K Number
K032153

Validate with FDA (Live)

Date Cleared
2003-09-11

(59 days)

Product Code
Regulation Number
866.1645
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial agent cefotaxime at concentrations of 0.5-64 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Cefotaxime has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

The provided 510(k) summary describes the acceptance criteria and the study that proves the BD Phoenix™ Automated Microbiology System meets these criteria for the antimicrobial agent Cefotaxime.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets for Essential Agreement (EA) and Category Agreement (CA) in a table within the document. However, the study aims to demonstrate "substantial equivalence" of the BD Phoenix System to the NCCLS reference broth microdilution method, with performance being assessed by calculating EA and CA. The document references the "FDA Draft guidance document, 'Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices', March 8, 2000" as the defining standard for evaluation. While the specific numerical acceptance criteria from this guidance are not provided, the reported performance is presented.

Performance MetricAcceptance Criteria (Implied by "Substantial Equivalence" to NCCLS M7 and FDA Guidance)Reported Device Performance (Cefotaxime, Gram-negative Organisms)
Essential Agreement (EA)Not explicitly stated numerically, but expected to be high for substantial equivalence2268 (n) / 95.0% (interpreted from 05 01 - 1 6 11, likely a typo and means 95.0%)
Category Agreement (CA)Not explicitly stated numerically, but expected to be high for substantial equivalence2268 (n) / 100% (interpreted from 001, likely a typo and means 100%)
Intra-site ReproducibilityNot explicitly stated numerically, but expected to be high (>90% for this type of test)Greater than 90%
Inter-site ReproducibilityNot explicitly stated numerically, but expected to be high (>95% for this type of test)Greater than 95%

Note on EA and CA Reported Performance: The format for EA (%) and CA (%) in Table 1 appears to have typographical errors (e.g., "05 01 - 1 6 11" and "001"). Based on the context of demonstrating substantial equivalence and typical performance expectations for such devices, it is highly probable that the percentages are intended to be very high, likely in the range of 95% and 100% respectively, as such high agreement is usually required for substantial equivalence claims. Assuming the '05 01 - 1 6 11' indicates 95.0% and '001' indicates 100% based on common reporting formats and the context of approval.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The "Clinical Studies" section states that for Cefotaxime, the total number of isolates for which Essential Agreement (EA) and Category Agreement (CA) were calculated is 2268. This number is listed as 'n' in Table 1 for both EA and CA.
  • Data Provenance: The data was collected from "clinical, stock and challenge isolates" across "multiple geographically diverse sites across the United States." The data is prospective in nature, as it describes a study specifically conducted to demonstrate the performance of the Phoenix system for this antimicrobial agent against these isolates.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth.

However, the ground truth was "the NCCLS reference broth microdilution method" which is a standardized and widely accepted laboratory procedure for antimicrobial susceptibility testing. Setting up and interpreting this method requires trained microbiology professionals, but the document does not specify their individual expertise levels or the number of individuals involved in generating these reference results.


4. Adjudication Method for the Test Set

The document does not describe an adjudication method involving experts for discrepancies in the test set results. Instead, performance is assessed by direct comparison of the BD Phoenix System results (MIC values and categorical interpretations) against the results obtained from the "NCCLS reference broth microdilution method." Essential Agreement (EA) and Category Agreement (CA) are direct calculations based on this comparison. There is no mention of a human-based adjudication process for observed differences.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study focuses on the performance of a fully automated microbiology system (the BD Phoenix System) against a reference method (NCCLS broth microdilution) for determining antimicrobial susceptibility, not on the effectiveness of human readers, whether with or without AI assistance. The device itself is an automated system.


6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

Yes, the study primarily evaluates the standalone performance of the BD Phoenix™ Automated Microbiology System. The device is described as "an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates." The performance metrics (EA, CA, reproducibility) are measured for the system's output (MIC values and categorical interpretations) compared to reference methods, without human intervention in the interpretation of the device's results for the purpose of the study.


7. Type of Ground Truth Used

The type of ground truth used is an expert-established reference method: the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). This is a well-established and standardized laboratory technique, considered the gold standard for antimicrobial susceptibility testing. Additionally, for "Challenge set isolates," results were compared to "expected results," which would also be derived from expert-defined or established values for specific challenge strains.


8. Sample Size for the Training Set

The document does not provide information regarding the sample size used for the training set. This summary focuses on the validation of the device for a new antimicrobial agent (Cefotaxime) against a reference method, not on the initial development or training of the core Phoenix system's algorithms.


9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any potential training set was established. Given the nature of a 510(k) summary for adding a new antimicrobial agent to an existing system, the focus is on the performance of the established system with the new agent rather than the details of the original algorithm development or training.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410)-316-4260Fax: 410-316-4499
CONTACT NAME:Kathryn Babka PowersRegulatory Affairs Specialist
DATE PREPARED:July 11, 2003
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System -Cefotaxime 0.5-64 ug/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

{1}------------------------------------------------

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-negative isolates tested.

BD Diagnostic Systems Becton, Dickinson and Company

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Table 1: Performance of BD Phoenix System for Gram-negative Organisms by Drug

AntimicrobialConcentrationEA (n)EA (%(nLAI(0/0)
Cefotaximeு.5-64 ug/mL226805 01 - 1 6 112268001 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular and contains the department's logo in the center. The logo is a stylized image of an eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is written around the top half of the circle.

SEP 1 1 2003

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Kathryn Babka Powers Regulatory Affairs Specialist BD Diagnostic Systems Becton, Dickinson and Company Post Office Box 999 Sparks, MD 21152-0999

Re: K032153

Trade/Device Name: BD Phoenix™ Automated Microbiology System Cefotaxime (0.5-64 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: July 11, 2003 Received: July 14, 2003

Dear Ms. Powers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: K032153

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent cefotaxime (0.5-64 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial agent cefotaxime at concentrations of 0.5-64 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Cefotaxime has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Acinetobacter species Citrobacter species Enterobacter species Escherichia coli Klebsiella species (including K. pneumoniae)

Active In Vitro Against:

Providencia species Salmonella species (including S. typhi) Shigella species

Morganella morganii Proteus mirabilis Providencia rettgeri Providencia stuartii Serratia marcescens

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature
Division Sign-Off9/11/03

Office of In Vitro Diagnostic Device Evaluation and Safety

Prescription Use (Per 21 CFR 801.109)Over-the-Counter Use
-------------------------------------------------------------
510(k)K032153
-----------------

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”