K Number
K030783

Validate with FDA (Live)

Date Cleared
2003-04-04

(23 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The indended use of the Frontier Power wheelchair is to provide mobility to persons limited to a sitting position, who have the capability of operating a power wheelchair

Device Description

The Frontier Power Wheelchair is a battery powered, electric motor driven device with the intended function of providing mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. It is a mid wheel drive format wheelchair as it's basic configuration with direct drive motors and four swivel castors. The wheelchair frame is of a non-folding type with a mid mounted drive wheels on individual, front pivoting extension arms ending in front mounted swivel castors. The rear of the main frame ends with a lateral pivoting beam arm with swivel castors mounted on each side. This frame design allows for all 6 wheels to contact the ground over uneven terrain.

The frame consists of a combination of 2" diameter round steel tube. 1 ¼" diameter round steel tube and 1 ¼" square tube. The wheelchair consists of the following basic components: frame with articulated drive wheels and casters, seat, armrests, and footrests.

AI/ML Overview

The provided document, K030783, describes a 510(k) premarket notification for the "Frontier Power Wheelchair." This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove acceptance criteria for an AI/ML powered medical device. Therefore, much of the requested information (like sample sizes for test/training sets, expert qualifications, and MRMC studies) is not applicable to this type of submission.

However, I can extract information related to the performance testing conducted and how it relates to general acceptance.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryStandard AppliedReported Device Performance
Performance TestingISO/CD 7176-21"In all instances, the Frontier Power Wheelchair met the required performance criteria and functioned as intended."
Performance TestingANSI/RESNA Vol.2 Section 21 Amendments for powered wheelchairs and motorized scooters"In all instances, the Frontier Power Wheelchair met the required performance criteria and functioned as intended."

2. Sample Size Used for the Test Set and Data Provenance

This information is not applicable as the document describes physical performance testing of a wheelchair, not an AI/ML algorithm requiring a test set of data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the document describes physical performance testing of a wheelchair, not an AI/ML algorithm requiring expert-established ground truth. The "ground truth" here would be the physical measurements and observations during the standardized tests.

4. Adjudication Method for the Test Set

This information is not applicable. The performance testing was conducted according to established international and national standards (ISO/CD 7176-21 and ANSI/RESNA Vol.2 Section 21 Amendments). These standards define specific test procedures and criteria, eliminating the need for an adjudication method in the context of expert review.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating human reader performance with and without AI assistance, which is not applicable to a power wheelchair.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. This device is a power wheelchair, a physical product, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance testing was established by objective measurements and observations according to the detailed methodologies outlined in the referenced international and national standards: ISO/CD 7176-21 and ANSI/RESNA Vol.2 Section 21 Amendments for powered wheelchairs and motorized scooters.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" in the context of a physical power wheelchair being tested for performance against established standards.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K030783

APR 0 4 2003

510(k) Summary Innovation In Motion 510(k) Premarket Notification Frontier Power Wheelchair

Applicant's Name, Address, Telephone, and Fax Numbers

Innovation In Motion 900 Growth Parkway PO Box 507 Angola, IN 46703 Phone 260.668.5677 Fax: 260.668.8967

Preparer's Name, Address, Telephone, Fax Number, Contact Name, and Date Prepared

Innovation In Motion 900 Growth Parkway PO Box 507 Angola, IN 46703 Phone 260.668.5677 Fax: 260.668.8967

Contact Person: Rick Michael North America Sales Manager

Date Prepared: February 2003

Manufacturer's Name, Address, Telephone, and Fax Numbers

Magic International Pty. Ltd. 2 / 16 Viewtech Place Rowville, Victoria 3178 Australia Phone: 011.613.9755.8100 Fax: 011.613.9755.8111

Name of Device and Name/Address of Sponsor:

Frontier Power Wheelchair

Innovation In Motion 900 Growth Parkway PO Box 507 Angola, IN 46703 Phone 260.668.5677 Fax: 260.668.8967

{1}------------------------------------------------

Common or Usual Name

Power Wheelchair

Classification Name Wheelchair, Powered

Predicate Devices

The product that is substantially equivalent to the Frontier Power Wheelchair is Invacare's Xterra GT Power Wheelchair (K012909, October 24, 2001).

Intended Use

The intended use of the Frontier Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair.

Technological Characteristics and Substantial Equivalence

Device Description:

The Frontier Power Wheelchair is a battery powered, electric motor driven device with the intended function of providing mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. It is a mid wheel drive format wheelchair as it's basic configuration with direct drive motors and four swivel castors. The wheelchair frame is of a non-folding type with a mid mounted drive wheels on individual, front pivoting extension arms ending in front mounted swivel castors. The rear of the main frame ends with a lateral pivoting beam arm with swivel castors mounted on each side. This frame design allows for all 6 wheels to contact the ground over uneven terrain.

The frame consists of a combination of 2" diameter round steel tube. 1 ¼" diameter round steel tube and 1 ¼" square tube. The wheelchair consists of the following basic components: frame with articulated drive wheels and casters, seat, armrests, and footrests.

Substantial Equivalence:

Both of these products are battery powered wheelchairs with the same intended function and use-the provision of mobility to persons limited to a sitting position that have the capability of operating a powered wheelchair. Similarities include large mid-wheels with attached motor/gearbox drive mechanisms, smaller pivoting casters for turning, and joystick operated motor controllers to engage system motion and steer the wheelchair. They are all constructed from the same basic materials, have the same basic operational principles, and all use DC batteries as their source of power.

The Frontier Power Wheelchair is substantially equivalent to Invacare's Xterra GT Power Wheelchair (ZX2) (K012909, October 24, 2001).

{2}------------------------------------------------

Performance Data:

As required by FDA's July 26, 1995 draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Application for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles", the Frontier Power Wheelchair was tested in accordance with ISO/CD 7176-21 and ANSI/RESNA Vol.2 Section 21 Amendments for powered wheelchairs and motorized scooters. In all instances, the Frontier Power Wheelchair met the required performance criteria and functioned as intended.

. . . .

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name encircling an abstract image of an eagle. The eagle is depicted with stylized lines, suggesting movement or flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 0 4 2003

Mr. Rick Michael North America Sales Manager Vestil Manufacturing Corporation 900 Growth Parkway Angola, IN 46703

Re: K030783

Trade/Device Name: Frontier Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: March 25, 2003 Received: March 26, 2003

Dear Mr. Michael:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Rick Michael

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours

Mark N. Mellem

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page_1__of__1_

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name:Frontier Power Wheelchair__

Indications For Use:

The indended use of the Frontier Power wheelchair is to provide mobility to persons limited to a sitting position, who have the capability of operating a power wheelchair

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for
(Division Sign-Off)
Division of General. Restorative
and Neurological Devices

510(k) NumberK030783
------------------------
OR Over-The-Counter Use ✓
------------------------------------------------------------------------------

ption Use
CFR 801.109)(Optional Format 1-2-96)

Prescrip (Per 21 C.


B 3

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).