K Number
K032128

Validate with FDA (Live)

Date Cleared
2003-08-01

(22 days)

Product Code
Regulation Number
870.4260
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Terumo AL8X Arterial Blood Filter for extra-corporeal service is indicated for use in all cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli, and aggregates composed of platelets, red blood cells, and other debris from the arterial line. The device may be used in procedures lasting up to 6 hours in duration.

Device Description

The filter is comprised of an outer housing that contains a smaller inner housing. The outer housing is cylindrical in shape and has a slight conical-shaped lid assembly with an air vent port. The blood inlet port is positioned along the upper-side axis of the outer housing and provides the entry point for blood. The base of the housing contains the blood outlet port. The inner housing of the device contains a screen filter through which blood will pass through for filtration of particulate matter. After the blood has been filtered, it then exits the assembly via the blood outlet port. The blood-contacting surfaces of the device are coated with Terumo's polymer coating solution.

AI/ML Overview

The Terumo AL8X Arterial Filter with X-Coating is a medical device intended for use in cardiopulmonary bypass procedures to remove micro-emboli (greater than 40 microns) from the arterial line. The submission K032128 establishes substantial equivalence to a predicate device, the Pall Medical AL8 Arterial Filter (K834380), rather than defining and meeting new acceptance criteria through a clinical study with a detailed ground truth establishment.

Here's an analysis based on the provided document, addressing your specific points:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of specific acceptance criteria with numerical thresholds. Instead, it demonstrates performance equivalence to an existing predicate device through comparative in-vitro evaluations. The primary "acceptance" is based on showing no clinically significant performance differences compared to the predicate.

Performance EvaluationPredicate Device (Pall AL8 Arterial Filter) StatusTerumo AL8X Arterial Filter with X-Coating PerformanceNotes
Filtration EfficiencyEstablished (40 µm filter mesh)EquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Air Handling CapabilitiesEstablished (centripetal force)EquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Device Effect Upon Cellular Blood ComponentsEstablishedEquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Pressure DropEstablishedEquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Mechanical IntegrityEstablishedEquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Static Priming VolumeEstablishedEquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
Connection of PVC Tubing to Inlet/Outlet PortsEstablishedEquivalentPerformed in-vitro, no specific numerical values provided for comparison beyond "equivalent".
BiocompatibilityEstablished (components)Meets applicable standardsBased on ISO 10993 guidelines for external communicating devices, circulating blood, limited exposure (≤ 24 hours). Polymer coating specifically evaluated in an in-vivo animal study (with an oxygenator device).
SterilizationEstablishedValidated to SAL of 10⁻⁶In accordance with AAMI guidelines.
Material CharacterizationEstablishedConducted for aged and non-aged devicesPhysico-chemical profiles.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify sample sizes for the in-vitro tests conducted. It generally states that "Terumo Cardiovascular Systems Corporation conducted several evaluations... to demonstrate its equivalence." The data provenance for these evaluations is described as "in-vitro performance tests," implying laboratory settings rather than human subject data. There is no information regarding the country of origin of the data or whether it was retrospective or prospective. An "in-vivo animal study (with an oxygenator device)" was mentioned for the polymer coating, but no details on size or provenance were given.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided. Since the study involved in-vitro performance evaluations and equivalence demonstration to a predicate device, the concept of "ground truth established by experts" in a clinical context (e.g., radiologists interpreting images) is not directly applicable here. The "ground truth" for the performance tests would be derived from physical measurements and established engineering/biomedical testing standards and methods.

4. Adjudication Method for the Test Set

This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation (e.g., image reading) to resolve discrepancies in expert opinions. Given the in-vitro nature of the performance evaluations, such adjudication methods were not employed.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study would involve human readers evaluating cases with and without AI assistance, which is not relevant to the evaluation of an arterial blood filter.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No, a standalone algorithm performance study was not done. The device is a physical medical device, not an AI algorithm. Its performance was evaluated through in-vitro tests and comparison to a predicate physical device.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluations was based on physical and biological measurements derived from in-vitro tests, and comparison to the established performance characteristics of the predicate device. For biocompatibility, it relied on established biological standards (ISO 10993) and an in-vivo animal study. For sterilization, it was based on AAMI guidelines. There was no "expert consensus," "pathology," or "outcomes data" ground truth in the sense of a diagnostic or prognostic device study.

8. The Sample Size for the Training Set

This information is not applicable. The Terumo AL8X Arterial Filter is a physical medical device, not a machine learning model, so there is no "training set" in the context of AI.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

K032128

Section II: 510(k) Summary and Certification Terumo Cardiovascular Systems Terumo AL8X Arterial Filter

Terumo AL8X Arterial Filter with X-Coating

Submitter Information:

AUG 01 2003

This submission was prepared in July 2003 by: Garry A. Courtney, MBA, RAC Sr. Regulatory Affairs Specialist Terumo Cardiovascular Systems Corp. 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-283-7866, Ext. 7420

Device Names:Proprietary Name:Terumo AL8X Arterial Filter with X-Coating
Common Name:Arterial Line Blood Filter
Device Classifications:Device Classification Name:Cardiopulmonary Bypass Arterial Line Blood Filter
Device Classification:The Terumo AL8X Arterial Filter with X-Coating is classified as a Class II device per 21 CFR § 870.4260
Regulation Number:21 CFR § 870.4260

Predicate Device:

The device submitted in this 510(k) maintains characteristics that are substantially equivalent in intended use, design, technology/principles of operation, materials and specifications to the following device:

Pall Medical AL8 Arterial Filter - (K834380).

Intended Use:

The Terumo AL8X Arterial Blood Filter for extra-corporeal service is indicated for use in all cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli, and aggregates composed of platelets, red blood cells, and other debris from the arterial line. The device may be used in procedures lasting up to 6 hours in duration.

{1}------------------------------------------------

Principles of Operation and Technology:

The Terumo AL8X Arterial Filter performs its functions using two basic forms of technology. As a filtration device, particulates in the blood stream are captured and removed from the blood flow as blood passes through the 40 um filter mesh material that is contained within the device housing. The 40 um filter mesh material establishes a physical barrier that entraps particulate matter and prevents it from moving downstream of the arterial filter assembly.

As an air-removal device, the Terumo AL8X Arterial Filter is designed so that air is removed from the blood stream as a result of centripetal force. The blood inlet port of the device is positioned on the upper-side axis of the polycarbonate housing, thereby creating a spiral blood flow pattern as blood enters the device. Because the blood flows through the device in a spiral motion, centripetal forces cause the air bubbles to migrate towards the top of the housing assembly. The top of the housing assembly has a slight conicalshape, thereby facilitating air movement towards the purge port located at the top of the assembly.

Design and Materials:

The design of the Terumo AL8X Arterial Filter is such that it will remove particulates while simultaneously facilitating the removal of air that might be in the blood flow. The blood-contacting surfaces of the device are coated with Terumo's polymer coating solution. The device accomplishes its intended use primarily due to its design characteristics.

The filter is comprised of an outer housing that contains a smaller inner housing. The outer housing is cylindrical in shape and has a slight conical-shaped lid assembly with an air vent port. The blood inlet port is positioned along the upper-side axis of the outer housing and provides the entry point for blood. The base of the housing contains the blood outlet port.

The inner housing of the device contains a screen filter through which blood will pass through for filtration of particulate matter. After the blood has been filtered, it then exits the assembly via the blood outlet port.

The materials of construction for the Terumo AL8X Arterial Filter are identical to the materials used in the predicate Pall AL8 Arterial Filter - except that the Terumo device contains a polymer coating which is not present on the predicate device. The difference in the materials do not raise any new issues of safety or effectiveness of the device, as the polymer coating has been demonstrated to be safe in many other Terumo devices that have been cleared by the United States Food and Drug Administration.

{2}------------------------------------------------

Performance Evaluations:

Terumo Cardiovascular Systems Corporation conducted several evaluations of the Terumo AL8X Arterial Filter to demonstrate its equivalence to the Pall AL8 Arterial Terumo conducted the following in-vitro performance tests to demonstrate Filter. equivalence:

  • Filtration Efficiency .
  • . Air Handling Capabilities
  • Device Effect Upon Cellular Blood Components .
  • Pressure Drop .
  • . Mechanical Integrity
  • Static Priming Volume .
  • Evaluation of the Connection of PVC Tubing to Inlet/Outlet Ports. .

{3}------------------------------------------------

Substantial Equivalence Comparison:

The Terumo AL8X Arterial Filter is substantially equivalent to the predicate Pall AL8 Arterial Filter device as follows:

  • The intended uses of the Terumo AL8X Arterial Filter and the Intended Use: . predicate Arterial Filter are exactly the same. Each of the Arterial Filter devices is for extra-corporeal service in all cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli, and aggregates composed of platelets, red blood cells, and other debris from the arterial line. The devices are also exactly the same in that each device may be used in procedures lasting up to 6 hours in duration.
    There are no differences in Intended Use between the Terumo AL8X Arterial Filter and the Pall Medical AL8 Arterial Filter.

  • Principles of Operation and Technology: The operation and technology of the Terumo . AL8X Arterial Filter and the predicate device are identical. The devices operate in a manner where two basic forms of technology are utilized. As filtration devices, particulates in the blood stream are captured and removed from the blood flow as blood passes through a 40 um filter mesh material that is contained within the device housing. The 40 um filter mesh material establishes a physical barrier that entraps particulate matter and prevents it from moving downstream of the arterial filter assembly.
    As air-removal devices, each of the filters is designed so that air is removed from the blood stream as a result of centripetal force. The blood inlet port of each device is positioned on the upper-side axis of the polycarbonate housing, thereby creating a spiral blood flow pattern as blood enters the devices. Because the blood flows through the devices in a spiral motion, centripetal forces cause the air bubbles to migrate towards the top of the housing assembly. The top of the housing assembly has a slight conical-shape, thereby facilitating air movement towards the purge port located at the top of the assembly.

There are no differences in operation or technology between the Terumo AL8X Arterial Filter and the Pall Medical AL8 Arterial Filter.

  • Design and Materials: The design and the materials of the Terumo AL8X Arterial ● Filter and the predicate device are essentially the same. The design of the two devices is identical in that they are each a polycarbonate housing with an inner filter - and intended to remove particulate matter and air bubbles.
    The devices are manufactured with the exact same materials. These materials include polycarbonate, polyester, polypropylene, acrylonitrile-butadiene-styrene (ABS) and polyvinyl chloride.

The Terumo AL8X Arterial Filter contains X-Coating, which is a biocompatible surface coating that reduces platelet adhesion to the device. The use of X-Coating has

{4}------------------------------------------------

been demonstrated as safe and raises no new issues of safety and/or effectiveness. The X-Coating is not present on the predicate Pall Medical AL8 Arterial Filter.

  • Performance: Comparisons of the performance of the Terumo AL8X Arterial Filter . and the predicate device were conducted. The comparisons demonstrated that there were no clinically significant performance differences between the devices.

Substantial Equivalence Summary:

In summary, the Terumo AL8X Arterial Filter and the predicate Pall AL8 Arterial Filter are substantially equivalent in intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the subject device and the predicate devices do not raise new issues of safety and effectiveness.

Additional Safety Information:

  • Sterilization conditions have been validated in accordance with AAMI guidelines to . provide a Sterility Assurance Level (SAL) of 10°.
  • . Biocompatibility studies have been conducted on the materials that are used in the construction of the Terumo AL8X Arterial Filter device - as recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. All blood-contacting materials have been found to meet the applicable biocompatibility standards.
  • . Terumo conducted studies for materials characterization, including physico-chemical profiles of aged and nonaged devices.
  • . The polymer coating material that is applied to the blood-contacting surfaces of the device was also evaluated in an in-vivo animal study (with an oxygenator device). No adverse conditions were noted.

Conclusion:

In summary, the Terumo AL8X Arterial Filter is substantially equivalent in intended use, principles of operation and technology, design and materials, and performance to the predicate Pall AL8 Arterial Filter (K834380).

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with two snakes entwined around it.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Terumo Cardiovascular Systems Corporation c/o Mr. Garry A. Courtney, MBA, RAC Senior Regulatory Affairs Specialist 125 Blue Ball Road

Re: K032128

Elkton, MD 21921

Trade Name: Terumo AL8X Arterial Filter with X-Coating Regulation Number: 21 CFR 870.4260 Regulation Name: Cardiopulmonary bypass arterial line blood filter Regulatory Class: Class II (two) Product Code: DTM Dated: July 8, 2003 Received: July 10, 2003

Dear Mr. Courtney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

AUG - 1 2003

{6}------------------------------------------------

Page 2 - Mr. Garry A. Courtney, MBA, RAC

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dr. R. Zukerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Number (if known):

Terumo AL8X Arterial Filter with X-Coating Device Name:

Indications For Use:

The Terumo AL8X Arterial Blood Filter for extra-corporeal service is indicated for use in all cardiopulmonary bypass procedures for the removal of micro-emboli greater than 40 microns in size, including gas emboli, fat emboli, and aggregates composed of platelets, red blood cells, and other debris from the arterial line. The device may be used in procedures lasting up to 6 hours in duration.

Gary A. Courtney, MBA, RAC
Terumo Cardiovascular Systems

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseXOROver-Th
(Per 21 CFR 801.109)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) NumberK032128
------------------------

e-Counter Use

§ 870.4260 Cardiopulmonary bypass arterial line blood filter.

(a)
Identification. A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”