K Number
K030960

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-07-16

(111 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACMI Electronic Video CystoNephroscope (ECN) System is a flexible endoscopic surgical video system consisting of an endoscope, CMOS video sensor and light source mounted in the endoscope, and a Controller unit. This system is intended for use to examine body cavities, hollow organs and canals in the body, in the urinary tract, and can be used percutaneously to examine the interior of the kidney; and, using additional accessories, can be used to perform various diagnostic and therapeutic procedures

Device Description

The ECN Video Cystonephroscope is a flexible endoscope that incorporates video sensor technology to capture the endoscopic image, replacing the fiber optic image bundle typically used in most endoscopes. The ECN Video Cystonephroscope can be introduced either through the urethra into the bladder or through a percutaneous tract into the abdominal cavity or kidney

AI/ML Overview

This 510(k) premarket notification for the ACMI ECN Video Cystonephroscope does not contain the detailed information required to answer the questions about acceptance criteria and a study that proves the device meets those criteria.

The document is a submission to the FDA for substantial equivalence to legally marketed predicate devices, not a report on a clinical or performance study with defined acceptance criteria. It primarily focuses on comparing the new device's technological characteristics and intended use to existing devices.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, expert qualifications, or MRMC studies.

Here's why the information is missing:

  • 510(k) Premarket Notification: This type of submission aims to demonstrate that a new device is as safe and effective as a legally marketed predicate device, primarily through comparison of technological characteristics and intended use. It often does not require new clinical trials or performance studies with specific statistical endpoints like those requested in your prompt, unless there are significant differences from the predicate that raise new questions of safety or effectiveness.
  • Focus on Substantial Equivalence: The document explicitly states its purpose is to show "Technological Characteristics and Substantial Equivalence." It lists predicate devices and highlights how the ECN Flexible Video Cystonephroscope utilizes existing flexible endoscope technology, video sensor technology (CMOS), and video processing technology found in those predicates.
  • Lack of Performance Data: The provided text does not include any performance metrics, success criteria, or results from a study designed to quantify the device's accuracy, sensitivity, specificity, or inter-reader agreement.

In summary, the provided document is a regulatory filing for market clearance based on substantial equivalence, not a detailed performance study with acceptance criteria.

{0}------------------------------------------------

Premarket Notification ECN Video Cystonephroscope

JUL 16 2003

SECTION 10

K 030960 pg 1 of 2

510(k) SUMMARY ACMI ECN Video Cystonephroscope

    1. ACMI Corporation Sponsor: 136 Turnpike Road Southborough, MA 01771-2104
      Contact: Frank J. Fucile. Telephone: 508-804-2632 Date: March 25, 2003
    1. Device Name:
      Proprietary Name: Common/Usual Name: Classification Name:

ACMI ECN Video CystoNephroscope System Endoscope, Video Camera and accessories Endoscope, Surgical Camera and accessories

    1. Predicate Devices:
      ACMI USA Series™ ACN™-2 Flexible CystoNephroscope (K904797)

ACMI USA Series™ APN™ -2 Flexible Choledochoscope (K012951)

Dyonics Vision 111 Mobile Video Unit, Smith & Nephew, Inc., (K010489)

Visera Cystovideoscope, Olympus (K020174)

Pentax ECY-1530 Video Choledoconephroscope, Pentax America, (probably covered under K951191 and internal letter to file adding video)

  • Device Description: 3.
    The ECN Video Cystonephroscope is a flexible endoscope that incorporates video sensor technology to capture the endoscopic image, replacing the fiber optic image bundle typically used in most endoscopes. The ECN Video Cystonephroscope can be introduced either through the urethra into the bladder or through a percutaneous tract into the abdominal cavity or kidney

    1. Intended Use:
      The ACMI Electronic Video CystoNephroscope (ECN) System is a flexible endoscopic surgical video system consisting of an endoscope with a CMOS video sensor and light source mounted in the endoscope, and a Controller unit. This system is intended for use to examine body cavities, hollow organs and canals in the body, in the urinary tract, and can be used percutaneously to examine the interior of the kidney; and, using additional accessories, can be used to perform various diagnostic and therapeutic procedures

{1}------------------------------------------------

K 0309 60
pg 2 of 2

Technological Characteristics and Substantial Equivalence: న.

The ECN Flexible Video Cystonenhroscope is substantially equivalent to features incorporated into the following legally marketed predicate devices:

The ECN Flexible Video Cystonephroscope utilizes the flexible endoscope technology (design and materials) of the following ACMI flexible endoscopes:

  • ACMI USA Series™ ACN-2 Flexible CystoNephroscope (K904797)

  • l ACMI USA Series™ APN -2 Flexible Choledochoscope (K012951)

The ECN Flexible Video Cystonephroscope incorporates the video sensor technology located in the distal tip of the endoscope similar to that utilized in the following devices:

  • Visera Cystovideoscope CYF Type V/VA, Olympus America, Inc., (K021074) 해
  • 1 Pentax ECY-1530 Video Choledoconephroscope, Pentax America, (probably covered under K951191 and internal letter to file adding video)

The ECN Flexible Video Cystonephroscope incorporates the CMOS sensor technology similar to that used in the Dyonics Vision 111 Mobile Video Unit, Smith & Nephew, Inc., (K010489).

Finally, the ECN Flexible Video Cystonephroscope uses video processing technology (hardware and software) similar to that used in the ACMI MicroDigital® IP 6.2 Controller (510K exempt).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines above a wavy line.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 16 2003

Mr. John A. DeLucia VP. OA/RA/CA ACMI Corporation 136 Turnpike Road SOUTHBOROUGH MA 01772

Re: K030960

Trade/Device Name: ACMP Electronic Video CystoNephroscope (ECN) System Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Product Codes: 78 FAJ and FGA Regulation Number: 21 CFR §878.4160 Regulation Name: Surgical camera and accessories Product Code: 80 FWF Dated: June 25, 2003 Received: June 27, 2003

Dear Mr. DeLucia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SECTION 8

K030960

Indications for Use Statement

030960 510(k) Number (if known):

Device Name: ACMI® Electronic Video CystoNephroscope (ECN) System

Indications For Use:

The ACMI Electronic Video CystoNephroscope (ECN) System is a flexible endoscopic surgical video system consisting of an endoscope, CMOS video sensor and light source mounted in the endoscope, and a Controller unit. This system is intended for use to examine body cavities, hollow organs and canals in the body, in the urinary tract, and can be used percutaneously to examine the interior of the kidney; and, using additional accessories, can be used to perform various diagnostic and therapeutic procedures

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use or Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Per 21 CFR 801.109)

Nancy C. Hogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices

510(k) Number R030760

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.