K Number
K033234

Validate with FDA (Live)

Date Cleared
2003-12-12

(67 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DPC IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does IMMULITE 2000. The intent of the system is to impart the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as IMMULITE 2000. Examples of the array of assays to be used with the DPC IMMULITE 2500 are two for which data in support of a claim of substantial equivalence have been submitted to FDA: hCG (product code: DHA) and TSH (JLW). The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000.

Device Description

The IMMULITE® 2500 is an Automated Immunoassay Analyzer. It is a discrete photometric chemistry analyzer for clinical use intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. The IMMULITE 2500 uses ¼ inch polystyrene antibody coated beads and assay specific antibody or antigen labeled with alkaline phosphatase. The chemiluminescent detection system is a phosphate ester stabilized dioxetane. Cleavage of the phosphate ester by alkaline phosphatase results in the decomposition of the dioxetane and the emission of photons, which are quantified by a luminometer and are proportional to the quantity of analyte present.

AI/ML Overview

The IMMULITE® 2500 Automated Immunoassay Analyzer is an automated immunoassay system that intends to assay the same broad range of analytes in patient samples as the IMMULITE 2000. The device is intended to provide the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as the IMMULITE 2000.

The 510(k) summary only mentions two analytes for which data was submitted to FDA in support of a claim of substantial equivalence: hCG (human chorionic gonadotropin) and TSH (thyroid-stimulating hormone). However, it does not explicitly state the acceptance criteria for these tests nor does it provide a table of performance data. The FDA's substantial equivalence determination letter refers to "indications for use stated in the enclosure" but this enclosure is not provided in the document.

Therefore, specific acceptance criteria and detailed device performance data are not available in the provided text.

Based on the limited information available:

1. A table of acceptance criteria and the reported device performance:

This information is not provided in the text. The document refers to "data in support of a claim of substantial equivalence have been submitted to FDA" for hCG and TSH, but the acceptance criteria and the actual performance data (e.g., sensitivity, specificity, accuracy, precision) are not detailed within the provided summary or the FDA letter.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not specified.
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as this is an in-vitro diagnostic device measuring analytes, not relying on expert interpretation of images or clinical findings for ground truth in the typical sense. The ground truth for analytical performance would typically be established by reference methods or gravimetric methods for controls and calibrators, and clinical outcome for patient samples. However, no details on how "ground truth" was established for the comparison studies are provided.

4. Adjudication method for the test set:

Not applicable for an in-vitro diagnostic device like this, which assesses analytical performance rather than diagnostic interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is an automated immunoassay analyzer, not a device requiring human readers or AI assistance in interpretation. The comparison is between the performance of the IMMULITE 2500 and its predicate device, IMMULITE 2000, in terms of analytical output.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, the IMMULITE 2500 is an automated immunoassay system, implying standalone performance. The intent is "to impart the same automation to the same array of immunoassays" as its predicate, the IMMULITE 2000, with "safe and effective performance when used by medical laboratory personnel." This indicates a standalone performance study against the predicate device.

7. The type of ground truth used:

While not explicitly stated, for an immunoassay analyzer, the "ground truth" for demonstrating substantial equivalence would typically involve:

  • Reference methods: Comparing results from the IMMULITE 2500 to established, validated reference methods for measuring hCG and TSH.
  • Comparison to predicate device: Direct comparison of results from patient samples and control materials analyzed on the IMMULITE 2500 against those analyzed on the IMMULITE 2000 (the predicate device). This would involve assessing correlation, bias, and agreement.
  • Known concentrations: Using control materials with known concentrations of analytes to verify accuracy and precision.

8. The sample size for the training set:

Not applicable. This device is an immunoassay analyzer and does not utilize a "training set" in the context of machine learning. Its performance is based on its established biochemical and detection technology.

9. How the ground truth for the training set was established:

Not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

K033234

IMMULITE® 2500 Automated Immunoassay Analyzer 510(k) Summary of Safety and Effectiveness

This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.

Name:Diagnostic Products Corporation
Address:5700 West 96th StreetLos Angeles, California 90045-5597
Telephone Number:(310) 645-8200
Facsimile Number:(310) 645-9999
Contact Person:Edward M. Levine, Ph.D.Director, Clinical Affairs
Date of Preparation:October 2, 2003
Device Name:Trade:IMMULITE® 2500
Catalog Number:IM5LITE
CFR:A discrete photometric chemistry analyzer forclinical use is a device intended to duplicate manualanalytical procedures by performing automaticallyvarious steps such as pipetting, preparing filtrates,heating, and measuring color intensity. This deviceis intended for use in conjunction with certainmaterials to measure a variety of analytes. Differentmodels of the device incorporate variousinstrumentation such as microanalysis apparatus,double beam, single, or dual channel photometers,and bichromatic 2-wavelength photometers. Somemodels of the device may include reagent-containing components that may also serve asreaction units.
Common:Discrete photometric chemistry analyzer for clinicaluse.
Classification:Class I device, JJE (21 CFR 862.2160)
Panel:Clinical Chemistry

{1}------------------------------------------------

Manufacturer:Corporate Headquarters:Diagnostic Products Corporation5700 West 96th StreetLos Angeles, CA 90045
Instrument Manufacturer:DPC Instrument Systems Division62 Flanders Bartley RoadFlanders, NJ 07836(Formerly DPC Cirrus Inc.)
Establishment Registration Number:DPC Headquarters Registration Number: 2017183DPC Instrument Systems Division Registration Number: 2247117
Substantially Equivalent Predicate Device:IMMULITE 2000 (K970227)
Description of Device:IMMULITE 2500 Automated Immunoassay Analyzer

Intended Use of the Device:

These modifications do not change the indications for use, nor the intended use from the IMMULITE 2000 to the IMMULITE 2500. The IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does IMMULITE 2000. The intent of the system is to impart the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as IMMULITE 2000. The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000.

Technology:

The IMMULITE 2500 uses ¼ inch polystyrene antibody coated beads and assay specific antibody or antigen labeled with alkaline phosphatase. The chemiluminescent detection system is a phosphate ester stabilized dioxetane. Cleavage of the phosphate ester by alkaline phosphatase results in the decomposition of the dioxetane and the emission of photons, which are quantified by a luminometer and are proportional to the quantity of analyte present.

Conclusion:

The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for the IMMULITE 2500 Automated Immunoassay Analyzer.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 2 2003

Edward Levine, Ph.D. Director, Clinical Affairs Diagnostics Products Corporation 5700 West 96th Street Los Angeles, CA 90045

Re: K033234 Trade/Device Name: IMMULITE® 2500 Automated Immunoassay Analyzer Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: DHA; JLW; JJE Dated: October 3, 2003 Received: October 6, 2003

Dear Dr. Levine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincercly yours,

Steven Putman

Steven 1. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): _ KO 33 2 3 4 IMMULITE® 2500 Automated Immunoassay Analyzer Device Name:

Indications For Use:

The DPC IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does IMMULITE 2000. The intent of the system is to impart the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as IMMULITE 2000. Examples of the array of assays to be used with the DPC IMMULITE 2500 are two for which data in support of a claim of substantial equivalence have been submitted to FDA: hCG (product code: DHA) and TSH (JLW). The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation i

Carol C Benson for Jean Cooper, DVM
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033234
OR10(k).
Over-The-Counter Use

Prescription Use
(Per 21 CFR 801.109)

(Optional Format 1-2-96)

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.