K Number
K023648

Validate with FDA (Live)

Date Cleared
2003-05-06

(188 days)

Product Code
Regulation Number
862.3870
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Visualine®Cup VI for Drugs of Abuse Test is used for qualitative testing for the presence of Cocaine and it's metabolites at or above 300 ng/ml, Cannabinoids and it's metabolites at or above 50 ng/ml, Morphine and it's metabolites at or above 300 ng/ml, Methamphetamine at or above 1,000 ng/ml, Amphetamine at or above 1,000 ng/ml, Phencyclidine at or above 25 ng/ml, and Benzodiazepines and it's metabolites at or above 300 ng/ml in human urine samples. This test provides only a preliminary screening result; a more specific alternative method should be used to confirm the test result.

Device Description

The Visualine® CupVI for Drugs of Abuse Test is based on the principle of antigen-antibody complexation and is used for the analysis of Cocaine, Cannabinoids, Morphine, Methamphetamine or Amphetamines, Phencyclidine, and Benzodiazepines and their corresponding metabolites in human urine samples. The assay utilizes a competitive immunochromatographic technique involving a sample of test urine delivered through a wicking action as the test panel (holding the porous membrane) is dipped into the urine sample. The drug in the sample competes for the limited antibody sites on the colored microspheres. When an adequate amount of drug is present, it will fill the limited antibody binding sites. This will prevent attachment of the colored microspheres to the probe site on the membrane. Therefore, a positive urine sample will inhibit the formation of precipitin at the probe site.

A reference or control line with a secondary antibody reaction is added to the membrane strip to check for proper sample migration on the membrane. This control line should always be present. A negative urine sample will produce two colored lines and a positive urine sample will show only one, the control line.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Visualine® Cup VI for Drugs of Abuse Test, based on the provided text:

Acceptance Criteria and Device Performance

AnalyteAcceptance Criteria (Cutoff Level)Reported Device Performance (Positive Agreement)Reported Device Performance (Negative Agreement)Reported Device Performance (Overall Agreement)
Cocaine300 ng/ml97.7% (42/43)>99% (381/381)>99% (423/424)
Cannabinoids50 ng/ml96% (46/48)>99% (376/376)>99% (424/424)
Morphine300 ng/ml>99% (48/48)>99% (375/376)>99% (424/424)
Methamphetamine1000 ng/ml>99% (47/47)>99% (377/377)>99% (424/424)
Phencyclidine25 ng/ml>99% (46/46)>99% (378/378)>99% (424/424)
Benzodiazepines300 ng/ml>99% (55/55)>99% (369/369)>99% (424/424)
Amphetamines1000 ng/ml95% (40/42)>99% (382/382)>99% (424/424)

Precision Acceptance Criteria:

  • Within run and run to run: >99%
  • Within day and day to day: >99%
  • Within lot and lot to lot: >99%

Stability Acceptance Criteria:

  • A urine specimen containing 0 ng/ml of the analyte of interest will render two distinct magenta lines (one test line and one control line).
  • Samples containing +25% of cutoff analyte levels will show positive results >99% of the time, therefore yielding only the control line.

Study Information

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 424 specimens for each drug category.
    • Data Provenance: Presumptive positive samples were provided by Redwood Toxicology Laboratory (Santa Rosa, California). Negative samples consisted of UTAK Laboratories Drug Free Urine pool #5488 and in-house negative urines. The study was conducted at Redwood Toxicology Laboratory and Sun Biomedical Laboratories. This indicates a mix of external and internal data, likely retrospective as existing samples were used.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the number or qualifications of experts used to establish ground truth. It states that presumptive positive samples were "assayed on the Hitachi at Redwood." This implies a laboratory instrument provided the reference standard, rather than human experts interpreting the results for ground truth.
  3. Adjudication method for the test set:

    • Not applicable. The ground truth was established by laboratory instrument (Hitachi), not through human interpretation requiring adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is an in-vitro diagnostic test for direct visual interpretation (presence/absence of lines) and does not involve AI or human reader interpretation in the context of diagnostic imaging or similar fields where MRMC studies are typical. The comparison was between the new device and existing reference devices.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The performance data presented is for the standalone performance of the Visualine® Cup VI test. It's a qualitative immunoassay that produces a visual result directly. There isn't an "algorithm" in the typical AI sense; the test itself is the "device" being evaluated for its direct output. The user simply interprets the visual lines.
  6. The type of ground truth used:

    • Laboratory Instrument Reference Standard: The ground truth for positive samples was established by assaying them on a Hitachi instrument. For negative samples, known drug-free urine pools (UTAK Laboratories Drug Free Urine pool #5488) and in-house negative urines were used. This is a laboratory-based, objective reference standard.
  7. The sample size for the training set:

    • The document does not explicitly mention a "training set" in the context of machine learning or AI. This device is a biochemical immunoassay, not an algorithmic device that requires a distinct training phase with a dataset. The development and optimization of the immunoassay itself would have involved internal validation and formulation studies, but these are not described as a "training set" in the common sense.
  8. How the ground truth for the training set was established:

    • As there's no explicit "training set" described for an AI/ML model, this question is not applicable. The development of the immunoassay (e.g., antibody selection, membrane formulation) would have been based on established chemical and biological principles and internal validation, rather than a "ground truth" derived from a specific dataset in the way an AI model is trained.

{0}------------------------------------------------

K023648

510(k) CONTENT SUMMARY

1. Name of Manufacturer:

Sun Biomedical Laboratories Inc. 604 VPR Center 1001 Lower Landing Road Blackwood, NJ 08012

MAY 0 6 2003

    1. Trade Name; Visualine® Cup VI for Drugs of Abuse Test for Qualitative determination of Cocaine, Cannabinoids, Morphine, Methamphetamine or Amphetamines, Phencyclidine and Benzodiazepines and/or their metabolites 1 Human Urine Samples.
    1. Common Name:

An in-vitro immunoassay, multi-drug test panel using visual color comparison for the detection of Cocaine, Cannabinoids, Morphine, Methamphetamine or Amphetamines, Phencyclidine, and Benzodiazepines in human urine samples.

    1. Regulation # and Classification:
      Reg. #862-3170, Class II Device
    1. Test Description:
      The Visualine® CupVI for Drugs of Abuse Test is based on the principle of antigen-antibody complexation and is used for the analysis of Cocaine, Cannabinoids, Morphine, Methamphetamine or Amphetamines, Phencyclidine, and Benzodiazepines and their corresponding metabolites in human urine samples. The assay utilizes a competitive immunochromatographic technique involving a sample of test urine delivered through a wicking action as the test panel (holding the porous membrane) is dipped into the urine sample. The drug in the sample competes for the limited antibody sites on the colored microspheres. When an adequate amount of drug is present, it will fill the limited antibody binding sites. This will prevent attachment of the colored microspheres to the probe site on the membrane. Therefore, a positive urine sample will inhibit the formation of precipitin at the probe site.

A reference or control line with a secondary antibody reaction is added to the membrane strip to check for proper sample migration on the membrane. This control line should always be present. A negative urine sample will produce two colored lines and a positive urine sample will show only one, the control line.

{1}------------------------------------------------

6. Comparison of Two Test Systems for Correlation Studies:

The Visualine® Cup VI for Drugs of Abuse Test for Qualitative Determination of Cocaine, Cannabinoids, Morphine, Methamphetamine or Amphetamines, Phencyclidine, and Benzodiazepines assay is correlated to the Sub Biomedical's existing Visualine® V Drugs of Abuse Test and the Visualine® II Benzodiazepine assay and Visualine® II Amphetamine drugs of abuse assay. The following table illustrates the similarities and differences between the assays.

Visualine®V Drugs of AbuseTest for QualitativeDetermination of Cocaine,Cannabinoids, Morphine,Methamphetamine, and theVisualine® II Drugs of AbuseAssay for Benzodiazepines andAmphetaminesVisualine® Cup VI for Drugsof Abuse Test for QualitativeDetermination of Cocaine,Cannabinoids, Morphine,Methamphetamine orAmphetamines, Phencyclidineand Benzodiazepines
Test PrincipleCompetitive bindingCompetitive binding
Sample/Sample SizeApprox. 200 ul urine/test stripApprox. 200 ul urine/test strip
AntibodyPolyclonal and MonoclonalPolyclonal and Monoclonal
TracerAb Colloidal GoldAb Colloidal Gold
Detection MethodVisual color precipitin formationVisual color precipitinformation
Test Run Time5 minutes5 minutes
StorageRequirements2-10°C (36-86°F)2-10°C (36-86°F)
Detection Levels:
Cocaine300 ng/ml300 ng/ml
Cannabinoids50 ng/ml50 ng/ml
Morphine300 ng/ml300 ng/ml
Methamphetamine1000 ng/ml1000 ng/ml
Phencyclidine25 ng/ml25 ng/ml
Benzodiazepines300 ng/ml300 ng/ml
Amphetamines1000 ng/ml1000 ng/ml
Ancillary Equipmentnonenone

{2}------------------------------------------------

    1. Visualine® Cup IV Test Performance Characteristics:
    • A. Correlation studies between Visualine® V Drugs of Abuse Test and Visualine® II drugs of Abuse Test for Cocaine and it's metabolites, Cannabinoids and it's metabolites, Morphine and it's metabolites, Methamphetamine or Amphetamines and their metabolites, Phencyclidine, and Benzodiazepines and it's metabolites were conducted at Redwood Toxicology Laboratory, Santa Rosa California and at Sun Biomedical Laboratories. The presumptive positive samples were provided by Redwood and assayed on the Hitachi at Redwood followed by testing by the Visualine® V Test, Visualine® Cup VI, and Visualine® II Benzodiazepine and Amphetamine tests. Negative samples consisted of UTAK Laboratories Drug Free Urine pool #5488, and in-house negative urines tested with Sun Biomedical's Visualine® II product for the targete: I drugs of abuse. The following information represents the correlation between the new Visualine® Cup VI Test and the existing Visualine® V Test, the Visualine® II Benzodiazepine and Amphetamine Tests.
Cocaine:Positive Agreement42/4397.7%
Negative Agreement381/381>99%
Overall Agreement423/424>99%
CannabinoidsPositive Agreement46/4896%
Negative Agreement376/376>99%
Overall Agreement424/424>99%
Morphine:Positive Agreement48/48>99%
Negative Agreement375/376>99%
Overall Agreement424/424>99%
Methamphetamine:Positive Agreement47/47>99%
Negative Agreement377/377>99%
Overall Agreement424/424>99%
Phencyclidine:Positive Agreement46/46>99%
Negative Agreement378/378>99%
Overall Agreement424/424>99%
Benzodiazepines:Positive Agreement55/55>99%
Negative Agreement369/369>99%
Overall Agreement424/424>99%
AmphetaminesPositive Agreement40/4295%
Negative Agreement382/382>99%
Overall Agreement424/424>99%

{3}------------------------------------------------

  • B. Specificity and Substances Detected:
    The individual tests are specific to the labeled drug of abuse or structurally: related compounds. The test detects Cocaine and it's metabolites at 300 ng/ml, Cannabinoids and it's metabolites at 50 ng/ml, Morphine and it's metabolites at 300 ng/ml, Methamphetamine or Amphetamines at 1000 ng/ml. Phencyclidine at 25 ng/ml, and Benzodiazepine (Oxazepam) and it's metabolites at 300 ng/ml.

  • C. Precision:
    Reproducibility studies for Visualine® Cup VI for Cocaine, Cannabinoids, Morphine, Methamphetamine, Amphetamine, PCP, and Benzodiazepine: indicate:

Within run and run to run>99%
Within day and day to day>99%
Within lot and lot to lot>99%

D. Stability Statement:

Visualine® Cup VI Test stability has been studied. The urine drugs test: are tested every three months and reviewed for acceptance by the Qualit; Control Manager, also every three months, for up to a period of over two years. The acceptance criteria are as follows: A urine specimen containing 0 ng/ml of the analyte of interest will render two distinct magenta lines, one test line and one control line. Samples containing +25% of cutoff analyte levels will show positive results >99% of the tirr >. therefore vielding only the control line. Visualine® Cup VI Tests are stable within their marked expiration date and under the storage conditions as described in the package insert.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 0 6 2003

Ming Sun, Ph.D. Sun Biomedical Laboratories Inc. 604 VPR Center 1001 Lower Landing Road Blackwood, NJ 08012

K023648 Re:

Trade/Device Name: Visualine® Cup VI for Drugs of Abuse Test Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid test system Regulatory Class: Class II Product Code: LDJ; DIO; DJG; DJC; LCM; JXM; Dated: January 30, 2003 Received: March 5, 2003

Dear Dr. Sun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: This image shows the logo and contact information for Sun Biomedical Laboratories, Inc. The logo features a black circle with a horizontal line running through it. The text includes the company name, address (604 VPR CENTER, 1001 LOWER LANDING ROAD, BLACKWOOD, NJ 08012), telephone number (856-401-1080), and fax number (856-401-1090).

510 (k) Number: Ko 336 48

DEVICE NAME: Visualine®Cup VI for Drugs of Abuse Test

INDICATIONS FOR USE: The Visualine®Cup VI for Drugs of Abuse Test is used for qualitative testing for the presence of Cocaine and it's metabolites at or above 300 ng/ml, Cannabinoids and it's metabolites at or above 50 ng/ml, Morphine and it's metabolites at or above 300 ng/ml, Methamphetamine at or above 1,000 ng/ml, Amphetamine at or above 1,000 ng/ml, Phencyclidine at or above 25 ng/ml, and Benzodiazepines and it's metabolites at or above 300 ng/ml in human urine samples. This test provides only a preliminary screening result; a more specific alternative method should be used to confirm the test result. This test is intended for use by medical professionals.

Jean Cooger
(Division Sign-Off)

Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number X023648

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter--Use

(Optional Format 1-2-96)

OR

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).