K Number
K032882

Validate with FDA (Live)

Date Cleared
2004-01-28

(135 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Recom Model 100 battery-operated Ambulatory, Digital Wireless ECG Monitor System is a true 12 lead ambulatory ECG monitor system capable of recording up to 48 hours of ECG data for the purpose of cardiac monitoring and diagnosis by a medical professional where ambulatory (Holter) monitoring is prescribed by a physician. The system includes recording of ECG data from each of the 12 leads, RF transmission of the recorded data to a PDA and storage.

The stored ECG data is then transferred from the PDA to another device, the Phillips Medical TraceMaster ECG Analysis System for ECG analysis. The Recom Model 100 battery-operated, Ambulatory, Digital, Wireless ECG Monitor System is not intended to sound any physiological alarms.

The Recom Model 100 battery-operated, Ambulatory, Digital Wireless ECG Monitor System is only to be used by or on the order of a physician.

Device Description

The device is a true 12 lead, battery-operated, ambulatory (Holter), digital, wireless, ECG monitor. The device acquires an ECG signal from 10 patient surface contact electrodes. Each electrode is connected to a cable that is connected to the Recom patient worn monitor/transmitter. The ECG signal is transmitted from the electrodes via the cable to a battery-operated ECG monitor unit worn by the patient where the signal is processed utilizing the company's signal processing technology and transmitted via radio frequency (Bluetooth) to a hand held Personal Digital Assistant (PDA) capable of recording up to 48 hours of ECG data.

The PDA converts the signal stream into a specific data format where it is stored on a compact flash card for transmission to and analysis by another device, the Phillips Medical TraceMaster ECG Analysis System K032103.

Monitoring data is not real time but is recorded and stored for further analysis by the Phillips System. The ECG data produced consists of Leads I, II, III, aVR, aVL, aVF and V1 through V6. The Recom Model 100 Ambulatory, Digital, Wireless monitor is not intended to alarm.

The fundamental technology of the Recom battery-operated, Ambulatory. Digital ECG Monitor System is the same as that of the predicate devices. The Recom Ambulatory, Digital, Wireless ECG Monitor System relies on the company's signal processing technology to minimize noise in the ECG signal created by the ambulatory nature of the signal source.

AI/ML Overview

The provided text describes a 510(k) summary for the Recom Model 100 ECG Monitor System, which is a Class II medical device. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than detailed performance studies with acceptance criteria in the way a novel, high-risk device might.

Here's an analysis of the provided text with respect to your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state quantitative acceptance criteria for a performance study evaluating the device's diagnostic accuracy or a specific clinical endpoint. Instead, the "acceptance criteria" appear to be related to the device's ability to accurately and reliably produce ECG tracings that can be analyzed by a separate, already cleared system (Phillips Medical TraceMaster ECG Analysis System).

Acceptance Criteria (Inferred)Reported Device Performance
Compliance with pertinent recognized industry standards for Ambulatory, Digital, Wireless Electrocardiograph devices (e.g., ANSI/AAMI EC-38, ANSI/AAMI EC-11, IEC 60601-1-2, IEC 60601-1-2-27, UL 2601-1, FCC Part 15)."Prior to marketing, the proposed Recom Ambulatory, Digital Wireless. ECG Monitor System will comply with all pertinent recognized industry standards for Ambulatory, Digital, Wireless Electrocardiograph devices." (Implied compliance, not direct performance data in this document.)
Ability to accurately and reliably produce ECG tracings for analysis by the Phillips Medical TraceMaster ECG Analysis System."Performance testing demonstrates that ECG tracings can be accurately and reliably produced by the Phillips Medical Systems TraceMaster ECG Management System from the file stored on the storage memory card of the PDA."
Device operates similarly to predicate devices without raising new questions of safety or effectiveness."The fundamental technology of the Recom battery-operated, Ambulatory. Digital ECG Monitor System is the same as that of the predicate devices.""An examination of the differences between the Recom Model 100 battery-operated, Ambulatory, Digital, Wireless ECG Monitor System does not raise new questions of safety or effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for a test set for performance or clinical accuracy. The "performance testing" mentioned is described as demonstrating the accurate and reliable production of ECG tracings, but details about the number of cases or subjects in this testing are not provided. There is no information about the country of origin of data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. Since the device is a data acquisition system and not a diagnostic algorithm itself, the testing described appears to focus on data fidelity and transfer rather than direct diagnostic accuracy requiring expert ground truth in a clinical trial setting. The "diagnosis by a medical professional" occurs after the data is analyzed by the Phillips system.

4. Adjudication Method for the Test Set

This information is not provided.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC study or any AI component in the Recom Model 100. This device is an ambulatory ECG monitor system for data acquisition and transfer, not an AI-powered diagnostic tool. The analysis is done by a separate Phillips Medical TraceMaster ECG Analysis System, which is also not described as AI-based in this document.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the Recom Model 100 is an ECG monitoring device, not a standalone diagnostic algorithm. Its function is to record and transmit data for subsequent analysis by another system (and ultimately, a human physician).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The document does not describe the use of specific clinical "ground truth" (like expert consensus or pathology) for the performance testing mentioned. The testing appears to confirm the technical fidelity of the recorded ECG data and its compatibility with the Phillips analysis system. The "ground truth" for cardiac monitoring and diagnosis is implicitly established by the "medical professional where ambulatory (Holter) monitoring is prescribed by a physician" after the data is analyzed by the Phillips system.

8. The Sample Size for the Training Set

This is not applicable. The Recom Model 100 is a hardware device for data acquisition and transmission, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established

This is not applicable as there is no training set for this device.

{0}------------------------------------------------

JAN 2 8 2004

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

1. Sponsor:

Recom Managed Systems, Inc. 4705 Laurel Canyon Blvd. #203 Valley Village, CA 91607 Phone (818) 432-4560 Facsimile (818) 432-4566

Contact Person: William R. Matthews, Consultant

Date Prepared: September 15, 2003

Recom Model 100, battery-operated Ambulatory, Digital, Wireless 2. Device Name: ECG Monitor System

Trade or Proprietary Name: Recom Model 100, battery-operated, Ambulatory, Digital, Wireless ECG Monitor System

Common or Usual Name: Ambulatory Electrocardiograph Recorder

Classification Name:

  • 21 CFR 870.2800 Electrocardiograph, Ambulatory . (without analysis)
  • 21 CFR 870.2910 Radiofrequency Physiological Signal . Transmitter and Recciver

3. Predicate Devices:

The Recom Model 100, battery-operated, Ambulatory, Digital, Wireless ECG Monitor is equivalent to:

  • CardioNet Ambulatory ECG Monitor, CardioNet, Inc., . K003707
  • . Mortara Ambulatory X-12 Telemetry Module, Mortara Instruments, Inc. K974149

4. Intended Use:

The Recom Model 100 battery-operated Ambulatory, Digital Wireless, ECG Monitor System is a true 12 lead ambulatory ECG monitor system capable of recording up to 48 hours of ECG data for the purpose of cardiac monitoring and diagnosis by a medical professional where ambulatory (Holter) monitoring is prescribed

{1}------------------------------------------------

Recom Managed Systems, Inc 510(k) Summary of Safety and Effectiveness

by a physician. The system includes recording of ECG data from each of the 12 leads, RF transmission of the recorded data to a PDA and storage.

The stored ECG data is then transferred from the PDA to another device, the Phillips Medical TraceMaster ECG Analysis System for ECG analysis. The Recom Model 100 battery-operated, Ambulatory, Digital, Wireless ECG Monitor System is not intended to sound any physiological alarms.

The Recom Model 100 battery-operated, Ambulatory, Digital Wireless ECG Monitor System is only to be used by or on the order of a physician.

5. Device Description:

The device is a true 12 lead, battery-operated, ambulatory (Holter), digital, wireless, ECG monitor. The device acquires an ECG signal from 10 patient surface contact electrodes. Each electrode is connected to a cable that is connected to the Recom patient worn monitor/transmitter. The ECG signal is transmitted from the electrodes via the cable to a battery-operated ECG monitor unit worn by the patient where the signal is processed utilizing the company's signal processing technology and transmitted via radio frequency (Bluetooth) to a hand held Personal Digital Assistant (PDA) capable of recording up to 48 hours of ECG data.

The PDA converts the signal stream into a specific data format where it is stored on a compact flash card for transmission to and analysis by another device, the Phillips Medical TraceMaster ECG Analysis System K032103.

Monitoring data is not real time but is recorded and stored for further analysis by the Phillips System. The ECG data produced consists of Leads I, II, III, aVR, aVL, aVF and V1 through V6. The Recom Model 100 Ambulatory, Digital, Wireless monitor is not intended to alarm.

The fundamental technology of the Recom battery-operated, Ambulatory. Digital ECG Monitor System is the same as that of the predicate devices. The Recom Ambulatory, Digital, Wireless ECG Monitor System relies on the company's signal processing technology to minimize noise in the ECG signal created by the ambulatory nature of the signal source.

{2}------------------------------------------------

Ko32882 3 of 3

6. Basis for Substantial Equivalence:

The Recom battery-operated, Ambulatory, Digital, Wireless ECG Monitor, the CardioNet Ambulatory ECG Monitor and the Mortara X-12 Telemetry Module are all intended to collect Ambulatory ECG data in a digital format by a wireless means for analysis.

Like the predicate devices, the Recom Model 100 Ambulatory, Digital, Wireless ECG Monitor System records ECG data in an ambulatory setting and transmits the data by radiofrequency means for analysis at another location.

An examination of the differences between the Recom Model 100 battery-operated, Ambulatory, Digital, Wireless ECG Monitor System does not raise new questions of safety or effectiveness.

Prior to marketing, the proposed Recom Ambulatory, Digital Wireless. ECG Monitor System will comply with all pertinent recognized industry standards for Ambulatory, Digital, Wireless Electrocardiograph devices. These include ANSI/AAMI EC-38, ANSI/AAMI EC-11, IEC 60601-1-2, IEC 60601-1-2-27 and UL The communications of the monitor will meet the 2601-1. applicable requirements of FCC Part 15.

Performance testing demonstrates that ECG tracings can be accurately and reliably produced by the Phillips Medical Systems TraceMaster ECG Management System from the file stored on the storage memory card of the PDA. The technical differences between the proposed Recom device and the predicate devices do not raise any new concerns of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 8 2004

Recom Managed Systems, Inc. c/o Mr. William R. Matthews Regulatory Affairs Consultant 4705 Laurel Canyon Blvd. #203 Valley Village, CA 91607

Re: K032882

Trade Name: Recom Model 100, Battery-Operated, Ambulatory, Digital Wireless ECG Monitor System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: MWJ Dated: November 5, 2003 Received: November 5, 2003

Dear Mr. Matthews:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. William R. Matthews

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s Issualite of a substitution with other requirements of the Act
that FDA has made a determination that your device of the requirements of the Act that FDA has made a decemination that your ateries by other Federal agencies. You must or any Federal statutes and regulations administers of registration and listing (21 l
eomply with all the Act's requirements, including, but not is requirements as set comply with all the Act's requirements, including, barticle requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements, the elect CFR Part 807); labeling (21 CFX Fall 801); good manatt820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Path 2001-1050 forth in the quality systems (QS) regulation (21-51-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Action Section 5 product radiation control provisions (Sections 951 device as described in your Section 510(k)
This letter will allow you to begin marketing your device of your device to a le This letter will allow you to begin marketing your actrice of your device to a legally
premarket notification. The FDA finding of substantial equivalence of this nermits vo premarket notification. The FDA iniding of substantial equive and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rstilled, If you desire specific advice for your ac 110 metro at 100 please note the regulation entitled, contact the Office of Compilance at (301) 574 10 10. FR Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtai "Misbranding by reference to prematics noutheation (ce Act from the Division of Small other general information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-fremenain html Manufacturers, International and Consumer Fiscossaft and consumers.html

Sincerely yours,

Whane R. bichner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Recom Model 100, battery-operated, Ambulatory, Digital Wireless Device Name: ECG Monitor System

Indications for Use:

The Recom Model 100 battery-operated Ambulatory, Digital Wireless ECG Monitor System is a true 12 lead ambulatory ECG monitor system capable of recording up to 48 hours of ECG data for the purpose of cardiac monitoring and diagnosis by a medical professional where ambulatory (Holter) monitoring is prescribed by a physician. The system includes recording of ECG data from each of the 12 leads, RF transmission of the recorded data to a PDA and storage.

The stored ECG data is then transferred from the PDA to another device, the Phillips Medical TraceMaster ECG Analysis System for ECG analysis. The Recom Model 100 battery-operated, Ambulatory, Digital, Wircless ECG Monitor System is not intended to sound any physiological alarms.

The Recom Model 100 battery-operated, Ambulatory, Digital Wircless ECG Monitor System is only to be used by or on the order of a physician.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. Licliner
Division of Cardiovascular & Respiratory Devices
510(k) Number K032882

Prescription Use X (Per 21 C.F.R. 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).