K Number
K023913

Validate with FDA (Live)

Date Cleared
2003-08-25

(273 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for prescriptive use.

Indications for Use:

  • Transcutaneous electrical nerve stimulation for the symptomatic relief and . management of chronic and intractable pain in the lower back region. The device has no curative values and should only be used in conjunction with medical supervision.
  • . Lumbar support.
Device Description

The BAXOLVE™ TENS device is a self-contained battery operated transcutaneous electrical muscle stimulator. The adhesive electrodes are mounted on a belt offering lumbar support to the user during therapy delivery.

The device is intended to provide a non-invasive prescriptive therapy.

AI/ML Overview

The provided text does not contain detailed information about specific acceptance criteria related to device performance in a clinical study for the NeuroTech® BAXOLVE™, Type 294. The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device through technological comparison and non-clinical tests (electrical outputs and compliance with standards). It does not present a clinical study with acceptance criteria and reported device performance in the same manner one might expect for a diagnostic or therapeutic device requiring such evidence for market clearance.

Therefore, many of the requested elements cannot be extracted from the given document.

Here's a breakdown based on the available information:

1. A table of acceptance criteria and the reported device performance

No explicit acceptance criteria for device performance in a clinical setting are provided, nor is there a table of reported device performance. The 510(k) emphasizes:

  • Technological Comparison: The BAXOLVE™ is similar to the Smart-TENS in its delivery of the stimulation signal and similar parameter settings.
  • Non-clinical Tests: Comparisons of electrical outputs for the two devices show similar results. Both were designed and independently tested to specific IEC and EN standards (e.g., IEC 60601-1:1990, IEC 60601-2-10, IEC 601-1-1, IEC 601-1-2, IEC 61000-4-2, IEC 61000-4-3, DD ENV 50204, EN 55011).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The document describes non-clinical engineering and safety compliance tests, not a clinical study on human subjects with a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical study with a test set and ground truth establishment is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a Transcutaneous Electrical Nerve Stimulator (TENS) and a truncal orthosis, not an AI-assisted diagnostic imaging device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI algorithm. The device itself is the standalone product. The non-clinical tests focus on the device's electrical output and safety compliance, not its performance as an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" was adherence to established electrical output specifications and internationally recognized safety and electromagnetic compatibility standards (e.g., IEC 60601 series, EN 55011).

8. The sample size for the training set

Not applicable. This device does not involve machine learning or a training set in the AI context.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

K023913

AUG 2 5 2003

510(k) Summary Page 1 of 2

Image: BMR logo and Bio-Medical Research Ltd address
510 (k) Summary of Safety and Effectiveness.
This summary is submitted in accordance with 21 CFR 807.92
a)1Submitted byBio-Medical Research LtdBMR HouseParkmore Business Park, WestGalwayRepublic of Ireland
Establishment RegistrationNumberContact PersonPhoneFaxe-mail8020867Michelle Sawyer+353 91 774361+353 91 773302msawyer@des.bmr.ie
TitleDate of PreparationRegulatory Affairs ManagerNovember 2002.
2Trade Name of DeviceNeuroTech© BAXOLVETM. Type 294.
Common NameNeuroTech© BAXOLVETM
Classification nameTranscutaneous Electrical NerveStimulator.(882.5810)
3Identification of predicatedeviceNeuroTech© Smart-TENS , Type 456,

{1}------------------------------------------------

K023913

Page 2 of 2 510(k) Summary

Description of Device 4

The BAXOLVE™ TENS device is a self-contained battery operated transcutaneous electrical muscle stimulator. The adhesive electrodes are mounted on a belt offering lumbar support to the user during therapy delivery.

The device is intended to provide a non-invasive prescriptive therapy.

ട Intended Use

The device is indicated for:

  • · Transcutaneous Electrical Nerve Stimulation (TENS), which provides the symptomatic relief and manaqement of chronic lower back pain.
  • · Lumbar support.

Technological Comparison ട

The BAXQLVE™ is similar to the Smart-TENS in it's delivery of the stimulation signal and has similar parameter settings. Both products utilise a LCD screen with user compliance logging.

Non- clinical Tests 7

Comparisons of electrical outputs for the two devices show similar results. They have both been designed and independently tested to the following requirements;

  • . IEC 60601-1:1990 Medical electrical equipment - Part 1: General requirements for safety.
  • . IEC 60601-2-10
  • IEC 601-1-1 and appendices A1:1991,A2:1995 . IEC 601-1-2: EMC requirements
  • . IEC 61000-4-2:1995: Electromagnetic compatibility
  • IEC 61000-4-3:1997: Electromagnetic compatibility
  • DD ENV 50204:1996: Electromagnetic compatibility .
  • EN 55011:1998: radiated emissions. .

Bio-Medical Research Ltd, (BMR), adheres to recognised and established industry practice, and all devices are subject to final performance testing.

A hazard analysis, a risk analysis and a failure mode effects analysis have been carried out for the device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Public Health Service

AUG 2 5 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Michelle Sawyer Regulatory Affairs Manager Bio-Medical Research Limited BMR House Parkmore Business Park, West Galway Republic of Ireland

Re: K023913

Trade/Device Name: NeuroTech® BAXOLVE™, Type 294 Regulation Number: 21 CFR 882.5890; 890.3490 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Truncal orthosis Regulatory Class: II Product Code: GZJ, IQE Dated: May 22, 2003 Received: May 27, 2003

Dear Ms. Sawyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Michelle Sawyer

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Mark n Mellem

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Not yet available

Device Name:

NeuroTech© BAXOLVE™ , type 294

Sponsor Name:

Bio-Medical Research Ltd.

KDZ3913

The device is intended for prescriptive use.

Indications for Use:

  • Transcutaneous electrical nerve stimulation for the symptomatic relief and . management of chronic and intractable pain in the lower back region. The device has no curative values and should only be used in conjunction with medical supervision.
  • . Lumbar support.

Do Not Write Below This Line - Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use Over-The-Counter Use

Mark N. Milliken

eral, Restorative

510(k) Number.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).