K Number
K031029

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-06-19

(79 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution, Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization.

Device Description

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar, high frequency electrosurgical system consisting of three components: an electrosurgical generator called the Controller; a disposable, bipolar, single use Loop Electrode; and a reusable Patient Cable.

AI/ML Overview

PMA/510(k) summary does not contain the performance data. No acceptance criteria mentioned.

{0}------------------------------------------------

JUN 1 9 2003

510(k) Summary

ArthroCare Corporation ArthroCare® Bipolar Loop Electrosurgery System

General Information

Submitter Name/Address:ArthroCare Corporation680 Vaqueros AvenueSunnyvale, CA 94085-2936
Phone Number:(408) 736-0224
Contact Person:Valerie Defiesta-NgDirector, Regulatory Affairs
Date Prepared:March 31, 2003

Device Description

Trade Name:ArthroCare Bipolar Loop ElectrosurgerySystem
Generic/Common Name:Electrosurgical Device and Accessories
Classification Name:Electrosurgical Cutting and CoagulationDevice and Accessories (21 CFR 878.4400)

Predicate Devices

ArthroCare Bipolar LoopElectrosurgery SystemK955531, K010568, K020832 and K022543
PlasmaKinetic EndourologyGeneratorK003569

Product Description

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar, high frequency electrosurgical system consisting of three components: an electrosurgical generator called the Controller; a disposable, bipolar, single use Loop Electrode; and a reusable Patient Cable.

{1}------------------------------------------------

Intended Uses

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution. Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization.

Substantial Equivalence

Based on the indications for use statement and technological characteristics to the predicate devices, the ArthroCare Bipolar Loop Electrosurgery System has shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.

Summary of Safety and Effectiveness

The Bipolar Loop Electrosurgery System, as described in this 510(k), is substantially equivalent to the predicate device with respect to the indications for use and technological characteristics. The modification of the indications for use statement does not raise any new issues of safety or efficacy. There are no changes or modifications being made to the technology, principle of operation, materials, packaging and sterilization parameters that were previously cleared 510(k)s.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

Public Health Service

JUN 1 9 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Valerie Defiesta-Ng Director, Regulatory Affairs ArthroCare Corporation 680 Vaqueros Avenue SUNNYVALE CA 94085-2936

Re: K031029

Trade/Device Name: ArthroCare Bipolar Electrosurgery System, Model TBD Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Codes: 78 FAS and KNS Dated: March 31, 2003 Received: May 13, 2003

Dear Ms. Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

ArthroCare® Bipolar Loop Electrosurgery System Device Name

510(k) Number: K 03/029

Indications for Use:

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution, Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-the-Counter Use

(Per 21 CFR 801.109)

David R. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devi-

X

510(k) Number.

ix

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).