K Number
K023989
Manufacturer
Date Cleared
2003-02-26

(86 days)

Product Code
Regulation Number
886.5925
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Device Description

The CONTAFLEX 55 Spherical Soft Contact Lenses are fabricated from methafilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The ionic lens material. (methafilcon A) is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid, cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% methafilcon A and 55% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility-handling tint, [phthalocyaninato (2-)] copper. The methafilcon A name has been adopted by the United States Adopted Names Council (USAN).

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (methafilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 55% water by weight. The physical properties of the lens are:

Refractive Index 1.405 (hydrated)
Light Transmission (clear) greater than 94%
Light Transmission (tinted) greater than 94%
Water Content 55 % ± 2%
Specific Gravity 1.067
Oxygen Permeability 19.6 X 10-11 (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method).

AI/ML Overview

The provided text describes the CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lens and its substantial equivalence to a predicate device. However, the document does not describe a study involving acceptance criteria in the typical sense of a clinical trial or algorithm performance study.

Instead, the submission for CONTAFLEX 55 focuses on establishing substantial equivalence to an already marketed device ("BENZ methafilcon A"). This is a regulatory pathway where the new device is compared to a legally marketed predicate device to demonstrate it is as safe and effective. The "acceptance criteria" here are essentially the demonstration that the new device shares fundamental technological characteristics and intended use with the predicate device, and does not raise new questions of safety or effectiveness.

Here's an analysis based on your requested information, acknowledging the nature of this 510(k) submission:


Acceptance Criteria and Device Performance (based on Substantial Equivalence)

Since this is a substantial equivalence submission, the "acceptance criteria" are implied by the comparison to the predicate device. The performance is demonstrated by showing similar characteristics.

Acceptance Criteria (Implied by Predicate)Reported Device Performance (CONTAFLEX 55)
Intended Use: Correct visual acuity in aphakic/non-aphakic persons with non-diseased eyes (myopia/hyperopia), astigmatism ≤ 0.75 D.Indicated for daily wear for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
Functionality: Act as a refractive medium.After machining from the optical blank, the contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina.
Indications: Daily wear, Soft (hydrophilic) contact lens.Daily wear, Soft (hydrophilic) contact lens.
Production Method: Lathe-cut.Lathe-cut.
FDA Group #: Group # 4 (>50% Water, ionic Polymers).Group # 4 (>50% Water, ionic Polymers).
USAN name: methafilcon A.methafilcon A.
Water Content: 55.0%.55.0% ± 2% (reported in device description, 55.0% in table).
Oxygen Permeability: Comparable to 18.3 X 10-11 (cm²/sec).19.6 X 10-11 (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method).
Specific Gravity: Comparable to 1.128.1.067.
Toxicology: Non-cytotoxic, non-systemically toxic, non-ocularly irritating.Meets requirements of Agarose Overlay Method (cytotoxicity). Considered non-toxic (systemic). Produced no ocular irritation.
Shelf Life: Satisfied.Satisfied via referencing rights granted to Contamac Ltd. for 510(k) 973597.

The "study" that proves the device meets these "acceptance criteria" is the pre-clinical equivalency and safety testing performed to demonstrate its characteristics align with the predicate device.


Detailed Study Information:

  1. Sample size used for the test set and the data provenance:

    • The document describes pre-clinical testing (in-vitro cytotoxicity, systemic toxicity, acute ocular irritation) rather than a clinical "test set" with human subjects for performance evaluation like an AI device would have. Specific sample sizes for these preclinical tests are not provided in the summary.
    • Data provenance: Not explicitly stated, but these are laboratory-based preclinical tests conducted on the device material itself, likely in a controlled lab environment.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This question is not applicable to this type of submission. There was no "ground truth" to establish in the context of expert review of images or clinical outcomes. The "ground truth" for the preclinical tests would be the established scientific methods and standards (e.g., ISO 10993-5 for cytotoxicity).
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This submission does not involve adjudication of expert opinions or discrepancies in a clinical test set.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI device. This section is entirely not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI device. This section is entirely not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the preclinical tests, the "ground truth" refers to the established scientific and regulatory standards for material safety (e.g., ISO 10993-5 for cytotoxicity, and generally accepted methods for systemic toxicity and ocular irritation). For material properties (water content, oxygen permeability), the ground truth is derived from standardized laboratory measurement methods (e.g., revised Fatt method for oxygen permeability).
  7. The sample size for the training set:

    • Not applicable. This contact lens is not an AI/ML device that requires a training set.
  8. How the ground truth for the training set was established:

    • Not applicable. This contact lens is not an AI/ML device that requires a training set.

{0}------------------------------------------------

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:K023989
-----------------------------------------

Applicant information:

Date Prepared:

November 25, 2002

Name:CONTAMAC Ltd.
AddressBearwalden Business ParkSaffron WaldenEssex England CB11 4JX
Contact Person:Robert McGregor
Phone number:44-1799 542 000
US Agent:Medvice Consulting, Inc.
Martin Dalsing
Phone number(970) 243-5490
Fax number(970) 243-5501

Device Information:

Device Classification:Class II
Classification Number:LPL
Classification Name:Lenses, Soft Contact, Daily Wear
Trade Name:CONTAFLEX 55 (methafilcon A) Spherical

Lens for Daily Wear (clear and tinted, lathe-cut)

Soft Contact

{1}------------------------------------------------

Equivalent Devices:

The CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lenses are substantially equivalent to the following predicate device

"BENZ methafilcon A" Predicate device: Manufactured/distributed by Benz Research and Development. 510(k) number: K980418

Device Description:

The CONTAFLEX 55 Spherical Soft Contact Lenses are fabricated from methafilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The ionic lens material. (methafilcon A) is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid, cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% methafilcon A and 55% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility-handling tint, [phthalocyaninato (2-)] copper. The methafilcon A name has been adopted by the United States Adopted Names Council (USAN).

In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (methafilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 55% water by weight. The physical properties of the lens are:

Refractive Index Light Transmission (clear) Light Transmission (tinted) Water Content Specific Gravity Oxygen Permeability

1.405 (hydrated)
greater than 94%
greater than 94%
55 % ± 2%
1.067
19.6 X 10-11 (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt
method).

{2}------------------------------------------------

Intended Use:

The CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

Description of Safety:

A series of preclinical testing were performed to demonstrate the safety and effectiveness of the CONTAFLEX 55 Soft Contact Lens material. The results of all testing demonstrated that the safety and effectiveness of the CONTAFLEX 55 Soft Contact Lens is equivalent to the currently marketed Benz methafilcon A contact lens material. A summary of these results from the preclinical studies is presented below.

Toxicology:

In-Vitro Cytotoxicity: ISO 10993-5 was conducted in accordance with standards on test article. The test article meets the requirements of the Agarose Overlay Method.

Systemic Toxicity: The lens material meets the requirements of the systemic injection test and is considered non-toxic.

Acute Ocular Irritation: Acute ocular irritation test was performed and produced no ocular irritation.

Shelf Life

Shelf life requirements are satisfied via referencing rights granted to Contamac Ltd. for 510(k) 973597. The data presented supports substantial equivalence of this CONTAFLEX 55 Soft Contact Lens material to the already marketed Benz methafilcon A.

Substantial Equivalence:

The CONTAFLEX 55 Soft Contact Lens is substantially equivalent to the predicate device as depicted in the following table, and does not raise different questions of safety and effectiveness than the predicate device identified previously.

The following table depicts the pre-clinical characteristics of the CONTAFLEX 55 material, as well as the predicate device.

{3}------------------------------------------------

Substantial Equivalence table

Pre-Clinicalequivalency / DeviceCONTAFLEX 55(methafilcon A)BENZ methafilcon A(methafilcon A)
1.)Intended UseIndicated for daily wear for thecorrection of visual acuity inaphakic and not aphakic personswith non-diseased eyes withmyopia or hyperopia.Indicated for daily wear for thecorrection of visual acuity inaphakic and not aphakic personswith non-diseased eyes withmyopia or hyperopia.
2.)FunctionalityAfter machining from the opticalblank, the contact lenses act as arefractive medium that focus lightrays from near and distant objectson the retina.After machining from the opticalblank, the contact lenses act as arefractive medium that focus lightrays from near and distant objectson the retina.
3.)IndicationsDaily wear, Soft (hydrophilic)contact lensDaily wear, Soft (hydrophilic)contact lens
4.)Production MethodLathe-cutLathe-cut
5.)FDA Group #Group # 4 >50% Water, ionicPolymersGroup # 4 >50% Water, ionicPolymers
6.)USAN namemethafilcon Amethafilcon A
7.)Water Content55.0%55.0%
8.)Oxygen Permeability19.6 X 10-11 (cm2/sec)(ml 02/ml x mm Hg @ 35 degreesC), (revised Fatt method).18.3 X 10-11 (cm2/sec)(ml 02/ml x mm Hg (@ 35 degreesC), (revised Fatt method).
9.)Specific Gravity1.0671.128

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

FEB 2 6 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Contamac Ltd. c/o Martin Dalsing Medvice Consulting, Inc. Regulatory Consultant and U.S. Agent for Contamac. Ltd. 623 Glacier Drive Grand Junction, CO 81503

Re: K023989

Trade/Device Name: CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lens for Daily Wear (clear and tinted, lathe-cut) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL Dated: November 25, 2002 Received: December 2, 2002

Dear Mr. Dalsing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Martin Dalsing

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Kalgi Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Device Name: CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lens for Daily Wear (clear and tinted, lathe-cut)

INDICATIONS FOR USE:

The CONTAFLEX 55 (methafilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected using a chemical disinfecting system.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.

(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises

Prescription Usex (Per 21 CFR 801.109)
510(k) NumberK023989
Over-The-Counter Use

(Optional Format 1-2-96)

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.