Intended Use

The ArthroCare System is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist)
Ablation and Debridement
• ACL/PCLKnee
• AcromioplastyShoulder
• Articular CartilageAll Joints
• BursectomyAll Joints
• ChondroplastyAll Joints
• FaciaAll Joints
• LigamentAll Joints
• NotchplastyKnee
• Scar TissueAll Joints
• Soft TissueAll Joints
• Subacromial DecompressionShoulder
• SynovectomyAll Joints
• TendonAll Joints
Excision and Resection
• Acetabular LabrumHip
• Articular LabrumAll Joints
• CapsuleAll Joints
• Capsular ReleaseKnee
• Cartilage FlapsKnee
• CystsAll Joints
• Discoid MeniscusKnee
• Frozen Shoulder ReleaseShoulder
• Glenoidale LabrumShoulder
• Lateral ReleaseKnee
• LigamentAll Joints
• Loose BodiesAll Joints
• Meniscal CystectomyKnee
• MeniscectomyKnee
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee
Coagulation
• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist
Device Description

The ArthroCare System is a bipolar, high frequency electrosurgical system consisting of three components: an electrosurgical generator called the Controller; a family of disposable, bipolar, single use Wands; and a reusable Patient Cable.

AI/ML Overview

This document is a 510(k) summary for the ArthroCare System, an electrosurgical device. It details the device's intended use, its substantial equivalence to a predicate device, and confirms that modifications do not significantly affect safety or efficacy. However, the provided text does not contain any information about acceptance criteria or a study proving that the device meets specific acceptance criteria related to its performance metrics.

The document discusses:

  • Device Description: A bipolar, high-frequency electrosurgical system comprising a Controller, disposable Wands, and a reusable Patient Cable.
  • Intended Uses: Resection, ablation, and coagulation of soft tissue, and hemostasis of blood vessels in various arthroscopic and orthopedic procedures across multiple joints.
  • Substantial Equivalence: The current submission (K032504) proposes modifications to performance specifications, dimensional specifications, and labeling of an already cleared device (K030551). It explicitly states that "The indications for use, technology, principle of operation, materials, packaging, and sterilization parameters of the ArthroCare System remain the same as in the predicate cleared 510(k)." The document concludes that the modifications are not substantial and do not significantly affect safety or efficacy.

Therefore, based on the provided text, I cannot extract the requested information regarding acceptance criteria and a study demonstrating the device meets those criteria. The document focuses on regulatory approval based on substantial equivalence, not on a detailed performance study with acceptance criteria.

{0}------------------------------------------------

AUG 2 1 2003

Image /page/0/Picture/1 description: The image shows the logo for ArthroCare Corporation. The logo features a stylized letter "A" followed by the text "rthroCare" in a serif font. Below the word "ArthroCare" is the word "CORPORATION" in a smaller, sans-serif font.

Page ① of 3 pages

510(k) Summary

ArthroCare Corporation ArthroCare System

K 032504

2951580

General Information

Submitter Name/Address:

ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-3523

Director, Regulatory Affairs

Establishment Registration Number:

Contact Person:

Date Prepared:

Device Description

Trade Name:

Generic/Common Name:

Classification Name:

ArthroCare® System

Valerie Defiesta-Ng

August 12, 2003

Electrosurgical Device and Accessories

Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400)

Predicate Devices ArthroCare® System

K030551

Product Description

The ArthroCare System is a bipolar, high frequency electrosurgical system consisting of three components: an electrosurgical generator called the Controller; a family of disposable, bipolar, single use Wands; and a reusable Patient Cable.

{1}------------------------------------------------

Page ② of 3 Pages

Intended Uses

The ArthroCare System is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
Ablation and Debridement
• ACL/PCLKnee
• AcromioplastyShoulder
• Articular CartilageAll Joints
• BursectomyAll Joints
• ChondroplastyAll Joints
• FaciaAll Joints
• LigamentAll Joints
• NotchplastyKnee
• Scar TissueAll Joints
• Soft TissueAll Joints
• Subacromial DecompressionShoulder
• SynovectomyAll Joints
• TendonAll Joints
Excision and Resection
• Acetabular LabrumHip
• Articular LabrumAll Joints
• CapsuleAll Joints
• Capsular ReleaseKnee
• Cartilage FlapsKnee
• CystsAll Joints
• Discoid MeniscusKnee
• Frozen Shoulder ReleaseShoulder
• Glenoidale LabrumShoulder
• Lateral ReleaseKnee
• LigamentAll Joints
• Loose BodiesAll Joints
• Meniscal CystectomyKnee
• MeniscectomyKnee

{2}------------------------------------------------

Continued

Arthroscopic and Orthopedic ProceduresJoint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist)
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee

Coagulation

• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist

Substantial Equivalence

This Special 510(k) proposes a modification in the performance specifications. dimensional specifications, and labeling for the ArthroCare System, which was previously cleared in K030551 on March 7, 2003. The indications for use, technology, principle of operation, materials, packaging, and sterilization parameters of the ArthroCare System remain the same as in the predicate cleared 510(k).

Summary of Safety and Effectiveness

The modified ArthroCare System, as described in this Special 510(k), is substantially equivalent to the predicate device. The proposed modifications in performance specifications, dimensional specifications, and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the System.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird with its wings spread, formed by three human profiles facing to the right. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 1 2003

Ms. Valerie Defiesta-Ng Director, Regulatory Affairs ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, California 94085

Re: K032504

Trade/Device Name: ArthroCare® System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 12, 2003 Received: August 13, 2003

Dear Ms. Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Valerie Defiesta-Ng

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

for Mark N. Milliken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

Page ① 7 Page 6

ArthroCare System Device Name

K 032504 510(k) Number:

Indications for Use:

The ArthroCare System is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist)
• Ablation and Debridement
• ACL/PCLKnee
• AcromioplastyShoulder
• Articular CartilageAll Joints
• BursectomyAll Joints
• ChondroplastyAll Joints
• FaciaAll Joints
• LigamentAll Joints
• NotchplastyKnee
• Scar TissueAll Joints
• Soft TissueAll Joints
• Subacromial DecompressionShoulder
• SynovectomyAll Joints
• TendonAll Joints
Excision and Resection
• Acetabular LabrumHip
• Articular LabrumAll Joints
• CapsuleAll Joints
• Capsular ReleaseKnee
• Cartilage FlapsKnee
• CystsAll Joints
• Discoid MeniscusKnee
• Frozen Shoulder ReleaseShoulder
• Glenoidale LabrumShoulder
• Lateral ReleaseKnee
• LigamentAll Joints
• Loose BodiesAll Joints
• Meniscal CystectomyKnee
• MeniscectomyKnee
  • Mills f General. R
    logical Dev ion Sign-

(k) Number

{6}------------------------------------------------

K 0 3 2 5 0 4

Page ② of Page 2

Continued

Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee
Coagulation
• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801.109)

OR

Over-the-Counter Use

Mark N. Milburn

Division Sign-Off) Division of General, Restorative ୍ୟୁ Neurological Devices

510(k) Number

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.