K Number
K032085

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-10-03

(88 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nidek Green Laser Photocoagulator Model GYC-1000 is indicated for use in ophthalmic surgical procedures, including retinal and macular photocoagulation, iridotomy and trabeculoplasty.

Device Description

The Nidek Green Laser Photocoagulator Model GYC-1000 is a frequency doubled diode pumped solid state (DPSS) laser ophthalmic photocoagulation system that produces a 532 nm (green) wavelength light as the treatment beam. The Model GYC-1000 uses a diode laser (red) with a wavelength of 635 nm as the aiming beam. The green laser beam is aligned with the red aiming laser beam in the optical system inside the unit and gathers them in a fiber-optic cable. The laser beam is led to the delivery unit via the fiber-optic cable, shaped into the specified spot size in the optical system, and emitted to the affected area (the emission areas of both the therapeutic laser beam and the aiming beam are the same).

The Model GYC-1000 can be used to coagulate the target tissue efficiently and safely, and the system can be applied to transpupillary photocoagulation procedures using a slit-lamp or indirect ophthalmoscope and intraocular photocoagulation procedures using an endophotocoagulation probe. The Model GYC-1000 is a smaller, modified version of the Models GYC-1500 and GYC-2000, which were the subject of premarket notification number K980547.

A protective filter may be attached to each delivery unit in the observation optical path so that the operator's eye can be protected from the laser beam if it is reflected from a patient's eve or contact lens during laser emission. The Model G Y C-1000 has a number of delivery units available for use in a variety of ophthalmic procedures:

  • Endophotocoagulation Delivery Unit .
  • NIDEK Slit-Lamp Delivery Unit
  • ZEISS Slit-Lamp Attachable Delivery Unit ●
  • HAAG Attachable Slit-Lamp Delivery Unit .
  • Binocular Indirect Ophthalmoscope Delivery Unit .
  • Combination Delivery Unit for Nidek Nd: YAG laser .
AI/ML Overview

The Nidek Green Laser Photocoagulator Model GYC-1000 acceptance criteria and the study that proves the device meets them are described below based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is primarily assessed through its adherence to specified safety and performance standards for medical electrical equipment and laser products. The acceptance criteria are implicit in meeting the requirements of these standards.

Acceptance Criteria (Implied by Standards Compliance)Reported Device Performance
Conformance to Electrical Safety Standard (EN 60601-1)The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with EN 60601-1 and was found to meet all requirements.
Conformance to Electromagnetic Compatibility (EN 60601-1-2)The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with EN 60601-1-2 and was found to meet all requirements.
Conformance to Programmable Electrical Medical Systems (EN 60601-1-4)The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with EN 60601-1-4 and was found to meet all requirements.
Conformance to Safety of Diagnostic and Therapeutic Laser Equipment (IEC 60601-2-22, JIS T1204, JIS C6802)The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with IEC 60601-2-22, JIS T1204, and JIS C6802, and was found to meet all requirements of the standards.
Substantial Equivalence to Predicate DeviceA comparison of technological characteristics was performed. The device was concluded to have the same basic technological characteristics as the predicate device (Nidek GYC-1500/2000, K980547) and to be equivalent. Differences were deemed insignificant and not affecting safety or effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

The provided text only mentions "System and component testing" without specifying a sample size in terms of the number of devices tested. It also does not specify the provenance of any data (e.g., country of origin, retrospective/prospective clinical data). The testing appears to be primarily laboratory/bench testing against engineering specifications and international standards, rather than clinical trials with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The study described is primarily technical testing against established engineering and safety standards, not a study requiring expert-established ground truth for medical diagnoses or outcomes.

4. Adjudication Method for the Test Set

Not applicable. The study described involves compliance testing against technical standards, not a process of adjudicating medical interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is a laser photocoagulator, not an AI software intended for image interpretation or diagnostic assistance. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical medical instrument (laser photocoagulator), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is compliance with established international and Japanese electro-medical and laser safety standards (EN 60601-1, EN 60601-1-2, EN 60601-1-4, IEC 60601-2-22, JIS T1204, and JIS C6802) and product specifications derived from a risk analysis. There is no mention of expert consensus, pathology, or outcomes data as "ground truth" in this context.

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical instrument and does not involve AI or machine learning algorithms that require a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

KO32087 510(k) SUMMARY NIDEK GREEN LASER PHOTOCOAGULATOR MODEL GYC-1000

SUBMITTER INFORMATION 1.

A.Company Name:Nidek Incorporated
B.Company Address:47651 Westinghouse Drive.Fremont, CA 94539-7474
C.Company Phone:(510) 353-7719
Company Fax:(510) 226-5750
D.Contact Person:Mr. Hiro MatsuzakiRegulatory and QA ManagerNidek Incorporated
E.Date Summary Prepared:October 1, 2003

2. DEVICE IDENTIFICATION

A.Classification Name:Ophthalmic Laser, and Powered Surgical LaserInstrument
B.Trade/Proprietary Name:Nidek Green Laser PhotocoagulatorModel GYC-1000
C.Device Classification:Class II (per 21 CFR 886.4390 and 878.4810)
D.Product Code:HQF and GEX

SUBSTANTIAL EQUIVALENCE 3.

The Nidek Green Laser Photocoagulator Model GYC-1000 is of comparable type and is substantially equivalent to the following predicate device:

Predicate DeviceManufacturer510(k) No.Date Cleared
Nidek Laser PhotocoagulatorModel GYC-1500/2000Nidek Co. Ltd.JapanK980547May 13, 1998

{1}------------------------------------------------

The fundamental technical characteristics and device specifications of the Nidek Green Laser Photocoagulator Model GYC-1000 are the same as those of the predicate device. The Model GYC-1000 and the predicate device are frequency doubled solidstate lasers with a wavelength of 532 nm. The Model GYC-1000 and the predicate device use a variety of delivery systems, including slit lamps, indirect ophthalmoscopes, endoprobes, and Nd: Y AG lasers to deliver the laser beam. The Model GYC-1000 and the predicate device are indicated for use in macular and retinal photocoagulation, and glaucoma procedures such as iridotomy and trabeculoplasty.

4. DEVICE DESCRIPTION

The Nidek Green Laser Photocoagulator Model GYC-1000 is a frequency doubled diode pumped solid state (DPSS) laser ophthalmic photocoagulation system that produces a 532 nm (green) wavelength light as the treatment beam. The Model GYC-1000 uses a diode laser (red) with a wavelength of 635 nm as the aiming beam. The green laser beam is aligned with the red aiming laser beam in the optical system inside the unit and gathers them in a fiber-optic cable. The laser beam is led to the delivery unit via the fiber-optic cable, shaped into the specified spot size in the optical system, and emitted to the affected area (the emission areas of both the therapeutic laser beam and the aiming beam are the same).

The Model GYC-1000 can be used to coagulate the target tissue efficiently and safely, and the system can be applied to transpupillary photocoagulation procedures using a slit-lamp or indirect ophthalmoscope and intraocular photocoagulation procedures using an endophotocoagulation probe. The Model GYC-1000 is a smaller, modified version of the Models GYC-1500 and GYC-2000, which were the subject of premarket notification number K980547.

A protective filter may be attached to each delivery unit in the observation optical path so that the operator's eye can be protected from the laser beam if it is reflected

{2}------------------------------------------------

from a patient's eve or contact lens during laser emission. The Model G Y C-1000 has a number of delivery units available for use in a variety of ophthalmic procedures:

  • Endophotocoagulation Delivery Unit .
  • NIDEK Slit-Lamp Delivery Unit
  • ZEISS Slit-Lamp Attachable Delivery Unit ●
  • HAAG Attachable Slit-Lamp Delivery Unit .
  • Binocular Indirect Ophthalmoscope Delivery Unit .
  • Combination Delivery Unit for Nidek Nd:YAG laser .

5. INDICATIONS FOR USE

The Nidek Green Laser Photocoagulator Model GYC-1000 is indicated for use in ophthalmic surgical procedures, including retinal and macular photocoagulation, iridotomy and trabeculoplasty.

TECHNOLOGICAL CHARACTERISTICS 6.

A comparison of the technological characteristics of the Nidek Green Laser Photocoagulator Model GYC-1000 and the predicate device has been performed, and the results are summarized in the table below. The results of this comparison demonstrate that the Nidek Green Laser Photocoagulator Model GYC-1000 has the same basic technological characteristics as the predicate device and is equivalent to the marketed predicate device. The differences between the Nidek Green Laser Photocoagulator Model GYC-1000 and the predicate device are insignificant and do not affect the safety or effectiveness of the device.

{3}------------------------------------------------

Nidek Model GYC-1000Nidek GYC-1500/2000K980547 (Predicate Device)
Treatment LaserLaser diode pumped,frequency doubled Nd:YAGsolid state laserModel GYC-1500: Arclamp(Krypton) pumped, frequencydoubled Nd:YAG solid state laserModel GYC-2000: Laser diodepumped, frequency doubledNd:YAG solid state laser
Wavelength532 nm532 nm
OutputContinuous WaveContinuous Wave
Power Output50 to 1700mW(50 to 500mW in 10mWincrements,500 to 1700mW in 50mWincrements)50 to 1500mW in 10mWincrements (Model I)or50 to 2000mW in 10mWincrements (Model II)
Exposure Time0.01 to 0.10 sec. in 0.01 sec.increments0.10 to 0.50 sec. in 0.05 sec.increments0.50 to 1.00 sec. in 0.10 sec.increments1.00 to 3.00 sec. in 1.00 sec.increments0.02 to 3.00 sec.
Repetition ModeInterval time 1.0 to 0.1 sec. in0.1 sec. increments1.0, 0.6, 0.4 and 0.2 sec.
Cooling MethodAir cooledInternal water cooling system
MicroprocessorTwo CPU's: Master CPU forcontrolling the device, and SubCPU for monitoring theMaster CPUOne CPU
Aiming Laser
TypeDiode LaserDiode Laser
Wavelength635 nm (red)633 nm (red)
Output TypeContinuous WaveContinuous Wave
Power Output0.2 to 0.4mW0.2 to 0.8mW
Accessories
Endophotocoagulation DeliveryUnitStraight and angled probes,which are single-usedisposableStraight and angled probes, whichare single-use disposable
Slit Lamp Delivery UnitNidek SL-1600 TypeNidek SL-1600 Type
Slit Lamp AdaptorsZeiss Type•Zeiss Type•
PREDICATE DEVICE COMPARISON TABLE
Nidek Model GYC-1000Nidek GYC-1500/2000K980547 (Predicate Device)
Binocular IndirectOphthalmoscope (B.I.O.)Delivery UnitHaag Streit Type Heine Omega 180 TypeHaag Streit Type Heine Omega 180 Type
Combination Delivery UnitFor use with Nidek Nd:YAGlasersFor use with Nidek Nd:YAGlasers
Spot Size at Target50 to 500 µm (parfocal) for Combination Unit andSlit Lamp Units 180 to 520 µm for B.I.O. 400 µm for Endo probes50 to 500 µm (parfocal) forCombination Unit and SlitLamp Units 180 to 520 µm for B.I.O. 400 µm for Endo probes
Operating Microscope FiltersZeiss Möeller Wedel Topcon WildZeiss Möeller Wedel Topcon Wild
Power Requirements90-264 VAC, 50/60 Hz,200 VA100/120/230 VAC, 50/60 Hz,1500 VA (continuous)

{4}------------------------------------------------

7. PERFORMANCE DATA

System and component testing was completed based on product specifications and hazard effects determined from the risk analysis. The following testing was performed on the Nidek Green Laser Photocoagulator Model GYC-1000 to demonstrate that it meets all specified requirements and is equivalent to the predicate device:

A. Electrical Safety Testing & Electromagnetic Compatibility

The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with EN 60601-1 and EN 60601-1-2, and was found to meet all requirements of both standards.

{5}------------------------------------------------

  • Programmable Electrical Medical Systems B.
    The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with EN 60601-1-4 and was found to meet all requirements of the standard.

  • C. Requirements For The Safety Of Diagnostic And Therapeutic Laser Equipment
    The Nidek Green Laser Photocoagulator Model GYC-1000 was tested in accordance with IEC 60601-2-22, JIS T1204, and JIS C6802, and was found to meet all requirements of the standards.

8. CONCLUSIONS

Nidek Incorporated has demonstrated through its evaluation of the Nidek Green Laser Photocoagulator Model GYC-1000 that the device is equivalent to the predicate device with respect to intended use, technological characteristics, and safety and effectiveness.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 3 2003

Mr. Hiro Matsuzaki Regulatory and QA Manager Nidek Incorporated 47651 Westinghouse Drive Fremont, CA 94539-7474

Re: K032085

Trade/Device Name: Nidek Green Laser Photocoagulator Model GYC-1000 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatiology Regulatory Class: II Product Code: GEX Dated: July 1, 2003 Received: July 7, 2003

Dear Mr. Matsuzaki:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 - Mr. Hiro Matsuzaki

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE
---------------------
510(k) Number:K032085 (To Be Assigned By FDA)
Device Trade Name:Nidek Green Laser Photocoagulator Model GYC-1000
Indications For Use:The Nidek Green Laser Photocoagulator Model GYC-1000 is indicated for use in ophthalmic surgical procedures, including retinal and macular photocoagulation, iridotomy and trabeculoplasty.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)


Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

Miriam C Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K032085

4

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.