(15 days)
The WIDE 2MP TFT LCD Monitor System IF1902A ™s intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners.
The WIDE 2MP TFT LCD Monitor System IF1902A ™ is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system used to display PACS medical images.
This document is a 510(k) summary for the WIDE 2MP TFT LCD Monitor System IF1902A™ and does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria.
The 510(k) summary primarily focuses on:
- Device Description: A flat panel hi-resolution LCD monitor system for displaying digital medical images.
- Intended Use: Displaying and viewing digital medical images for review and analysis by trained medical practitioners.
- Technological Characteristics: An image display system consisting of computer software and components; does not contact the patient or control life-sustaining devices.
- Predicate Device: CORNIS 2MP MEDICAL FLAT PANEL DISPLAY SYSTEM (K023322).
- Conclusion: The device is substantially equivalent to the predicate device.
- Compliance: States manufacturing in accordance with voluntary and safety standards (UL, FCC, CE, VCC, etc.).
There is no mention of:
- A table of acceptance criteria and reported device performance.
- Sample sizes or data provenance for any test set.
- Number or qualifications of experts used to establish ground truth.
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study.
- Standalone algorithm performance.
- Type of ground truth used.
- Sample size for a training set.
- How ground truth for a training set was established.
This document serves as a regulatory submission demonstrating substantial equivalence to a predicate device, rather than a clinical study report detailing performance against specific acceptance criteria.
{0}------------------------------------------------
OCT - 8 2003
Image /page/0/Picture/1 description: The image shows the word "WIDE" in large, bold, white letters against a black background. Below the word "WIDE" is smaller text that reads "World Innovative Display Engineering". The text is also in white. The image has a slightly grainy or textured appearance.
K$\phi$329 b$\phi$
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: August 4, 2003
Submitter's Information: 21 CFR 807.92(a)(1) WIDE Corporation Mr. YS Lim, President & CEO 576-5 Miwon-Ri, Miwon-Myun, Cheongwon-Gun Chung, Korea 363-874
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) WIDE 2MP TFT LCD Monitor System IF1902A ™ Trade Name: Picture Archiving Communications System Common Name: Device Classification: 892.2050 System, Image Processing Name:
Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL |
|---|---|
| Regulation Number | 892.2050 |
| 510(k) Number | K023322 |
| Device Name | CORNIS 2MP MEDICAL FLAT PANEL DISPLAYSYSTEM |
| Applicant | BARCO NVP.O. BOX 12038LA JOLLA, CA 92039 2038 |
| Product Code | LLZ |
Device Description: 21 CFR 807 92(a)(4)
The WIDE 2MP TFT LCD Monitor System IF1902A ™ is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system used to display PACS medical images.
Indications for Use: 21 CFR 807 92(a)(5)
The WIDE 2MP TFT LCD Monitor System IF1902A ™s intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "WIDE" in large, bold, white letters against a black background. Below the word "WIDE" is smaller text that reads "World Innovative Leading Engineering". The text is also white against the black background. The image appears to be a logo or branding element.
K$\phi$3296$\phi$
Technological Characteristics: 21 CFR 807 92(a)(6)
The device is an image display system consisting of computer software and components. The device does not contact the patient, nor does it control any life sustaining devices. A physician or trained medical practitioner provides ample opportunity for competent human intervention to interpret images and information being displayed.
Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for the WIDE 2MP TFT LCD Monitor System IF1902A ™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
The WIDE 2MP TFT LCD Monitor System IF1902A ™will be manufactured by WIDE Corporation in accordance with the voluntary and safety standards, i.e. Safety / Immunity UL2601-1/EN60601-1 / IEC601-1, FCC Class B, CE, VCC, UL 950, cUL2601-1, CE Mark EMC/IEC = VCCI, CE, MIC FCC Class B digital device, pursuant to Part 15.
The submission contains the results of a hazard analysis and the potential hazards have been classified as Minor.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the top half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 8 2003
WIDE Corporation % Mr. Ned Devine Responsible Third Party Entela, Inc. 3033 Madison Ave. SE GRAND RAPIDS MI 49548 Re: K032960
Trade/Device Name: WIDE 2MP TFT LCD Monitor Systems IF 1902A™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: August 25, 2003
Received: September 23, 2003
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|---|---|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of -1
(Indications for Use Form)
510(k) Number:
K∅3296∅
Device Name: WIDE 2MP TFT LCD Monitor System IF1902A ™
Indications for Use:
The 2MP TFT LCD Monitor System IF1902A ™ is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Narieyt Brogdon
(Division Sign Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number
K032960
N/A