K Number
K032169

Validate with FDA (Live)

Date Cleared
2003-09-22

(68 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Open and closed fracture fixation, nonunion, precise control of bone segments location including angulation, rotation, translation, lengthening, and shortening. The External Fixator also aids correction of bone deformities or defects associated with fractures and other pathological conditions of bone. The limb function is preserved if the External Fixator is properly applied.

Device Description

CyberOrthology Carbon Fiber Composite Circular Fixation (CIRfix) Device (Half Ring). Part of external fixator aiding trophism in the correction of bone deformities/defects. Same physical design as predicate, except aesthetically and tactual more pleasing with improved impression of superior strength and safe mobility. Radiolucent Carbon Fiber Composite utilizing Randomly orientated, pre-impregnated carbon strands resulting in a carbon fiber content of better than 62%.

AI/ML Overview

The CyberOrthology Carbon Fiber Composite Circular Fixation (CIRfix) Device (Half Ring) was evaluated against its predicate device, the Ilizarov External Fixation System (K962808). The study focused on demonstrating comparative performance rather than setting specific acceptance criteria for absolute performance.

Here's a breakdown of the information based on the provided document:

1. Acceptance Criteria and Reported Device Performance

The document describes a comparative study, implying that the "acceptance criteria" were met by demonstrating superiority in certain mechanical tests compared to the predicate device.

Acceptance Criteria (Implied)Reported Device Performance (CyberOrthology CIRfix)
Mechanical Performance Equal or Superior to Predicate Device15%-20% superiority compared to the predicate device in comparative testing. Specific tests mentioned include: Compression Stiffness (ASTM F-1746), 3-Point Bending, Cantilever Bending, and Wire Pull-out Test.
Similar Indication for UseSame as the predicate device: Part of external fixator aiding trophism in the correction of bone deformities/defects.
Similar Target PopulationSame as the predicate device: Human of any gender and at the advanced walking age.
Material PropertiesRadiolucent Carbon Fiber Composite utilizing Randomly orientated, pre-impregnated carbon strands resulting in a carbon fiber content of better than 62%. (Predicate features pre-determined fiber orientations and typically 55% carbon fiber content).
SterilityShipped non-sterile & device may be sterilized as required by any method. (Same as predicate).
BiocompatibilityComponent is a non-invasive external device, not for implantation or contact with skin/soft tissues. (Same as predicate).
Mechanical SafetyTest procedures and requirements allow for appropriate rigidity and stability. (Same as predicate).
Chemical SafetyIndicated for use in clinical and common patient living environments, not exposed to harmful chemical elements. (Same as predicate).
Compatibility with other DevicesCompatible with all appropriate predicate devices. (Predicate compatible with specially designed frames, clamps, rods, couplings, pins, posts, bolts, washers, nuts, & others).

2. Sample Size and Data Provenance

The document does not specify the precise sample size used for the comparative mechanical tests. It only states "Comparatively tested." The data provenance is not mentioned (e.g., country of origin) nor whether it was retrospective or prospective, but given it's a mechanical performance comparison, it would likely be laboratory-based testing.

3. Number of Experts and Qualifications for Ground Truth

This information is not applicable as the study described is a mechanical performance comparison of a medical device, not a study involving human interpretation of data where expert ground truth would be established (e.g., in medical image analysis).

4. Adjudication Method for the Test Set

This information is not applicable for the reasons stated above.

5. MRMC Comparative Effectiveness Study

An MRMC (Multi Reader Multi Case) comparative effectiveness study was not conducted. The study described is a direct mechanical comparison of a new physical device against an existing physical device, not an evaluation of human reader performance with or without AI assistance.

6. Standalone Performance Study

A standalone performance study in the context of patient outcomes or clinical efficacy was not explicitly described for the CyberOrthology CIRfix. The "performance" described is a standalone mechanical performance comparison against a predicate device in a laboratory setting. The document highlights mechanical superiority of the new device (15-20% superiority in various tests) when compared to the predicate device.

7. Type of Ground Truth Used

The "ground truth" for this study was established through mechanical testing standards (e.g., ASTM F-1746 for Compression Stiffness) and direct comparison of measurable physical properties against a predicate device. This is not expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set

This information is not applicable. The device is a physical medical device, not a machine learning algorithm that requires a "training set" in the computational sense.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the reasons stated above.

{0}------------------------------------------------

SEP 2 2 2003

510(K) SUMMARY (as required by 807.92(c))

032169
page 1 of 3

Submitter of 510(k):Cyber Orthology, Inc.6301 Hughes Dr.Sterling Heights, MI 48312Phone: 586-264-9544Fax: 586-264-9566
Contact Person:Dr. Djoldas Kuldjanov
Date of Summary:June 30, 2003
Trade Name:CyberOrthology Carbon Fiber Composite Circular Fixation(CIRfix) Device (Half Ring)
Classification Name:APPLIANCE, FIXATION, NAIL/BLADE/PLATECOMBINATION, MULTIPLE COMPONENT
Classification Product Code:KTT
Predicate Device:Ilizarov External Fixation SystemK962808
External Fixation SystemK870961
Intended Use:

Open and closed fracture fixation, nonunion, precise control of bone segments location including angulation, rotation, translation, lengthening, and shortening. The External Fixator also aids correction of bone deformities or defects associated with fractures and other pathological conditions of bone. The limb function is preserved if the External Fixator is properly applied.

Conclusion:

This device has equivalent intended use, has similar promotional claims, conforms to similar standards, and has equivalent technological characteristics to predicate devices.

{1}------------------------------------------------

Device Comparison Chart

Image /page/1/Picture/1 description: The image shows handwritten text on a white background. The text at the top reads "K032169", and the text below reads "page 2 of 3". The handwriting is in black ink and appears to be cursive.

Comparison ItemsCyberOrthology CarbonFiber Composite CircularFixation (CIRfix) deviceIlizarov External FixationSystem K962808
Indication for UsePart of external fixator aidingtrophism in the correction ofbone deformities/defects.Same
Target populationHuman of any gender and atthe advanced walking ageSame
DesignSame physical design, exceptaesthetically and tactual morepleasing with improvedimpression of superiorstrength and safe mobilityCircular half-ring withoverlapping ends to form aperfect ring whenjoined together by bolts andnuts
MaterialRadiolucent Carbon FiberComposite utilizingRandomly orientated, pre-impregnated carbonstrands resulting in acarbon fiber content ofbetter than 62%.Radiolucent Carbon FiberComposite featuringpre-determined fiberorientations and typically 55%carbon fiber content.
PerformanceComparatively tested,indicating 15%-20%superiority.Compression Stiffness,(ASTM F-1746). 3-PointBending. Cantilever Bending.Wire Pull-out Test.
SterilityShipped none-sterile & devicemay be sterilized as requiredby any method.Same
BiocompatibilityThe component is a noneinvasive external device andwill not be used forimplantation or contact withskin or soft tissues.Same
Mechanical SafetyThe test procedure&requirements allow for theappropriate amount of rigidityand stability.Same
Chemical SafetyThe device is indicated for usein clinical and common patientliving environment not beingexposed toharmful chemical elementsSame
Compatibility with otherDevicesCompatible with allappropriate predicate devicesCompatible with speciallydesigned frames, clamps, rods,couplings, pins, posts, bolts,washers, nuts, & others for themanagement of appropriateorthopedic surgeries

{2}------------------------------------------------

Device Comparison Chart

Image /page/2/Picture/1 description: The image shows handwritten text on a white background. The top line reads 'K032169', with the 'k' being lowercase and the other characters being numbers. The second line reads 'page 3 of 3', indicating that this is the third page of a three-page document.

Where usedThe device is being used in hospitals and in patients surroundings and environments
Same

.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal. The seal is circular and contains the words "HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle are three stylized lines that resemble a bird in flight. The lines are thick and curved, and they are arranged in a way that suggests movement.

Public Health Service

SEP 2 2 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CyberOrthology, Inc. c/o Mr. Arthur Ward AJW Technology Consultants, Inc. 962 Allegro Lane Apollo Beach, FL 33572

Re: K032169

Trade/Device Name: CyberOrthology Carbon Fiber Composite Circular Fixation (CIRfix) Device (Half Ring) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: LXT Dated: June 30, 2003 Received: July 21, 2003

Dear Mr. Ward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Lisa M. Boyle

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Mark M. Wilkerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

Device Name: CyberOrthology Carbon Fiber Composite Circular Fixation (CIRfix) Device (Half Ring)

Indications For Use:

Open and closed fracture fixation, nonunion, precise control of bone segments location including angulation, rotation, translation, lengthening, and shortening. The External Fixator also aids correction of bone deformities or defects associated with fractures and other pathological conditions of bone. The limb function is preserved if the External Fixator is properly applied.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark A. Milkman

(Division of General, Restorative and Neurological I

Number K032169

510(k) Number -

OR

Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use

(Optional Format 1-2-96)

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.