K Number
K031785

Validate with FDA (Live)

Date Cleared
2003-09-04

(86 days)

Product Code
Regulation Number
870.3470
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MacroPore Cardio-Wrap (TS) is indicated for use as a pericardium replacement device in patients that may require re-operation within six months.

Device Description

MacroPore Cardio-Wrap (TS) is a resorbable implant in sheet form manufactured from poly lactic acid (PLA). MacroPore Cardio-Wrap (TS) can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Cardio-Wrap (TS) to the desired shape or size. MacroPore Cardio-Wrap (TS) Sheet is fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPore Cardio-Wrap (TS) can be used either alone or in conjunction with soft tissue fixation devices such as resorbable sutures, which can also serve to fixate the MacroPore Cardio-Wrap (TS) and prevent dislocation. The MacroPore Cardio-Wrap (TS) may be used in conjunction with various MacroPore manual instruments. MacroPore Cardio-Wrap (TS) is provided in various shapes such as rectangles, ovals, and circles and will be provided in other shapes and sizes as needed for particular surgical procedures. MacroPore Cardio-Wrap (TS) is provided in sheets of 10mm to 500mm x 500mm and will be provided in other shapes and sizes as needed for particular surgical procedures. The thickness of the MacroPore Cardio-Wrap (TS) ranges from 0.05 mm to 1.0 mm according to the region to be treated. The MacroPore Cardio-Wrap (TS) is provided in solid sheets. The borders of the sheets may be aligned with holes to attach suture material.

AI/ML Overview

The provided text describes the MacroPore Cardio-Wrap (TS) device for 510(k) clearance, focusing on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria in a clinical study. Therefore, many of the requested elements for a study proving device meeting acceptance criteria (like sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) are not included in this type of submission.

However, I can extract the information provided regarding the testing and the basis for its equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance AspectReported Device Performance (Proxy for Acceptance)
Material CompositionFabricated from polylactic acid (PLA) – Met (consistent with description)
Inherent Viscosity (after heating)Viscosity stayed within an appropriate range over 120 minutes when heated to 60°C in saline. – Met
Aging (Strength)Demonstrated to be strong enough for the indications for use. – Met
Mechanical Strength (Tensile & Suture Pull-out)Substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions. – Met
Safety and Efficacy (In Vivo)Animal studies demonstrated that the materials are safe and efficacious for the indications for use. – Met
Indications for Use (Equivalence)Shares identical indications for use principles with predicate devices for the same surgical procedures. – Met
Physical Design (Equivalence)Substantially equivalent to predicate devices (thin semi-rigid sheets, contourable when heated). – Met
Thickness (Equivalence)Thinnest MacroPore Cardio-Wrap (TS) thickness (0.5mm) is virtually identical to predicate (1.0mm). – Met
Dimensions (Equivalence)Comparable to predicate devices (rectangular sheets, several centimeters in size). – Met

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size:
    • In Vitro Testing: Not specified for each specific test (e.g., how many viscosity samples, how many mechanical tests).
    • Animal Study: Not specified how many animals were used.
  • Data Provenance: Not explicitly stated (e.g., country of origin), but implied to be conducted by the manufacturer for the 510(k) submission. These would be considered prospective tests conducted specifically for this regulatory submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as this is a device for surgical repair, and the "ground truth" is based on physical and biological performance rather than expert interpretation of data. The ground truth for equivalence is established by direct comparison of physical and mechanical properties and biological response in animal models.

4. Adjudication method for the test set:

  • Not applicable for the types of in-vitro and in-vivo (animal) studies described, as these are objective measurements and observations rather than expert adjudication of clinical outcomes.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a surgical implant, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a surgical implant; there is no "algorithm" in the sense of a software-driven process.

7. The type of ground truth used:

  • In Vitro Testing: Objective physical and chemical measurements (e.g., viscosity, mechanical strength, material composition).
  • In Vivo Testing (Animal Study): Safety and efficacy observations in an animal model, likely assessed by veterinarians/researchers.
  • Equivalence: Direct comparison to known properties and indications of established predicate devices.

8. The sample size for the training set:

  • Not applicable. This is not a machine learning or AI-driven device requiring a training set. The "training" here would be the scientific and engineering principles used to design and manufacture the device.

9. How the ground truth for the training set was established:

  • Not applicable for the reason stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The sequence reads 'K031785'. The characters are written in a bold, somewhat stylized manner, with varying stroke thicknesses. The overall impression is that of a handwritten code or identifier.

510(k) SUMMARY

MacroPore Cardio-Wrap (TS)

Page 1 of 3

ADMINISTRATIVE INFORMATION

Manufacturer Name: MacroPore Biosurgery, Inc. 6740 Top Gun Street San Diego, CA 92121 Official Contact: Kenneth K. Kleinhenz Director of Regulatory Affairs

DEVICE NAME

Classification Name:

Intracardiac Patch

Fax (858) 458-0994

Trade/Proprietary Name:

MacroPore Cardio-Wrap (TS)

Telephone (858) 458-0900

ESTABLISHMENT REGISTRATION NUMBER 2031733

DEVICE CLASSIFICATION AND PRODUCT CODE

As shown in 21CFR 870.3470, an Intracardiac Patch is a fabric device intended to be placed in the heart for use to repair septal defects, for patch grafting, and to repair tissue These devices are classified as Class II. Intracardiac Patches have been assigned Product Code DXZ.

INTENDED USE

The MacroPore Cardio-Wrap (TS) is indicated for use as a pericardium replacement device in patients that may require re-operation within six months.

{1}------------------------------------------------

DEVICE DESCRIPTION

Design Characteristics

MacroPore Cardio-Wrap (TS) is a resorbable implant in sheet form manufactured from poly lactic acid (PLA). MacroPore Cardio-Wrap (TS) can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Cardio-Wrap (TS) to the desired shape or size. MacroPore Cardio-Wrap (TS) Sheet is fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPore Cardio-Wrap (TS) can be used either alone or in conjunction with soft tissue fixation devices such as resorbable sutures, which can also serve to fixate the MacroPore Cardio-Wrap (TS) and prevent dislocation. The MacroPore Cardio-Wrap (TS) may be used in conjunction with various MacroPore manual instruments.

MacroPore Cardio-Wrap (TS) is provided in various shapes such as rectangles, ovals, and circles and will be provided in other shapes and sizes as needed for particular surgical procedures. MacroPore Cardio-Wrap (TS) is provided in sheets of 10mm to 500mm x 500mm and will be provided in other shapes and sizes as needed for particular surgical procedures. The thickness of the MacroPore Cardio-Wrap (TS) ranges from 0.05 mm to 1.0 mm according to the region to be treated. The MacroPore Cardio-Wrap (TS) is provided in solid sheets. The borders of the sheets may be aligned with holes to attach suture material.

Material Composition

The MacroPore Cardio-Wrap (TS) is fabricated from polylactic acid (PLA).

In Vitro Testing

The MacroPore Cardio-Wrap (TS) is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures. Therefore, testing was performed to determine the effect of prolonged heating in saline at 60°C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. The relatively brief exposure anticipated during the surgical preparation of MacroPore Cardio-Wrap (TS) is not expected to have a significant effect on its mechanical properties.

Aging testing was performed on MacroPore Cardio-Wrap (TS). Testing demonstrated that the MacroPore Cardio-Wrap (TS) is strong enough for the indications for use.

Mechanical testing was performed on the MacroPore Cardio-Wrap (TS) which determined the MacroPore Cardio-Wrap (TS) to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions,

{2}------------------------------------------------

In Vivo Testing

An animal study was conducted to demonstrate safety and efficacy of the MacroPore Cardio-Wrap (TS) material. The animal studies demonstrated that the MacroPore Cardio-Wrap (TS) materials are safe and efficacious for the indications for use.

EQUIVALENCE TO MARKETED PRODUCT

The MacroPore Cardio-Wrap (TS) shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to preamendment devices: Bio-Vascular Supple Peri-Guard. Sulzer Carbomedics Cardio Fix, and W.L. Gore Preclude Pericandial Membrane; Class II medical devices that were cleared for marketing in the United States under K983602, K993288, and K012098, respectively.

Indications For Use

The MacroPore Cardio-Wrap (TS) shares identical indications for use principles with the predicate devices as both the MacroPore Cardio-Wrap (TS) and the predicate devices are indicated for the same surgical procedures.

Design and Materials

The physical designs of MacroPore Cardio-Wrap (TS) and the predicate devices (Bio-Vascular Supple Peri-Guard, Sulzer Carbomedics Cardio Fix, and W.L. Gore Preclude Pericardial Membrane) are substantially equivalent, consisting of thin semi-rigid sheets. The MacroPore Cardio-Wrap (TS) and the predicates also share design features of allowing for contouring. The MacroPore Cardio-Wrap (TS) is fully contourable when heated to approximately 55℃. The thickness of the predicate devices and the MacroPore Cardio-Wrap (TS) are substantially equivalent as the thinnest MacroPore Cardio-Wrap (TS) thickness is virtually identical to the predicate (0.5mm vs 1.0mm). The dimensions of the predicate devices are also comparable to the MacroPore Cardio-Wrap (TS) as both devices are provided in rectangular sheets that are several centimeters in size. The mechanical characteristics of the MacroPore Cardio-Wrap (TS) are substantially equivalent to the predicate devices with respect to mechanical strength as measured by tensile and suture pull out testing. In addition to physical characteristics, both the predicate device and the MacroPore Cardio-Wrap (TS) can be cut to specific shapes and sizes by the end user.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged around the upper portion of the circle.

FEB 23 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Macropore Biosurgery, Inc. c/o Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs 6740 Top Gun Street San Diego, CA 92121

Re: K031785

Trade Name: MacroPore Cardio-Wrap (TS) Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac patch or pledget Regulatory Class: Class II (two) Product Code: OMH Dated: June 9, 2003 Received: June 10, 2003

Dear Mr. Kleinhenz:

This letter corrects our substantially equivalent letter of September 4, 2003.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Kenneth K. Kleinhenz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Dmna R. bohner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Device Name: MacroPore Cardio-Wrap (TS)

Indications for Use:

The MacroPore Cardio-Wrap (TS) is indicated for use as a pericardium replacement device in patients that may require re-operation within six months.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)

40.0~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

・・

Veleta

510(k) Num

ున

§ 870.3470 Intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene.

(a)
Identification. An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.(b)
Classification. Class II (performance standards).