K Number
K032783

Validate with FDA (Live)

Date Cleared
2003-11-13

(66 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powdered vinyl patient examination gloves, white (non-colored) that meets all of the requirements of ASTM standard D 5250-004.

AI/ML Overview

Acceptance Criteria and Study for Blue Sail (Brand) Powdered Vinyl Patient Examination Gloves

This document summarizes the acceptance criteria and the study conducted to demonstrate that the Blue Sail (Brand) Powdered Vinyl Patient Examination Gloves meet these criteria, as described in the provided 510(k) summary (K032783).

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-00e4Meets
Physical PropertiesASTM standard D 5250-00e4Meets
Freedom from pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-00e4Meets
Biocompatibility - Primary Skin IrritationPrimary Skin Irritation in rabbits (Passes: Not a Primary Skin Irritation)<10mg/dm2; Passes; Not a Primary Skin Irritation
Biocompatibility - Dermal SensitizationDermal sensitization in the guinea pig (Passes: Not a Dermal sensitization)Passes; Not a Dermal sensitization

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not specify the sample sizes used for the individual tests (e.g., number of gloves tested for dimensions, physical properties, or pinholes).

The data provenance is implicit: the tests were conducted by the manufacturer, Zibo Blue Sail Chemical Co., Ltd., presumably in P.R. China, to demonstrate compliance with the specified ASTM standards and FDA regulations. The studies appear to be retrospective in nature, as they involve testing a manufactured product against established standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable. The ground truth for this device is based on objective, quantifiable physical and chemical parameters defined by established industry standards (ASTM) and federal regulations (CFR). There is no mention of human experts being used to establish a "ground truth" for the performance of the gloves beyond interpreting the results of standardized tests.

4. Adjudication Method for the Test Set

Not applicable. Since the tests are based on objective measurements against defined standard limits, an adjudication method for conflicting expert opinions is not relevant. The pass/fail criteria are directly derived from the standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted for this device. This type of study is typically used for diagnostic or interpretive medical devices where human readers (e.g., radiologists) interpret images or data. Patient examination gloves are a physical barrier device, and their effectiveness is assessed through physical and biocompatibility testing, not through human interpretation of medical data.

6. Standalone Performance Study (Algorithm Only)

Not applicable. This device is a medical glove, not an algorithm or software. Therefore, the concept of a "standalone (algorithm only)" performance study does not apply. The performance evaluation focuses on the physical and chemical characteristics of the glove itself.

7. Type of Ground Truth Used

The ground truth used for evaluating this device is based on objective, quantitative standards and regulatory requirements. Specifically:

  • ASTM standards (D 5250-00e4 and presumably D6124-01): These standards define the acceptable ranges for physical properties (dimensions, tensile strength, elongation, puncture resistance) and powder amount.
  • 21 CFR 800.20: This regulation defines the acceptable quality level (AQL) for freedom from pinholes (water leak test).
  • ISO 10993-10: This international standard covers biological evaluation of medical devices, specifically for irritation and sensitization.

Essentially, the "ground truth" is defined by the pass/fail criteria specified within these published standards and regulations.

8. Sample Size for the Training Set

Not applicable. This device is a physical product (medical glove), not an AI/ML algorithm or system that requires a "training set" in the context of machine learning. The manufacturing process is subject to quality control, but there is no "training set" of data for an algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As established in point 8, there is no "training set" for this device. The manufacturing and testing processes follow established quality management systems and standard operating procedures to ensure consistent product quality, but this is distinct from establishing ground truth for machine learning training.

{0}------------------------------------------------

C Summary

NOV 1 3 2003

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: 1032783 ." (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :Zibo Blue Sail Chemical Co., Ltd
Submitter's address :Qilu Chemical Industrial Area, Zibo City, Shandong Province, 255414, P.R.China
Phone number :(86)533-7486199
Fax number :(86)533-7481177
Name of contact person:Liu Wenjing
Date the summary was prepared:3 September 2003

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powdered Vinyl Patient Examination Gloves, White (non-colored)
Proprietary/Trade name:Blue Sail (Brand) Powdered Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powdered vinyl patient examination gloves , white (non-colored) that meets all of the requirements of ASTM standard D 5250-0064.

Predicate device : powdered synthetic vinyl patient examination gloves, Shijiazhuang Eversharp Plastics Products Co., Ltd. K011884 .

[(a)(4)} A description of the device

Device Description : powdered vinyl patient examination gloves, white (non-colored) that meets all of the requirements of ASTM standard D 5250-004.

Page 1of 2 Section C

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains a sequence of handwritten alphanumeric characters. The characters are 'K032783'. The characters appear to be written with a dark ink or marker on a white background. The writing style is cursive.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: powdered vinyl patient examination glove, white (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powdered vinyl patient examination gloves, white (non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-00e4.Meets
Physical PropertiesASTM standard D 5250-00e4.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-00e4Meets
BiocompatabilityPrimary Skin Irritation in rabbits<10mg/dm2PassesNot a Primary Skin Irritation
Dermal sensitization in the guinea pigPassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powdered vinyl patient examination gloves , white (non-colored) meet requirements per ASTM D5250-0064, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Blue Sail (Brand) Powdered Vinyl Patient Examination Gloves , White (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 3 2003

Zibo Blue Sail Chemical Corporation Limited C/O Ms. Yuhong Chen Tuv Rheinland Beijing Office Rm 707, AVIC Bldg, No. 2 No. 2 Dong San Huan Nan Road Chaoyang District, Beijing, 100022, P.R. CHINA

Re: K032783

Trade/Device Name: Blue Sail (Brand) Powdered Vinyl Patient Examination gloves, White (Non-Colored) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 3, 2003 Received: September 8, 2003

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Yuhong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Chue L

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

3.0 Indications for Use Statement: Include the following or equivalent Indications for Use page. The information, data and labeling claims in the entire 510(k) submission must support and agree with the Indications for Use statement.

INDICATIONS FOR USE

Applicant: Zibo Blue Sail Chemical Co.,Ltd.

510(k) Number (if known): * KO32783 Blue Sail (Brand) Device Name: Powdered Vinyl Patient Examination Gloves, White(non-colored)

Indications For Use:

Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Susanna F. Bourud

nesthesiology, General Hospital, Infection Control. Dental

510(k) Number:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.