K Number
K030718

Validate with FDA (Live)

Date Cleared
2003-06-05

(90 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SCB Computer Motion Interface Controller is indicated for use with the Storz Communication Bus (SCB) manufactured by Karl Storz to connect HERMES-Ready™ devices to the SCB.

Device Description

The SCB Computer Motion Interface Controller (SCIC) connects devices that are controlled by Computer Motion's voice-control system to the Storz Communication Bus (SCB) manufactured by Karl Storz.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called the "SCB Computer Motion Interface Controller" (SCIC). It focuses on establishing substantial equivalence to a predicate device and demonstrating adherence to various electrical and safety standards.

There is no information in the provided document about:

  • Acceptance criteria for clinical performance or diagnostic accuracy. The document exclusively discusses electrical safety and functional equivalence.
  • A clinical study involving human patients or data to prove device performance in terms of diagnostic or clinical outcomes.
  • Sample sizes for test sets or training sets.
  • Data provenance (country of origin, retrospective/prospective).
  • Number of experts or their qualifications for establishing ground truth.
  • Adjudication methods.
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
  • Standalone algorithm performance.
  • Types of ground truth (pathology, outcomes data).

The "Device Testing" section only mentions "Hazard analyses and hardware/software validations procedures were provided in this submission." This indicates that the testing focused on engineering and functional validation, not clinical performance.

Based on the provided text, the only "acceptance criteria" and "device performance" described are related to compliance with electrical and medical device safety standards.

Here's a summary of what is available regarding acceptance criteria and performance:

1. Table of Acceptance Criteria and Reported Device Performance

Test / StandardTitleDevice Performance (Implied)
IEC 601-1International Standard for Medical Electrical EquipmentCompliant (Implied by submission for 510(k))
IEC 601-1-1International Standard for Medical Electrical EquipmentCompliant (Implied by submission for 510(k))
IEC 601-2-18International Standard for Medical Electrical EquipmentCompliant (Implied by submission for 510(k))
UL 2601-1Underwriters Laboratory (Standard for Medical Electrical Equipment)Compliant (Implied by submission for 510(k))
EN55011/A1 CISPR 11Conducted EmissionCompliant (Implied by submission for 510(k))
EN55011/A1 CISPR11Radiated EmissionCompliant (Implied by submission for 510(k))
EN61000-4-2Electrostatic DischargeCompliant (Implied by submission for 510(k))
EN61000-4-3 and EN50140RF ImmunityCompliant (Implied by submission for 510(k))
EN61000-4-4EFT/Bursts ImmunityCompliant (Implied by submission for 510(k))
EN61000-4-5Surge ImmunityCompliant (Implied by submission for 510(k))
EN61000-4-6Conducted ImmunityCompliant (Implied by submission for 510(k))
EN60601-1International Standard for Medical Electrical EquipmentCompliant (Implied by submission for 510(k))
EN60601-1-1General Requirements for Safety - Collateral StandardCompliant (Implied by submission for 510(k))
EN 60601-1-2Emissions and Immunity Test MeasurementsCompliant (Implied by submission for 510(k))
CAN/CSA-C22.2 No. 601.1Medical Electrical Equipment Part 1, General Requirements for Safety, General Instructions Part 1Compliant (Implied by submission for 510(k))
VA-19795CMI Environmental TestingCompliant (Implied by submission for 510(k))
Substantial Equivalence to HERMES® Port ExpanderFunctional equivalence in extending connectivity from a central control system for medical devices.Achieved (FDA determined substantially equivalent)
Hazard analyses and hardware/software validationsSafety and effectiveness of the device when connecting HERMES-Ready™ devices to the Storz Communication Bus.Demonstrated safety and effectiveness

Study Demonstrating Acceptance Criteria:

The "study" or rather, the evidence provided to demonstrate that the device meets its acceptance criteria, consists of the following:

  • Compliance with various international and national electrical and medical device safety standards (IEC, UL, EN, CISPR, CSA). The document explicitly lists these tests that the SCIC was subjected to.
  • Hazard analyses and hardware/software validation procedures. These were submitted as part of the 510(k) application.
  • Establishment of substantial equivalence to the predicate device (HERMES® Port Expander) in terms of intended use and technological characteristics (providing additional device connection ports). The FDA's letter confirms this determination.

2. Sample size used for the test set and the data provenance

  • Not Applicable. This device is an interface controller, not a diagnostic or clinical AI device that processes data from a test set of patient cases. The testing described relates to electrical safety and functional validation of the hardware and software components.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. See point 2.

4. Adjudication method for the test set

  • Not Applicable. See point 2.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not an AI device involved in diagnostic interpretation or human reader assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an algorithm or AI device. It's a hardware interface controller.

7. The type of ground truth used

  • Not Applicable. The "ground truth" for this device's validation is adherence to engineering specifications, electrical safety standards, and functional performance as an interface. There is no biological or diagnostic "ground truth" in the context of this submission.

8. The sample size for the training set

  • Not Applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

JUN - 5 2003

K030718

an 1 of V

SCB Computer Motion Interface Controller 510(k) Summary

In accordance with 21 CFR section 807.92 Computer Motion, Inc. (CMI) is submitting the following 510(k) Summary:

1) Submitter Information

Computer Motion. Inc. 130-B Cremona Drive Goleta, CA 93117

Contact: Cathy Stupak, Ph.D.

2) Name of Device:

Proprietary Name:SCB Computer Motion Interface Controller
Common Name:n/a
Classification Name:Laparoscope, General & Plastic Surgery
Regulation Number:876.1500
Product Code:GCJ
Class:Class II.

3) Substantial Equivalence:

This submission establishes the substantial equivalence of the SCB Computer Motion Interface Controller to Computer Motion's HERMES® Port Expander

4) Description of the Device:

The SCB Computer Motion Interface Controller (SCIC) connects devices that are controlled by Computer Motion's voice-control system to the Storz Communication Bus (SCB) manufactured by Karl Storz.

TestTitle
IEC 601-1International Standard for MedicalElectrical Equipment
IEC 601-1-1International Standard for MedicalElectrical Equipment
IEC 601-2-18International Standard for MedicalElectrical Equipment
UL 2601-1Underwriters Laboratory

The SCIC is tested to the following standards:

continued

{1}------------------------------------------------

/ C 3 0 7 / EONFIDENTIAL Pg 2 of 2

TestTitle
EN55011/A1 CISPR 11Conducted Emission
EN55011/A1 CISPR11Radiated Emission
EN61000-4-2Electrostatic Discharge
EN61000-4-3 and EN50140RF Immunity
EN61000-4-4EFT/Bursts Immunity
EN61000-4-5Surge Immunity
EN61000-4-6Conducted Immunity
EN60601-1International Standard for MedicalElectrical Equipment
EN60601-1-1General Requirements for Safety -Collateral Standard
EN 60601-1-2Emissions and Immunity TestMeasurements
CAN/CSA-C22.2 No. 601.1Medical Electrical Equipment Part 1,General Requirements for Safety, GeneralInstructions Part 1
VA-19795CMI Environmental Testing

5) Intended Use

The SCB Computer Motion Interface Controller is indicated for use with the Storz Communication Bus (SCB) manufactured by Karl Storz to connect HERMES-Ready™ devices to the SCB.

6) Technological Characteristics in Comparison to the Predicate

The predicate, the HERMES Port Expander, is an extension of the HERMES O.R. Control Center, providing 8 additional device connection ports for devices, whereas the SCIC is an extension of the Storz SCB, providing 8 additional device connection ports for devices.

8) Device Testing

Hazard analyses and hardware/software validations procedures were provided in this submission.

9) Conclusion drawn from the Device Testing

Data included in this submission demonstrate the safety and effectiveness of the SCB Computer Motion Interface Controller.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 5 2003

Mr. Keith Lowrey Regulatory Affairs Manager Computer Motion, Inc. 130-B Cremona Drive GOLETA CA 93117

Re: K030718

Trade/Device Name: SCB Computer Motion Interface Controller Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 GCJ Dated: March 6, 2003 Received: March 7, 2003

Dear Mr. Lowrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual fegistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Brogdon

Nancy C. Brogdon

Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

CONFIDENTIAL

INDICATION FOR USE STATEMENT

Ko30718 510(k) Number (if known): _

Device Name: SCB Computer Motion Interface Controller

The SCB Computer Motion Interface Controller is indicated for use with the Storz Communication Bus (SCB) manufactured by Karl Storz to connect HERMES-Ready™ devices to the SCB.

(PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CRDH, Office of Device Evaluation (ODE)

Prescription Use (per 21 CFR 801.109)

OR

Over-the-Counter Use (Optional Format 1-2-96)

":

David A. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices h10(k) Number

000023

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.