K Number
K033881

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-03-02

(78 days)

Product Code
Regulation Number
870.1650
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medrad Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The ISI module option is intended for the specific purpose of allowing an injector to interface with a CT scanner.

Device Description

The Medrad Stellant CT Injector System with ISI Module is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography procedures. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for computed the organy studies. The addition of the ISI Module option to the Stellant CT Injector System simply establishes an interface between the injector and the CT scanner. The Stellant CT Injector System with ISI Module is comprised of the same main components as the predicate device: an Injector Head, a Display Control Unit (DCU) and sterile disposables. The Imaging System Interface Module is added to the system as an accessory.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Medrad Stellant CT Injector System with Imaging System Interface Module), not an AI/ML device. Therefore, the provided text does not contain information about acceptance criteria, study details, sample sizes, expert involvement, or ground truth relevant to AI/ML performance.

The document focuses on demonstrating substantial equivalence to a predicate device (Medrad Stellant CT Injector System K023183) by comparing features and intended use, rather than proving performance against specific acceptance criteria through a study.

Therefore, I cannot provide the requested information.

Here's why each point cannot be answered:

  1. A table of acceptance criteria and the reported device performance: This document is about regulatory clearance (510(k) submission) based on substantial equivalence, not a performance study against specific acceptance criteria for a new AI/ML algorithm. The "reported device performance" in this context is the device's operational parameters and intended use, which are deemed equivalent to the predicate device.
  2. Sample sized used for the test set and the data provenance: Not applicable. There is no "test set" in the context of an AI/ML performance study described here.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no ground truth, test set, or expert review process described for an AI/ML algorithm.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. There is no test set or adjudication process for an AI/ML algorithm.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an injector system, not an AI/ML diagnostic tool that assists human readers.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI/ML algorithm.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. There is no ground truth for an AI/ML algorithm described.
  8. The sample size for the training set: Not applicable. There is no AI/ML training set.
  9. How the ground truth for the training set was established: Not applicable. There is no AI/ML training set.

{0}------------------------------------------------

MAR - 2 2004

510(k) Summary

OFFICIAL CONTACT:Troy A. JackRegulatory Affairs AnalystMedrad, Inc.One Medrad DriveIndianola, PA 15051(412) 767-2400 ext. 3305
CLASSIFICATION NAME:Injector with Syringe, Angiographic
COMMON NAME(S):Powered Injector with Syringe
PROPRIETARY NAME:Medrad Stellant CT Injector System with ImagingSystem Interface Module
PREDICATE DEVICE:Medrad Stellant CT Injector System (K023183)
INTENDED USE:The Medrad Stellant CT Injector System is intendedfor the specific purpose of injecting intravenouscontrast media into humans for diagnostic studies incomputed tomography (CT) applications. The ISImodule option is intended for the specific purpose ofallowing an injector to interface with a CT scanner.

DEVICE DESCRIPTION AND COMPARISON TO UNMODIFIED PREDICATE:

The Medrad Stellant CT Injector System with the ISI feature maintains the same intended use, similar operational parameters, similar labeling and is used in a manner intoniate ass, and in the Stellant CT Injector System with ISI Module is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography procedures. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for computed the organy studies. The addition of the ISI Module option to the Stellant CT Injector System simply establishes an interface between the injector and the CT scanner.

The Stellant CT Injector System with ISI Module is comprised of the same main components as the predicate device: an Injector Head, a Display Control Unit (DCU) and sterile disposables. The Imaging System Interface Module is added to the system as an Differences between the predicate device and the Stellant CT Injector accessory. System with ISI functionality are detailed in the following Comparison Matrix:

{1}------------------------------------------------

TABLE 1 - COMPARISON OF STELLANT INJECTOR SYSTEM with ISI MODULE AND STELLANT CT INJECTOR SYSTEM

FeaturePredicate Device:Stellant CT Injector System (K023183)Proposed Device:Stellant CT Injector System with ISIModule
Intended UseIntended to be used for the specific purpose ofinjecting intravenous contrast media intohumans for diagnostic studies in computedtomography (CT) applications. The ISI moduleoption is intended for the specific purpose ofallowing an injector to interface with a CTscanner.Same
Single or Dual Syringe SystemSingle and dual syringe modelsSame
Information DisplayColor LCDSame
Programming KeysNon-dedicated keys -- software determinedSame
Touch screenYesYes
Multi-Phase1 - 6 phases per injectionSame
Arming ModesSingleSame
Protocol Storage Capability32 protocolsSame
Hold Capability20 minutes max.Same
Scan Delay1 - 300 secondsSame
Safety Stop MechanismMulti-layered software stops with backupmonitoringSame
Syringe SystemSingle syringe model: 200 ml syringeDual syringe model: two 200 ml syringesSame
Programmed Volume1 to 200 mlSame
Volume Remaining ReadoutLED on injector head; graphical and numericon LCDSame
Fill RateVariable up to 10 ml/secSame
Flow Rate0.1ml/s to 10.0 ml/sSame
Programmable Pressure Limit325 PSI default, user settable 50 to 300 PSISame
PauseProgrammable - 1 sec to 900 sec in 1 secincrementsSame
AutofillFill rate 4 ml/secSame
Retract ControlYes (Automatic)Yes
Remote Start SwitchYesYes
Pressure GraphYesYes
Syringe SensingYesYes
AutoloadYesYes
Auto Dock/Retract/AdvanceYes; user-selectable autodock and advance;user-selectable auto-retractYes. Same functionality.
Protocol Lock / Remote ArmingYesYes
Remote Check for Air (fromHead)YesYes
Scan Delay1sec to 300 sec in 1sec incrementsSame
Store/Recall32 protocolsSame
Test InjectYesYes
Syringe Heat MaintainerYesYes
PackagingTyvek lid covering polystyrene traySame
SterilizationEtO sterilizationSame
Method of determining Pyrogen- FreenessLimulus Amebocyte Lysate test to USP requirementsSame
Luer FittingsISO 594-1 & ISO 594-2 compliant designSame
Syringe:
Barrel Material CompositionVoridian (Eastman) 7352 PETSame
Barrel Length7.504"Same
Barrel OD2.002"Same
Barrel ID1.844"Same
Plunger Material CompositionGE Lexan 141 PolycarbonateSame
Plunger CoverHelvoet Pharma Polyisoprene R2522-58Same
Plunger Silicone CoatingDow Corning 360Same
Barrel FlangeEasy-engage design for non-rotational orientation (no alignment necessary)Same
Syringe sensing featureGrooves at bottom of barrel to be optically identifiedSame
Connector Tubing Components:
Maximum Pressure400Same
Tube MaterialUnichem PVCSame
Tube Length60"Same
Bonding AgentCyclohexanoneSame
T-connectorMedical grade polycarbonateSame
Burron DP 1000 SpikeABSSame
Priming TubeLow density polyethyleneSame

{2}------------------------------------------------

TABLE 2 COMPARISON OF STERILE DISPOSABLES

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread. The eagle is facing to the right and has three stripes on its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 2 2004

Medrad, Inc. c/o Mr. Troy A. Jack Regulatory Affairs Analyst One Medrad Drive Indianola, Pa 15051

Re: K033881

K033881
Medrad Stellant CT Injector System with Imaging System Interface Module Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic injector and syringe Regulatory Class: Class II (two) Product Code: DXT Dated: December 12, 2003 Received: December 15, 2003

Dear Mr. Jack:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your because is reley pe device is substantially equivalent (for the indications felerenced above and nave determined ly marketed predicate devices marketed in interstate for use stated in the encrosure) to regars) and enactment date of the Medical Device Amendments, or to conninered pror to Milly 20, 1977) in accordance with the provisions of the Federal Food, Drug, de vices mat have been results in asse approval of a premarket approval application (PMA). and Costictle Act () tee) and to novice, subject to the general controls provisions of the Act. The I va may, merciolo, mainer of the Act include requirements for annual registration, listing of general controls providions practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Mr. Troy A. Jack

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

R R R

Bram D Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K033881

Device Name: Medrad Stellant CT Injector System with Imaging System Interface Module

Indications For Use:

The Medrad Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The ISI module option is intended for the specific purpose of allowing an injector to interface with a CT scanner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-off
Division of Cardiovascular Devices
510(k) Number:K033881

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.