Intended Use

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine. The designated Cutoff concentrations for these drugs are as follows: Barbiturates at 300 ng/mL, Benzodiazepines at 300 ng/mL, Methadone at 300 ng/mL, Methylenedioxymethamphetamine (Ecstasy) at 500 ng/mL, Opiates at 300 ng/mL and Tricyclic antidepressant at 1,000 ng/mL. They are intended for healthcare professionals including professionals at point of care sites.

Device Description

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous detection of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine samples. The test is based on the principle of antigenantibody immunochemistry. It utilizes mouse antibodies to selectively detect elevated levels of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine at Cutoff concentrations of 300 ng/mL (BZO), 1,000 ng/mL (TCA), 300 ng/mL (BAR), 500 ng/mL (MDMA), 300 ng/mL (MTD) and 300 ng/mL (MOP). These tests can be performed without the use of an instrument.

A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants at the concentrations below the designated cutoff levels will generate a colored-line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the ACON One Step Multi-Drug Multi-Line Screen Test meets them, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria for the ACON One Step Multi-Drug Multi-Line Screen Test were implicitly established by comparing its performance against GC/MS analysis, which is considered the gold standard for drug detection in urine. The predicate devices, previously FDA-cleared single drug tests, also served as a benchmark for substantial equivalence. The reported device performance is presented in the tables below for both the "Test Card" and "Test Device" variants.

ACON Test Card Performance vs. GC/MS Analysis (95% Confidence Interval)

DrugPositive AgreementNegative AgreementOverall Agreement
Barbiturates (BAR)127/133 = 95% (90-98%)160/160 = >99% (98-99%)287/293 = 98% (96-99%)
Benzodiazepines (BZO)128/131 = 98% (93-99%)160/160 = >99% (98-99%)288/291 = 99% (97-99%)
Methadone (MTD)120/129 = 93% (87-97%)177/177 = >99% (98-99%)297/306 = 97% (97-99%)
MDMA (Ecstasy)86/87 = 99% (94-99%)155/155 = >99% (98-99%)241/242 = 99% (98-99%)
Opiates (MOP)122/122 = >99% (97-99%)157/160 = 98% (95-99%)279/282 = 99% (97-99%)
Tricyclic Antidepressants34/34 = >99% (90-99%)181/192 = 94% (90-97%)215/226 = 95% (91-98%)

ACON Test Device Performance vs. GC/MS Analysis (95% Confidence Interval)

DrugPositive AgreementNegative AgreementOverall Agreement
Barbiturates (BAR)124/133 = 93% (88-97%)160/160 = >99% (98-99%)284/293 = 97% (94-97%)
Benzodiazepines (BZO)127/131 = 98% (92-99%)160/160 = >99% (98-99%)287/291 = 99% (97-99%)
Methadone (MTD)120/129 = 93% (87-97%)177/177 = >99% (98-99%)297/306 = 97% (97-99%)
MDMA (Ecstasy)87/87 = >99% (96-99%)154/155 = 99% (96-99%)241/242 = 99% (98-99%)
Opiates (MOP)122/122 = >99% (97-99%)158/160 = 99% (97-99%)280/282 = 99% (95-99%)
Tricyclic Antidepressants34/34 = >99% (90-99%)181/192 = 94% (90-97%)215/226 = 95% (91-98%)

The study aimed to demonstrate "substantial equivalency" to previously FDA-cleared tests and that the device is "safe and effective" in detecting the specified drugs at the given cutoff concentrations. The high percentages for positive, negative, and overall agreement across all drugs and both device forms indicate the device met these implicit acceptance criteria effectively.


Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: "Over 1,000 clinical specimens" were employed for the clinical evaluation. Approximately 10% of these samples had drug concentrations in the -25% to +25% Cutoff range. The exact breakdown per drug is available in the tables above, ranging from 226 to 306 total specimens per drug.
    • Data Provenance: The text states "clinical urine specimens" were used, without specifying the country of origin. The study appears to be retrospective, as it involved collected "clinical urine specimens" and comparison with "previously FDA-cleared Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants Single tests; as well as against data obtained from the customary GC/MS analysis."
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth was established by GC/MS analysis. This is a laboratory technique and does not involve human experts in the traditional sense of interpreting results. The personnel operating the GC/MS equipment would be trained laboratory technicians, but their qualifications are not specified nor are they referred to as "experts" for ground truth establishment in this context.
  3. Adjudication method for the test set:

    • No adjudication method (e.g., 2+1, 3+1) is described for the test set. The direct comparison was made between the ACON device results and the GC/MS analysis results.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or performed. This study compared the device's performance against a gold standard (GC/MS) and predicate devices, not primarily the improvement of human readers with or without AI assistance, as AI is not the primary technology here.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance evaluation was conducted. The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays that provide a visual, qualitative end result. The performance data presented in the tables are the results of the device alone in detecting the presence of drugs compared to GC/MS. The "physician's office laboratory POL study" mentioned at the end suggests use by "professionals at point-of-care sites," implying human interpretation of the visual result, but the core accuracy data are derived from the device's ability to correlate with GC/MS.
  6. The type of ground truth used:

    • The ground truth used was GC/MS analysis, which is a highly accurate and widely accepted laboratory method for confirming the presence and concentration of drugs in biological samples.
  7. The sample size for the training set:

    • The document does not specify a separate training set or its size. This is typical for in vitro diagnostic (IVD) devices that are not based on machine learning or AI algorithms requiring explicit training data. The device's design (antigen-antibody immunochemistry) is based on established scientific principles rather than learning from data.
  8. How the ground truth for the training set was established:

    • As no training set is described for this type of immunoassay device, the establishment of ground truth for a training set is not applicable.

{0}------------------------------------------------

8. 510(k) SUMMARY

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The Assigned 510(k) number is ________________________________________________________________________________________________________________________________________________

Submitter:

ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, California 92121

Tel.: 858-535-2030 Fax: 858-535-2038

Date: November 22, 2002

Contact Person: Edward Tung, Ph.D.

Product Names:

ACON One Step Multi-Drug Multi-Line Screen Test Card

ACON One Step Multi-Drug Multi-Line Screen Test Device

Common Name:

Immunochromatographic test for the simultaneously qualitative detection of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine.

Device Classification:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are similar to other FDA-cleared devices for the qualitative and simultaneous detection of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine specimens. These tests are used to provide a preliminary analytical result only. Benzodiazepines, Tricyclic Antidepressants, Barbiturates, MDMA, Methadone and Opiates test systems have been classified as Class II devices with moderate complexity.

Classification Name:

Barbiturate, Benzodiazepine, Methadone, MDMA, Opiate and Tricyclic Antidepressant test system

{1}------------------------------------------------

Intended Use:

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine samples. The designated cutoff concentrations for these drugs are as follows: Barbiturates at 300 ng/mL, Benzodiazepines at 300 ng/mL, Methadone at 300 ng/mL, Methylenedioxymethamphetamine (Ecstasy) at 500 ng/mL, Opiates at 300 ng/mL and Tricyclic antidepressant at 1,000 ng/mL. They are intended for healthcare professionals including professionals at point of care sites.

Description:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous detection of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine samples. The test is based on the principle of antigenantibody immunochemistry. It utilizes mouse antibodies to selectively detect elevated levels of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine at Cutoff concentrations of 300 ng/mL (BZO), 1,000 ng/mL (TCA), 300 ng/mL (BAR), 500 ng/mL (MDMA), 300 ng/mL (MTD) and 300 ng/mL (MOP). These tests can be performed without the use of an instrument.

A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants at the concentrations below the designated cutoff levels will generate a colored-line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

Predicate Devices:

ACON™ Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants Single Drug Test strips were used as the predicate devices for the ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device to compare their performance with clinical urine specimens.

The 510(k) Numbers for these predicate devices are.
ACON BAR One Step Barbiturates Test StripK012824
ACON BZO One Step Benzodiazepines Test StripK012300
ACON MTD One Step Methadone Test StripK012595
ACON MDMA One Step MDMA Test StripK022589
ACON MOP One Step Opiates Test StripK011353
ACON TCA One Step Tricyclic Antidepressant Test StripK021526

The 510/k) Numbers for these predicate devices are

{2}------------------------------------------------

Comparison to a Predicate Device:

A comparison of the features of the ACON™ One Step Multi-Drug Multi-Line Screen Test Card and Test Device versus the ACON™ Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants Single Tests is shown below:

  • Both tests are assays intended for the qualitative detection of Barbiturates, ● Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants in urine samples.
  • Both tests are intended as a screening method that provides a preliminary analytical test ● result.
  • . Both tests are immunochromatographic, lateral flow assays for the rapid detection of Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants with a visual, qualitative end result, while ACON Multi-Drug Multi-Line Test detects 2 to 6 of the above drugs simultaneously.
  • Both tests utilize the same basic immunoassav principles that rely on antigen/ antibody interactions to indicate a positive or negative result.
  • Both tests have the same Cutoff for each drug test.

Safety and Effectiveness Data:

Accuracy

A clinical evaluation was conducted using clinical urine specimens. This evaluation compared the test results between the ACON One Step Multi-Line Screen Test Card and the Test Device versus previously FDA-cleared Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants Single tests; as well as against data obtained from the customary GC/MS analysis. Over 1,000 clinical specimens were employed including approximately 10% of the samples with drug concentrations in the -25% to +25% Cutoff range. The comparisons of data obtained from this study yielded the following results:

{3}------------------------------------------------

Clinical study results of ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are compared to GC/MS analysis data:

The GC/MS Cutoff levels for each of the six drugs tested are as follows:

Barbiturates (BAR) at300 ng/mL
Benzodiazepines (BZO) at300 ng/mL
Methadone (MTD) at300 ng/mL
MDMA (Ecstasy) at500 ng/mL
Opiates (MOP) at300 ng/mL
Tricyclic Antidepressants (TCA) at1,000 ng/mL

Samples with drug concentration above the Cutoff level were considered presumptive positive and concentrations below the Cutoff are considered negative.

ACON Test CardPositive AgreementNegative AgreementOverall Agreement
BAR127/133 = 95%(90% - 98%)*160/160 = >99%(98% - 99%)*287 / 293 = 98%(96% - 99%)*
BZO128/131=98%(93% - 99%)*160/160=>99%(98% - 99%)*288/291= 99%(97% - 99%)*
MTD120/129 = 93%(87% - 97%)*177/177 = >99%(98% - 99%)*297 / 306 = 97%(97% - 99%)
MDMA86/87 = 99%(94% - 99%)*155/155 = >99%(98% - 99%)*241/242 = 99%(98% - 99%)*
MOP122/122 = >99%(97% - 99%)*157/160 = 98%(95% - 99%)*279/282 = 99%(97% - 99%)*
TCA34/34 = >99%(90% - 99%)*181/192 = 94%(90% - 97%)*215/226 = 95%(91% - 98%)*

ACON One Step Multi-Drug Multi-Line Screen Test Card vs. GC/MS Analysis

  • 95% Confidence Interval

{4}------------------------------------------------

ACON Test DevicePositive AgreementNegative AgreementOverall Agreement
BAR124/133 = 93%(88% - 97%)*160/160 = >99%(98% - 99%)*284 / 293 = 97%(94% - 97%)*
BZO127/131=98%(92% - 99%)160/160=>99%(98% - 99%)287/291= 99%(97% - 99%)*
MTD120/129 = 93%(87% - 97%)*177/177 = >99%(98% - 99%)*297 / 306 = 97%(97% - 99%)
MDMA87/87 = >99%(96% - 99%)*154/155 = 99%(96% - 99%)*241/242 = 99%(98% - 99%)*
MOP122/122 = >99%(97% - 99%)*158/160 = 99%(97% - 99%)*280/282 = 99%(95% - 99%)*
TCA34/34 = >99%(90% - 99%)*181/192 = 94%(90% - 97%)*215/226 = 95%(91% - 98%)*

A CON One Step Multi-Drug Multi-Line Screen Test Device vs. GC/MS Analysis

  • 95% Confidence Interval

Conclusion:

Clinical study results demonstrate the substantial equivalency between the ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device and the Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants single tests, which have already been cleared by FDA and marketed in the United States. It is also demonstrated that these tests are safe and effective in detecting Barbiturates, Benzodiazepines, Methadone, MDMA, Opiates and Tricyclic Antidepressants at the following Cutoff concentrations: Barbiturates 300 ng/mL, Benzodiazepines 300 ng/mL, Methadone 300 ng/mL, MDMA 500 ng/mL, Opiates 300 ng/mL and Tricyclic Antidepressants 1,000 ng/mL. The physician's office laboratory POL study demonstrated that these tests are also suitable for use by professionals at point-of-care sites.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top and left side. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FEB 0 5 2003

Edward Tung, Ph.D. Director of Regulatory Affair ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121

Re: K023946 Trade/Device Name: ACON® One Step Multi-Drug Multi-Line Screen Test Card ACON® One Step Multi-Drug Multi-Line Screen Test Device Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: DJC; DJG; DJG; JXM; DJR; LFG Dated: November 25, 2002 Received: November 27, 2002

Dear Dr. Tung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE 10.

510(k) Number: K023946

ACON® One Step Multi-Drug Multi-Line Screen Test Card Device Name:

ACON® One Step Multi-Drug Multi-Line Screen Test Device

Indications for Use:

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine. The designated Cutoff concentrations for these drugs are as follows:

AnalyteAbbreviationCalibratorCutoff Conc.
BarbituratesBARSecobarbital300 ng/mL
BenzodiazepinesBZOOxazepam300 ng/mL
MethadoneMTDMethadone300 ng/mL
MDMA (Ecstasy)MDMAMDMA500 ng/mL
OpiatesMOP or OPIMorphine300 ng/mL
Tricyclic AntidepressantsTCANortriptyline1,000 ng/mL

These tests are intended for healthcare professionals including professionals at the point of care sites.

(Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

Or

Over-The-Counter Use_

(Per 21 CFR 801.109)

Sean Compton

Division Sign-Off) Division of Clinical L 10k) Number

§ 862.3150 Barbiturate test system.

(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).