K Number
K030982

Validate with FDA (Live)

Date Cleared
2003-04-08

(11 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

syngo Colonography is a self-contained image analysis software package for evaluating CT volume data sets. This software package can also be utilized for evaluating suitable MR volume datasets. Combining enhanced commercially available digital image processing tools with optimized workflow and reporting tools, the software is designed to support the physician in studying the inside (intra-lumenal view), the wall and the outside (extra-lumenal view) of the colon. With the functionality to view datasets from both the prone and supine positions, it facilitates the detection of colonic lesions (eg. Polyps) in addition to the evaluation, documentation and follow-up of any such lesions using standard spiral CT or MR scanning. This evaluation tool allows for volumetric analysis of colonic polyps or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions. shape and position.

Due to all these capabilities the syngo Colonography software has the advantage of non-invasive evaluation of colonic lesions as compared to conventional colonoscopy.

Device Description

syngo Colonography is a self-contained image analysis software package for evaluating CT volume data sets. This software package can also be utilized for evaluating suitable MR volume datasets. Combining enhanced commercially available digital image processing tools with optimized workflow and reporting tools, the software is designed to support the physician in studying the inside (intra-lumenal view), the wall and the outside (extra-lumenal view) of the colon. With the functionality to view datasets from both the prone and supine positions, it facilitates the detection of colonic lesions (eg. Polyps) in addition to the evaluation, documentation and follow-up of any such lesions using standard spiral CT or MR scanning. This evaluation tool allows for volumetric analysis of colonic polyps or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.

AI/ML Overview

I apologize, but the provided text does not contain detailed information about acceptance criteria or specific study results that prove the device meets said criteria.

The document is a 510(k) summary for the Syngo Colonography software package, primarily focused on establishing substantial equivalence to previously cleared predicate devices. It describes the device's intended use, general safety and effectiveness concerns, and its relationship to other Siemens software.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I answer questions regarding:

  • Sample size used for the test set and data provenance.
  • Number of experts used to establish ground truth and their qualifications.
  • Adjudication method for the test set.
  • Multi-reader multi-case (MRMC) comparative effectiveness study results or effect size.
  • Whether a standalone performance study was done.
  • The type of ground truth used.
  • Sample size for the training set.
  • How ground truth for the training set was established.

This type of detailed study information is typically found in the full 510(k) submission, not necessarily in the public summary. The summary focuses on the regulatory aspects of equivalence rather than a detailed performance report against specific acceptance criteria.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

K030982

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

I. GENERAL INFORMATION

Device Name and Classification

Product Name:Syngo Colonography software package
Common Name3D Reconstruction Software
Classification Name:Accessory to Computed Tomography System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750 2050
Device Class:Class II
Product Code:90 JAK LLZ

Establishment:

Importer/Distributor:

Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

Registration Number: 2240869

Manufacturing Facility:

Siemens AG Medical Solutions Henkestrasse 127 D-91052 Erlangen, Germany syngo is a registered trademark of Siemens AG

Contact Person: Mr. Jamie Yieh Senior Technical Specialist Telephone: (610) 448-1785 Fax: (610) 448-1787

Date of Preparation of Summary: November 26th 2002

II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING SUBSTANTIAL THE EQUIVALENCE DETERMINATION

Device Description and Intended Use:

This premarket notification covers Siemens Syngo Colonography software package. It is based on Siemens syngo software platform.

syngo Colonography is a self-contained image analysis software package for evaluating CT volume data sets. This software package can also be utilized for evaluating suitable MR volume datasets. Combining enhanced commercially available digital image processing tools

{1}------------------------------------------------

with optimized workflow and reporting tools, the software is designed to support the physician in studying the inside (intra-lumenal view), the wall and the outside (extra-lumenal view) of the colon. With the functionality to view datasets from both the prone and supine positions, it facilitates the detection of colonic lesions (eg. Polyps) in addition to the evaluation, documentation and follow-up of any such lesions using standard spiral CT or MR scanning. This evaluation tool allows for volumetric analysis of colonic polyps or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions, shape and position.

Due to all these capabilities the syngo Colonography software has the advantage of noninvasive evaluation of colonic lesions as compared to conventional colonoscopy.

General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warning, to provide for safe and effective use of the device.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

Substantial Equivalence:

The Syngo Colonography software package, addressed in this premarket notification, is substantially equivalent to the following commercially available software package:

Predicate Device NameFDA ClearanceNumberFDA ClearanceDate
GE CT Colonography/Navigator 2WorkstationK01231308/07/01
Siemens Fly ThroughK97171709/03/97
Siemens RealTime 3D DiagnosticWorkstation (3D Virtuoso)K97301011/10/97

The Syngo Colonography software package described in this 510(k) has the same intended use and similar technical characteristics as the commercially available software listed above.

In addition, many of the image processing, display and evaluation components of syngo Colonography are currently available on software options like the Volume Rendering Technique option, K923524/S2, cleared on May 17th 1994 and workstations like the syngo Multimodality Workstation, K010938 cleared on 26th June 2001 wherein the Fly Thorough software algorithms were transferred over to the syngo software platform. syngo Colonography packages these image processing and image display components in an optimized workflow palette.

In summary, Siemens is of the opinion that Syngo Colonography software package does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate software components and the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three parallel lines that curve upwards and resemble wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 0 8 2003

Siemens Medical Solutions, Inc. % Mr. Heinz Joerg Steneberg Division Manager, Medical Division TUV Rheinland of North America 12 Commerce Road NEWTON CT 06470

Re: K030982

Trade/Device Name: Syngo Colongraphy Software Package Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: March 26, 2003 Received: March 28, 2003

Dear Mr. Steneberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for use

KO30982 510(k) Number (if known):

Device Name:

syngo Colonography Software Package

syngo Colonography is a self-contained image analysis software package for evaluating CT volume data sets. This software package can also be utilized for evaluating suitable MR volume datasets. Combining enhanced commercially available digital image processing tools with optimized workflow and reporting tools, the software is designed to support the physician in studying the inside (intra-lumenal view), the wall and the outside (extra-lumenal view) of the colon. With the functionality to view datasets from both the prone and supine positions, it facilitates the detection of colonic lesions (eg. Polyps) in addition to the evaluation, documentation and follow-up of any such lesions using standard spiral CT or MR scanning. This evaluation tool allows for volumetric analysis of colonic polyps or lesion size over time, helping the Physician to assess the changes in their growth. It is also designed to help the physician classify conspicuous regions of tissue unambiguously, with respect to their size, dimensions. shape and position.

Due to all these capabilities the syngo Colonography software has the advantage of non-invasive evaluation of colonic lesions as compared to conventional colonoscopy.

(Please do not write below this line - continue on another page if needed)

Concurrence of the CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal,

and Radiological Devices

510(k) NumberK030982
------------------------
Prescription Use (Per 21 CFR 801.109)OR Over-The-Counter Use
----------------------------------------------------------------

N/A