Intended Use

LabOne Micro-Plate Cotinine EIA is a competitive micro-plate immunoassay for the qualitative and semi-quantitative determination of cotinine in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device. LabOne Micro-Plate Cotinine EIA is used as an aid in the detection of cotinine after use of tobacco products or other products containing nicotine. The test is for in Vitro use only.

The LabOne Microplate Cotinine Calibrators are a device intended for medical purposes for use with the LabOne Microplate Cotinine assay to establish points of reference that are used in determination of values in the measurement of cotinine in saliva.

The LabOne Microplate Cotinine Controls are intended for use as an assayed quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for cotinine.

Device Description

LabOne Micro-Plate Cotinine EIA is a solid phase competitive enzyme immunoassay for the qualitative and semi-quantitative analysis of cotinine in oral fluid specimens collected with the OraSure® Oral Specimen Collection Device.

AI/ML Overview

The provided text describes a 510(k) summary for the LabOne Micro-Plate Cotinine EIA, which is a device for detecting cotinine in oral fluid. However, the document does NOT contain detailed information about specific acceptance criteria and the comprehensive study results to "prove" the device meets them in the way described by the requested table and points.

Specifically, the document states:
"Performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity showed that the LabOne Micro-Plate Cotinine EIA is substantially equivalent to the OraSure Micro-Plate Cotinine EIA."
And:
"Results tested from self-claimed patient specimens and diluted samples with both the LabOne Micro-Plate Cotinine EIA and the OraSure Micro-Plate Cotinine EIA also showed that the sample results from this two test systems are substantially equivalent when using self-claimed specimen results and urine results as references."

This indicates that studies were performed, but the actual acceptance criteria (e.g., minimum sensitivity, specificity, or precision values) and the specific reported device performance values for these criteria are not provided in the summary. The core of the submission is to demonstrate "substantial equivalence" to a predicate device (OraSure Micro-Plate Cotinine EIA, K974234), rather than proving it meets standalone, predefined performance metrics with detailed study data.

Therefore, many of your requested points cannot be directly extracted from the provided text.

Here's an attempt to answer what can be inferred from the document, along with explanations for what cannot:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred/Generic)Reported Device Performance (Inferred from "Substantially Equivalent")
Precision requirementsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
Analytical Sensitivity requirementsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
Interference limitsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
Antibody Cross-reactivity limitsDeemed "substantially equivalent" to OraSure Micro-Plate Cotinine EIA
Agreement with predicate device (OraSure Micro-Plate Cotinine EIA) for self-claimed patient specimens and diluted samples"Substantially equivalent" when using self-claimed specimen results and urine results as references.

Explanation: The document focuses on declaring substantial equivalence to the predicate device, K974234. It states that "performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity showed that the LabOne Micro-Plate Cotinine EIA is substantially equivalent" and that "sample results from this two test systems are substantially equivalent when using self-claimed specimen results and urine results as references." However, the specific quantitative acceptance criteria (e.g., "precision CV < X%") and the actual performance metrics achieved (e.g., "precision CV was Y%") are not reported in this summary.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated. The document mentions "self-claimed patient specimens and diluted samples" were tested, but the number of these samples is not provided.
  • Data Provenance: Not explicitly stated. However, given the context of a US FDA submission, it's highly likely the data would be derived from studies relevant to the US regulatory environment. The terms "self-claimed patient specimens" suggest human biological samples, but their origin (country, etc.) is not specified.
  • Retrospective or Prospective: Not explicitly stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not mentioned.
  • Qualifications of Experts: Not mentioned.
  • Explanation: For an in vitro diagnostic (IVD) device like this, ground truth would typically be established through a reference method or clinical status, rather than expert interpretation of images or other subjective data. So, the concept of "experts" in the context of radiologists, for example, is not applicable here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable/not mentioned. This type of adjudication is typically for subjective assessments (e.g., reading medical images). For an IVD like this, ground truth is usually objectively determined via a reference standard or clinical outcome.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This is an in vitro diagnostic device, not an imaging device requiring human reader interpretation in the same way. The device's performance is compared to a predicate device and reference methods, not human readers.
  • Effect size of human readers improvement: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Yes, in essence. The device itself (LabOne Micro-Plate Cotinine EIA) is a diagnostic test kit that measures cotinine. Its "performance" refers to how accurately it detects or quantifies cotinine, irrespective of human interpretation in the loop in the way AI applications are often discussed. The comparison data described ("Performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity") are effectively standalone performance evaluations against established analytical standards and the predicate device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: The document mentions "using self-claimed specimen results and urine results as references." This implies that the presence or absence of cotinine, or its concentration, was established using:
    • Self-claimed patient information: Patients' own statements about tobacco use.
    • Urine results: Testing of urine samples (likely using a validated reference method for cotinine) as an independent corroborator.
    • Additionally, the "performance characteristic studies" would rely on controlled samples with known cotinine concentrations or known interfering substances, which serve as a form of ground truth for analytical validity.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable/Not mentioned. For a traditional immunoassay kit like this, there isn't a "training set" in the sense of machine learning algorithms. The device's components (reagents, antibodies, etc.) are developed and optimized, but not "trained" on data in the same way.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable, as there isn't a "training set" in this context. The development of such a kit would involve significant R&D to select and optimize reagents that provide the desired biochemical specificity and sensitivity. The "ground truth" during this development would be the known concentration of cotinine (or other analytes) in characterization samples.

{0}------------------------------------------------

FEB 1 0 2004

510(k) Summary-LabOne Micro-Plate Cotinine ELA

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is K033601

Date of Summary:

August 25th, 2003

Correspondent:

Liuming Yu Name: 10101 Renner Boulevard Address: Lenexa, Kansas 66210-9752 Phone Number: 913-895-2308 Fax Number: 913-888-4186

Product Name: -

Common Name: Micro-Plate Cotinine EIA Trade Name: LabOne Micro-Plate Cotinine EIA Classification Number: 862.3220

Substantially Equivalent Device:

Product: OraSure Micro-Plate Cotinine EIA Manufactured by: OraSure® Technologies, Inc. 510(k) Number: K974234

Intended Use and Product Description:

LabOne Micro-Plate Cotinine EIA is a solid phase competitive enzyme immunoassay for the qualitative and semi-quantitative analysis of cotinine in oral fluid specimens collected with the OraSure® Oral Specimen Collection Device.

Comparison:

LabOne Micro-Plate Cotinine ELA is substantially equivalent to the OraSure Micro-Plate Cotinine EIA when used to qualitatively and semi-quantitatively test cotinine in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device

Comparison Performance Data:

Performance characteristic studies on precision, analytical sensitivity, interference and antibody cross-reactivity showed that the LabOne Micro-Plate Cotinine EIA is substantially equivalent to the OraSure Micro-Plate Cotinine EIA.

Results tested from self-claimed patient specimens and diluted samples with both the LabOne Micro-Plate Cotinine EIA and the OraSure Micro-Plate Cotinine EIA also showed that the

{1}------------------------------------------------

sample results from this two test systems are substantially equivalent when using self-claimed specimen results and urine results as references.

Conclusion:

Islour.
The Micro-Plate Cotinine EIA can be used to qualitatively and semi-quantitatively test cotinine in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device. .

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the symbol.

Public Health Service

FEB 1 0 2004

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

LabOne, Inc c/o Ms. Laura Danielson Responsible Third Party Official 510(k) Program Manager TUV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891

Re. K033601

Trade/Device Name: LabOne Micro-Plate Cotinine EIA Regulation Number: 21 CFR 862.3200 Regulation Name: Calibrator Regulatory Class: Class II Product Code: DLJ; MKU Dated: January 26, 2004 Received: January 28, 2004

Dear Ms. Danielson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Jean M. Cooper, MS, DVM.

Yean M. Cooper, MS. D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use Statement

510(k) Number (if known) K_

Device Name: LabOne Micro-Plate Cotinine EIA

Indication For Use:

LabOne Micro-Plate Cotinine EIA is a competitive micro-plate immunoassay for the qualitative and semi-quantitative determination of cotinine in oral fluid collected with the OraSure® Oral Fluid Specimen Collection Device. LabOne Micro-Plate Cotinine EIA is used as an aid in the detection of cotinine after use of tobacco products or other products containing nicotine. The test is for in Vitro use only.

The LabOne Microplate Cotinine Calibrators are a device intended for medical purposes for use with the LabOne Microplate Cotinine assay to establish points of reference that are used in determination of values in the measurement of cotinine in saliva.

The LabOne Microplate Cotinine Controls are intended for use as an assayed quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for cotinine.

i

்ப

Ovez - file - Conlen

Prescription

Allate C
Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033601

§ 862.3200 Clinical toxicology calibrator.

(a)
Identification. A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.