(90 days)
VariSeed 7.1 is a software application for planning and evaluating brachytherapy procedures for the treatment of prostate cancer that can be modeled according to AAPM TG-43. It facilitates pre-operative planning and post-operative evaluation as well as intra-operative planning and evaluation.
VariSeed 7.1 is a computer based software application for planning and evaluating prostate brachytherapy procedures. The application runs on standard Intel PCs under Microsoft Window® 32-bit operating systems. The application interfaces with video sources, printers, network sources, as well as DICOM and other image and data sources.
The provided text is a 510(k) Premarket Notification Summary for a medical device called VariSeed 7.1. This type of document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study of acceptance criteria and device performance in the way a clinical trial or a performance study for a novel device would.
Therefore, the information required to fully answer your request (especially regarding detailed acceptance criteria, sample sizes, expert qualifications, and specific performance metrics like those in MRMC studies) is not present in the provided text. The document primarily describes the device, its intended use, and its technological characteristics compared to predicate devices.
However, I can extract the available information and point out what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
- Acceptance Criteria: Not explicitly stated as quantifiable acceptance criteria with thresholds. The document relies on demonstrating "substantial equivalence" to predicate devices (VariSeed 7.0 and BrachyVision 6.0) by matching their functionality and adhering to specific industry standards.
- Reported Device Performance: No specific performance metrics (e.g., accuracy, sensitivity, specificity, or quantitative error rates) are reported in the text. The device performance is implicitly considered acceptable if it matches the functionalities and adheres to the standards of the predicate devices.
| Feature/Characteristic | Acceptance Criteria (Implicit from Predicate Devices) | Reported Device Performance (VariSeed 7.1) |
|---|---|---|
| Intended Use | Planning and evaluation of brachytherapy procedures for prostate cancer. | Broadened to include HDR image acquisition. Substantially equivalent to predicate devices. |
| Work Flows | Offline, real-time, and nomogram planning with/without ultrasound images. | All matched with predicate devices. |
| HDR Image Acquisition | Utility for acquiring images and inputting pre-plan artifacts (present in BrachyVision). | Added; uses existing VariSeed 7.0 capabilities. Export to DICOM RT. |
| Dose Calculation Standard | AAPM TG-43 support for Anisotropy Constant and Factor. | All matched with predicate devices. |
| Anisotropy Function | Support for Anisotropy Function (present in BrachyVision). | Added; clinicians have option to represent dose distributions per TG-43 standard. |
| Real-time Isodose Calc. | Real-time calculation and display of isodoses over patient anatomy. | Matched with predicate devices. |
| Automatic Source Decay | Automatic source decay calculation. | Matched with predicate devices. |
| Industry Conformance | Conforms to AAPM TG-43 - Dosimetry of interstitial brachytherapy sources. | Matched with predicate devices. |
| DICOM Export | Image and structure data export to DICOM (VariSeed 7.0), Plan export (BrachyVision). | Image, structure, and plan data export to DICOM. |
| Level of Concern | Moderate (implicit from predicate devices' classification and function). | Moderate. Not intended for diagnostics or control of plan delivery. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size for Test Set: Not mentioned or applicable in this type of submission. Performance testing data relevant to specific sample sizes is not provided.
- Data Provenance: Not mentioned.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Number of Experts: Not mentioned.
- Qualifications of Experts: Not mentioned.
4. Adjudication Method for the Test Set
- Adjudication Method: Not mentioned. This type of submission would typically involve verification and validation testing against known physics models or established clinical results, rather than expert adjudication of "ground truth" on a test set as might be done for diagnostic AI.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- MRMC Study: No, an MRMC study was not done or reported. This device is a treatment planning software, not a diagnostic AI system intended to assist human readers in interpreting medical images.
- Effect Size: Not applicable as no MRMC study was conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: The document does not describe standalone performance in the typical sense of an algorithm making decisions. The software is the algorithm, performing calculations (like dose distribution) based on user input. The "performance" is implicitly tied to its adherence to standards like AAPM TG-43 and its ability to replicate the functionality of predicate devices. There wouldn't be a separate "standalone" study from its core function. It's a tool for planning, not an autonomous agent.
7. The Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for this device's functionality would likely be based on:
- Physics Models/Standards: Conformance to AAPM TG-43 dosimetry standards for dose calculation.
- Predicate Device Functionality: The functions and outputs of the previously cleared VariSeed 7.0 and BrachyVision 6.0 are the primary "ground truth" for demonstrating substantial equivalence.
- User Requirements/Specifications: The software was developed to meet defined clinical needs for brachytherapy planning.
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. This is not a machine learning device that relies on a "training set" in the common sense. It's a rule-based software application for calculations and data management.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no training set in the machine learning context. The "ground truth" for its development would be the established physics principles and clinical requirements for brachytherapy planning.
{0}------------------------------------------------
K$\phi$3$\phi$534
Premarket Notification [510K] Summary as required by 21 CFR 807.92
Date Summary was prepared:
February 14, 2003
Submitter's Name:
Varian Medical Systems 3100 Hansen Way E-110 Palo Alto, CA 94304
Contact Person:
Linda S. Nash Corporate Director, Regulatory Affairs and Quality Assurance Phone (650) 424-6990 FAX (650) 842-5040 E-mail lind.nash@varian.om
Device Name:
VariSeed 7.1
Classification Name:
Source, Brachytherapy, Radionuclide
Predicate Device:
VariSeed 7.0 (K982821), BrachyVision 6.0 (K992762)
Product Description:
VariSeed 7.1 is a computer based software application for planning and evaluating prostate brachytherapy procedures.
Hardware Platform and Operating System
The application runs on standard Intel PCs under Microsoft Window® 32-bit operating systems.
{1}------------------------------------------------
Kp3ø534
Peripherals and Accessories
The application interfaces with video sources, printers, network sources, as well as DICOM and other image and data sources.
Software Features
-
- Image Acquisition/Import: The ability to acquire patient data from which a plan or evaluation is constructed
-
- Structure Contouring: The ability to define patient structures within the image space.
-
- Source Placement & Dosimetry: The ability to plan source placement within the image space and to calculate the resulting dosimetry.
-
- 2D and 3D visualization: The ability to visualize the resulting dose profile and structure in 2D and 3D.
-
- Reporting: The ability to produce hard and soft copy reports to facilitate the delivery of the plan, to document the plan and to document the post-procedure evaluation.
-
- Brachy Source Specification: The ability for the expert user to modify the profile of seed source specifications and device specifications.
-
- Database functions: The ability to manage the patient data in the application database including archiving, deleting and restoring data.
-
- Licensing: The ability to license the system by application function and interface.
-
- Interface: The ability to interface with other planning systems.
Application Development
The application was developed for 32-bit Microsoft Windows® operating system using Microsoft Visual C++.
Intended Use:
VariSeed 7.1 is a software application used for planning and evaluation of brachytherapy procedures.
Technological Characteristics:
The VariSeed 7.1 treatment planning system is substantially equivalent to the brachytherapy functions of the Varian Medical Systems, Inc. VariSeed 7.0 (K982821) and Varian Medical System, Inc. Brachy Vision 6.0 (K992762) treatment planning systems.
The table below compares the feature sets of the three devices. The new software features for VariSeed 7.1 are in bold, all other VariSeed features are supported in a currently approved product.
{2}------------------------------------------------
Kp3p534
The intended use and indications for use of VariSeed 7.1 have been broadened to account for HDR image acquisition functionality. The intent of this functionality is more clearly stated below.
VariSeed 7.1 is substantially equivalent to the two predicate devices offering functionality for planning and evaluating brachytherapy procedures for the treatment of prostate cancer that can be modeled according to AAPM TG-43.
Significant functionality improvements to VariSeed 7.1 include anisotropy function and a utility for acquiring images and inputting pre-plan artifacts for high dose rate (HDR) brachytherapy planning. Each is available in the predicate BrachyVision device. With the addition of anisotropy function as a dose calculation method clinicians can have the option of representing dose distributions according to another model supported by the TG-43 standard. The HDR capture utility uses existing capabilities in the VariSeed 7.0 device to provide a simplified interface for acquiring ultrasound images and inputting pre-plan artifacts that can be exported according to the DICOM RT standard. The level of concern for VariSeed 7.1 is moderate. VariSeed 7.1 is not intended for diagnostic purposes nor will it control the delivery of the plan. Plan implementation will continue to be performed manually.
| VariSeed 7.0(Predicate) | BrachyVision(Predicate) | VariSeed 7.1(New) |
|---|---|---|
| Work Flows | ||
| Off-line planning with or withoutultrasound images | Off-line planning with or withoutultrasound images | Off-line planning with or withoutultrasound images |
| Real-time planning using imagesacquired from ultrasound | Real-time planning using imagesacquired from ultrasound | Real-time planning using imagesacquired from ultrasound |
| Nomogram planning | Nomogram planning | Nomogram planning |
| HDR Image Acquisition utility | HDR Image Acquisition utility-The HDR capture utility usesexisting capabilities in the VariSeed7.0 device to provide a simplifiedinterface for acquiring ultrasoundimages and inputting pre-planartifacts that can be exportedaccording to the DICOM RTstandard. | |
| Calculation | ||
| AAPM TG-43 Support for AnisotropyConstant | AAPM TG-43 Support for AnisotropyConstant | AAPM TG-43 Support for AnisotropyConstant |
| AAPM TG-43 Support for AnistropyFactor | AAPM TG-43 Support for AnistropyFactor | AAPM TG-43 Support for AnistropyFactor |
| AAPM TG-43 Support for AnisotropyFunction | AAPM TG-43 Support forAnisotropy Function-With the addition of anisotropyfunction as a dose calculationmethod clinicians can have theoption of representing dosedistributions according to another | |
| model supported by the TG-43standard. | ||
| Real-time calculation of isodoses anddisplay over patient anatomy asisodose lines. | Real-time calculation of isodoses anddisplay over patient anatomy asisodose lines. | Real-time calculation of isodoses anddisplay over patient anatomy asisodose lines. |
| Automatic source decay | Automatic source decay | Automatic source decay |
| Industry Conformance | ||
| Conforms to AAPM TG-43 - Dosimetryof interstitial brachytherapy sources | Conforms to AAPM TG-43 - Dosimetryof interstitial brachytherapy sources | Conforms to AAPM TG-43 - Dosimetryof interstitial brachytherapy sources |
| General | ||
| Independent of the radiation deliverysystem | Independent of the radiation deliverysystem | Independent of the radiation deliverysystem |
| Hardcopy reports of all views | Hardcopy reports of all views | Hardcopy reports of all views |
| Planning support for all solid sourcedelivery systems - line, seed, andseed train | Planning support for all solid sourcedelivery systems - line, seed, andseed train | Planning support for all solid sourcedelivery systems - line, seed, andseed train |
| User defined dose points | User defined dose points | User defined dose points |
| Manual structuring | Manual structuring | Manual structuring |
| Manual source localization | Manual source localization | Manual source localization |
| Window level tools | Window level tools | Window level tools |
| Multiple plan variations | Multiple plan variations | Multiple plan variations |
| Ultrasound import from DICOM | Ultrasound import from DICOM | Ultrasound import from DICOM |
| Included source management utilities | Included source management utilities | Included source management utilities |
| Included patient management utilities | Included patient management utilities | Included patient management utilities |
| Automated prostate contouring. | Automated prostate contouring. | Automated prostate contouring. |
| Contour editing | Contour editing | Contour editing |
| Contour interpolation | Contour interpolation | Contour interpolation |
| Contour margining | Contour margining | Contour margining |
| Point and line landmarks. | Reference points and lines | Point and line landmarks. |
| Source location export | Plan export | Source location export |
| Volume editing | Volume editing | Volume editing |
| Image and structure data export toDICOM | Image and structure data export toDICOM | Image, structure, and plan data exportto DICOM |
| Dosimetric quality alerts | Unacceptable dose values are flaggedin the Plan Organizer. | Dosimetric quality alerts |
| Image set co-registration | Image set co-registration | Image set co-registration |
{3}------------------------------------------------
Kp3ø524/
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of an eagle or bird with three wing-like shapes and a flowing tail.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 21 2003
Ms. Linda S. Nash Corporate Director, Regulatory Affairs and Ouality Assurance Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304-1038
Re: K030534
Trade/Device Name: VariSeed 7.1 Regulation Number: 21 CFR §892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: 90 KXK Dated: February 14, 2003 Received: February 20, 2003
Dear Ms. Nash:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|---|---|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on vour responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy Cbrogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Statement of Indications for Use
(if known): K$\phi$3$\phi$534
510(k) Number (if known):
Device Name: VariSeed 7.1
Indications for use:
VariSeed 7.1 is a software application for planning and evaluating brachytherapy procedures for the treatment of prostate cancer that can be modeled according to AAPM TG-43. It facilitates pre-operative planning and post-operative evaluation as well as intra-operative planning and evaluation.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Usef OR (Per 21 CFR 801.109)
Over-The-Counter Use (Optional Format 1-2-96)
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K030534
N/A