K Number
K024012

Validate with FDA (Live)

Date Cleared
2003-03-20

(106 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OEC 9800 E/CV+ Digital Mobile Imaging System is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. It includes features designed specifically for diagnostic and interventional cardiac imaging procedures, and is also intended for use in cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may also be used for other imaging applications at the physician's discretion.

Device Description

The OEC 9800 E/CV+ is comprised of two mobile units: a C-arm stand and a workstation. The C-arm stand supports the high-voltage generator, x-ray controls, and a "C" shaped apparatus, which supports an x-ray tube on one end and an image intensifier on the other. The C-arm is designed to perform linear and rotational motions that allow the user to position the x-ray imaging components at various and distances with respect to the patient. The mobile workstation supports image display monitors, image processing and recording devices.

AI/ML Overview

The provided text is a 510(k) summary for the GE OEC 9800 E/CV+ Digital Mobile Imaging System. This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving that a device meets specific acceptance criteria through a study with detailed performance metrics.

Therefore, the input document does not contain the information requested in the prompt regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set information.

The document focuses on:

  • Device Description: What the device is and its components.
  • Indications for Use: The medical procedures for which the device is intended.
  • Substantial Equivalence: A claim that the device is similar to previously cleared devices (GE OEC 9800 Plus Mobile Digital Imaging System and Siemens Medical Solutions Powermobil).
  • Standards Compliance: A list of safety and electrical standards the device adheres to.

Conclusion based on the provided text:

  1. Table of acceptance criteria and reported device performance: Not provided. The document states substantial equivalence to predicate devices, but no specific performance criteria or results are listed.
  2. Sample size used for the test set and the data provenance: Not applicable, as no performance study against acceptance criteria is described.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a fluoroscopic imaging system, not an AI-powered diagnostic tool for interpretation assistance.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

MAR 2 0 2003

510(k) Summary

K0240/2

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Date:

November 27, 2002

Name of Submitter:

GE OEC Medical Systems, Inc. 384 Wright Brothers Drive Salt Lake City, UT 84116 801-328-9300

Corresponding Official:

Jeff Wagner. Manager, Regulatory Affairs

Device Proprietary Name:

OEC 9800 E/CV+ Digital Mobile Imaging System

Classification Name:

System, X-ray, Fluoroscopic, Image-Intensified - or System, X-ray, Mobile

{1}------------------------------------------------

Common/Usual Names:

Mobile C-arm, Fluoroscopic Imaging System

Substantial Equivalence:

The OEC 9800 E/CV+ is substantially equivalent to the following devices that are currently marketed:

  • GE OEC Medical Systems OEC 9800 Plus Mobile Digital Imaging System .
  • Siemens Medical Solutions Powermobil .

These devices are mobile C-arm type x-ray systems intended for fluoroscopic imaging. The systems all include a high-voltage x-ray generator and control, rotating anode x-ray tube, image intensifier, video image displays, digital image processing and image storage capability, as well as conventional spot-film capability.

Device Description:

Indications for Use

The OEC 9800 E/CV+ is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and is also intended for cholanqiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may be used for other imaging applications at the physician's discretion.

User Characteristics

The device is used by health care professionals such as physicians, surgeons, cardiologists, radiologists and technologists in hospitals, out-patient clinics and other clinical environments. It is expected that the device will be used on a daily basis. Users are trained in the proper use of the device by GE OEC applications specialists and/or qualified site personnel. The device labeling stipulates that only properly trained persons operate this equipment.

General Description

The OEC 9800 E/CV+ is comprised of two mobile units: a C-arm stand and a workstation. The C-arm stand supports the high-voltage generator, x-ray controls, and a "C" shaped apparatus, which supports an x-ray tube on one end and an image intensifier on the other. The C-arm is designed to perform linear and rotational motions that allow the user to position the x-ray imaging components at various and distances with respect to the patient. The mobile workstation supports image display monitors, image processing and recording devices.

Interfaces are provided for optional peripheral devices such as thermal or laser printers and VCRs. Video outputs are compatible with RS-170 format for domestic markets. CCIR format for international markets, and DICOM 3.0. An auxiliary connection is provided for a Medrad angjographic injector system to facilitate synchronized acquisition of angiographic images during contrast media injection.

{2}------------------------------------------------

Standards:

The OEC 9800 E/CV+ is designed in accordance with product safety requirements established in the following standards:

  • · Federal Performance Standard for Diagnostic X-ray Systems (21 CFR 1020.30-32)
  • · ANSI/NFPA 70 & 99 National Electrical Code and Standard for Health Care Facilities
  • · UL 2601 Medical Electrical Equipment
  • · CSA-C22.2 No. 601.1-M90 Medical Electrical Equipment
  • · IEC 60601-1, Medical Electrical Equipment, General Requirements for Safety
  • · IEC 60601-1-2, Medical Electrical Equipment, General Requirements for Safety, Electromagnetic Compatibility
  • · IEC 60601-1-3, Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray Equipment
  • · IEC 60601-1-4, General requirements for safety, Programmable electrical medical systems.
  • · IEC 60601-2-7, Medical Electrical Equipment, Safety of HV/X-ray Generators
  • · IEC 60601-2-32, Medical Electrical Equipment, Safety of Associated X-ray Equipment
  • · IEC 60601-2-43 Particular requirements for safety of X-ray equipment for interventional procedures.
  • · 93/42/EEC Annex 1 Essential Requirements of the Medical Devices Directive

{3}------------------------------------------------

Conclusion:

The OEC 9800 E/CV+ does not raise new questions of safety and effectiveness and is substantially equivalent to current legally marketed devices.

This concludes this 510(k) summary.

Jeff Wagner

Wagner. Manger, Regulatory Affairs GE OEC Medical Systems, Inc.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles forming a bird-like shape.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jeff Wagner Manager, Regulatory Affairs GE OEC Medical Systems, Inc. General Electric Company 384 Wright Brothers Drive SALT LAKE CITY UT 84116-2862

MAR 2 0 2003

Re: K024012 Trade/Device Name: OEC 9800 E/CV+ Digital Mobil Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1720 Regulation Name: Mobil x-ray system Regulatory Class: II Product Code: 90 JAA and IZL Dated: February 28, 2003 Received: March 3, 2003

Dear Mr. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon.

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications For Use Statement

Applicant:GE OEC Medical Systems, Inc.
510(k) No. (if known):K024012
Device name:OEC 9800 E/CV+ Digital Mobile Imaging System
Indications for use:The OEC 9800 E/CV+ Digital Mobile Imaging System is designed toprovide fluoroscopic and spot-film imaging of the patient duringdiagnostic, surgical and interventional procedures. It includes featuresdesigned specifically for diagnostic and interventional cardiac imagingprocedures, and is also intended for use in cholangiography,endoscopic, urologic, orthopedic, neurologic, vascular, critical care andemergency room procedures. It may also be used for other imagingapplications at the physician's discretion.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter

(Optional Format 1-2-96)

Nancy C brogdon

(Division Sign-Off)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K024012

N/A