K Number
K032541

Validate with FDA (Live)

Device Name
POLESTAR N-20
Date Cleared
2003-09-12

(25 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.

Anatomical regions: extremities and selected sections of the head.

Nuclei excited: H-1

Diagnostic uses: T1, T2, T2* and density weighted imaging.

The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.

Device Description

The PoleStar N-20 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N-20 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N-20 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.

AI/ML Overview

This document is a 510(k) summary for the Odin Medical Technologies PoleStar N-20, an intraoperative MRI device. The primary claim for substantial equivalence is based on a comparison to the predicate device, Odin's PoleStar N-10. This filing does not describe a study to prove acceptance criteria for device performance. Instead, it focuses on demonstrating that the PoleStar N-20 is substantially equivalent to an existing device (the PoleStar N-10) based on similar technical specifications and identical intended use.

Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and statistical analysis is not present in the provided text.

Here's a breakdown of what can be extracted and what is missing:

1. A table of acceptance criteria and the reported device performance:

The document doesn't provide explicit "acceptance criteria" for performance that would typically be seen in a clinical validation study for a new diagnostic device. Instead, it offers a "Substantial Equivalence Comparison Table" that lists technical specifications of the new device (PoleStar N-20) against its predicate (PoleStar N-10). The "performance" in this context refers to these specifications matching or being comparable to the predicate.

ParameterPoleStar N-10 (Predicate, K010850)PoleStar N-20 (New Device)Performance/Comparison
Clinical applicationExtremities and selected sections of the headExtremities and selected sections of the headIdentical
Magnet typePermanentPermanentIdentical
Field strength0.12T0.13TSlight increase, deemed comparable
5 gauss fringe field (radial/axial, m)1.52.2Minor difference, likely considered acceptable for similar use
ShimmingPassive, activePassive, activeIdentical
Gradient subsystem Strength mT/m2522Minor difference, likely considered acceptable
Gradient subsystem Rise time to 10mT/m msec<1<1Identical
Computer system - CPUPentium 586Pentium 586Identical
Computer system - Memory size [MB]6464Identical
Array processor4xDSP C44 TI4xDSP C44 TIIdentical
Array processor - Memory size [MB]40004000Identical
Storage mediaMagnetic disk, floppy diskMagnetic disk, floppy diskIdentical
Number of images stored50005000Identical
Imaging modesSingle, multislice, volume studySingle, multislice, volume studyIdentical
Reconstruction time - single slice, sec<3/slice<3/sliceIdentical
Reconstruction time - multislice, sec<3/slice<3/sliceIdentical
Reconstruction time - volume sec<20/volume<20/volumeIdentical
Cardiac gating (ECG/peripheral)NoNoIdentical
Respiratory gatingNoNoIdentical
AngiographyOptionalOptionalIdentical
SpectroscopyNoNoIdentical
Imaging - pulse sequenceSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DIdentical
Imaging - repetition time, msec10-5000 increments of 110-5000 increments of 1Identical
Imaging - echo time, msec3-1503-150Identical
Imaging - inversion time, msecN/AN/AIdentical
Imaging - slice thickness, mm2-102-10Identical
Imaging - scan orientationTransverse, coronal, sagittal, obliqueTransverse, coronal, sagittal, obliqueIdentical
Imaging - measuring matrix64x64 to 256x256 steps of 1 in phase encoding64x64 to 256x256 steps of 1 in phase encodingIdentical
Imaging - display matrix1024x7681024x768Identical
Imaging - pixel intensity0-40950-4095Identical
Surface coilsKnee, extremity, headKnee, extremity, headIdentical
Bore diameter or WxH, cm24.5x3925.2x42Minor difference, likely considered acceptable
Bore featuresOpen access to patientOpen access to patientIdentical
Cooling system typeClosed loop water cooling (Gradients only)Closed loop water cooling (Gradients only)Identical
Cryogen useNoNoIdentical
Magnet weight, kg330400Minor difference, likely considered acceptable
HxWxD, cm145x96x120153x97x120Minor difference, likely considered acceptable
Field Of View (FOV), cm5-185-20Minor difference, likely considered acceptable
Dicom 3.0 interfaceYesYesIdentical
Power requirements3x208V, 16 Kva, <10000 BTU/hr A/C3x208V, 15 Kva, <10000 BTU/hr A/CMinor difference, likely considered acceptable

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Not Applicable. This document does not describe a clinical study with a test set of data. The substantial equivalence is based on technical specifications and intended use comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. No ground truth establishment is described as there is no clinical test set in this filing. The filing mentions that "When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis," which refers to the eventual use of the device, not a study performed for this 510(k).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. No test set or adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This document describes an MRI diagnostic device, not an AI or CAD system, and therefore, an MRMC study related to AI assistance would not be relevant or expected for this device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This is a hardware device (MRI scanner) which requires human operation and interpretation. The concept of standalone algorithm performance does not apply.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. As no clinical study or test set is described, no ground truth was established for this 510(k filing for the PoleStar N-20.**

8. The sample size for the training set:

  • Not Applicable. This document does not describe a machine learning or AI-based device that would require a training set.

9. How the ground truth for the training set was established:

  • Not Applicable. No training set or ground truth for a training set is described.

In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence of the PoleStar N-20 to its predicate, the PoleStar N-10, by comparing technical specifications and intended use. It does not include information about performance acceptance criteria validated through a clinical study with a test set, expert ground truth establishment, or AI-related metrics.

{0}------------------------------------------------

SEP 1 2 2003

Image /page/0/Picture/2 description: The image shows the logo for Odin Medical Technologies. The logo features the word "ODIN" in bold, sans-serif font, with the words "Medical Technologies" in a smaller font size underneath. A stylized, incomplete circle or ring is positioned to the right of the word "ODIN", with a small, solid circle placed at the top of the ring, giving the impression of a planet with a ring system. The overall design is clean and modern.

Special 510(k) Summary of Safety and Effectiveness

The Following Special 510(k) Summary of Safety and Effectiveness has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR § 807.92(a).

807.92(a)(1) - Submitter Details:

Submitter name:Address:Phone:Fax:E-mail:Einav Shlomovitz -- Quality and Regulatory DirectorP.O. Box 548, Yokneam Elit 20698, Israel+972-4-959-1010+972-4-959-1011einav@odin.co.il
Contact Person:Einav Shlomovitz -Quality and Regulatory Director
Date:August 4, 2003

807.92(a)(2) - Device Details:

Trade Name and Common Name:PoleStar N-20 - Magnetic ResonanceDiagnostic Device
Classification:21 CRF 892.1000 Magnetic ResonanceDiagnostic Device.
Class:IIMRDD were reclassified by FDA fromClass III to Class II effective July 28,1998.
Product Code:LNH – Magnetic Resonance ImagingSystem

{1}------------------------------------------------

807.92(a)(3) -- Predicate Devices:

The PoleStar N-20 is comparable to Odin's PoleStar N-10

MedicalDevice NameApplicant Name510(k)NumberClassification
PoleStar N-10 Odin Medical Thechnologies Ltd.K010850l Class Il device

Additional Substantial Equivalence Information is provided in the attached Substantial Equivalence Comparison Table.

807.92(a)(4) - Device Description:

The PoleStar N-20 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N-20 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N-20 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.

{2}------------------------------------------------

807.92(a)(5) - Device Intended Use:

The general purpose of the device as defined in 21 CFR 892.1000:

The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.

  • . Anatomical regions:
    Diagnostic uses:

extremities and selected sections of the head.

  • . Nuclei excited:
    .

T1, T2, T2* and density weighted imaqing.

The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.

H-1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in bold, sans-serif font, with the "o" slightly larger than the other letters. Below "odin" is the phrase "Medical Technologies" in a smaller, italicized font. A circular design element is to the right of the word "odin".

807.92(a)(6) – Substantial Equivalence Comparison Table:

ModelOdinOdin
parameterPoleStar N-10 (K010850)PoleStar N-20
Clinical applicationExtremities and selectedsections of the headExtremities and selectedsections of the head
Magnet typePermanentPermanent
Field strength0.12T0.13T
5 gauss fringe field(radial/axial, m)1.52.2
ShimmingPassive, activePassive, active
Gradient subsystem
Strength mT/m2522
Rise time to 10mT/m<1<1
msec
Computer system
- CPU:Pentium 586Pentium 586
- Memory size [MB]6464
array processor4xDSP C44 TI4xDSP C44 TI
- Memory size [MB]40004000
storage mediamagnetic disk, floppy diskmagnetic disk, floppy disk
number of images stored50005000
Imaging modes:
- singleYesYes
- multisliceYesYes
- volume studyYesYes
- otherNoNo
Reconstruction time:
- single slice, sec<3/slice<3/slice
- multislice, sec<3/slice<3/slice
- volume sec<20/volume<20/volume
ModelOdinOdin
parameterPoleStar N-10 (K010850)PoleStar N-20
Cardiac gating(ECG/peripheral)NoNo
Respiratory gatingNoNo
AngiographyOptionalOptional
SpectroscopyNoNo
Imaging:
- pulse sequenceSpin Echo, Fast Spin Echo,Gradient Echo, 2D 3DSpin Echo, Fast Spin Echo,Gradient Echo, 2D 3D
- repetition time, msec10-5000 increments of 110-5000 increments of 1
- echo time, msec3-1503-150
- inversion time, msecN/AN/A
- slice thickness, mm2-102-10
- scan orientationTransverse, coronal,sagittal, obliqueTransverse, coronal,sagittal, oblique
- measuring matrix64x64 to 256x256 steps of1 in phase encoding64x64 to 256x256 steps of1 in phase encoding
- display matrix1024x7681024x768
- pixel intensity0-40950-4095
Surface coils:
- spineNoNo
- kneeYesYes
- neckNoNo
- TMJNoNo
- extremityYesYes
- headYesYes
- breastNoNo
- shoulderNoNo
- othersNoNo
ModelOdin
parameterPoleStar N-10 (K010850)PoleStar N-20
Bore diameter or WxH, cm24.5x3925.2x42
Bore featuresOpen access to patientOpen access to patient
Cooling system typeClosed loop water cooling(Gradients only)Closed loop water cooling(Gradients only).
Cryogen useNoNo
Magnet weight, kg330400
HxWxD, cm145x96x120153x97x120
Filed Of View (FOV),cm5-185-20
Dicom 3.0 interfaceYesYes
Power requirements:
- line voltage, V3x208 (3 phase)3x208 (3 phase)
- Kva1615
- A/C, BTU/hr<10000<10000

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in bold, black letters, with the words "Medical Technologies" in a smaller font below. There is a circular design to the right of the word "odin".

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in large, bold, sans-serif font, with a stylized circle above the "i". Below "odin" are the words "Medical Technologies" in a smaller, sans-serif font. A partial circle is behind the word "odin".

·

{6}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 2 2003

Mr. Einav Shlomovitz Ouality and Regulatory Director Odin Medical Technologies Ltd. P.O. Box 548 Yokneam Elit, 20698 ISRAEL

Re: K032541

Trade/Device Name: PoleStar N-20 Regulatory Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: August 4, 2003 Received: August 18, 2003

Dear Mr. Shlomovitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301).594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

510(k) Number (if known):

Device Name: PoleStar N-20

Indication For Use:

The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.

Anatomical regions: extremities and selected sections of

the head.

Nuclei excited:H-1
Diagnostic uses:T1, T2, T2* and density weighted
imaging.

The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.

(PLEASE DO NOT WRITE BELOW THE LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Daniel A. Bergman

(Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number

N/A