K Number
K030167

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-04-02

(76 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MIERU (etafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The MIERU (etafilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may have astigmatism of 7.00D or less.

The MIERU (etafilcon A) Multifocal Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic. The lens may be worn by persons who exhibit no more than 2.00 D of astigmatism and can obtain satisfactory visual acuity in a power range of +4.00D to -5.00D and have near add requirements up to 3.00D.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Device Description

The MIERU (etafilcon A) Soft (hydrophilic) Contact Lens is available as a single vision spherical lens, aspherical multifocal lens and as a back surface astigmatic (toric) lens. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The ionic lens material, (etafilcon A) is a co-polymer of 2- hydroxyethyl methacrylate and methacrylic acid cross-linked with 1,1,1 - trimethylol propane trimethacrylate glycol demethacrylate. It consists of 42% etafilcon A and 58% water by weight when immersed in buffered saline solution. The lens polymer contains a UV absorbing compound and is available clear or with a blue visibility-handling tint, color additive 'Reactive Blue 19', 21 CFR part 73.3121. The (etafilcon A) name has been adopted by the United States Adopted Names Council (USAN).

In the MIERU Contact Lens with UV Blocker, a Benzophenone UV absorbing monomer is used to block UV radiation. The UV blocking for MIERU averages > 99% in the UVB range of 280nm - 315nm and 83% in the UVA range of 316 - 380nm.

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out. it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight. The physical properties of the lens are: Refractive Index 1.4050 (wet), Light Transmission (clear) greater than 95%, Light Transmission (tinted) greater than 91%, Water Content 58 %, Specific Gravity 1.017 (hydrated), Oxygen Permeability 19.9 X 10-11 (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt method).

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the MIERU (etafilcon A) Daily Wear Contact Lens, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

This 510(k) submission doesn't explicitly state "acceptance criteria" in terms of performance metrics like sensitivity or specificity. Instead, the study focuses on demonstrating substantial equivalence to existing predicate devices based on technological characteristics and non-clinical testing. The acceptance criteria are implicitly met if the new device's properties are comparable to or within acceptable ranges of the predicate devices.

Here's a table summarizing the key technological characteristics considered for equivalence:

Pre-Clinical Equivalency / DeviceClearLab800 (etafilcon A) (New Device)Frequency 58 (etafilcon A) (Predicate)AcuVue 2 (etafilcon A) (Predicate)Igel 55 UV (methafilcon A) (Predicate)Reported Device Performance (ClearLab800)
Intended UseCorrection of visual acuity in aphakic/non-aphakic persons with non-diseased eyes, myopia/hyperopia.Correction of visual acuity in aphakic/non-aphakic persons with non-diseased eyes, myopia/hyperopia.Correction of visual acuity in aphakic/non-aphakic persons with non-diseased eyes, myopia/hyperopia.Correction of visual acuity in aphakic/non-aphakic persons with non-diseased eyes, myopia/hyperopia.Matches intended use of predicates
FunctionalityActs as a refractive medium to focus light.Acts as a refractive medium to focus light.Acts as a refractive medium to focus light.Acts as a refractive medium to focus light.Matches functionality of predicates
IndicationsDaily wear, Soft (hydrophilic) contact lensDaily wear, Soft (hydrophilic) contact lensDaily wear, Soft (hydrophilic) contact lensDaily wear, Soft (hydrophilic) contact lensMatches indications of predicates
Production MethodCast-moldedCast-moldedCast-moldedCast-moldedCast-molded
FDA Group #Group # 4 >50% Water, Ionic PolymersGroup # 4 >50% Water, Ionic PolymersGroup # 4 >50% Water, Ionic PolymersGroup # 4 >50% Water, Ionic PolymersGroup # 4 >50% Water, Ionic Polymers
USAN nameetafilcon Aetafilcon Aetafilcon Amethafilcon Aetafilcon A
Water Uptake (%)58.0%58.0%58.0%55.0%58.0%
Oxygen Permeability19.9 x 10^-11 (cm²/sec) (ml O²/ml x mm Hg @ 35°C), (revised Fatt method)20.2 x 10^-11 (cm²/sec) (ml O²/ml x mm Hg @ 35°C), (revised Fatt method)19.9 x 10^-11 (cm²/sec) (ml O²/ml x mm Hg @ 35°C), (revised Fatt method)19.6 x 10^-11 (cm²/sec) (ml O²/ml x mm Hg @ 35°C), (revised Fatt method)19.9 x 10^-11 (cm²/sec)
Specific Gravity1.0171.0541.0211.0621.017
UV Blocking (UVB)Averages > 99%Not explicitly stated in tableNot explicitly stated in tableNot explicitly stated in table> 99%
UV Blocking (UVA)83%Not explicitly stated in tableNot explicitly stated in tableNot explicitly stated in table83%

Summary of Non-Clinical Performance Data (Used to demonstrate equivalence):

The device underwent the following non-clinical tests as recommended by the Premarket Notification 510(k) guidance document for Daily Wear Contact Lenses, revised May 1994:

  • Toxicology testing:
    • Cytotoxicity
    • USP Ocular Irritation
    • USP Systemic Injection
  • Leachability / Residual monomer Studies
  • Physicochemical property testing

The conclusion drawn from these studies was that the MIERU Soft Contact Lens is substantially equivalent to the predicate devices and does not raise different questions of safety and effectiveness.


Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a separate "test set" in the context of clinical trials. The data presented for the new device (ClearLab800 etafilcon A) are physical and chemical properties derived from laboratory testing of the device itself.
    • Data Provenance: The data refers to the characteristics of the device (ClearLab800 (etafilcon A)) itself, rather than test data from patients. The origin of the raw materials for manufacturing is not specified, nor is there a mention of specific countries for testing. This is a non-clinical submission, not a clinical trial. The data is prospective in the sense that these properties were measured on the newly developed device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable (N/A). This submission relies on non-clinical data (physical and chemical properties, toxicology) to demonstrate substantial equivalence to predicate devices, not on expert-adjudicated clinical outcomes for a "ground truth." The "ground truth" here is the established safety and effectiveness of the predicate devices based on their own prior approvals and the standardized methods used for material characterization and toxicology.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • None. As there's no clinical test set requiring expert adjudication for a diagnosis or outcome, no adjudication method was used.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/CAD device. It is a contact lens. Therefore, no MRMC study was performed, and this question is not applicable.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence. The "device performance" in this context refers to the inherent physical and chemical properties of the contact lens itself, measured directly in a laboratory setting, independent of human interaction or a human-in-the-loop scenario. The non-clinical tests (toxicology, leachability, physicochemical properties) are standalone assessments of the device's characteristics.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" in this submission is indirectly established by the legally marketed predicate devices and the objective measurements of the new device's physical, chemical, and biological properties through standardized laboratory testing (e.g., water content, oxygen permeability, specific gravity, toxicology tests). The concept of "ground truth" derived from expert consensus or pathology, as typically used in AI/CAD imaging, does not apply here. The benchmark is the established safety and performance profile of the predicate devices.
  7. The sample size for the training set:

    • Not applicable (N/A). This is a submission for a medical device (contact lens), not an AI algorithm requiring a training set.
  8. How the ground truth for the training set was established:

    • Not applicable (N/A). As stated above, this is not an AI algorithm, so there is no training set or associated ground truth establishment process in that context.

{0}------------------------------------------------

510(k) SUMMARY of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

K030167

Applicant information:

Date Prepared:January 14, 2003
Name:AddressClearLab PTE, Ltd.139, Joo Seng Raod #06-01ATD Centre, Singapore 368362
Contact Person:Phone number:Fax number:Steve NewmanCountry code 65+ 67491090Country code 65+ 62848534
Parent Company:1800 CONTACTS, Inc.66 E. Wadsworth Park Drive 3rd. FloorDraper, UT 84020
FDA US Agent/Official Correspondent:Phone numberFax numberMedvice Consulting, Inc.Martin Dalsing(970) 243-5490(970) 243-5501
Device Information:

Device Classification: Class II Classification Number: LPL Classification Name: Lenses, Soft Contact, Daily Wear MIERU (etafilcon A) Daily Wear Contact Lens clear and Trade Name: visibility tint, with UV blocker.

{1}------------------------------------------------

Equivalent Predicate Devices:

The MIERU (etafilcon A) Soft Contact Lenses are substantially equivalent to the following predicate devices:

    1. IGEL 55UV (methafilcon A), K020855, Manufactured by IGEL Visioncare Pte. Ltd.
    1. Frequency 58 (etafilcon A), K980634, manufactured by Aspect Vision Care.
    1. AcuVue 2 (etafilcon A), K962804, manufactured by Vistakon.

Device Description:

The MIERU (etafilcon A) Soft (hydrophilic) Contact Lens is available as a single vision spherical lens, aspherical multifocal lens and as a back surface astigmatic (toric) lens. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The ionic lens material, (etafilcon A) is a co-polymer of 2- hydroxyethyl methacrylate and methacrylic acid cross-linked with 1,1,1 - trimethylol propane trimethacrylate glycol demethacrylate. It consists of 42% etafilcon A and 58% water by weight when immersed in buffered saline solution. The lens polymer contains a UV absorbing compound and is available clear or with a blue visibility-handling tint, color additive 'Reactive Blue 19', 21 CFR part 73.3121. The (etafilcon A) name has been adopted by the United States Adopted Names Council (USAN).

In the MIERU Contact Lens with UV Blocker, a Benzophenone UV absorbing monomer is used to block UV radiation. The UV blocking for MIERU averages > 99% in the UVB range of 280nm - 315nm and 83% in the UVA range of 316 - 380nm.

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly bevond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out. it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight. The physical properties of the lens are:

Refractive Index1.4050 (wet)
Light Transmission (clear)greater than 95%
Light Transmission (tinted)greater than 91%
Water Content58 %
Specific Gravity1.017 (hydrated)
Oxygen Permeability19.9 X 10-11 (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt method)

{2}------------------------------------------------

Intended Use:

The MIERU (etafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The MIERU (etafilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may have astigmatism of 7.00D or less.

The MIERU (etafilcon A) Multifocal Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic. The lens may be worn by persons who exhibit no more than 2.00 D of astigmatism and can obtain satisfactory visual acuity in a power range of +4.00D to =5.00D and have near add requirements up to 3.00D.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

{3}------------------------------------------------

Technological Characteristics:

The technological characteristics of the MIERU contact lens as compared to the technological characteristics of the predicate devices are illustrated in the following table.

Pre-Clinicalequivalency /DeviceClearLab800(etafilcon A)new deviceFrequency 58(etafilcon A)predicate deviceAcuVue 2(etafilcon A)predicate deviceIgel 55 UV(methafilcon A)predicate device
Intended UseIndicated for dailywear for the correctionof visual acuity inaphakic and notaphakic persons withnon-diseased eyes withmyopia or hyperopia.Indicated for dailywear for the correctionof visual acuity inaphakic and notaphakic persons withnon-diseased eyes withmyopia or hyperopia.Indicated for dailywear for the correctionof visual acuity inaphakic and notaphakic persons withnon-diseased eyes withmyopia or hyperopia.Indicated for dailywear for the correctioof visual acuity inaphakic and notaphakic persons withnon-diseased eyes witmyopia or hyperopia.
FunctionalityThe contact lenses actas a refractive mediumthat focus light raysfrom near and distantobjects on the retina.The contact lenses actas a refractive mediumthat focus light raysfrom near and distantobjects on the retina.The contact lenses actas a refractive mediumthat focus light raysfrom near and distantobjects on the retina.The contact lenses acas a refractive mediumthat focus light raysfrom near and distantobjects on the retina.
IndicationsDaily wear, Soft(hydrophilic) contactlensDaily wear, Soft(hydrophilic) contactlensDaily wear, Soft(hydrophilic) contactlensDaily wear, Soft(hydrophilic) contactlens
ProductionMethodCast-moldedCast-moldedCast-moldedCast-molded
FDA Group #Group # 4 >50%Water, Ionic PolymersGroup # 4 >50%Water, Ionic PolymersGroup # 4 >50%Water, Ionic PolymersGroup # 4 >50%Water, Ionic Polymers
USAN nameetafilcon Aetafilcon Aetafilcon Amethafilcon A
WaterUptake(%)58.0%58.0%58.0%55.0%
OxygenPermeability19.9 x 10-11 (cm2/sec)(ml O2/ml x mm Hg@ 35 degrees C),(revised Fatt method).20.2 x 10-11 (cm2/sec)(ml O2/ml x mm Hg@ 35 degrees C),(revised Fatt method).19.9 x 10-11 (cm2/sec)(ml O2/ml x mm Hg@ 35 degrees C),(revised Fatt method).19.6 x 10-11 (cm2/sec)(ml O2/ml x mm Hg@ 35 degrees C),(revised Fatt method).
SpecificGravity1.0171.0541.0211.062

{4}------------------------------------------------

Summary of Non-Clinical Performance Data:

The following non-clinical tests were conducted as recommended by the Premarket Notification 510(k) guidance document for Daily Wear Contact Lenses, revised May 1994.

    1. Toxicology testing
    • a. Cytotoxicity
    • b. USP Ocular Irritation
    • c. USP Systemic Injection
    1. Leachability / Residual monomer Studies
    1. Physiochemical property testing

Clinical Studies:

No clinical data is required for this submission.

Conclusions Drawn from the Studies:

The MIERU Soft Contact Lens is substantially equivalent to the predicate devices and does not raise different questions of safety and effectiveness than that of the predicate devices.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged in a stacked formation, creating a sense of depth. Encircling the profiles is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Druq Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 2 2003

ClearLab PTE, Ltd. c/o Martin Dalsing Medvice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503

Re: K030167

Trade/Device Name: MIERU (etafilcon A) Daily Wear Contact Lens clear and visibility tint, with UV blocker Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: March 17, 2003 Received: March 24, 2003

Dear Mr. Dalsing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 - Martin Dalsing

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

A. halpi Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE STATEMENT

MIERU (etafilcon A) Daily Wear Contact Lens clear and visibility tint. with Device Name: UV blocker.

INDICATIONS FOR USE:

The MIERU (etafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.

The MIERU (etafilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and may have astigmatism of 7.00D or less.

The MIERU (etafilcon A) Multifocal Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic. The lens may be worn by persons who exhibit no more than 2,00 D of astigmatism and can obtain satisfactory visual acuity in a power range of +4.00D to -5.00D and have near add requirements up to 3.00D.

Evecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning. disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

or

Over-The-Counter Use

(Optional Format +2-96)

signature

ision of Ophthalmic Ear. Nose and Throat Dev

510(k) Number K030167

N/A