K Number
K030422

Validate with FDA (Live)

Date Cleared
2003-04-14

(63 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A single use retrieval bag designed to temporarily contain organs or stones and facilitate their removal from the patient without contamination during laparoscopic surgery such as the appendix, an ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, fibroid tumors, small sections of bowel and other tissue structures.

Device Description

The LiNA-Bag is a laparoscopic instrument consisting of a collapsible pouch attached and placed within a long slender multi-part tube allowing for introduction of said pouch through a laparoscopic trocar. Opening and closing of the pouch is activated by manual manipulation in a push and pull motion of the attached tube that has a corresponding nylon drawstring attached to the pouch.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the LiNA-Bag. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a study. Therefore, much of the requested information (such as study details, sample sizes, expert qualifications, and MRMC studies) is not present in the provided document.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state formal acceptance criteria with numerical targets. The "performance" being demonstrated is primarily substantial equivalence to predicate devices. The implicit criteria for substantial equivalence are based on:

Acceptance Criteria (Implicit from 510(k) Process)Reported Device Performance (as stated in document)
Intended Use Equivalence: Device's intended use is the same as predicate devices.The LiNA-Bag is a sterile, single-use bag designed to contain tissue, small organic matter temporarily in order to facilitate the removal of this matter without wound contamination during laparoscopic procedures. This is consistent with the general intended use of specimen retrieval bags.
Technological Characteristics Equivalence: Device has similar materials, design, energy source, or other features as predicate devices.The description indicates a laparoscopic instrument with a collapsible pouch, multi-part tube, and manual manipulation for opening/closing, activated by a nylon drawstring. This design is consistent with laparoscopic specimen retrieval bags. The device is also single-use and sterile. Specific material compositions are not detailed but are implied to be similar to legally marketed devices.
Performance Data (if applicable): While no specific performance study is detailed, the demonstration of substantial equivalence implies that the device is expected to perform safely and effectively for its intended use, similar to its predicates.No specific performance study data (e.g., strength, burst pressure, tear resistance, retrieval rates) are provided in this summary. The 510(k) relies on the inherent safety and effectiveness of the existing predicate devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The document describes a 510(k) submission based on substantial equivalence, not a clinical or performance study with a test set. There are no test subjects or data provenance mentioned for performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This information pertains to a study or clinical trial, which is not described in this 510(k) summary. Ground truth establishment for a test set is not part of this type of submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As no test set performance study is detailed, there is no mention of an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This type of study (MRMC, AI assistance) is for comparative effectiveness of diagnostic tools, which is not relevant to a laparoscopic specimen retrieval bag.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical surgical instrument, not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. No ground truth was established as this is not a diagnostic device or a study involving such. The "ground truth" for the 510(k) process is the safety and effectiveness of the legally marketed predicate devices.

8. The sample size for the training set

Not applicable. This document does not describe a training set for an algorithm or AI model.

9. How the ground truth for the training set was established

Not applicable. As there is no training set described, the method for establishing its ground truth is also not applicable.

{0}------------------------------------------------

K030422

APR 1 4 2003

510(K) Summary as required by 21 CFR section 807.92 for LiNA-BAG Prepared February 7, 2003

Submitted by:LiNA Medical CML ApSFormervangen 5GlostrupDenmark DK-2600
Contact Person:Device Trade Name:Common Name:Classification:Mr. Niels Kornerup, PresidentLiNA-BagLaparoscopic Specimen Retrieval DeviceLaparoscopic, General & Plastic Surgery, 876.1500 Class II
Predicate Devices:Patton Endo-bag (K0122150), distributed by Patton Medical,1000 Westbank Drive, Suite 5A200, Austin, TX 78746ENDOPOUCH Specimen Retrieval Bag (K933104), manufactured by Ethicon,Inc., P.O. Box 151, Somerville, NJ 08876Auto Suture ENDO CATCH Single Use Specimen Pouch(K922123), manufactured by U.S. Surgical, Division of TYCO, 150 Glover Ave.,Norwalk, CT 06856

Description of Device:

The LiNA-Bag is a laparoscopic instrument consisting of a collapsible pouch attached and placed within a long slender multi-part tube allowing for introduction of said pouch through a laparoscopic trocar. Opening and closing of the pouch is activated by manual manipulation in a push and pull motion of the attached tube that has a corresponding nylon drawstring attached to the pouch.

Intended Use of Device:

The LiNA-Bag is a sterile, single-use bag designed to contain tissue, small organic matter temporarily in order to facilitate the removal of this matter without wound contamination during laparoscopic procedures.

Substantial Equivalence to Predicate Device:

The LiNA-Bag is substantially equivalent to the Patton Endo-bag (K0122150), distributed by Patton Medical, 1000 Westback Drive, Suite 5A200, Austin, TX 78746 and the ENDOPOUCH Specimen Retrieval Bag (K933104), manufactured by Ethicon, Inc., P.O. Box 151, Somerville, NJ 08876 and the Auto Suture ENDO CATCH Single Use Specimen Pouch (K922123), manufactured by U.S. Surgical, Division of TYCO, 150 Glover Ave, Norwalk, CT 06856.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing, ribbon-like shape at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 4 2003

Mr. Niels Kornerup President LiNa Medical CML ApS Formervangen 5 Glostrup, Denmark DK-2600

Re: K030422

Trade/Device Name: LiNa-Bag Regulation Number: 21 CFR 876.1500 Regulation Names: Endoscope and accessories Regulatory Class: II Product Codes: GCJ Dated: February 7, 2003 Received: February 10, 2003

Dear Mr. Kornerup:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Niels Kornerup

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Sincerely yours,

Mark N Millkern

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

NR. 999 S. 3

Indications for Use

13.5 - Found sures I ruy 1 diministration ar Jos Destens and Radiological Health

CDRH Home | Search | A-Z Index | Feedback

Page 1 of 1

Side I af 2

510(k) Number (if known): K030422

LiNA-Bag, model LB-60 & LB-100 Device Name:

Indications for Use:

A single use retrieval bag designed to temporarily contain organs or stones and facilitate their removal from the patient without contamination during laparoscopic surgery such as the appendix, an ectopic pregnancy, gallbladder, gallstones, lymph nodes, ovaries, fibroid tumors, small sections of bowel and other tissue structures.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark R. Milleson
(Division Sign-Off)

Restorative Divis anc

510(k) Number K030926

http://www.fda.gov/cdrh/ode/indicate.html

21-03-03

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.