AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pasco MIC and MIC/ID panels are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement of category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

This 5.10(k) notification is for the antimicrobial Daptomycin at concentrations of 0.06 - 8 mcg/ml to Pasco Panels. Daptomycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobic.

Active In Vitro and in Clinical Infectious Against:

Aerobic Gram-positive microorganisms

Enterococcus faecalis (vancomycin-susceptible strains only) Staphylococcus aureus (including methicillin-resistant strains)

Active In Vitro but their clinical significance is unknown

Aerobic Gram-positive microorganisms

Enterococcus faecalis (vancomycin-resistant strains) Enterococcus faecium (including vancomycin-resistant strains) Staphylococcus epidermidis (including methicillin-resistant strains) Staphylococcus haemolyticus

Device Description

Pasco Panels are used for quantitatively measuring the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms. Varying concentrations of antimicrobial agents (usually in two-fold dilutions) are dispensed into the Pasco microdilution panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours and panels are then observed for visible growth or color changes as described in the package insert.

The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Changes in pH and production of specific metabolites from growth in biochemical substrates are interpreted as described in the package insert for conventional tubed media.

AI/ML Overview

The provided document describes the acceptance criteria and the study conducted for the Pasco MIC and MIC/ID Panels with the inclusion of Daptomycin.

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Essential Agreement (EA)Not explicitly stated as a numerical threshold, but implied to be high based on the "Class II Special Controls Guidance Document".99.6% for 413 challenge and clinical Staphylococcus spp. and 365 challenge and clinical Enterococcus spp.
QC EndpointsAcceptable for NCCLS recommended OC organisms (S. aureus ATCC 29213 and E. faecalis ATCC 29212) from both reference and test methodology.Acceptable
Reproducibility (Inter-site)Not explicitly stated as a numerical threshold.100% for all three sites
Reproducibility (Intra-site)Not explicitly stated as a numerical threshold.100% for all three sites

Study Information

  1. Sample size used for the test set and the data provenance:

    • Test Set Size: 413 challenge and clinical Staphylococcus spp. and 365 challenge and clinical Enterococcus spp. (Total = 778 isolates).
    • Data Provenance: The document states "Challenge strains, fresh clinical isolates, stock clinical isolates and OC strains were tested concurrently". It also mentions "Testing was conducted at three test sites." The country of origin is not explicitly stated. The nature of "fresh clinical isolates" and "stock clinical isolates" suggests a mix of retrospective and prospective data. "Challenge strains" and "OC strains" are likely laboratory-derived.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.
    • The ground truth was established by "reference methodology." While the document doesn't detail who performs this or their qualifications, it implies a standard, established method, likely overseen by qualified laboratory personnel.
  3. Adjudication method for the test set:

    • Not applicable as the study compares the device's results against a "reference methodology" rather than relying on multiple expert readings for ground truth establishment.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an Antimicrobial Susceptibility Test (AST) system, which typically provides objective measurements (MIC values) rather than interpretation requiring human readers in the same way an imaging AI would. This study focuses on the agreement of the device with a reference method, not on human interpretive improvement with AI assistance.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone study. The device (Pasco MIC and MIC/ID Panels) is a diagnostic test system that produces a result (MIC values). The study assessed the performance of this system ("Pasco methodology") independently against a "reference methodology." There is no mention of a human-in-the-loop component for the device's primary function or for the evaluation.
  6. The type of ground truth used:

    • The ground truth was established using reference methodology. For Antimicrobial Susceptibility Testing, this typically involves a standardized, validated laboratory method (e.g., agar dilution or broth microdilution following CLSI/NCCLS guidelines) that is considered the gold standard for determining MIC values.
  7. The sample size for the training set:

    • The document does not explicitly mention a "training set" in the context of an algorithm or AI. This study is for a traditional diagnostic panel, not an AI-driven device that requires training. The described isolates are for the "test set" or samples used for validation.
  8. How the ground truth for the training set was established:

    • As there is no mention of a "training set" for an algorithm, this question is not applicable. For the isolates used in the performance study, the ground truth was established by "reference methodology" as described in point 6.

{0}------------------------------------------------

NOV 1 8 2003

K033119

2

DATE:September 29, 2003
CONTACT PERSON:Linda K. DillonChuck LakelPasco Laboratories12750 West 42nd AvenueWheat Ridge, Co 80033303-390-3240
TRADE NAME OF DEVICE:Pasco MIC and MIC/ID Panels
COMMON NAME:Antimicrobial Susceptibility Test
CLASSIFICATION NAME:Class II Antimicrobial Susceptibility TestMicrobiology Panel #83

SUBSTANTIAL EQUIVALENCE:

In review of previous 510(k) notifications for the Pasco MIC and MIC/ID panel: K03259, August 19, 2003 RE: Gatifloxacin (Strep); K032518, September 15, 2003 RE: Gemifloxacin (Strep); K031727, July 30, 2003 RE: Gemifloxacin; K031205, June 13, 2003 RE: Linezolid: K031103, June 12, 2003 RE: Ertapenem; K030933, May 1, 2003 RE: Moxifloxacin: K030620, April 14, 2003 RE: Gatifloxacin; K011116, April 24, 2001 RE: ESBL Confirmatory Test; K010508, April 23, 2001 RE: ESBL Screen Test; K020331, March 20, 2002 RE: Ertapenem; K001953, August 10, 2000 RE: Amoxicillin; K001887, August 9, 2000 RE: Ampicillin; K001721, August 4, 2000 RE: Clarithromycin: K001612, July 18, 2000 RE: Linezolid: K001516, July 12, 2000 RE: Moxifloxacin: K992853, November 4, 1999 RE: Cefdinir, K992726, November 3, 1999 RE: Synercid (non-fastidious); K992717, November 2, 1999 RE: Synercid; K992646, October 19, 1999 RE: Penicillin: K992647, October 19, 1999 RE: Erythromycin; K992593, October 14, 1999 RE: Chloramphenicol; K992562, October 13, 1999 RE: Ceftriaxone; K992568, October 14, 1999 RE: Cefotaxime; K992507, October 18, 1999 RE: Trovafloxacin; K992546, October 12, 1999 RE: Meropenem; K992420, September 27, 1999 RE: Sparfloxacin; K992296, September 21, 1999 RE: Vancomycin; K992297, September 3, 1999 RE: Levofloxacin; K992143, September 16, 1999 RE: Clindamycin; K992108, September 3, 1999 RE: Cefepime; K992076, August 30, 1999 RE: Cefuroxime; K992059, August 30, 1999 RE: Imipenem; K992077, September 3, 1999 RE: Offloxacin; K991925, August 20, 1999 RE: Amoxicillin/Clavulanic Acid; and K946126, January 17, 1995 RE: Detection of resistant pneumococci), the FDA has determined the Pasco panels to be substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments.

{1}------------------------------------------------

510(k) SUMMARY (page 2 of 3)

3

The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of

substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.

DESCRIPTION OF THE DEVICE:

Pasco Panels are used for quantitatively measuring the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms. Varying concentrations of antimicrobial agents (usually in two-fold dilutions) are dispensed into the Pasco microdilution panels and the panels are then frozen. Panels are thawed prior to use, inoculated with the test organisms, incubated the traditional 16-24 hours and panels are then observed for visible growth or color changes as described in the package insert.

The lowest concentration of each antimicrobial agent with no apparent visible growth of the test organism is recorded as the minimum inhibitory concentration (MIC). Changes in pH and production of specific metabolites from growth in biochemical substrates are interpreted as described in the package insert for conventional tubed media.

INTENDED USE FOR THE PASCO MIC AND MIC/ID PANELS:

PASCO MIC AND MIC/ID PANELS are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement or category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

SUMMARY/CONCLUSION OF SUBSTANTIAL EQUIVALENCE TESTING: Challenge strains, fresh clinical isolates, stock clinical isolates and OC strains were tested concurrently using both Pasco methodology and reference methodology in panels that contained Daptomycin at concentrations ranging from 0.06 - 8 mcg/ml. Testing was conducted at three test sites.

Test results of 413 challenge and clinical Staphylococcus spp. and 365 challenge and clinical Enterococcus spp. demonstrated an Essential Agreement (EA) of 99.6%.

{2}------------------------------------------------

510(k) SUMMARY (page 3 of 3)

4

OC endpoints for the NCCLS recommended OC organisms (S. aureus ATCC 29213 and E. faecalis ATCC 29212) from panels using both the reference and test methodology were acceptable.

Reproducibility testing of 10 organisms at each site on three separate days in triplicate demonstrated inter-site and intra-site reproducibility of MIC results of 100% for all three sites.

The results of the clinical testing, reproducibility testing and QC performance testing supports Substantial Equivalence as outlined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes resembling birds in flight, stacked on top of each other.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV 1 8 2003

Ms. Linda K. Dillon R&D Manager BD Diagnostics Systems Pasco Laboratories 12750 W. 42nd Avenue Wheat Ridge, CO 80033-2440

Re: K033119

Trade/Device Name: PASCO MIC and MIC/ID Panels Daptomycin, 0.06-8 mcg/ml Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Regulatory Class: Class II Product Code: JWY Dated: September 29, 2003 Received: September 30, 2003

Dear Ms. Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: PASCO MIC and MIC/ID Panels - Daptomycin 0.06 - 8 mcg/ml

Indications For Use: Inclusion of Daptomycin

Pasco MIC and MIC/ID panels are used for quantitatively measuring (with the exception of the Breakpoint/ID panel which provides qualitative measurement of category results) the susceptibility of rapidly growing aerobic and facultative anaerobic bacterial pathogens to a battery of antimicrobial agents and determining the biochemical identification of those organisms.

This 5.10(k) notification is for the antimicrobial Daptomycin at concentrations of 0.06 - 8 mcg/ml to Pasco Panels. Daptomycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobic.

Active In Vitro and in Clinical Infectious Against:

Aerobic Gram-positive microorganisms

Enterococcus faecalis (vancomycin-susceptible strains only) Staphylococcus aureus (including methicillin-resistant strains)

Active In Vitro but their clinical significance is unknown

Aerobic Gram-positive microorganisms

Enterococcus faecalis (vancomycin-resistant strains) Enterococcus faecium (including vancomycin-resistant strains) Staphylococcus epidermidis (including methicillin-resistant strains) Staphylococcus haemolyticus

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CRF 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

Judith Pool

on Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safe

510(k) K03 3115

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).