(54 days)
The RT-PLUS Knee System is a tricompartmental rotating hinged prosthesis o the total condylar type. The system consists of femoral, tibial and patellar components. It is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.
The RT-PLUS Knee and RT-PLUS Modular Knee are hinged tri-compartmental prostheses with femoral, tibial, and patella components and are of the cemented total condylar type. They are intended to be used for replacement of the total knee ioint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. Both of these knee systems provide joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band. Both systems include polyethylene tibial augmentation blocks. The Modular knee includes modular femoral and tibial stems.
The modifications described in this submission include:
- Changes to the outside surfaces of the ultra high molecular weight polyethylene (ASTM F648) 1. augmentation blocks and addition of blocks for Size 2 and 10 tibial baseplates. (Original cleared on K003504)
- Additional titanium alloy (ASTM F136) stem sizes (Original cleared on K003504) 2.
- Addition of CoCrMo (ASTM F75) augmentation blocks 3.
- Changes to femoral and tibial baseplate components to accept CoCrMo blocks (Original cleared on 4. K003504)
-
- Addition of size 2 and 10 femoral and tibial baseplate components (Non-modular cleared on K013340)
This 510(k) summary describes modifications to an existing knee implant device, not a new AI/ML-powered device. Therefore, the questions related to AI/ML device performance, ground truth, expert adjudication, and sample sizes for training/testing are not applicable because the provided text is about a traditional medical device (a knee prosthesis).
However, I can still extract information about the acceptance criteria and the study conducted for this traditional medical device as described in the provided document.
Acceptance Criteria and Study for RT-PLUS and RT-PLUS Modular Knee Systems
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Standard / Test) | Reported Device Performance / Compliance |
|---|---|
| ISO 14789-1: "Implants for surgery -- Total knee-joint prostheses -- Part 1: Determination of endurance properties of knee tibial trays" | Complied with (as stated in "Summary Nonclinical Tests") |
Explanation of the Acceptance Criteria (ISO 14789-1):
This international standard specifies methods for determining the endurance properties of knee tibial trays. This typically involves fatigue testing to assess the strength and durability of the tibial component under simulated physiological loads over a large number of cycles to ensure it can withstand the forces experienced in the human body over its intended lifespan without failure.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified in the provided document. The standard ISO 14789-1 generally requires testing multiple samples to demonstrate compliance, but the exact number for this submission is not given.
- Data Provenance: Not applicable in the context of clinical data for AI/ML. The "data" here refers to the results of mechanical non-clinical testing performed on the device components. This testing would have been conducted in a laboratory setting by the manufacturer or a contracted testing facility.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
Not applicable. This is not an AI/ML device where expert ground truth is established for a test set. The "ground truth" for a mechanical test like ISO 14789-1 is the physical failure or endurance of the device under specified test conditions, observed directly by engineers/technicians.
4. Adjudication Method for the Test Set
Not applicable. There is no adjudication method in the context of mechanical endurance testing. The results are objective measurements against the standard's requirements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
Not applicable. This is not an AI/ML device, and no MRMC study would be conducted for a knee prosthesis.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done
Not applicable. This is not an AI/ML device.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by the physical performance of the device components during mechanical testing according to the ISO 14789-1 standard. This involves objective measurements of endurance, fatigue, and structural integrity under simulated conditions.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is not an AI/ML device that requires a training set.
Summary of the Device and Submission:
The submission describes modifications to existing RT-PLUS and RT-PLUS Modular Knee Systems. The modifications primarily involve changes to materials, sizes, and components (e.g., augmentation blocks, stem sizes). The claim for substantial equivalence is based on the devices having the "same indications, materials and fundamental scientific technology" as the previously cleared predicate devices. The non-clinical testing performed to support the modifications and substantial equivalence claim involved compliance with ISO 14789-1 for "endurance properties of knee tibial trays." This indicates that the mechanical integrity and durability of the updated tibial components were assessed according to established industry standards.
{0}------------------------------------------------
K023(667
Amended 510(k) Summary
Date November 26, 2002
Submitter
PLUS ORTHOPEDICS 6055 Lusk Blvd San Diego, CA 92121
Contact person
J.D. Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-4694
Common name
Hinged knee
Classification name
Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer (per 21 CFR section 888.3510)
Equivalent Device
The RT-PLUS and RT-PLUS Knee systems are equivalent to the RT-PLUS Knee (K013340) and RT-PLUS Modular Knee (K003504).
Device Description
The RT-PLUS Knee and RT-PLUS Modular Knee are hinged tri-compartmental prostheses with femoral, tibial, and patella components and are of the cemented total condylar type. They are intended to be used for replacement of the total knee ioint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. Both of these knee systems provide joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band. Both systems include polyethylene tibial augmentation blocks. The Modular knee includes modular femoral and tibial stems.
The modifications described in this submission include:
- Changes to the outside surfaces of the ultra high molecular weight polyethylene (ASTM F648) 1. augmentation blocks and addition of blocks for Size 2 and 10 tibial baseplates. (Original cleared on K003504)
- Additional titanium alloy (ASTM F136) stem sizes (Original cleared on K003504) 2.
- Addition of CoCrMo (ASTM F75) augmentation blocks 3.
- Changes to femoral and tibial baseplate components to accept CoCrMo blocks (Original cleared on 4. K003504)
-
- Addition of size 2 and 10 femoral and tibial baseplate components (Non-modular cleared on K013340)
Intended Use
The RT-PLUS Knee System is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.
Summary of Technological Characteristics Compared to Predicate Device
The RT-PLUS and RT-PLUS Modular Knee systems with modifications have the same indications, materials and fundamental scientific technology as the RT-PLUS Knee (K013340) and RT-PLUS Modular Knee (K003504).
Summary Nonclinical Tests
ISO 14789-1 "Implants for surgery -- Total knee-joint prostheses -- Part 1: Determination of endurance properties of knee tibial trays" was complied with.
DEC 2 4 2002
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 4 2002
PLUS Orthopedics c/o Mr. J.D. Webb 1001 Oakwood Road Round Rock, Texas 78681
Re: K023667
Trade/Device Name: RT-PLUS Solution and RT-PLUS Modular Solution Knee Regulation Number: 21 CFR §888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer
Constrained Cemented Prosthesis
Regulatory Class: II Product Code: KRO Dated: November 29, 2002 Received: December 2, 2002
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. J.D. Webb
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
ia Witten, Ph.D., M.D. Ce Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) number (if known):
K023(6-)
Device Name: RT-PLUS & RT-PLUS Modular Knee Systems
Indications for Use:
RT-PLUS™ & RT-PLUS™ Modular Knee Systems Indications for Use .
The RT-PLUS Knee System is a tricompartmental rotating hinged prosthesis o the total condylar type. The system consists of femoral, tibial and patellar components. It is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Prescription Use _X (per 21 CFR 801.109)OROver-the-Counter Use ____(Optional format 1-2-96)
Gella Welle (Division Sign-off)
Division of General, Neurological and Restorative Devices
sion of General, Neurological and Restorative Devices
510(k) Number KO23667
N/A