K Number
K033894
Date Cleared
2004-03-11

(86 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Panasonic Wrist Blood Pressure Monitor, Models EW 3031 and EW 3032 are devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist.

For the EW 3032, display of Normal (green), Prehypertensive (yellow) or Hypertensive (orange) are based on blood pressure values classified in the paper: "JNC Express, The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure" , U.S. Department of Health and Human Services, National Institutes of Health, National Heart, Lung, and Blood Pressure Institute, National High Blood Pressure Education Program, NIH Publication No 03-5233, May 2003. The display values are generally known, but not proven, to be an indicator of your blood pressure.

The EW 3032 is not intended to be a diagnostic device. Contact your physician if prehypertensive or hypertensive values are indicated.

Device Description

The Panasonic Wrist Blood Pressure Monitor, Models 3031/3032 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Panasonic Wrist Blood Pressure Monitor, Models EW 3031/EW 3032, based on the provided text:

Acceptance Criteria and Device Performance

The provided 510(k) summary primarily focuses on the substantial equivalence of the new device to a predicate device and adherence to industry standards, rather than specific numerical acceptance criteria for blood pressure measurement accuracy.

Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit in text)Reported Device Performance
Blood Pressure AccuracyMet all relevant requirements of ANSI/AAMI SP10-1992Device met all relevant requirements of ANSI/AAMI SP10-1992 for electronic or automated sphygmomanometers.
Pulse Rate MeasurementAcceptable performance for pulse rate measurementEvaluation of Pulse Rate Measurement was performed and deemed satisfactory.
Electrical SafetyMet all relevant requirements of IEC 60601-1Device met all relevant requirements of IEC 60601-1.
Mechanical PerformanceMet requirements for reliability (drop, storage, vibrating)Reliability tests (operation, drop, storage, vibration) were conducted and met relevant requirements.
Environmental PerformanceMet requirements for EMCEMC Test was conducted and met relevant requirements.
General Safety/EffectivenessMet requirements of FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions"Device met all relevant requirements of the FDA Reviewer Guidance.

Important Note: The document states that the devices met "all relevant requirements" of the ANSI/AAMI SP10-1992 standard. This standard provides specific limits for accuracy (e.g., mean difference and standard deviation between automated and reference measurements). However, the specific numerical results from the study (e.g., the actual mean difference and standard deviation) are not provided in this 510(k) summary. The summary only confirms compliance with the standard itself.

Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not specified in the provided document.
    • Data Provenance: Not specified. It's implied to be prospective as it's a clinical test performed on the device. No country of origin is mentioned for the clinical study participants.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • Number of Experts: Not specified. ANSI/AAMI SP10-1992 typically requires comparison against trained observers using a reference method (e.g., auscultation with a mercury manometer), but the number of observers and their specific qualifications are not detailed in this summary.
    • Qualifications of Experts: Not specified.
  3. Adjudication Method for the Test Set:

    • Not specified. (For blood pressure measurements, ground truth is typically established by trained observers taking repeated measurements, which is then compared against the device. Any discrepancies or methods for resolving them are not detailed.)
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study assesses how AI assistance changes human performance. The Panasonic device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.
  5. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, a standalone study was performed. The clinical testing performed according to ANSI/AAMI SP10-1992 assesses the accuracy of the device (algorithm + hardware) in measuring blood pressure directly against a reference standard, without human intervention in the measurement process itself. The document explicitly states, "Discussion of Clinical Tests Performed: ANSI/AAMI SP10-1992 'National Standard for Electronic or Automated Sphygmomanometers' testing was performed. All relevant sections were addressed and testing conducted. The EW 3031 and EW 3032 met all relevant requirements of this standard."
  6. Type of Ground Truth Used:

    • Expert Consensus/Reference Standard: The ground truth for the clinical study (adhering to ANSI/AAMI SP10-1992) would typically be established by simultaneous or sequential measurements by trained human observers using a validated reference method, such as auscultation with a mercury manometer. The standard provides guidelines for this process.
  7. Sample Size for the Training Set:

    • Not applicable/Not specified. This device uses the "oscillometric method," which is a well-established technique for blood pressure measurement. While internal development and calibration would involve data, the document does not mention a distinct "training set" in the context of a machine learning algorithm. The oscillometric method relies on analyzing pressure oscillations in the cuff, not on a machine learning model trained on a large dataset in the way modern AI is.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable/Not specified. As explained above, a dedicated "training set" for a machine learning algorithm is not relevant here. The oscillometric method's principles are based on known physiological responses and signal processing, not on a ground truth established for a neural network.

{0}------------------------------------------------

K033894
page 1 of 3

EXHIBIT # 1

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is:

1. Submitter's Identification:

Matsushita Electric Works, Ltd. 33 Okamachi Hikone, Shiga 522-85 Japan TEL: +81-(0) 749-27-2963 FAX: +81-(0)-749-23-0349

Date Summary Prepared: December 15, 2003

Name of the Device: 2.

Panasonic Wrist Blood Pressure Monitor, Models EW 3031/EW 3032

3. Predicate Device Information:

The Panasonic Wrist Blood Pressure Monitor, Models 3031/3032 are substantially equivalent to the MECA Blood Pressure Meter, Model EW 243/254, K991458.

4. Device Description:

The Panasonic Wrist Blood Pressure Monitor, Models 3031/3032 are designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

Intended Use: 5.

The Panasonic Oscillometric Wrist Blood Pressure Monitors, Models EW3031 and EW 3032 are devices intended to measure systolic and diastolic blood

{1}------------------------------------------------

K033894.
page 2 of 3

pressure rate of an adult individual by using a pressurized cuff on the left wrist. The devices are not intended for use on infants or children. The devices are designed for home use only and not for ambulatory measurements.

Comparison to Predicate Devices: 6.

Both devices use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. A wrist cuff is inflated automatically; deflate rate is controlled but a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Each device uses a similar capacitance-type pressure sensor to translate the pressure variations to electrical signals that can be interpreted by an integrating circuit. Once the reading is determined each unit operates a solenoid valve to release the pressure to zero. Our Wrist Blood Pressure Monitor, Model EW 3031/3032 , differs from the predicate device in the cuff application part, display function, and memory function.

Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:

Testing information demonstrating safety and effectiveness of the in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements. The following testing was conducted:

  • Reliability Test Operation Conditions a.
  • Reliability Test Drop Testing b.
  • Reliability Test Storage C.
  • Reliability Test Vibrating Testing ದೆ.
  • EMC Test e.
  • IEC 60601-1 Safety Test f.
  • Evaluation of Pulse Rate Measurement g.

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the Panasonic Wrist Blood Pressure Monitor, Models 3031/3032 tested met all relevant requirements of the aforementioned tests.

8. Discussion of Clinical Tests Performed:

ANSI/AAMI SP10-1992 "National Standard for Electronic or Automated Sphygmomanometers" testing was performed. All relevant sections were addressed and testing conducted. The EW 3031 and EW 3032 met all relevant requirements of this standard.

{2}------------------------------------------------

<033874
page 3 of 3

9. Conclusions:

We have demonstrated that the Panasonic Model EW 3031 and EW 3032 areas safe and effective as the predicate based on electrical, mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-1992 testing results.

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 1 2004

Matsushita Electric Works, Ltd. c/o Ms. Carolann Kotula Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021

Re: K033894

Trade Name: Panasonic Wrist Blood Pressure Monitor, Models EW 3031 and EW 3032 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: February 12, 2004 Received: February 13, 2004

Dear Ms. Kotula:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Carolann Kotula

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsma/dsmamain.html

Sincerely yours,

Drian R. Kochner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K03389415

Exhibit B

Page 1

510(k) Number (if known): ¥033894

Device Name: Panasonic Wrist Blood Pressure Monitor, Models EW 3031 and EW 3032

Indications For Use: The Panasonic Wrist Blood Pressure Monitor, Models EW 3031 and EW 3032 are devices intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist.

For the EW 3032, display of Normal (green), Prehypertensive (yellow) or Hypertensive (orange) are based on blood pressure values classified in the paper: "JNC Express, The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure" , U.S. Department of Health and Human Services, National Institutes of Health, National Heart, Lung, and Blood Pressure Institute, National High Blood Pressure Education Program, NIH Publication No 03-5233, May 2003. The display values are generally known, but not proven, to be an indicator of your blood pressure.

The EW 3032 is not intended to be a diagnostic device. Contact your physician if prehypertensive or hypertensive values are indicated.

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use X (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Donna R. Lochner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number K033894

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).