K Number
K023519

Validate with FDA (Live)

Date Cleared
2003-01-16

(87 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VariLink 2 MMLC is intended to assist the radiation oncologist team in the delivery of radiation to well defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. With Radionics' Treatment Planning Software, the MMLC enables static conformal treatments to be performed with finely shaped field patterns. The Varilink 2 MMLC can be used as part of IMRT (Intensity Modulated Radiotherapy) or general conformal radiation.

Device Description

The device is an optional accessory for the Radionics Mini-Multileaf Collimator (MMLC), mounted on a Varian C-Series Linear Accelerator (Linac). VariLink 2 allows Intensity Modulation Radiation Therapy (IMRT) functionality to be used on a Varian Linac. The Radionics' Treatment Planning software generates treatment plans with segmented beams, in which the intensity within the irradiation field is not constant (as in conventional treatments) but varies across the field. This requires the LINAC to suspend the beam between segments to allow the MMLC to change fields. VariLink 2 uses the Varian gating board to suspend the beam to allow MMLC to change the shape of the field.

AI/ML Overview

The provided text does not contain detailed acceptance criteria and the study that proves the device meets them. The document is a 510(k) summary for the Radionics VariLink 2, which focuses on establishing substantial equivalence to a predicate device rather than detailing specific performance studies with acceptance criteria.

The relevant section {1} states: "System testing verifies that the device is ready for clinical use. A rigorous test of the functionality of VariLink 2 was conducted using a Varian C-Series Linac with the Gating Board Interface."

This statement confirms that some testing was performed to verify functionality, but it does not provide any specific acceptance criteria, performance metrics, or details about the study design or results. Without further information beyond what is present in this 510(k) summary, it is impossible to complete the requested tables and answer the questions.

Therefore, I cannot provide the requested information for the following reasons:

  1. No Acceptance Criteria or Reported Performance: The document does not list any specific numerical acceptance criteria (e.g., accuracy, precision, sensitivity, specificity thresholds) or the actual reported device performance against such criteria.
  2. No Details on Study Design: The summary mentions "system testing" and a "rigorous test of the functionality" but provides no information on the methodology, sample size, type of data used, ground truth establishment, expert involvement, or adjudication methods.
  3. No Standalone or MRMC Study Details: There is no mention of whether a standalone algorithm performance study or a multi-reader, multi-case (MRMC) comparative effectiveness study was conducted.
  4. No Training Set Information: The document does not describe a training set, how its ground truth was established, or its sample size.

In summary, the provided text confirms that functional testing was done but lacks the specific details required to answer your questions comprehensively.

{0}------------------------------------------------

K023519

12 510(k) Summary

JAN 1 6 2003

Radionics VariLink 2 510(k) Summary

This summary of the 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. 807.92.

1.0 The submitter of this premarket notification is:

Kevin J. O'Connell Senior Regulatory Associate Radionics, a division of Tyco Healthcare Group LP 22 Terry Avenue Burlington, MA 01803 Tel .: (781) 272-1233 Fax: (781) 272-2428

This summary was prepared on October 16, 2002.

  • 2.0 The name of the device is the Radionics VariLink 2. The common name is Radiotherapy beam shaping block, and its classification name is Medical charged-particle radiation therapy system.
  • 3.0 The above device is substantial equivalent to the Radionics ConforMAX MMLCVR1 was cleared via 510(k), K993594, on December 15, 1999.
  • 4.0 The device is an optional accessory for the Radionics Mini-Multileaf Collimator (MMLC), mounted on a Varian C-Series Linear Accelerator (Linac). VariLink 2 allows Intensity Modulation Radiation Therapy (IMRT) functionality to be used on a Varian Linac. The Radionics' Treatment Planning software generates treatment plans with segmented beams, in which the intensity within the irradiation field is not constant (as in conventional treatments) but varies across the field. This requires the LINAC to suspend the beam between segments to allow the MMLC to change fields. VariLink 2 uses the Varian gating board to suspend the beam to allow MMLC to change the shape of the field.
  • 5.0 The device is intended for use in stereotactic, conformal, computer planned, LINAC (linear accelerator) based radiation. The indications for use are: The VariLink 2 MMLC is intended to assist the radiation oncologist team in the delivery of radiation to well defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. With Radionics' Treatment Planning Software, the MMLC enables static conformal treatments to be performed with finely shaped field patterns.

{1}------------------------------------------------

  • 6.0 System testing verifies that the device is ready for clinical use. A rigorous test of the functionality of VariLink 2 was conducted using a Varian C-Series Linac with the Gating Board Interface.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three lines that form the shape of a snake wrapped around a staff.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 6 2003

Mr. Kevin J. O'Connell Senior Regulatory Associate Radionics. A Division of Tyco Healthcare 22 Terry Avenue BURLINGTON MA 01803-2516 USA

Re: K023519

Trade/Device Name: Radionics VariLink 2 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system

Regulatory Class: II Product Code: 90 IYE Dated: October 17, 2002 Received: October 21, 2002

Dear Mr. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

ODE Indications for Use Statement

510(k) Number (if known): K023519

Device Name: Radionies VariLink 2 MMLC

indicalions for Use:

The Varilink 2 MMLC is intended to assist the radiation oncologist team in the The Varithe 2 Wirise 15 method target volumes while sparing surrounding normal the world critical organs from excess radiation. The Varilink 2 MMLC can be used s part of IMKT (Intensity Modulated Radiotherapy) or general conformal radiation цていたけどのし

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEBDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prevengstion Use (ther 21 CFR $ 801.109)

OR

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

David A. Seymon
(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.