K Number
K032397

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-05-27

(297 days)

Product Code
Regulation Number
878.4810
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dual Yellow laser, operating at 578 nm, is indicated in dermatology, plastic surgery, and general surgery for treatment of benign cutaneous vascular lesions including but not limited to: Treatment of wrinkles, Periocular wrinkles, Periorbital wrinkles. Facial and leg telangiectasia, Rosacea, Cherry angiomas, Port wine stains, Hemangiomas and venous lakes, Angioma, Spider angioma, and Poikiloderma of Civatte, Inflammatory Acne Vulgaris, Verrucae/Warts, Scars. Striea, and Psoriasis. Podiatry - for benign cutaneous lesions and warts. The Dual Yellow lase, operating at 511 nm, is indicated in dermatology, plastic surgery, and general surgery for treatment of benign cutancous pigmented lesions including but not limited to: Lentigines, Solar keratoses. Adenoma serabaceum, Skin tabs, Trichoepitheliomas (benign lesions similar to skin tags) Naevi, Keratoses, Syringomas Seborrhoeic keratoses Verrucae vulgaris, and Warts.

Device Description

The Norseld Dual Yellow D10B Laser is a copper bromide laser which emits it energy at 511 and 578 nm. The device consists of an cabinet, fiber optic delivery system, and a user/software interface.

AI/ML Overview

The provided text describes a 510(k) summary for the Norseld Dual Yellow D10B laser, which is a medical device. As such, the concept of "acceptance criteria" and "device performance" in the context of this submission differs significantly from that of AI/ML-based diagnostic devices.

For a medical device like a laser, the "acceptance criteria" are related to its intended use and safety through substantial equivalence to predicate devices, rather than statistical performance metrics like sensitivity, specificity, or AUC against a ground truth. The "study" proving it meets acceptance criteria is primarily an a demonstration of safety and effectiveness based on this substantial equivalence.

Here's how to interpret the provided information in the context of the requested table, focusing on the available details:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from 510(k) context)Reported Device Performance (from 510(k) context)
Safety and Effectiveness for Benign Cutaneous Vascular LesionsThe Dual Yellow laser, operating at 578 nm, is indicated for the treatment of various benign cutaneous vascular lesions including, but not limited to: Treatment of wrinkles, Periocular wrinkles, Periorbital wrinkles, Facial and leg telangiectasia, Rosacea, Cherry angiomas, Port wine stains, Hemangiomas and venous lakes, Angioma, Spider angioma, Poikiloderma of Civatte, Inflammatory Acne Vulgaris, Verrucae/Warts, Scars, Stria, and Psoriasis. Podiatry - for benign cutaneous lesions and warts. Performance was supported by: The results of a published clinical study for the treatment of inflammatory acne vulgaris (578 nm). Equivalency to legally marketed predicate devices for other 578 nm indications.
Safety and Effectiveness for Benign Cutaneous Pigmented LesionsThe Dual Yellow laser, operating at 511 nm, is indicated for the treatment of various benign cutaneous pigmented lesions including, but not limited to: Lentigines, Solar keratoses, Adenoma serabaceum, Skin tabs, Trichoepitheliomas (benign lesions similar to skin tags), Naevi, Keratoses, Syringomas, Seborrhoeic keratoses, Verrucae vulgaris, and Warts. Performance was supported by: Supporting literature and articles. Equivalency to legally marketed predicate devices for 511 nm indications.
Substantial Equivalence to Predicate DevicesThe device was determined to be "substantially equivalent to the cited legally marketed predicates for the indications listed above." Predicate devices listed include ICN (and SLS BIOPHILE) NLite Laser Systems, Candela Vbeam, Norseld Dual Yellow Laser (K023899), Asclepion-Meditec YelloStar, COSMOS COMPACT KTP, Dio-Light/DioLite, Laserscope Lyra, and Viridis Laser.

Since this is a 510(k) submission for a laser device, the terms "sample size," "test set," "training set," "ground truth," and "experts" are not directly applicable in the way they are for AI/ML performance studies. The primary "study" is the demonstration of substantial equivalence to already approved predicate devices.

Here's an attempt to address the other points based on the provided text, noting where information is not available or not applicable:

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated as a "test set" in the context of AI/ML. The only mention of a specific study is for "inflammatory acne vulgaris." The number of subjects in that study is not provided.
  • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The text only mentions "a published clinical study" for one indication.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not applicable/not stated. The concept of "ground truth" established by experts for a test set, as in AI/ML performance, is not discussed. The evaluation relies on clinical outcomes and comparison to predicate devices, where expert judgment might have been involved in the original studies of the predicate devices or in the single clinical study mentioned.
  • Qualifications of Experts: Not applicable/not stated.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable/not stated.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: Not applicable. This is not an AI-assisted diagnostic device, but a therapeutic laser.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Standalone Performance: Not applicable. This is a therapeutic laser device, not an algorithm. Its performance is tied to its physical operation and clinical use by a human.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: For the "published clinical study" mentioned for inflammatory acne vulgaris, the "ground truth" would likely be clinical outcomes data (e.g., reduction in lesion count, improvement in severity scores) based on dermatologist assessment. For other indications, the "ground truth" relies on the established efficacy and safety of the predicate devices for those conditions.

8. The sample size for the training set

  • Sample Size: Not applicable. This is not an AI/ML device that uses a "training set."

9. How the ground truth for the training set was established

  • Ground Truth Establishment: Not applicable.

{0}------------------------------------------------

MAY 27 2004

SECTION 9

510(k) SUMMARY

KO32597

This 510(k) summary of safety and effectiveness for Norseld Dual Yellow D10B laser is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.

  • Applicant: Norseld Pty., Ltd.
    :

  • 9 Claxton St Address: Adelaide South Australia 5000

  • Manufacturer: Norseld Pty, Ltd. 9 Claxton St Adelaide South Australia 5000

  • Contact Person: Mr. Peter Davis Managing Director

  • Telephone Number: 011-8-8231 9000 Fax Number: 011-8-8231 9009

Preparation Date: January 2004 (of the Summary)

Dual Yellow Laser, Dual Yellow D10B Laser Device Name:

Common Name: Laser surgical device

Classification: Laser surgical device Class II medical device 21 CFR 878.4810

Product Code: GEX Panel: 79

Predicate devices: ICN (and SLS BIOPHILE) NLite Laser Systems (K000811, K020729, K013461, K014130); Candela Vbeam (K021180), and Norseld Dual Yellow Laser (K023899); Asclepion-Meditec YelloStar(K013940); COSMOS COMPACT KTP (K983020); Dio-Light/DioLite (K981447, K980201, K964074); Laserscope Lyra (K990903); and Viridis Laser (K001784).

{1}------------------------------------------------

K032397 Page 2 of 3

  • Device description: The Norseld Dual Yellow D10B Laser is a copper bromide laser which emits it energy at 511 and 578 nm. The device consists of an cabinet, fiber optic delivery system, and a user/software interface.
    The Dual Yellow laser is indicated for the treatment of benign pigmented Indications: and cutaneous vascular lesions.

The Dual Yellow laser, operating at 578 nm, is indicated in dermatology, plastic surgery, and general surgery for treatment of benign cutaneous vascular lesions including but not limited to:

Treatment of wrinkles. Periocular wrinkles, Periorbital wrinkles. Facial and leg telangiectasia, Rosacea, Cherry angiomas, Port wine stains, Hemangiomas and venous lakes, Angioma, Spider angioma, and Poikiloderma of Civatte, Inflammatory Acne Vulgaris, Verrucae/Warts, Scars. Strica, and Psoriasis.

Podiatry - for benign cutaneous lesions and warts.

The Dual Yellow lase, operating at 511 nm, is indicated in dermatology, plastic surgery, and general surgery for treatment of benign cutancous pigmented lesions including but not limited to:

Lentigines, Solar keratoses. Adenoma serabaceum, Skin tabs, Trichoepitheliomas (benign lesions similar to skin tags) Naevi, Keratoses, Syringomas Seborrhocic keratoses Verrucae vulgaris, and Warts.

{2}------------------------------------------------

K032397 Page 3 of 3

Supporting literature and articles were provided in support of the Performance Data: indications for use of the 511 nm wavelength.

The results of a published clinical study were submitted in support of the indication for the treatment of inflammatory acne vulgaris using the 578 nm wavelength. Indications for the 578 nm wavelength were based on equivalency to cited legally marketed products.

  • Based on the information in the notification Norseld Pty., Ltd. believes CONCLUSION: that Dual Yellow D10B Laser is substantially equivalent to the cited legally marketed predicates for the indications listed above..
    rev. 2/2004

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the HHS logo in the center. The HHS logo consists of a stylized caduceus, which is a symbol of medicine, with three wavy lines representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the perimeter of the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 7 2004

Norseld Pty., Ltd. c/o Mr. Roger Barnes 342 Sunset Bay Road Hot Springs, Arkansas 71913

Re: K032397 Trade/Device Name: Norseld Pty., Ltd. Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in gencral and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 24, 2004 Received: March 1, 2004

Dear Mr. Barnes:

We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosurc) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Roger Barnes

This letter will allow you to begin marketing your device as described in your Section 510(k) rmarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincercly vours.

Mark A Millburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

SECTION 7

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K032397_________________________________________________________________________________________________________________________________________

Device Name: Norseld Pty., Ltd. Dual Yellow (Dual Yellow D10B Laser)

Indications for Use Statement:

The Dual Yellow laser, operating at 578 nm, is indicated in dermatology, plastic surgery, The Daar Fenrery for treatment of benign cutaneous vascular lesions including but not limited to:

Treatment of wrinkles, Periocular wrinkles, Periorbital wrinkles. Facial and leg telangicctasia, Rosacea, Cherry angiomas, Port wine stains, Hemangiomas and venous lakes, Angioma, Spider angioma, and Poikiloderma of Civatte, Inflammatory Acne Vulgaris, Verrucae/Warts, Scars. Striea, and Psoriasis.

Podiatry - for benign cutaneous lesions and warts.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation
--------------------------------------------------
Prescription Use X(Per 21 CFR 801.109)OR Over The Counter Use _____
---------------------------------------------------------------------------
(Division Sign-Off)Division of General, Restorative,and Neurological Devices
--------------------------------------------------------------------------------------
510(k) NumberK032397
------------------------

5

{6}------------------------------------------------

Page 2 of 2

INDICATIONS FOR USE (continued):

The Dual Yellow lase, operating at 511 nm, is indicated in dermatology, plastic surgery, The Duar Tenow last, opening of benign cutaneous pigmented lesions including but not limited to:

Lentigines, Solar keratoses, Adenoma serabaceum, Skin tabs, Trichoepitheliomas (benign lesions similar to skin tags) Naevi, Keratoses, Syringomas Seborrhoeic kcratoses Verrucac vulgaris, and Warts.

Mark A Mellums

rev. 2/2004

Division Division of General, Restorative, and Neurologica. Devices

510(k) Number K032347

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.