Intended Use

The EBI® SpineLink® Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation.

Device Description

The EBI® SpineLink® Spinal Fixation System is a spinal fixation device that utilizes interconnecting links, screws (fixed and polydirectional designs), transverse connectors, hooks, endcaps, locknuts, and link ties. This submission is for a modification to the Indications for Use to include degenerative disc disease (ddd).

AI/ML Overview

The provided text describes a 510(k) Summary for the EBI® SpineLink® Spinal Fixation System, which is a spinal fixation device. This document is a pre-market notification to the FDA to demonstrate substantial equivalence to legally marketed predicate devices, not a study proving the device meets specific acceptance criteria in the sense of a clinical trial or performance study determining diagnostic accuracy or similar statistical measures.

Therefore, many of the requested categories for a study proving acceptance criteria are not applicable in this context. The "acceptance criteria" here refer to the regulatory requirement of demonstrating substantial equivalence to predicate devices, primarily through comparison of intended use, materials, and function, supported by mechanical testing.

Here's an analysis based on the information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied by 510(k) process)Reported Device Performance
Intended Use Equivalence: Device intended use aligns with predicate devices.The EBI® SpineLink® Spinal Fixation System's intended use, including the expanded indication for degenerative disc disease, is compared to predicate devices: Spinal Concepts BacFix Spinal Fixation System, Sofamor Danek CD Spinal System, and Sofamor Danek Townley Pedicle Screw Plating System. The document states: "It is substantially equivalent to the predicate devices in regards to intended use."
Materials Equivalence: Device materials are equivalent to predicate devices.Components are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136. The document states: "It is substantially equivalent to the predicate devices in regards to...materials."
Functional Equivalence/Performance: Device functions similarly and meets applicable standards, comparable to predicate devices.Mechanical testing was performed comparing the EBI® SpineLink® Spinal Fixation System to two predicate systems. The document states: "Mechanical testing comparing the System to two predicate systems demonstrates that the device complies with applicable standards and meets all of its functional requirements."

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the document. The document refers to "mechanical testing" but does not specify the sample size of tested devices or the details of the test set. Given this is a device for human implantation, the "test set" would refer to the physical devices undergoing mechanical tests, not patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided and is generally not applicable to the type of regulatory submission described. Mechanical testing typically involves engineers and lab technicians applying standardized tests, not medical experts establishing "ground truth" on cases.

4. Adjudication Method for the Test Set

This information is not provided and is not applicable. Mechanical testing results are typically objectively measured against pre-defined engineering and material standards, not adjudicated by experts in the context of diagnostic performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging or AI devices where human readers interpret medical images. The EBI® SpineLink® Spinal Fixation System is a physical medical device (spinal implant), and MRMC studies are not applicable.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone performance study (in the context of an algorithm's performance) was not done. This is not an AI/algorithm-based device. The "performance" assessment here refers to the physical device's mechanical integrity as determined by testing.

7. Type of Ground Truth Used

The "ground truth" for demonstrating substantial equivalence and functional requirements in this context is based on:

  • Established engineering and material standards: ASTM F136 for titanium alloy.
  • Mechanical test results: Comparing the device's performance against industry standards and the performance of similar predicate devices.
  • Regulatory definitions: Such as the definition of degenerative disc disease for the expanded indication.

8. Sample Size for the Training Set

This information is not applicable and not provided. As this is a physical medical device and not an AI or machine learning algorithm, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided for the same reasons as #8.

{0}------------------------------------------------

JUL 2 2 2 2003

K031424
page 1 of 2

510(k) Summary

This 510(k) Summary for the EBI® SpineLink® Spinal Fixation System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.

  • Contact Person: Peter Allan 1. Submitter: EBI, L.P. 100 Interpace Parkway Telephone: (973) 299-9300, ext.3329 Parsippany, NJ 07054
    Date prepared: July 10, 2003
2. Proprietary Name:EBI® SpineLink® Spinal Fixation System
Common Name:Spinal Fixation Device
Classification Names:Spinal Intervertebral Body Fixation Orthosis

3. Predicate or legally marketed devices that are substantially equivalent:

  • Spinal Concepts BacFix Spinal Fixation System .
  • . Sofamor Danek CD Spinal System
  • Sofamor Danek Townley Pedicle Screw Plating System .
    1. Description of the device: The EBI® SpineLink® Spinal Fixation System is a spinal fixation device that utilizes interconnecting links, screws (fixed and polydirectional designs), transverse connectors, hooks, endcaps, locknuts, and link ties. This submission is for a modification to the Indications for Use to include degenerative disc disease (ddd).
    1. Intended Use: The EB1® SpineLink® Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The

{1}------------------------------------------------

K031424
page 2 of 2

device is indicated for all of the following indications regardless of use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation.

    1. Materials: The components of the System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136.
    1. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the EBI® SpineLink® Spinal Fixation System and other legally marketed spinal systems. It is substantially equivalent* to the predicate devices in regards to intended use, materials, and function. Mechanical testing comparing the System to two predicate systems demonstrates that the device complies with applicable standards and meets all of its functional requirements.
  • Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or reclassification and is not interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 2 2003

Mr. Peter Allan Regulatory Affairs Associate EBI, L.P. 100 Interpace Parkway Parsippany, New Jersey 07054

Re: K031424

Trade/Device Name: EBI® Spinelink® Spinal Fixation System Regulatory Number: 21 CFR 888.3070(b), 888.3050, 888.3060 Regulation Name: Pedicle screw spinal system, Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis Regulatory Class: III Product Code: NKB, MNI, MNH, KWP, KWQ Dated: May 5, 2003 Received: May 6, 2003

Dear Mr. Allan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Peter Allan

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark N Millener

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Statement of Indications for Use

510(k) Number: K031424

Page I of

Device Name: EBI® SpineLink® Spinal Fixation System

Indications For Use:

The EBI® SpineLink® Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacraViliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

Mark N. Millburn

ivision of General, Restorative and Neurological Devices

510(k) Number K031424