K Number
K023889

Validate with FDA (Live)

Date Cleared
2004-02-11

(447 days)

Product Code
Regulation Number
866.3870
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wiener Laboratory enzyme-linked Immunosorbent assay (ELISA) recombinante V. 3.0 test system is a manual and automated instrument assay for the qualitative detection of total antibodies (IgG and IgM) to Trypanosoma cruzi in human serum and plasma (EDTA, heparin, or Citrate) using recombinant antigens of T. cruzi. Reactive results are presumptive evidence of present or past infection with Trypanosoma cruzi.

Device Description

In this qualitative technique for the detection of antibodies anti-T. Cruzi, the sample is diluted in the wells in which recombinant antigens starting from specific proteins from the epimastigote and trypomastigote stages of the T. cruzi corresponding to highly conserved zones are immobilized. These antigens are proteins with aminoacid sequences repeated in tandem. SAPA (shed acute phase antigen) antigens detect antibodies in 93% of the patients' sera during the acute phase of the infection. It comes from the trypomastigote-bloodstream form of the parasite; #1, #2 and #30 antigens detect antibodies in chronic patients; #13 and #36 specially detect antibodies in sera both from acute and chronic patients.
If the sample contains Chagas' antibodies, they bind to the antigens bound to the support. The unbound antigens and antibodies are removed by washing, after which anti-human immunoglobulin antibodies conjugated to peroxidase are added. If a reaction was produced in the first step of the process, the conjugate is bound. After a new washing step and the addition of a chromogenic substrate and stopping reagent, specimens containing antibodies to T. cruzi produce a color reaction which can be read with a standard ELISA plate reader.

AI/ML Overview

Here's an analysis of the provided text to extract information about acceptance criteria and the study that proves the device meets them:

Device Name: Chagatest ELISA recombinante v.3.0 (also referred to as Wiener lab Chagatest ELISA rec. v.3.0)

Intended Use: Qualitative detection of antibodies to Trypanosoma cruzi (causative agent of Chagas' disease) in human serum and plasma. Useful in establishing prior exposure to T. cruzi and as an aid in diagnosis.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in a dedicated section. However, it presents "Performance Characteristics" including "Relative Sensitivity," "Relative Specificity," and "Relative Agreement," with specified confidence intervals. It also highlights an overall relative sensitivity of 97.9% and relative specificity of 97.8% (after resolution of discordant samples) or 97.9% and 97.8% (before resolution of discordant samples) from a combined analysis of multiple sites.

Given the context of a 510(k) summary seeking substantial equivalence, implicit acceptance criteria would likely involve demonstrating high relative sensitivity and specificity that are comparable to or better than predicate devices and meet general diagnostic accuracy expectations for such assays. While not explicitly stated as "acceptance criteria," the reported performance values are what would be used by the FDA to determine substantial equivalence.

Performance MetricReported Device Performance (after resolution unless specified)
Relative Sensitivity97.9% (95% CI: 95.6 - 99.1%)
Relative Specificity97.8% (95% CI: 97.0 - 98.5%)
Relative Agreement97.9% (95% CI: 97.1 - 98.5%)
Intra-assay Precision (%CV) - Manual4.8% to 19.7%
Intra-assay Precision (%CV) - Automated7.7% to 14.0%
Inter-assay Precision (%CV) - Manual7.4% to 26.4%
Inter-assay Precision (%CV) - Automated15.7% to 29.2%

2. Sample Sizes Used for the Test Set and Data Provenance

The test set consisted of multiple studies performed at different sites. The total sample size and data provenance are:

Study/SiteSample SizeData Provenance
1.1 Comparison with MERIDIAN (U.S. low-risk)500U.S. (Iowa), low-risk individuals (retrospective, assumed)
1.2 Comparison with MERIDIAN (Mexican Blood Bank)83Mexican Blood Bank (Guadalajara and Iowa) (retrospective, assumed)
1.3 Comparison with ABBOTT (Argentina)790Argentina (Rosario, SF) (retrospective, assumed)
1.4 Comparison with ABBOTT (Argentina)286Argentina (Buenos Aires, Hospital de Clínicas) (retrospective, assumed)
1.5 Comparison with ABBOTT and MERIDIAN90Unspecified; panel of sera (retrospective, assumed)
2.1 Specificity (U.S. low-risk)500U.S. (Iowa Hospitals, Iowa City, USA) (retrospective, assumed)
2.2 Sensitivity (Endemic Area, Argentina)118Argentina (Centro de Enfermedades y Patología Regional, Santiago del Estero), selected reactive specimens (retrospective, assumed)
2.3 Sensitivity (Chile)51Chile (Instituto de Ciencias Biomedical, Facultad de Medicina, Universidad de Chile, Santiago de Chile) (retrospective, assumed)
2.4 ISP Evaluation (Chile)150Chile (ISP, Santiago de Chile) (retrospective, assumed)
2.5 High Prevalence (Salta, Argentina)58Argentina (San Carlos, Provincial de Salta), blood bank specimens (retrospective, assumed)
2.5 High Prevalence (Salta, Argentina) - IHA52Argentina (Salta), IHA-reactive specimens (retrospective, assumed)
2.6 Specificity (Buenos Aires, Argentina)286Argentina (Hospital de Clínicas, Buenos Aires) (retrospective, assumed)
2.7 Blood Donors (Sao Paulo, Brazil)1236Brazil (Hemocentro de Sao Paulo, Sao Paulo), blood donors (retrospective, assumed)
2.7 Blood Donors (Sao Paulo, Brazil) - Panels300Brazil (Sao Paulo), selected reactive/non-reactive panels (retrospective, assumed)
2.8 Sensitivity Panel (Sao Paulo, Brazil)188Brazil (Centro de Imunologia e Imunogenetica, Sao Paulo), sensitivity panel (retrospective, assumed)
2.9 Blood Donors (Sao Paulo, Brazil)400Brazil (Centro de Imunologia e Imunogenetica, Sao Paulo), blood donors (retrospective, assumed)
2.10 Study (Rio de Janeiro, Brazil)914Brazil (Fundação Oswaldo Cruz, Rio de Janeiro), reactive and non-reactive specimens (retrospective, assumed)
2.11 Xenodiagnosed Positive70Unspecified, xenodiagnosed positive specimens (retrospective, assumed)
5. Patient Populations (Salta, Argentina)52Argentina (Salta), IHA-reactive specimens (retrospective, assumed)
5. Patient Populations (Santiago del Estero, Argentina)118Argentina (Santiago del Estero), IHA/IFA/ELISA-reactive specimens (retrospective, assumed)
5. Patient Populations (Xenodiagnosed)70Unspecified, xenodiagnosed positive individuals (retrospective, assumed)
5. Pregnant Women (Santiago del Estero, Argentina)368Argentina (Santiago del Estero), pregnant women (retrospective, assumed)

Total combined sample size for comparative effectiveness/performance studies appears to be over 5000 specimens across various studies. All data appears to be retrospective based on the nature of the studies (testing existing specimens, panels, or comparing against established methods). The countries of origin include Argentina, Mexico, Chile, and Brazil, with some studies also including low-risk individuals from the U.S.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number or specific qualifications of individual experts. However, it describes that ground truth was established using:

  • Comparison methods: Predicate devices (MERIDIAN Premier™ Chagas' IgG ELISA, ABBOTT Chagas Antibody EIA) were used as primary comparators.
  • Confirmatory/Reference tests: Radio Immuno Precipitation Assay (RIPA), Indirect Hemagglutination Assay (IHA), Immunofluorescence Assay (IFA), and xenodiagnosis were used as reference methods, particularly for resolving discordant results or establishing the status of selected panels.
  • Clinical context: Specimens were often from "patients with confirmed Chagas' disease by xenodiagnosis, IHA and IFA," "blood donors," or "patients from an endemic area."
  • The reference tests (IHA, IFA, RIPA, xenodiagnosis) are typically performed and interpreted by experienced laboratory personnel or medical professionals.

4. Adjudication Method for the Test Set

Adjudication methods are described for discordant results:

  • Discordant results between the Wiener lab ELISA and predicate devices were often adjudicated using reference methods like RIPA, IHA, and/or IFA.
    • For example: "Out of 5 specimens with discordant results, all 5 were reactive by Wiener lab ELISA and a RIPA, but non-reactive by Meridian." (Section 1.2)
    • "Of the 12 specimens with discordant results... 6 specimens were reactive by Wiener lab Chagatest ELISA rec.v.3.0 and non-reactive by Abbott Chagas Antibody EIA. Out of those 6 specimens, 3 were reactive by IFA." (Section 1.3)
    • "7 specimens were reactive by Wiener lab ELISA, reactive by IHA / IFA / RIPA and/or Meridian ELISA and non-reactive by Abbott ELISA." (Section 1.5a)

The specific process for how these reference results were used to adjudicate a final ground truth (e.g., majority vote, hierarchy of tests) is not explicitly detailed but implied that the consensus or superior reference test determined the final status.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not explicitly described. This is an in vitro diagnostic device for laboratory use, and its performance is typically assessed by comparing its numerical output (e.g., signal/cut-off ratio) to a threshold and then to gold standard/reference methods, rather than human readers interpreting images or data for diagnosis. The studies focused on comparing the device's diagnostic accuracy (sensitivity, specificity, agreement) to other assays and established reference methods.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the studies reported are standalone performance evaluations of the Wiener lab Chagatest ELISA recombinante v.3.0. The "algorithm" here refers to the ELISA assay itself, which generates quantitative results (absorbance, s/c ratio) that are then interpreted against a predefined cutoff. The performance metrics (relative sensitivity, specificity, agreement) measure the device's accuracy in identifying positive or negative samples independently, without direct human intervention in the interpretation of each test result beyond comparing it to the cutoff. The studies compare the device's output to other test systems.


7. The Type of Ground Truth Used

The ground truth for the test sets was established using a combination of:

  • Expert Consensus/Reference Methods: Primarily through established serological tests like Indirect Hemagglutination Assay (IHA), Immunofluorescence Assay (IFA), Radio Immuno Precipitation Assay (RIPA), and in some cases, previous ELISA results from other reference systems.
  • Pathology/Outcomes Data (Indirectly): Xenodiagnosis, a biological method for detecting T. cruzi infection, was used for some "confirmed Chagas' disease" cases (e.g., Section 2.3, 2.11, 5). This can be considered a strong indicator of true infection.
  • Clinical context: Patients from endemic areas and those with confirmed Chagas' disease were used to ensure the presence of antibodies, while low-risk individuals were used for specificity studies.

Therefore, the ground truth is a composite ground truth, derived from multiple established diagnostic methods and clinical confirmation strategies.


8. The Sample Size for the Training Set

The document does not specify a separate "training set" or "validation set" in the context of machine learning model development. This is an ELISA kit, which is a biochemical assay, not a machine learning algorithm that requires a training phase. The "studies" described are performance evaluations against established predicate devices and reference methods using various clinical samples. The "training" data for such a kit would implicitly be the data used during its development and optimization, which is not detailed here. The studies provided demonstrate the device's performance on various patient populations, which serves as a clinical validation.


9. How the Ground Truth for the Training Set Was Established

As noted above, this is an ELISA assay, not a machine learning model, so the concept of a "training set" ground truth in the AI context does not directly apply. The establishment of ground truth for the samples used in the performance evaluations (analogous to a test set in ML) is described in point 7 (composite ground truth from various reference methods and clinical confirmation).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a circular seal with the text "Wiener lab. SISTEMA DE CALIDAD CERTIFICADO" around the edge. In the center of the seal, there is the text "ISO 9001" above a logo that appears to be "TUV". The seal is black and white and appears to be a stamp.

Image /page/0/Picture/3 description: The image shows the logo for Wiener lab. The logo consists of a stylized letter W inside of a circle on the left, and the text "Wiener lab." on the right. Below the text is the phrase "Especialidades para Laboratorios Clinicos".

WIENER LABORATORIOS S.A.I.C. - Riobamba 2944 - 2000 Rosario - Argentina
Phone +54 (341) 432-9191/6 - Fax +54 (341) 432-5454/5555
Internet: http://www.wiener-lab.com.ar

Section 6 – Summary

510(k) Summary "This summary of 510(k) safety and effectiveness information is being

"This summary of 510(k) safety and effectiveness informants hary of OTO(R) ourordance with the requirements of SMDA 1990 and 21CFR 807.92"

"The assigned 510(k) number is: _____________________________________________________________________________________________________________________________________________

IntroductionAccording to the requirements of 21 CFR 866.3870, the following information provides sufficient details to understand the basis of a determination of substantial equivalence.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
6-1 Submitter Name, Address, ContactWiener Laboratorios S.A.I.C.Riobamba 29442000 - Rosario - ArgentinaContact person: Viviana CétolaDate Prepared: November 5, 2002
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
6-2 Device Name
Proprietary name:Chagatest ELISA recombinante v.3.0
Common name:Trypanosoma cruzi serological reagent.
Classification name:Enzyme linked Immunosorbent assay Trypanosoma cruzi
Device ClassI

{1}------------------------------------------------

We claim substantial equivalence to the currently marketed 6-3 Predicate WE Claim Cabblankia' IgG ELISA and ABBOTT Chagas Device Antibody EIA test systems. In this qualitative technique for the detection of antibodies anti-6-4 Device In this qualitat. The sample is diluted in the wells in which Description 1. Cruzi, the campio re immobilized. These antigens are recombinant antigens artisting es starting from specific proteins from the epimastigote and trypomastigote stages of the T. cruzi corresponding to highly conserved zones stages of the 7. State sen streep are proteins with aminoacid among annoront one in tandem. SAPA (shed acute phase sequences Topoutou in tath 93% of the patients' sera during antigen) antigense of the infection. It comes from the the addite prider of parasite; #1, #2 and #30 antigens trypomastigoto-blags in chronic patients; #13 and #36 specially dotest antibodies in sera both from acute and chronic patients. pationis.
If the sample contains Chagas' antibodies, they bind to the If the sumplo contains ound to the support. The unbound antigens and Tomain boo washing, after which anti-human immunoglobulin antibodies conjugated to peroxidase are mimanoglobalin annia was produced in the first step of the added. If a readjugate is bound. After a new washing step process, the conjugua chromogenic substrate and stopping and the adultions containing antibodies to T. cruzi produce reagent, opsoimeneaction which can be read with a standard ELISA plate reader. The Wiener lab. Chagatest ELISA recombinante v.3.0 test 6-5 Intended Use

system is an in vitro diagnostic system intended to be used in the qualitative detection of antibodies to Trypanosoma cruzi, the causative agent of Chagas' disease in human serum and the causalive ugent according to instructions, the kit is useful in establishing prior exposure to T. Cruzi and as an aid in the diagnosis of Chagas' disease.

Chagas' disease, an acute form of trypanosomiasis in children, most seriously affects the central nervous system and heart muscle.

{2}------------------------------------------------

6-6 Equivalencies and differences

The WIENER LAB. Chagatest ELISA recombinante v.3.0 test system is substantially The WENER LAB. Chagatest ELIOA Tecomminte on intended for similar use. Most
equivalent to other products in commercial distribution interest MERICAN Promier" equivalent to biner products in Commercial natures narketed MERIDIAN Premier™
notably it is substantially equivalent to the currently marketed MERIDIAN Premier™ nolably it is "substantially" equiridian" is "antibody ElA test systems.

The following table illustrates the similarities and differences between the The Tollowing table Indistrates the children is 3.0 test system and the WIENEN LAB: Onagatos: ELICA Freas' IgG ELISA test system.

MERIDIANTest SystemWIENER LAB. TestSystem
Intended useQualitative detection ofantibody to Trypanosomacruzi, the causativeagent for Chagas'disease in human serum.Qualitative detection ofantibody to Trypanosomacruzi, the causativeagent for Chagas'disease in human serumor plasma.
Test principleELISA test employingpurified T. cruzi antigensbound to ELISA platewells, ALP-labeledconjugate, pNPP assubstrate and NaOH asstopping solution.ELISA test employingrecombinant T. cruziantigens bound to ELISAplate wells, POD-labeledconjugate, TMB assubstrate and sulfuricacid as stopping solution.
EssentialComponentsPurified T. cruzi antigensRecombinant T. cruziantigens
SampleHuman serum.Human serum andplasma (heparinized,EDTA, citrated)
Wavelength ofreading.405 nm450 nm

{3}------------------------------------------------

The following table illustrates the similarities and differences between The WIENER LAB. Chagatest ELISA recombinante v.3.0 test system the WIENER EAB Onagatest ABBOTT Chagas Antibody EIA test system.

ABBOTTTest SystemWIENER LAB. TestSystem
Intended useQualitative detection ofantibody toTrypanosoma cruzi, thecausative agent forChagas' disease inhuman serum orplasma.Qualitative detection ofantibody to Trypanosomacruzi, the causative agentfor Chagas' disease inhuman serum or plasma.
Test principleELISA test employingpurified T. cruzi antigensbound to beads, POD-labeled conjugate, OPDas substrate and sulfuricacid as stoppingsolution.ELISA test employingrecombinant T. cruziantigens bound to ELISAplate wells, POD-labeledconjugate, TMB assubstrate and sulfuricacid as stopping solution.
EssentialComponentsPurified T. cruziantigensRecombinant T. cruziantigens
SampleHuman serum andplasma (EDTA,potassium oxalate,heparin and citratebased anticoagulants).Human serum andplasma (heparin, EDTAand citrate basedanticoagulants)
Wavelength ofreading.492 nm450 nm

Equivalence is demonstrated by the following comparative results:

PERFORMANCE CHARACTERISTICS

  1. Studies and comparison

1.1 Comparison of WIENER LAB and MERIDIAN Chagas kits.

In a study, 300 serum and 200 plasma specimens from random U.S. low-risk individuals, in a Study, 000 Scram and 200 Wiener lab Chagatest ELISA rec. v.3.0. and the Premier from fowa, were evaluation Diag.). Any of these specimens (0%) was reactive for antibodies to T. cruzi.

{4}------------------------------------------------

Premier Chagas' IgG ELISA (Meridian Diag.)
Wiener lab ChagatestELISA rec v.3.0ReactiveNon-reactiveDiscordant results
05000

1.2 Comparison of WIENER LAB and MERIDIAN Chagas kits.

1.2 Ooklpanson of theremens was tested from a Mexican Blood Bank (Guadalajara and A population of 60 SposilViener lab Chagatest ELISA rec. v.3.0 and Premier Chagas' IgG ELISA (Meridian Diag.).

Certain specimens were also tested by RIPA (Radio Immuno Precipitation Assay). The following results were obtained:

Premier Chagas' IgG ELISA (Meridian Diag.)
Wiener lab ChagatestELISA rec v.3.0ReactiveNon-reactiveDiscordant results
9695a

Out of 5 specimens with discordant results, all 5 were reactive by Wiener lab ELISA and a RIPA, but non-reactive by Meridian.

1.3 Comparison of WIENER LAB and ABBOTT Chagas kits.

1.5 Ooklpansof of Villand from Argentina (Rosario, SF, Argentina) was tested in A population of 100 chagatest ELISA rec. v.3.0 and Abbott Chagas Antibody ElA. puration betwoon were also tested by Wiener lab Indirect Hemagglutination Assay (IHA) and an in-house Immunofluorescence assay (IFA).

The following results were obtained:

Abbott Chagas Antibody ElA
Wiener lab ChagatestELISA rec v.3.0ReactiveNon-reactiveDiscordant results
22655212

Of the 12 specimens with discordant results: a

6 specimens were reactive by Wiener lab Chagatest ELISA rec.v.3.0 and non-reactive by Abbott Chagas Antibody EIA. Out of those 6 specimens, 3 were reactive by IFA.

  • 6 specimens were non reactive by Wiener lab Chagatest ELISA rec.v.3.0 and reactive by Abbott Chagas Antibody EIA. Out of those 6 specimens, 1 was reactive by IFA and IHA.
    1.4 Comparison of WIENER LAB and ABBOTT Chagas kits.

A population of 286 specimens was tested in Buenos Aires (Hospital de Clínicas, Buenos Aires, Argentina) by Wiener lab Chagatest ELISA rec. v.3.0, Polychaco Indirect Hemagglutination Assay(IHA) and Abbott Chagas Antibody EIA. The following results were obtained:

{5}------------------------------------------------

Abbott Chagas Antibody EIA
Wiener lab ChagatestELISA rec v.3.0ReactiveNon-reactiveDiscordant results
02851a

Reactive by Wiener lab Chagatest ELISA rec. v.3.0 and non-reactive by Abbott Chagas ನ Antibody EIA and IHA.

1.5 Comparison of WIENER LAB, with ABBOTT and MERIDIAN Chagas kits

1.5 Collipanson of VVLNEN DND, With ADD Min -reactive for the presence of Anti- T.Cruzi A pariel of Sela Typined as Teactive and Tierrosination Assay (IHA), an in-house Immunofluorescence assay (IFA) and/or RIPA, it was also tested by the following 3 ELISA (Maridion ifrimundolescence assay (if // androi Android Android Chagas' IgG ELISA (Meridian Diag.) and Abbott Chagas Antibody EIA.

The following results were obtained:

1 5a. Comparison Wiener lab vs Abbott ELISA kits

Abbott Chagas Antibody EIA
ReactiveNon-reactiveDiscordant results
Wiener lab ChagatestELISA rec v.3.0473310a
  • Out of 10 specimens with discordant results: a
  • Out of 10 specimens with discordant rount.
    7 specimens were reactive by Wiener lab ELISA, reactive by IHA / IFA / RIPA and/or -Meridian ELISA and non-reactive by Abbott ELISA
  • Mendian ELISA and non-reactive by Kiener lab ELISA and Meridian ELISA, but reactive । 2 Specimens were non rousine by other by other two specimens are equivocals for Wiener lab's ELISA kit.
  • Wiener lab s LLOA Kit.
    1 specimen was non-reactive by Wiener lab ELISA, non-reactive by all other methods but reactive by Abbott ELISA.

1,5b. Comparison Wiener lab vs Meridian ELISA kits

Premier Chagas' IgG ELISA (Meridian Diag.)
ReactiveNon-reactiveDiscordant results
Wiener lab ChagatestELISA rec v.3.0413613b

Of the 13 specimens with discordant results, all were reactive by Wiener lab. ELISA, b reactive by IHA / IFA / RIPA and/or Abbott ELISA, but non-reactive by Meridian ELISA.

Data summary

Relative sensitivity

I Crative Schildry Osing the allemative mountals, besctive, 2 were equivocal and 7 were non-reactive using the Wiener lab Chagatest ELISA rec. v.3.0. The relative sensitivity was 97.9%.

{6}------------------------------------------------

Relative Specificity

Relative Specificity
Using the alternative methods, 1507 specimens tested non-reactive using the Using the alternative methods, 1307 Specifican toological mon-reactive using the antibodies. Of these specimens, "52 were roustrol and 1988.
Wiener lab Chagatest ELISA rec. v. 3.0. The relative specificity was 97.8%.

Relative agreement

Relative agreement
Using the above results, the relative agreement was 97.9%

Note: samples giving equivocal results were not included in the calculation of relative Note: samples gring our relative agreement.

Relative Sensitivity, Specificity and Agreement before resolution of discordant samnies

samples.SiteNRelativeSensitivity (%)RelativeSpecificity (%)RelativeAgreement (%)
1500N/A100.0 (500/500)100.0 (500/500)
283100.0 (9/9)93.2 (69/74)94.9 (78/83)
379097.4 (226/232)98.9 (552/558)98.5 (778/790)
4286N/A99.7 (285/286)99.7 (285/286)
5a9097.5 (47/48)82.5 (33/40)90.9 (80/88)
5b90100.0 (41/41)73.5 (36/49)85.5 (77/90)
Total183997.9 (323/330)97.8 (1475/1707)97.9 (1798/1837)

Relative refers to a direct comparison of Wiener lab Chagatest ELISA rec. v. 3.0 results to Relative refers to a direct companson of victire lab correlate with disease presence of that of a similar test. No attempt has been made to convision in the predicate assay's accuracy to predict Chagas' disease.

Predicate Chagas' EIA kits
ReactiveEquivocalNon-ReactiveTotal
Wiener labChagatestELISA rec. v.3.0Reactive323-32355
Equivocal2--2
Non-Reactive7-14751482
Total332 (330)-15071839 (1837)

Relative Sensitivity = 97.9% (323/330), 95% Cr (Confidence Interval)= 95.6 = 99.1% Relative Sensitivity = 37.8% (1475/1507), 95% Cl = 97.0 - 98.5%)
Relative Specificity = 97.8% (1475/1507), 95% Cl = 97.0 - 98.5% Relative Specificity - 07.9% (1798/1837), 95% Cl = 97.1 - 98.5%

*95% Confidence Intervals (CI) calculated by the Exact Method.

2. Specificity and Sensitivity

At present, there is no recognized standard method for establishing the presence of At present, there is no recognized other more includes to testing serum and antibodies to 7. cruzi in numan blood. Relative opeoment geographical regions of the American continent.

Sensitivity for T. cruzi antibodies was calculated based on testing of xenodiagnosed Sensitivity for 1. crazi antibodies was calculated data assay (IHA), Indirect

{7}------------------------------------------------

Immunofluorescence Assay (IFA) and/or Enzyme-linked Immunosorbent Assay (ELISA) reactive specimens.

The performed studies show that:

  • 2.1 Relative specificity based on an assumed zero prevalence of antibody to T. cruzi in Relative specimenty based on an accument (Clinical Laboratories of lowards) of lowards Hospitals, lowa City, USA) is estimated to be 100% (500/500) with a 95.0% Confidence Interval (CI) of 99.2 - 100.0%.
  • 2.2 Relative sensitivity of the Wiener lab. Chagatest ELISA rec.v.3.0 in a selected Relative Johns reactive specimens (previously selected as reactive by IHA, population of 110 Onages road.ro op from 1 month to 84 years old) from an endemic IFA and/or ELIOA and Tanging In Eganss y Patología Regional, Santiago del Estero, area (Ochto de Enformisdad as 31% (117/118) with a 95.0% Confidence Interval of 95.3 -- 99.98%.
IHA / IFA / ELISA
Wiener labChagatestELISA rec.v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE1170117
Non-Reactive101
TOTAL1180118
  • 2.3 Relative sensitivity of the Wiener lab. Chagatest ELISA rec.v.3.0 in 2 selected Relative Scholly of the Wenchive specimens from Chile (Instituto de Ciencias populations of ST Onagas Touch) Voisersidad de Chile, Santiago de Chile, Chile) is Diomedical, Fucultud do Mouteirta, Criver Confidence Interval of 93.0 - 100.0 %.
Group1: 25 sera from patientswith confirmed Chagas' diseaseby xenodiagnosis, IHA and IFA.Wiener lab Chagatest ELISA rec. v.3.0Reactive Results25Sensitivity: 100%
Group 2: 26 sera from blooddonors, with at least 2 reactiveresults of serological tests forChagas' diseaseReactive Results26
  • 2.4 In an evaluation performed by the ISP (Instituto de Salud Pública, Santiago de Chile, In an evaluation performent previously tested by IFA and ELISA, were assayed by Wiener Onlic), 100 spoolinens provide of the sensitivity of 100% (65/65) with a 95.0% Confidence Interval of 94.5 - 100.0% and a specificity of 98.8% (84/85) with a 95.0% Confidence Interval of 93.62 -- 99.97%.
IFA /ELISA
Wiener labChagatest ELISArec. v.3.0REACTIVEREACTIVENon-ReactiveTOTAL
REACTIVE65166
Non-Reactive08484
TOTAL6585150

Patient with an autoimmune disease.

{8}------------------------------------------------

  • 2.5 In a population of high prevalence of Chagas' disease (San Carlos, Provincial de Salta, Argentina), 58 blood bank specimens were studied, analyzed in parallel between Algentifa), 50 blood bank Specimens word an include Indirect Hemagglutination Viener Tab Criagatest ELEA 100. tv 100% (8/8) and a specificity of 98% (49/50) Assay(ITA). It is obtained a sonstitutif of 89.95%. The individual -reactive with With a 95.0% Connique interval or 80.58 - on-reactive for IHA - presented an Wierel Tab Chagatest EEFOR TOU: For and Territic sign of acute phase from Thagas' disease. Three weeks later, when a new specimen was drawn, the sample Chagas disease. Thise weeks later, when a lowen this individual is removed from showed reachvily for ight in A and Torin ( tooks with a 95.0% Confidence Interval of 92.75 - 100.0%. (Prevalence (9/58): 15.5%).
IHA
Wiener labChagatest ELISArec. v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE819
Non-Reactive04949
TOTAL85058

Acute case

From the same geographic region (Salta, Argentina) 52 specimens previously From the Same geographic Tegion (Outla) Assay (IHA) were studied, teactive by an' in-nouse indicat Thernaggious with a 95.0% Confidence Interval of 93.15 - 100.0%.

IHA
Wiener labChagatest ELISArec. v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE52052
Non-Reactive000
TOTAL52052

2.6 The relative specificity was studied in a panel of 286 specimens from a population of The relative Specificity was stadiou Buenos Aires, Argentina). These specimens were modelate fisk ((riospital de Gilhiodo, Bachoonination Assay(lHA) and Abott Chagas hori-feactive by Forychaco Indicol Homaggiathation Research, 1996)
Antibody ElA. A relative specificity of 99.65% is obtained. (285/286) with a 95.0% Confidence Interval of 98.1 - 99.99%.

IHA / ELISA
Wiener labChagatest ELISArec. v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE011
Non-Reactive0285285
TOTAL0286286
  • 2.7 In a population of moderate risk (Hemocentro de Sao Paulo, Sao Paulo, Brasil), 1236 in a population of moderate fish (Homocon.co tested, together with the IHA, IFA and Specificity of blood donors from the the breaklence of 0.32% and a relative EEIOA. A Sensitively of 1031/122) with a 95.0% Confidence Interval of 99.55 - 100.0%.

{9}------------------------------------------------

IHA / IFA / ELISA
REACTIVENon-ReactiveTOTAL
Wiener labChagatest ELISArec. v.3.0REACTIVE415
Non-Reactive012311231
TOTAL412321236

The relative specificity and sensitivity were also studied in the same site, using sera panels selected by their reactivity with II A. In A and Room (1995.0% Confidence interval of specimens a sensitivity of 100% (100% (100% (100%) is specificity of 99.5% (199/200) is
96.38 - 100%. For the panel of non-reactive 100.00% 96.38 - 100%. For the panol of the panel of 97.25 - 99.99%.

Sera panel
REACTIVENon-ReactiveTOTAL
Wiener labChagatest ELISArec. v.3.0REACTIVE1001101
Non-Reactive0199199
TOTAL100200300

2.8 A sensitivity panel from a Brazilian was studied (Centro de Imunologia e
ralian and from a Brazilian samarilative samaitive sanaitivity of 100% (188188) A sensitivity panel Trom a Brazilan population Mac Sensitivity of 100% (188/188)
Imunogenetica, Sao Paulo, Brasil), os in 100,000 who 188 secondive speciments for Imunogenetica, Sao Paulo, Brasily, Obtailing on 0%, when 188 reactive specimens for with a 95.0% Confidence interval of 96.1 - 100.0%, which 108 February in-house RIPA
anti-T.cruzi antibodies previously tested by IHA, IFA, ELISA and an in-house RIPA were assayed.

IHA / IFA / ELISA / RIPA
REACTIVENon-ReactiveTOTAL
Wiener labChagatest ELISArec. v.3.0REACTIVE1880188
Non-Reactive000
TOTAL1880188
  • 2.9 In a population of 400 specimens from blood donors of Brazil (Centro de Imunologia e In a population of 400 specifiens from blood using the UFA, IFA and ELISA, and ELISA, and Imunogenetica, Sao Faulo, Drasily Similateously of 0.5% and a relative specificity sensitivity of 100% (2/2) is obtains with a 95.0% Confidence Interval of 97.1 – 99.6%.
IHA / IFA / ELISA
REACTIVENon-ReactiveTOTAL
Wiener labChagatest ELISArec. v.3.0257
0393393
2398400
  • In a study carried out in Brazil (Fundação Oswaldo Cruz, Rio de Janeiro, Brasil), 2.10 J in a study camed out in Brazil (r and go of disease were tested with Wiener lab
    reactive and non-reactive specimens for Chagas' disease were tested with Viener lab reactive and not-reactive specificher or Shagao Vere obtained: from 118 reactive Chagatest ELISA TEC. V.S.O. The Tonoming Toe the relative sensitivity is estimated to be 100% (118/118) with a 95.0% Confidence Interval of 96.92 - 100.0%, while from 796 100% (110/110) with a 95.0% Ochildthous ince systems, the relative specificity is non-leactive "Specimens" (ested" by "the "Same" of the "Same" of 98.9 - 99.92%.

{10}------------------------------------------------

IHA / IFA / ELISA / RIPA
REACTIVENon-ReactiveTOTAL
Wiener labChagatest ELISArec. v.3.0REACTIVE1183121
Non-Reactive0793793
TOTAL118796914

In a selected population of 70 xenodiagnosed Chagas positive specimens, all 70 2 11 -1 - In a selected population of 70 xenodagnosou chagas posses
(100%) specimens were repeatedly reactive for antibodies to T. cruzi by Wiener lab Chagatest ELISA rec. v.3.0.

Note: Relative refers to a direct comparison of Chagatest ELISA rec. v.3.0 that similari
and no Note: Relative refers to a direct companson or onlying or absence, and no
assays. No attemp has been made to correlate secures geence or absence, and no assays. No attemp has been made to conference with alsocures process.
judgment can be made regarding the predicates assay's accuracy to predict Chagas' disease.

Analytical Reactivity 3

3 Analytical Reactivity
Two reactive specimens were serially diluted in T. cruzi antibody non-reactive speciments we Two reactive specimens were senally and in 17 crear and 3 . 3 . Premier Chagas'lyG
tested by three methods (Wiener lab Chagatest ELISA rec. v. 3 .0 . Premier Chagas')gG tested by three methods (Viener lab Chagates Leon Percial South American Indirect
ELISA (Meridian Diag.) and Wiener lass submitted detectable pactivity in the Wiener ELISA (Meridian Diag) and Wieher Tab This (Sellinaecan e creativity in the Wiener
Hemagglutination Assay kit). One specimen exhibited detectable Hemagglutination Assay Kit). One specificer Childing the other one exhibited detectable
lab Chagatest ELISA rec v.3.0 at the sixh dillusing the other one exhibited detectable lab Chagatest ELISA "Tec V.S.S at the elish "Links" in the following table:
reactivity at the fourth dilution. The results are shown in the following table:

Wiener lab Chagatest ELISA rec.v.3.0. PERFORMANCE ON SERIAL DILUTIONS OF Wiener lab Chagatest ELISA PEC.V.S.O. TECH ORIFINGE GARS (GG ELISA (MERIDIAN
REACTIVE SPECIMENS. COMPARISON WITH Premier Chagas' (gG ELISA (MERIDIAN) REACTIVE SPECIMENS. COMFANISON VATT Promation (IHA).

Specimen Nr.1 (706007)Wiener lab ChagatestELISA rec v.3.0Premier Chagas'IgG ELISA(Meridian Diag.)Wiener labIHA
DILUTIONAbsorbance450/650 nms/c ratioAbsorbance405 nms/c ratioResult
Neata>3.000>9.871.4142.45+
1:2>3.000>9.871.2832.22+
1:4>3.000>9.871.1331.96+
1:8>3.000>9.870.9741.69+
1:162.3417.700.7351.27+
1:321.2784.200.593b1.03+
1:640.469b1.540.3130.54+
1:1280.0890.290.2140.37+
1:2560.0050.020.1210.21±b
1:5120.0040.010.0880.15-
cutoff0.3040.577

cimen Nr 1 (706007)

Serial dilutions in normal human serum, seronegative for T. cruzi

Highest dilution considered reactive. b

s/c ratio, signal/cut-off ratio

{11}------------------------------------------------

Specifieri NI. Z (201012)Wiener lab ChagatestELISA rec v.3.0Premier Chagas'IgG ELISA(Meridian Diag.)Wiener labIHA
DILUTIONAbsorbance450/650 nms/c ratioAbsorbance405 nms/c ratioResult
Neat®>3.000>9.871.0901.89+
1:2>3.000>9.870.8341.44+
1:42.9649.750.704°1.22+
1.81.7515.760.530°0.924.6
1:160.777"2.550.3810.66l
1:320.308€1.010.2870.50-
1:640.0060.020.2150.37-
1:1280.0040.010.1090.19-
1:2560.0040.010.0810.14ﻬﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ
1:5120.0040.010.0390.07
cutoff0.3040.577

1778/ מסוקס/ מחוז מח

Serial dilutions in normal human serum, seronegative for T. cruzi

ﻣ Highest dilution considered reactive.

ט Equivocal s/c ratio, signal/cut-off ratio

4 Precision study

  • 4 I Intra-assay precision study was determined by assaying three specimens 20 times in intra-assay producibility was determined using both manual (microplate reader from one Turi. Reproducibility Was actorni (Labotech**) assay methods. The intra-assay Molecular Deviation (SD) and Percent coefficient of Variation (%CV) were calculated.

Manual Intra-assay Precision study

Sample Nr.nMeanSD (O.D.)%CV
1202.118± 0.1760± 8.3%
2202.775± 0.1330± 4.8%
3200.004± 0.0008± 19.7%

Automated Intra-assay Precision study

Sample Nr.nMeanSD (O.D.)%CV
1201.798$\pm$ 0.2380$\pm$ 13.2%
2202.220$\pm$ 0.1710$\pm$ 7.7%
3200.005$\pm$ 0.0007$\pm$ 14.0%

4.2 Inter-assay precision study was determined by assaying three specimens 20 times in three consecutive runs. Reproducibility was determined using both manual (microplate thee conseculive rans: Roproadeling and automated (Labotech**) assay methods. The inter-assay Standard Deviation (SD) and Percent Coefficient of Variation (%CV) were calculated.

{12}------------------------------------------------

Manual Inter-assav Precision study

Sample Nr.nMeanSD (O.D.)%CV
1602.178$\pm$ 0.3439$\pm$ 15.7%
2602.785$\pm$ 0.2060$\pm$ 7.4%
3600.0034$\pm$ 0.0009$\pm$ 26.4%

Automated Inter-assay Precision study

Sample Nr.nMeanSD (O.D.)%CV
1601.893$\pm$ 0.3021$\pm$ 15.9%
2602.375$\pm$ 0.3730$\pm$ 15.7%
3600.0048$\pm$ 0.0014$\pm$ 29.2%
  • TM Molecular Devices Corp.

** TM Biochem Immunosystems, Inc

Reactivity in patient populations 5

5 - Reactivity in patient populations.
The reactivity of the Wiener lab Chagatest ELISA rec v.3.0 was determined by testing The reactivity of the Wiener lab Shage of patients initially determined to be positive by specifient trom pations in two group of xenodiagnosed positive specimens and in a group of pregnant women from an endemic area of Argentina.

A population of 52 specimens (from Salta, Argentina) from patients with Chagas' disease A population of 52 Specifiche (Iron Odita, Argenanalism Assay (IFA) was studied.
Previously reactive by an in-house Indirect Hemagglutination Assay (IFA) (2006) (50/50) previously reactive by an hagatest ELISA rec v.3.0 a relative sensitivity of 100% (52/52) with a 95.0% Confidence Interval of 93.15 - 100.0% was obtained.

IHA
Wiener labChagatestELISA rec.v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE52052
Non-Reactive000
TOTAL52052

Relative sensitivity of the Wiener lab Chagatest ELISA rec v.3.0 in a selected patient Relative sensitivity of the Vilener labsitive by IHA, IFA and/or ELISA population of 110 specifically (provisually bears old) from an endemic area (Centro de and Tanging In Lago Tratología Regional, Santiago del Estero, Argentina) was estimated to be 99.15% (117/118) with a 95.0% Confidence Interval of 95.3 - 99.98%.

IHA / IFA / ELISA
Wiener labChagatestELISA rec.v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE1170117
Non-Reactive101
TOTAL1180118

Specimens from 70 xenodiagnosed positive individuals, all 70 (100%) specimens were Opeomons Trom To xenutibodies to T. cruzi by Wiener lab Chagatest ELISA rec. v.3.0. A sensitivity of 100% with a 95.0% Confidence Interval of 94.87 - 100.0% was obtained.

{13}------------------------------------------------

Xenodiagnosis
Wiener labChagatestELISA rec.v.3.0REACTIVENon-ReactiveTOTAL
REACTIVE70070
Non-Reactive000
TOTAL70070

Specimens from 368 pregnant women from an endemic area (Santiago del Estero, Speciments Trom 300 pregnant women Than an EliusA rec v.3.0 and Premier Chagas'
Argentina), were assayed by Wiener lab Chagatest ELISA rec v.3.0 and Premier Chagas' Argentifia), were assayed by Thener lab Chargated.
IgG ELISA and the results were confirmed by RIPA. The following results were obtained:

Premier Chagas' IgG ELISA (Meridian Diag.)
Wiener lab ChagatestELISA rec v.3.0ReactiveNon-reactiveDiscordant results
2033414b
  • All 20 specimens were reactive by Wiener lab ELISA, Meridian ELISA and RIPA. a
  • All 20 specimens with discordant results were reactive by RIPA, however:
  • riese 14 specimens with disocraant rocation wore reach of and equivocal by ー Meridian ELISA.
  • Mendian ELISA.
    8 specimens were reactive by Wiener lab Chagatest ELISA rec.v.3.0 and non-reactive । by Meridian ELISA.
  • by Mendian ELISA.
    1 specimen was equivocal by Wiener lab Chagatest ELISA rec.v.3.0 and reactive by -Meridian ELISA.
  • Mendian ELISA.
    1 specimen was non-reactive by Wiener lab Chagatest ELISA rec.v.3.0 and equivocal by Meridian ELISA.

6-7 Conclusion Based on the data above mentioned, we believe that the Dased on the data above substantial equivalence to extended oldims commercial distribution intended for similar use

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes on its wing, representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Dr. Viviana Cetola OC/OA Manager Wiener Laboratorios S.A.I.C 2944 Riobamba 2000 Rosario (Santa Fe) Argentina

FEB 1 1 2004

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

K023889 Re:

Trade/Device Name: Chagatest ELISA rec. v.3.0 Regulation Number: 21 CFR § 866.3870 Regulation Name: Trypanosoma spp. Serological reagents Regulatory Class: I Product Code: MIU Dated: November 11, 2003 Received: November 13, 2003

Dear Dr. Cetola:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{15}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lot notification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of of quebecome on atte promise Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Saqartys

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Indications for Use

510(k) Number (if known):K023889

Device Name:_Chagatest ELISA rec. v.3.0

Indications For Use:

The Wiener Laboratory enzyme-linked Immunosorbent assay (ELISA) recombinante V. 3.0 test The Wieher Laboratory onlynned (using the Biochem Immunosystems, Inc Labotech System 1s a mandar and action of total antibodies (IgG and IgM) to Trypanosoma instrument abouty for the quantal (EDTA, heparin, or Citrate) using recombinant antigens of T. numan ocram and placer are presumptive evidence of present or past infection with cruzi. Trypanosoma cruzi.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

In Vitro Diagnostic Devices (OIVD) Concurrence of

Office of In Vitro Diagnostic Device Evaluation and Safety

510/k

§ 866.3870

Trypanosoma spp. serological reagents.(a)
Identification. Trypanosoma spp. serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toTrypanosoma spp. in serum. The identification aids in the diagnosis of trypanosomiasis, a disease caused by parasitic protozoans belonging to the genusTrypanosoma. Trypanosomiasis in adults is a chronic disease characterized by fever, chills, headache, and vomiting. Central nervous system involvement produces typical sleeping sickness syndrome: physical exhaustion, inability to eat, tissue wasting, and eventual death. Chagas disease, an acute form of trypanosomiasis in children, most seriously affects the central nervous system and heart muscle.(b)
Classification. Class I (general controls).