K Number
K033927

Validate with FDA (Live)

Date Cleared
2004-01-15

(28 days)

Product Code
Regulation Number
888.1100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthroscopy Pump A115 is intended to provide fluid distention and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities during diagnostic and operative arthroscopic procedures.

Device Description

The Arthroscopy Pump A115 is a microprocessor controlled single roller pump designed to provide fluid distention and irrigation of joint cavities during diagnostic and operative arthroscopy. The pump functions according to the peristaltic principle and consists of the following main components: a housing, a power supply, a roller wheel, a pump head, various setting keys and display elements. The pump is used with special designed irrigation tubings, a pressure remove control and a foot pedal. A constant performed redundant pressure control assures conformity of the actual pressure in the joint cavity with the pre-set nominal pressure.

AI/ML Overview

This medical device is an "Arthroscopy Pump" rather than an AI/ML device, so the requested information (acceptance criteria, study details, human reader improvement with AI, standalone performance, ground truth, training set size, etc.) is not applicable in the context of the provided documentation.

The 510(k) summary for the Arthroscopy Pump A115 explicitly states:

VIII. Performance/Clinical Data:
"No performace or clinical data is provided."

Therefore, the document does not contain information on acceptance criteria, a study proving the device meets those criteria, or any of the detailed AI/ML specific questions. The submission primarily relies on substantial equivalence to predicate devices (other arthroscopy pumps) and compliance with voluntary standards for electrical safety and electromagnetic compatibility.

{0}------------------------------------------------

Ko 33927 1/2

'JAN 1 5 2004

510(K) SUMMARY Arthroscopic Pump A115

Submitter's Name, Address, Telephone Number, Contact Person, I. Sublishment Registration Number and Date Summary was Prepared:

W.O.M. WORLD OF MEDICINE AG Alte Poststraße 11 96337 Ludwigsstadt Germany

Phone number:001 49 30 399 81 610
FAX number:001 49 30 399 81 592

Establishment registration number: 8043980

Contact Person:Susanne Raab
368 North St. Asaph Street
Alexandria, VA 22314
Phone number :703 299 0523
Fax number:703 299 0523
Date Prepared :Dezember 9, 2003

II. Device Names:

  • Arthroscope and Accessories Classification Name: 1.
  • Arthroscopic Pump, Tubing Sets and Common or Usual Name: 2.
  • Accessory Arthroscopic Pump A115 Proprietary Name: 3.

Product Classification: III.

Product Code:HRX
C.F.R. Section:21 C.F.R. § 888.1100
Device Class:II

Predicate Devices: IV.

  • Arthroscopy Pump A107 (K030402) manufactured by W.O.M. WORLD . OF MEDICINE AG
  • Arthro-Surgimat- 2000 ECU (K990443) manufactured by W.O.M. WORLD . OF MEDICINE AG
  • Arthro-Surgimat-1500 (K983910) manufactured by W.O.M. WORLD OF . MEDICINE AG

{1}------------------------------------------------

510(K) SUMMARY Page -2- / -2-

Intended Use: V.

The Arthroscopy Pump A115 is intended to provide fluid distention and irrigation The Arthroscopy I amp Arre ankle and wrist joint cavities during diagnostic and operative arthroscopic procedures.

Device Description: VI.

The Arthroscopy Pump A115 is a microprocessor controlled single roller pump I no Arch obbopy of the micronand irrigation of joint cavities during designed to provide throscopy. The pump functions according to the thaghostic and operative anaists of the following main components: a housing, a portsuitio principler wheel, a pump head, various setting keys and display power supply, a roner whose used with special designed irrigation tubings, a clemonts. The and foot pedal. A constant performed redundant pressure remove control and 1000 peonformity of the actual pressure in the joint cavity with the pre-set nominal pressure.

Substantial Equivalence: VII.

The Arthroscopy Pump A115 described in this notification is similar in design and technological characteristics to the Arthroscopy Pump A107 (K030402), to the Arthro-Surgimat-2000 ECU (K990443) and Arthro-Surgimat-1500 (K983910) manufactured by W.O.M. WORLD OF MEDICINE AG. In addition, both the manufactured by Pump A115 and the predicate devices are intended to provide fluid distention and irrigation of knec, shoulder, hip, elbow, ankle and wrist joint cavities during diagnostic and operative arthroscopic procedures.

The differences between the Arthroscopy Pump A 1 15 and the predicate devices are minor and raise no new questions of safety and effectiveness. Accordingly, W.O.M. WORLD OF MEDICINE AG believes that the Arthroscopy Pump A 115 is substantially equivalent to the predicate devices currently on the market.

VIII. Performance/Clinical Data:

No performace or clinical data is provided.

IX. Voluntary Standards:

The device complies with the International Standard IEC 60601-1 (Flectrical Safety) and IEC 60601-1-2 (Electromagnetic Compatibility) and bears the CE mark in accordance with the Medical Device Directive 93/42/FEC.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services (HHS) in the USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 5 2004

W.O.M. World of Medicine AG c/o Ms. Susanne Raab 368 North Saint Asaph Street Alexandria, Virginia 22314

Re: K033927

Trade/Device Name: Arthroscopy Pump A115 Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope and accessories Regulatory Class: II Product Code: HRX Dated: December 16, 2003 Received: December 18, 2003

Dear Ms. Raab:

We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or (o devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Susanne Raab

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

fed Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATION FOR USE STATEMENT

510(k) Number (if known): KO53927

Device Name: Arthroscopy Pump A115

Indications for Use:

The Arthroscopy Pump A115 is intended to provide fluid distention and irrigation of knee, shoulder, hip, elbow, ankle and wrist joint cavities during diagnostic and operative arthroscopic procedures.

(PLFASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Miriam C. Provost

Sivision Sign-Oil) Division of General, Restorative and Neurological Devices

KO 33927 l (h) Nameru . .

Prescription Use
Use
(Per 21 C.F.R. 801.109)

OR

Over-The-Counter

(Optional Format 3-10-98)

§ 888.1100 Arthroscope.

(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.