The Telecare®D.R. is intended to be used in conjunction with 510(k) certified devices that upon prescription by an authorized healthcare provider for patients provides a means of receiving and storing audio, video, and patient biometric data including vital signs information received over standard phone lines or high speed connection between patients, typically at home, and health professionals at the health care provider's site.
The Telecare DR System consists of multiple software and hardware components. Utilizing this system of components, healthcare professionals are empowered with the ability to conduct real-time, secure audio/visual communication and measurement collection encounters from the patient's home to the healthcare facility and presented for secure Internet browser review.
Typically, the Telecare D.R. resides on a health facility's server and connects to the HCV NurseStation. Using the HCV NurseStation software component, the healthcare professional instantiates audio/visual communication with the patient using a standalone videophone system (i.e., CS-126S, CS-126, or CS-146 videophone devices) that is equipped with the capability to transmit and receive data from a variety of external vital signs and medical measurement devices. The CS 126S also includes an integrated digital stethoscope that provides for both local and remote auscultation functions and an optional accessory plug-in pulse oximetry sensor unit.
Upon connection with a standalone videophone, the HCV NurseStation, located at a healthcare professional's office or facility, identifies all the external vital signs and medical measurement devices (i.e., weight scale, glucometer, vital signs monitor, NIBP, pulse oximetry and spirometer) connected to the standalone videophone. Using the HCV NurseStation, the healthcare professional may proceed to communicate using standard phone line or high speed connection through internal modems and transmit real-time video, audio, and data between them. The real-time video and audio communications allow the patient and caregiver to view and speak with each other.
Using the HCV NurseStation component, the healthcare professional is capable of monitoring the data from the measurement devices in real-time during collection, as well as, storing these measurements to the centralized data repository using the Telecare D.R. server software. To terminate the encounter, the healthcare professional disconnects the video call via the HCV NurseStation software component. This event triggers the HCV Conduit software component, which completes remaining data inserts/updates, broadcasts notifications and prepares data for secure browser viewing via the secure HCV Rendezviewer software component.
With existing legally marketed vital signs measurement devices attached to the patient's system, the health professional and caregiver may monitor the patient's blood pressure. pulse rate, temperature, weight, blood oxygen saturation, blood glucose level, breath peak flow, images, and transmit this data to the caregiver's system. Vital signs measurement devices used with the HCV NurseStation and Telecare D.R. are FDA approved devices and are used for the same purposes which they received 510(k) approval or are devices that are exempt under applicable 21 CFR sections. The data may be captured, displayed, and/or saved on Telecare D.R.
Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Telecare D.R. device:
Based on the provided 510(k) summary, the Telecare D.R. is a software system intended as an accessory to existing 510(k) cleared medical devices for remote patient monitoring. Crucially, the document explicitly states that "Clinical study information was not submitted for the purpose of demonstrating substantial equivalence to the predicate devices." Therefore, there is no information about clinical acceptance criteria or studies to directly prove device performance in a clinical setting provided in this document.
The focus of the submission is to demonstrate substantial equivalence to predicate devices through non-clinical testing of the software's functional performance and its compatibility with approved vital signs measurement devices.
Here's a breakdown of the requested information based on the document:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Functional Equivalence | Software performs equivalently when used with Telecare D.R. as when operated as independent devices. | "Testing was performed to validate the functional performance of the Telecare D.R. Specifically, performance testing of the software was conducted to show that performance exceeds and thus meets the substantial equivalency of the predicate devices.""Further testing was performed with each vital signs measurement device to show that they operate equivalently when used with the Telecare D.R. as when operated as independent devices." |
| Data Transmission & Storage | Capability to receive and store audio, video, and patient biometric data (vital signs) over standard phone lines or high-speed connections. | "The Telecare D.R. is intended to be used...to provide a means of receiving and storing audio, video, and patient biometric data including vital signs information received over standard phone lines or high speed connection...""The main functional difference between the systems are the Telecare D.R. is the data server/Web server that permits the storage, retrieval, and presentation of patient-specific vital sign data using standard phone lines and high speed connections over the Internet." |
| Compatibility with Approved Devices | Device interfaces correctly with listed 510(k) cleared vital signs measurement devices (e.g., Criticare Vital Signs Monitor, NIBP Monitor UA-767PC, etc.). | "Further testing was performed with each vital signs measurement device to show that they operate equivalently when used with the Telecare D.R. as when operated as independent devices." (Implies successful integration and performance). |
Study Details (Based on Non-Clinical Testing Section)
-
Sample size used for the test set and the data provenance:
- Sample Size: Not specified. The document only mentions "performance testing of the software" and "testing was performed with each vital signs measurement device." This likely refers to a technical verification and validation activity rather than a clinical sample size.
- Data Provenance: Not applicable in the context of a clinical test set, as no clinical studies were performed. For the functional testing, the data would be generated from the connected devices and the Telecare D.R. system itself.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. As no clinical studies were conducted, there was no need for expert-established ground truth in a clinical sense. The ground truth for functional testing would be the expected outputs and behaviors of the system and connected devices as per their specifications.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. No adjudication method for a test set is mentioned as no clinical studies were performed.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. The document explicitly states: "Clinical study information was not submitted for the purpose of demonstrating substantial equivalence to the predicate devices." This means no MRMC study was performed.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, in a sense. The document describes "performance testing of the software" and testing to ensure connected devices "operate equivalently when used with the Telecare D.R. as when operated as independent devices." This suggests standalone functional testing of the software and its interaction with the devices, without necessarily involving interpretation by human readers in a diagnostic capacity. The device itself is not a diagnostic AI, but a communication and data management system.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical functional testing, the ground truth would be the expected functional behavior and data integrity as defined by the system's design specifications and the known outputs of the connected FDA-approved vital signs monitors. It's not clinical ground truth like pathology or expert consensus.
-
The sample size for the training set:
- Not applicable. The Telecare D.R. is described as a software system for communication and data management, not an AI or machine learning algorithm requiring a "training set" in the conventional sense for diagnostic or predictive purposes.
-
How the ground truth for the training set was established:
- Not applicable, as there is no mention of a training set or AI/ML components requiring one.
{0}------------------------------------------------
HealthCare Vision, Inc.
Telemedicine Solutions
Section E: 510K Summary
Submitter
HealthCare Vision, Inc. 2601 Scott Avenue, Suite 600 Fort Worth, Texas 76103 Telephone: (817) 531-8992 Facsimile: (817) 531-8999
Contact Person
Craig A. Walker HealthCare Vision, Inc. Tel: 817.531.8992 ext 320 Tel: 512.832.6710 Fax: 817.531.8999 Fax: 512.832.5521
Date Prepared
September 29, 2003
Software Information
Trade Name: Telecare D.R.
Common Name: Telemedicine Communications Module
Classification Name: Powered Communication System
Software Description
The Telecare DR System consists of multiple software and hardware components. Utilizing this system of components, healthcare professionals are empowered with the ability to conduct real-time, secure audio/visual communication and measurement collection encounters from the patient's home to the healthcare facility and presented for secure Internet browser review.
Typically, the Telecare D.R. resides on a health facility's server and connects to the HCV NurseStation. Using the HCV NurseStation software component, the healthcare professional instantiates audio/visual communication with the patient using a standalone videophone system (i.e., CS-126S, CS-126, or CS-146 videophone devices) that is equipped with the capability to transmit and receive data from a variety of external vital
{1}------------------------------------------------
signs and medical measurement devices. The CS 126S also includes an integrated digital stethoscope that provides for both local and remote auscultation functions and an optional accessory plug-in pulse oximetry sensor unit.
Upon connection with a standalone videophone, the HCV NurseStation, located at a healthcare professional's office or facility, identifies all the external vital signs and medical measurement devices (i.e., weight scale, glucometer, vital signs monitor, NIBP, pulse oximetry and spirometer) connected to the standalone videophone. Using the HCV NurseStation, the healthcare professional may proceed to communicate using standard phone line or high speed connection through internal modems and transmit real-time video, audio, and data between them. The real-time video and audio communications allow the patient and caregiver to view and speak with each other.
Using the HCV NurseStation component, the healthcare professional is capable of monitoring the data from the measurement devices in real-time during collection, as well as, storing these measurements to the centralized data repository using the Telecare D.R. server software. To terminate the encounter, the healthcare professional disconnects the video call via the HCV NurseStation software component. This event triggers the HCV Conduit software component, which completes remaining data inserts/updates, broadcasts notifications and prepares data for secure browser viewing via the secure HCV Rendezviewer software component.
With existing legally marketed vital signs measurement devices attached to the patient's system, the health professional and caregiver may monitor the patient's blood pressure. pulse rate, temperature, weight, blood oxygen saturation, blood glucose level, breath peak flow, images, and transmit this data to the caregiver's system. Vital signs measurement devices used with the HCV NurseStation and Telecare D.R. are FDA approved devices and are used for the same purposes which they received 510(k) approval or are devices that are exempt under applicable 21 CFR sections. The data may be captured, displayed, and/or saved on Telecare D.R.
Substantial Equivalence
The Telecare D.R. system and the predicate devices listed above have the same intended use and very similar principles of operations and technological characteristics. The Telecare D.R. and it predicate systems have the same general use to provide the capability for health professionals to monitor remote patient vital signs and/or breath and heart sounds and store the information on a data server/web server over the Internet and over standard phone line. The Telecare D.R. has no internal medical devices and it is a software system that performs as an accessory to 510K medical devices. Communication occurs through a plural of modalities include using standard network connection (H.323), standard phone line (H.324), and wireless communications to connect to devices for recording vital sign data (i.e., blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultations of the chest and abdominal cavities). These devices function independently in accordance with their own individual specifications and operation. None of the systems are intended solely for
{2}------------------------------------------------
diagnostic purposes. In general, basic operation for the Telecare D.R. consists of 1) receiving patient vital sign data, images and sounds of the heart, lung, or bowel cavities 2) presenting the data in a graphical and visual format, and 3) providing Internet access to the data by assigned health professionals.
In addition, the following devices can be used with the Telecare D.R. system to provide the external vital signs monitoring functions: Criticare Vital Signs Monitor (K884216) by Criticare Systems Inc., NIBP Monitor UA-767PC (K982481) by A&D Engineering, Inc., Digital Weight Scale UC-321 (exempt under 21 CFR 800.2700) by A&D Engineering. Inc., Ferraris KOKO Peak Pro (K013489) by PDS Healthcare Products Inc., Freestyle Blood Glucose Monitor (K000582) by Therasense. With regards to the transmission of patient vital signs information from external measurement devices, it is equivalent to the Carecompanion Nurse Station/Carecompanion Patient Station (K020584). With respect to the pulse oximetry accessory, it is equivalent to the Zymed Easi-View Telemetry System (K001308).
The main difference from the predicate devices and platforms is that Telecare D.R. is a software component to the predicate devices and systems. Using off-the-shelf hardware and patented technology, Telecare D.R. provides more interface options with approved 510(k) vital sign devices and systems. In addition, the Telecare D.R. uses standard database architecture for managing the vital sign data received from the devices. Telecare D.R. and its predicate devices and systems have the same general use: to provide the capability for health care professionals to monitor blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultation sounds of their patients from remote locations. The main functional difference between the systems are the Telecare D.R. is the data server/Web server that permits the storage, retrieval, and presentation of patient-specific vital sign data using standard phone lines and high speed connections over the Internet.
Intended Use
The Telecare D.R. is intended to be used upon prescription of an authorized health care provider by patients to provide two-way audio/video communications, data transmission & communications, transmission of vital signs information over standard telephone lines and high speed telecommunications between the patient, typically at home, and a health professional at the health care professional's site.
The information includes heart, lung, and bowel sounds, blood oxygen saturation, pulse rate, blood pressure, temperature, blood glucose, weight, and breath peak flow measurements. The information is collected upon the request and direction of the health care provider.
The software system is a diagnostic aid that is an accessory to 510(k) devices. Clinical judgment and experience are required to check and interpret the information transmitted. The software system is not intended as a substitute for medical care. The software is
{3}------------------------------------------------
contraindicated for patients requiring direct medical supervision or emergency intervention.
Brief Description of Non-Clinical Testing
Testing was performed to validate the functional performance of the Telecare D.R. Specifically, performance testing of the software was conducted to show that performance exceeds and thus meets the substantial equivalency of the predicate devices. Further testing was performed with each vital signs measurement device to show that they operate equivalently when used with the Telecare D.R. as when operated as indenendent devices.
Brief Desciption of Clinical Testing
Clinical study information was not submitted for the purpose of demonstrating substantial equivalence to the predicate devices.
Conclusion
The indications for use of the Telecare D.R. software as an accessory of 510(k) devices is consistent with that in labeling for telemedicine communications modules legally marked in the United States.
The results of testing Telecare D.R. with approved 510(k) devices for monitoring or recording vital sign data (i.e., blood pressure, pulse, heart rate, oxygen saturation, glucose, weight, ECG, temperature breath peak flow and auscultations of the chest and abdominal cavities) indicate that the software is substantially equivalent to its predicates and does not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of a bird.
APR 1 3 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Craig Walker Vice President of Public Policy & Development HealthCare Vision, Inc. 2601 Scott Avenue Suite 600 Fort Worth, TX 76103
Re: K033133
Trade Name: Telecare D.R. Server Software System Regulation Number: 21 CFR 870.2300 Regulation Name: Patient Physiological Monitor (Without Arrhythmia Detection or Alarms) Regulatory Class: Class II (two) Product Code: MWI Dated: January 28, 2004 Received: January 29, 2004
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Craig Walker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed prodicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Duana R. Luchner
(A) Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510K Pending File: K033133 Healthcare Vision, Inc. Telecare D.R. Server Software System
Section D: Indications For Use - Revised 3-30-2004
510(k) Number (if known): K033133
Device Name: Telecare® D.R.
Indications For Use:
The Telecare®D.R. is intended to be used in conjunction with 510(k) certified devices that upon prescription by an authorized healthcare provider for patients provides a means of receiving and storing audio, video, and patient biometric data including vital signs information received over standard phone lines or high speed connection between patients, typically at home, and health professionals at the health care provider's site.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dana R. Vachner
(Division Sign-Off) Jivision of Cardiovascular Devices
25 510(K) Number_K0 331
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).