K Number
K033275

Validate with FDA (Live)

Date Cleared
2003-11-06

(27 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIFEPAK 500, LIFEPAK CR Plus, and the LIFEPAK 12 and LIFEPAK 20 in automated external defibrillation mode are intended for use on patients in cardiac arrest.

LIFEPAK 12 and LIFEPAK 20 Defibrillator/Monitor/Pacemakers: The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm.

LIFEPAK 500 Automated External Defibrillator: The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing no signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient's ECG rhythm.

LIFEPAK CR Plus Automated External Defibrillator: The LIFEPAK CR Plus defibrillator is indicated for use on patients in cardiac arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement).

Device Description

The LIFEPAK 500 and LIFEPAK CR Plus automated external defibrillators are small, portable, battery operated devices intended for treatment of cardiac arrest. Both devices use a patented software algorithm to analyze the patient's electrocardiogram (ECG) to determine if a shockable rhythm is present. The LIFEPAK 500 and the semi-automatic version of the LIFEPAK CR Plus will then inform the operator if it detects a shockable rhythm and the operator can then press the shock button to deliver energy. The energy is delivered via disposable defibrillation electrodes applied to the chest. The LIFEPAK CR Plus is also available in a fully automatic version which does not require operator interaction to charge and discharge.

The LIFEPAK 12 and LIFEPAK 20 devices use the same shock advising software algorithm as the devices described above. Features available in addition to automated external defibrillation include manual external defibrillation, noninvasive pacing, ECG monitoring, pulse oximetry, and synchronized cardioversion. The LIFEPAK 12 also offers noninvasive blood pressure monitor, invasive blood pressure monitor, and end-tidal CO2 monitor features.

AI/ML Overview

The provided text does not contain detailed information about specific acceptance criteria and a study proving a device meets these criteria. It is a 510(k) summary for Medtronic Physio-Control defibrillators, focusing on substantial equivalence to previously cleared devices rather than a de novo study with explicit acceptance criteria.

However, based on the general nature of such devices, we can infer some aspects and highlight what information is missing.

Here's an attempt to answer the questions, categorizing what can be inferred and what is explicitly not available in the provided text:


1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Inferred)Reported Device Performance (Inferred/Not explicitly stated)
Accuracy of ECG analysis for shockable rhythm detection (e.g., Sensitivity, Specificity, Positive Predictive Value, Negative Predictive Value for identifying Ventricular Fibrillation (VF) and Ventricular Tachycardia (VT) suitable for defibrillation, and correctly identifying non-shockable rhythms)The document states, "Both devices use a patented software algorithm to analyze the patient's electrocardiogram (ECG) to determine if a shockable rhythm is present." It also mentions, "The LIFEPAK 12 and LIFEPAK 20 devices use the same shock advising software algorithm as the devices described above." The implication is that this algorithm's performance is acceptable and consistent with previously cleared predicate devices. Specific performance metrics (e.g., sensitivity, specificity) and numerical acceptance thresholds are NOT provided in this document.
Safety and Effectiveness (General)The 510(k) process determines "substantial equivalence" to legally marketed predicate devices, implying that the new devices are as safe and effective as the older ones. The letter from FDA states, "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent... You may, therefore, market the device, subject to the general controls provisions of the Act." Specific safety and effectiveness data or metrics are NOT detailed.
Device Performance - Intended Use Population (e.g., appropriate rhythm analysis for patients in cardiac arrest)The indications for use specify the target population: "patients in cardiac arrest," "unconscious, pulseless, and not breathing spontaneously." The device is intended "to analyze the patient's ECG rhythm." Specific performance within these populations (e.g., how often it correctly identifies shockable rhythms in actual cardiac arrest patients) is NOT quantified.
Functional Features (e.g., manual external defibrillation, noninvasive pacing, ECG monitoring, pulse oximetry, synchronized cardioversion, blood pressure, CO2 monitoring)The document describes these features as available on LIFEPAK 12 and 20. For the new submission, it states, "The features and functions of the LIFEPAK 12, LIFEPAK 20, LIFEPAK 500, and LIFEPAK CR Plus are the same as those of the currently marketed versions of each device." This implies functional equivalence, but no specific performance data or acceptance criteria for these ancillary functions are provided.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified.
  • Data Provenance: Not specified. The document focuses on substantial equivalence based on prior clearances and the shared algorithm, not new clinical study data details.
  • Retrospective or Prospective: Not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not specified. The document does not describe a study involving human experts establishing ground truth for evaluating the device's performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not specified. As no expert ground truth establishment for a test set is detailed, no adjudication method is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study is not mentioned. This document is for a 510(k) clearance based on substantial equivalence, not a new comparative effectiveness study with human readers and AI assistance. The devices include an "automated external defibrillator" (AED) mode, implying the algorithm is the primary decision-maker for shock advice, not an "AI assistance" for human readers in a diagnostic capacity.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The core of the AED functionality described relies on a "patented software algorithm to analyze the patient's electrocardiogram (ECG) to determine if a shockable rhythm is present." While some devices (LIFEPAK 500 and semi-automatic CR Plus) then require operator interaction to press a shock button, the analytical part of the decision-making is automated. The fully automatic CR Plus explicitly "does not require operator interaction to charge and discharge." The substantial equivalence argument rests on this existing, cleared algorithm's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The document does not explicitly state the type of ground truth used for the original validation of the "patented software algorithm." However, for AEDs, ground truth for shockable rhythms (Ventricular Fibrillation (VF) and rapid Ventricular Tachycardia (VT)) is typically established by:
    • Expert Consensus: Review of ECG waveforms by cardiologists or electrophysiologists.
    • Physiological Correlation: In some research contexts, correlation with actual cardiac activity or response to defibrillation.
    • Standardized Databases: Often, development and validation use large, annotated ECG databases derived from various clinical situations.

8. The sample size for the training set

  • Not specified. The document describes existing algorithms and devices, not the details of their initial training data.

9. How the ground truth for the training set was established

  • Not specified. As with the test set ground truth, the document refers to a "patented software algorithm" and its performance in previously cleared devices, but does not detail the methodology for establishing ground truth during its original development and training.

{0}------------------------------------------------

SECTION E: 510(k) SUMMARY

Submitter's Name and Address:

Medtronic Physio-Control Corp. 11811 Willows Road Northeast P.O. Box 97006 Redmond, WA 98073

Contact Person:

Michelle Ackermann (425) 867-4744

Date Summary Prepared:

September 22nd, 2003

Devices:

Medtronic Physio-Control LIFEPAK® 12, 20, 500, and CR Plus Defibrillators

Classification:

Low Energy DC-Defibrillator: Class II Automatic External Defibrillator: Class III

Substantial Equivalence:

The features and functions of the LIFEPAK 12, LIFEPAK 20, LIFEPAK 500, and LIFEPAK CR Plus are substantially equivalent to the previously cleared version of each device listed in the table below.

Device510(k) Information
Medtronic Physio-Control LIFEPAK 12DefibrillatorK991910 (09/03/99)K010918 (08/23/01)
Medtronic Physio-Control LIFEPAK 20DefibrillatorK012274 (02/05/02)
Medtronic Physio-Control LIFEPAK 500AEDK983393 (05/05/99)K012428 (09/28/01)
Medtronic Physio-Control LIFEPAK CRPlus AEDK011144 (12/03/01)

{1}------------------------------------------------

Description:

The LIFEPAK 500 and LIFEPAK CR Plus automated external defibrillators are small, portable, battery operated devices intended for treatment of cardiac arrest. Both devices use a patented software algorithm to analyze the patient's electrocardiogram (ECG) to determine if a shockable rhythm is present. The LIFEPAK 500 and the semi-automatic version of the LIFEPAK CR Plus will then inform the operator if it detects a shockable rhythm and the operator can then press the shock button to deliver energy. The energy is delivered via disposable defibrillation electrodes applied to the chest. The LIFEPAK CR Plus is also available in a fully automatic version which does not require operator interaction to charge and discharge.

The LIFEPAK 12 and LIFEPAK 20 devices use the same shock advising software algorithm as the devices described above. Features available in addition to automated external defibrillation include manual external defibrillation, noninvasive pacing, ECG monitoring, pulse oximetry, and synchronized cardioversion. The LIFEPAK 12 also offers noninvasive blood pressure monitor, invasive blood pressure monitor, and end-tidal CO2 monitor features.

Intended Use:

The LIFEPAK 500, LIFEPAK CR Plus, and the LIFEPAK 12 and LIFEPAK 20 in automated external defibrillation mode are intended for use on patients in cardiac arrest.

Technological characteristics of new and predicate devices:

The features and functions of the LIFEPAK 12, LIFEPAK 20, LIFEPAK 500, and LIFEPAK CR Plus are the same as those of the currently marketed versions of each device.

Summary of Design Controls:

This 510(k) includes a summary of design control activities and a declaration of conformity to design controls.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

NOV - 6 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic Physio-Control Corp. c/o Ms. Michelle Ackermann Regulatory Affairs Specialist 11811 Willows Road NE P.O. Box 97006 Redmond, WA 98073-9706

Re: K033275

Trade Name: LIFEPAK® 12, 20, 500 and CR Plus Defibrillators Regulation Number: 21 CFR 870.1025 Regulation Name: Automatic External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: October 9, 2003 Received: October 10, 2003

Dear Ms. Ackermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Michelle Ackermann

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Dng Leiseher for
Brent D. Zuckerman, M.D.

ram D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SECTION D: STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): Not yet assigned

Device Name: LIFEPAK 12 and LIFEPAK 20 Defibrillator/Monitor/Pacemakers

Indications for AED Use:

The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 12 and LIFEPAK 20 defibrillator/monitors are not intended for use on pediatric patients less than 8 years old.

Device Name: LIFEPAK 500 Automated External Defibrillator

Indications for Use:

The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing no signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient's ECG rhythm. With Infant/Child Reduced Energy Defibrillation Electrodes, the specially configured biphasic LIFEPAK 500 AED may be used on children up to 8 years old or 25kg (55 lb).

Device Name: LIFEPAK CR Plus Automated External Defibrillator

Indications for Use:

The LIFEPAK CR Plus defibrillator is indicated for use on patients in cardiac arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). With Infant/Child Reduced Energy Defibrillation Electrodes, the LIFEPAK CR Plus defibrillator may be used on children up to 8 years old of 25kg (55 lb).

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

fice of Device Evaluation (ODE)

(Optional Format 3-10-98)

Ana Luna
Division Sign-Off

5 iu(K) number K03275

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.