met all specified design and performance requirements.

Biocompatibility Information

All device matcrials / components were assessed for biocompatibility consistent with ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". All materials were found to be biocompatible and in compliance to ISO 10993-1.

Sterilization Validation

The SelectSite Model C304 Deflectable Catheter System will be sterilized using a validated 100% Ethylene Oxide (EtO) sterilization process.

Conclusion

Through the data and information presented, as well as similarities to legally marketed devices, Medtronic, Inc. considers the SelectSite Model C304-S59 and C304-L69 Deflectable Catheter Systems to be substantially equivalent to legally marketed predicate device(s).

Image /page/1/Picture/16 description: The image shows the Medtronic logo with the word "Confidential" to the right of it. The Medtronic logo consists of a stylized caduceus symbol to the left of the word "Medtronic." The word "Medtronic" is written in a bold, sans-serif font. The word "Confidential" is written in a smaller, lighter font.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three tail feathers, representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 2 2004

Medtronic, Inc c/o Ms. Kristy K. Mollner, M.S., RAC Senior Regulatory Affairs Specialist 7000 Central Avenue NE Mail Stop CW304 Minneapolis, MN 55432-3576

Re: K033989

Trade Name: Medtronic® SelectSite™ Model C304-S59 Deflectable Catheter System and Medtronic® SelectSite™ Model C304-L69 Deflectable Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: December 22, 2003 Received: December 23, 2003

Dear Ms. Mollner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Kristy K. Mollner, M.S., RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh.dsmaldsmamain.html.

Sincerely vours.

Q. Melli

$\mathbb{R}$

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

SelectSite™ Model C304 KXXXXXXX

INDICATIONS FOR USE

Page I of I

510(k) Number (if known): K 0 33 19 9

Device Name:

Medtronic® SelectSite™ Model C304-S59 Deflectable Catheter System

Medtronic® SelectSite™ Model C304-L69 Deflectable Catheter System

Indications for Use: The deflectable catheter system is indicated to provide a pathway through which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801.109)

OR

Over-the Counter Use

el.m.a.

(Division Sian-Off) Division of Cardiovascular Devices (Optional Format 1-2-96)

510(k) Number K033987

Image /page/4/Picture/18 description: The image shows the Medtronic logo with the word "Confidential" next to it. The Medtronic logo consists of a stylized image of a person inside of a circle. The word "Medtronic" is written in a bold, sans-serif font. The word "Confidential" is written in a smaller, sans-serif font.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).