K Number
K033276

Validate with FDA (Live)

Device Name
VPAP III ST-A
Manufacturer
Date Cleared
2004-06-07

(241 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VPAP III ST-A system is Intended to provide non-invasive ventilation for patients with respiratory insufficiency or obstructive sleep apnea (OSA), in the hospital or home.

Device Description

The VPAP III ST-A provides a mode of non-invasive positive pressure ventilation (NPPV) called Pressure Support with PEEP, which delivers two treatment pressures (bi-level ventilation). A higher pressure is applied when the patient inhales - IPAP or inspiratory positive airway pressure, and a lower pressure is applied when the patient exhales - EPAP or expiratory positive airway pressure, sometimes referred to as PEEP or positive end-expiratory pressure. The difference between the two treatment pressures represents the amount of pressure support provided to the patient.

AI/ML Overview

I am sorry, but the provided text does not contain the detailed information necessary to answer your request about acceptance criteria and the study proving the device meets them.

The text is a 510(k) premarket notification letter and summary for a medical device (Resmed VPAP III-STA). While it discusses the device's technical characteristics, its intended use, and states that performance data was collected, it does not provide the acceptance criteria or the specific detailed results of the study.

Here's what I can tell you based on the provided text, and what's missing:

What is present:

  • Device Name: Resmed VPAP III-STA
  • Predicate Device: VPAP II ST-A (K974417)
  • Intended Use: "The VPAP III ST-A system is intended to provide non-invasive ventilation for patients with respiratory insufficiency or obstructive sleep apnea (OSA), in the hospital or home."
  • Performance Data Mention: "The VPAP III ST-A was tested in accordance with the recommendations of the FDA Draft Reviewer Guidance for Ventilators (July 1995)." and "The VPAP III ST-A ventilator complies with the following standards: IEC 60601-1-2; IEC 60068-2-27/ -6/ -34; IEC 10993-1; ASTM F 1100-90."
  • Conclusion: The device is "substantially equivalent" to its predicate.

What is missing (and therefore I cannot provide answers to your specific questions):

  1. Table of acceptance criteria and reported device performance: The text mentions testing was done and standards were complied with, but it does not specify what the acceptance criteria were (e.g., specific pressure accuracy ranges, flow rate tolerances) or what the measured performance results were.
  2. Sample size used for the test set and data provenance: No information on the number of subjects, cases, or the origin (country, retrospective/prospective) of any clinical test data is provided.
  3. Number of experts used to establish ground truth and qualifications: There is no mention of experts, ground truth, or adjudication methods for any clinical study because the document focuses on substantial equivalence based on technological characteristics and compliance with standards, not necessarily a clinical outcomes study with human readers.
  4. Adjudication method: Not applicable as there is no mention of expert review or ground truth establishment in this context.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: This type of study is not mentioned. The filing is for a ventilator, not an AI or imaging diagnostic tool that typically uses MRMC studies.
  6. Standalone (algorithm only) performance: Not applicable as this is a physical medical device (ventilator), not a diagnostic algorithm.
  7. Type of ground truth used: Not specified, as detailed study results are absent.
  8. Sample size for the training set: Not applicable as this is a physical medical device, not an AI model that typically uses a training set for machine learning.
  9. How ground truth for the training set was established: Not applicable.

To obtain the information you're asking for, you would generally need to review the full submission to the FDA, which would include detailed test reports, or clinical study reports if they were conducted. The provided text is a high-level summary and approval letter, not the detailed technical and clinical evidence itself.

{0}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 4 2004

Mr. David D' Cruz Vice President Regulatory & Clinical Affairs US Resmed Corporation 14040 Danielson Street Poway, CA 92064-6857

Re: K033276 Trade/Device Name: Resmed VPAP III-STA Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: May 17, 2004 Received: May 19, 2004

Dear Mr. D'Cruz:

This letter corrects our substantially equivalent letter of June 7, 2004, regarding the VPAP III-STA. Our letter identified the regulation number as 868.5905 and the product code as BZD. This is in error; the regulation number is 868.5895 and the correct product code is MNS as indicated above.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 -Mr. David D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing regiotice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-5613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Clu

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Sponsor: ResMed Lid

VPAP III ST-A Special 510(k) Premarket Notification

INDICATIONS FOR USE 4

510(k) Number (if known):

KO33276

Device Name:

VPAP III -STA

Indications for Use:

The VPAP III ST-A system is Intended to provide non-invasive ventilation for patients with respiratory insufficiency or obstructive sleep apnea (OSA), in the hospital or home.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH: Office of Device Evaluation (ODE)
(Division Sign-Off)Division of Anesthesiology, General Hospital:Infection Control, Dental Devices
510(k) NumberK033276
OR
Prescription Use(Per 21 CFR 801.109)Over-The-Counter Use

(Optional Format 1-2-96)

Page 12

{3}------------------------------------------------

JUN - 7 2004

K033276

510(k) SUMMARY—VPAP III ST-A

Submitter Name:ResMed Corp.
Submitter Address:14040 Danielson Street, Poway CA 92064-6857USA
Contact Person:David D'Cruz, VP Regulatory & Clinical Affairs US
Phone Number:(858) 746 2238
Fax Number:(858) 746 2915
Date Prepared:June 3, 2004
Device Trade Name:VPAP III ST-A
Device Common Name/Classification Name:Variable Positive Air Pressure (VPAP) System /Continuous Ventilator, Non-life-support
Predicate Devices:VPAP II ST-A (K974417)
Device Description:The VPAP III ST-A provides a mode of non-invasive positivepressure ventilation (NPPV) called Pressure Support withPEEP, which delivers two treatment pressures (bi-levelventilation).A higher pressure is applied when the patient inhales - IPAPor inspiratory positive airway pressure, and a lower pressure isapplied when the patient exhales - EPAP or expiratory positiveairway pressure, sometimes referred to as PEEP or positiveend-expiratory pressure.The difference between the two treatment pressuresrepresents the amount of pressure support provided to thepatient.
Intended Use:The VPAP III ST-A system is intended to provide non-invasiveventilation for patients with respiratory insufficiency orobstructive sleep apnea (OSA), in the hospital or home.

Device Technological Characteristics and Comparison to Predicate Device(s):

The VPAP III ST-A is a flow-cycled, pressure-limited ventilator. A blower assembly generates airway pressure. A flow sensor and a pressure sensor in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the user to set and adjust certain clinical parameters.

The VPAP III ST-A has a CPAP mode in which a fixed pressure is delivered, and three bi-level operating modes which determine how the changes between IPAP and EPAP pressures are made: Spontaneous, Spontaneous/Timed and Timed.

{4}------------------------------------------------

The VPAP III ST-A is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

VPAP III ST-A is substantially equivalent to the VPAP II ST-A. The two ventilators have the same intended use, operating principle, technological characteristics and manufacturing process. The materials used in the air path are either predicate materials (previously cleared for the same intended use) or have been tested and found compliant with the biocompatibility requirements.

The main differences between the modified device VPAP III ST-A and the VPAP II ST-A are:

  • A new enclosure design for VPAP III ST-A comparable in size and shape with that of . the VPAP II ST-A:
  • Additional LCD display in the User Interface of the VPAP III ST-A. .
    • VPAP JI ST-A requires an external Universal Control Unit (UCU) in order to display therapy configuration and treatment data. VPAP III ST-A has the ability to display such data on the ventilator itself via an LCD.
  • A new power supply unit (PSU), which can be used across the ResMed product range; . the new PSU is a switching mode power supply, i.e., it has the same principle of operation as the VPAP II ST-A power supply.

The VPAP III ST-A can also be run from a 12V battery via a DC-DC (12V/ 24V) converter in the event of mains power not being reliable. The VPAP III ST-A is however not intended for transport use as defined in section 5.5 of the FDA Draft Reviewer Guidance for Ventilators (July 1995).

  • A new blower motor, which can supply operating pressure up to 30 cmH2O, having the . same principle of operation as the VPAP II ST-A blower motor;
  • New flow sensor with superior accuracy at low flow values, and new microprocessor . controller with superior processing power and more competitive price;

VPAP II ST-A has one flow sensor and two pressure sensors, the second pressure sensor having its own alarm microprocessor controller and acting as backup in the event that the first pressure sensor fails. VPAP III ST-A has a flow sensor and one pressure sensor, the microprocessor controller checks every 50ms that the pressure sensor works properly thus eliminating the need for a backup pressure sensor with its own microprocessor controller.

VPAP III ST-A compensates for pressure drop in the air tubing, whereas VPAP II ST-A does not.

  • When a humidifier is used, the pressure sensing point for VPAP II ST-A and VPAP III ST-. A is before the humidifier, except that when an integrated humidifier (H2i) is used with VPAP III ST-A the pressure sensing point was moved after the humidifier to compensate for the pressure drop in the humidifier and ensuring a more accurate control of the mask pressure.
  • VPAP II ST-A has an analog breath detection circuit and fixed thresholds for triggering . and cycling. The VPAP III ST-A breath detection is software-based, and the device has adjustable sensitivity levels - Low, Medium and High - for triggering and cycling thresholds.
  • VPAP II ST-A has a single generic vent pressure-flow compensation characteristic . estimated in hardware. VPAP III ST-A has a range of vent pressure-flow compensation characteristics for different mask types, calculated more accurately by software based on the known mask pressure.
  • Leak management for VPAP II ST-A is implemented in hardware. Leak in VPAP III ST-A . is calculated more accurately by software taking into consideration the fact that leak rate depends on pressure.

{5}------------------------------------------------

  • The rise time, i.e., the time taken to for the pressure to increase from EPAP to IPAP can . be adjusted in VPAP III ST-A for increased patient comfort. VPAP II ST-A does not have this feature.
  • Additional alarms for VPAP III ST-A; .

VPAP II ST-A has the following alarms: power fail, high pressure and low pressure. System faults, which might cause high or low pressures, and gross leak are reported either by the high-pressure alarm or by the low-pressure alarm, as applicable. VPAP III ST-A has the following alarms: power fail, high pressure and low pressure, mask-off condition, (gross) over pressure, over use (operation outside specification), system fault, non-vented mask, and low minute ventilation.

  • VPAP III ST-A incorporates additional compliance statistics compared to VPAP II ST-A. .
  • VPAP III ST-A has more flexibility in setting the ramp (or delay) time: from 5 to 45 . minutes in 5 minutes increments, compared to 5, 10 or 20 minutes for VPAP II ST-A.

Performance Data:

The VPAP III ST-A was tested in accordance with the recommendations of the FDA Draft Reviewer Guidance for Ventilators (July 1995).

The VPAP III ST-A ventilator complies with the following standards: IEC 60601-1-2; IEC 60068-2-27/ -6/ -34; IEC 10993-1; ASTM F 1100-90.

Conclusion:

The VPAP III ST-A is substantially equivalent to the VPAP II ST-A.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).