(21 days)
The Signa Ovation (formerly known as MFO/I) system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa Ovation system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa Ovation system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Due to the 'open' design of the system, the Signa Ovation may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display, and MR safe biopsy needles.
The Extra Large Body Coil is a modification to the Signa MFO/i MR System (K002611) which includes the addition of a flexible extra large body coil.
This 510(k) summary describes a modification to an existing device, the GE Signa MFO/i MR System (K002611), by adding an "Extra Large Body Coil." The document focuses on demonstrating substantial equivalence to the predicate device rather than providing a detailed study of the new coil's performance against specific acceptance criteria in a clinical setting.
Therefore, many of the requested elements for a study proving a device meets acceptance criteria are not explicitly present in the provided text. The document states that "Testing was performed to demonstrate that the design modifications to the Extra Large Body Coil meet predetermined acceptance criteria," but it does not elaborate on what those criteria were, the specific tests conducted, or the results.
Based on the provided text, here is what can be extracted and what cannot:
1. Table of acceptance criteria and the reported device performance
The document does not provide a table of acceptance criteria or specific reported device performance metrics for the Extra Large Body Coil against those criteria. It broadly states that "Testing was performed to demonstrate that the design modifications to the Extra Large Body Coil meet predetermined acceptance criteria." These acceptance criteria would likely relate to image quality (resolution, signal-to-noise ratio), safety, and compatibility with the existing MR system, but details are not given.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This information is not provided in the document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not provided. The study design described is not a clinical study involving expert interpretation of images for ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study is mentioned. This device is an MRI coil, not an AI-powered diagnostic tool, so an MRMC study comparing human readers with and without AI assistance would not be applicable here.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable, as this is hardware (an MRI coil), not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" in the context of clinical diagnostic accuracy studies is not directly applicable here. The testing would have focused on engineering and performance specifications of the coil, not diagnostic accuracy based on clinical outcomes or pathology.
8. The sample size for the training set
Not applicable, as this is a hardware device, not a machine learning algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable.
Summary of Study from the document:
The study described is a technical verification that the "design modifications to the Extra Large Body Coil meet predetermined acceptance criteria." The purpose of this study was to support the claim of substantial equivalence to the predicate device (GE Signa MFO/i MR system, K002611). The document concludes that "Usage of the Extra Large Body Coil does not result in any new potential hazards." This type of submission (510(k)) primarily focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than providing detailed clinical performance data against specific diagnostic acceptance criteria.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, enclosed within a circle. There are decorative swirls or flourishes around the circle, adding to the logo's design. The logo is black and white.
Page 1 of 2
GE Medical Systems
General Electric Company P.O. Box 414, Milwaukee, WI 53201
510(k) Summary
K02342
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE Medical SystemsPO Box 414Milwaukee, WI 53201 |
|---|---|
| Contact Person: | Larry A. Kroger Ph.D.Manager, Regulatory Programs |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | October 8 2002 |
Device Name:
Signa Ovation with Extra Large Body Coil Magnetic Resonance Coil, 21 CFR 892.1000, 90-MOS
Marketed Device:
The Signa Ovation system with Extra Large Body Coil is substantially equivalent to the currently marketed GE Signa MFO/i MR system (K002611).
Device Description:
The Extra Large Body Coil is a modification to the Signa MFO/i MR System (K002611) which includes the addition of a flexible extra large body coil.
Indications for Use:
The Signa Ovation (formerly known as MFO/I) system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa Ovation system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa Ovation system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Due to the 'open' design of the system, the Signa Ovation may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display, and MR safe biopsy needles.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are in black, contrasting with the white background. The logo is a well-known symbol representing the multinational conglomerate.
General Electric Company P.O. Box 414, Milwaukee, WI 53201
Comparison with Predicate Device:
GE Medical Systems intends to begin marketing the Signa Ovation with Extra Large Body Coil. The Extra Large Body Coil is a modification of the Signa MFO/i MR System (K002611) with the main differences being the addition of a flexible extra large body coil. Signa MFO/i MR System has the same basic technological characteristics, and, uses similar design, construction, and materials.
Summary of Studies:
Testing was performed to demonstrate that the design modifications to the Extra Large Body Coil meet predetermined acceptance criteria.
Conclusion:
It is the opinion of GE that the Signa Ovation with Extra Large Body Coil is substantially equivalent to the GE Signa MFO/i MR System (K002611). Usage of the Extra Large Body Coil does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white logo. The logo features a stylized bird with three curved lines representing its wings or feathers. The bird is facing to the right. The words "DEPARTMENT OF" are written vertically along the left side of the logo, following the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
・、
5 2002 NOV
Larry Kroger, Ph.D. Regulatory Programs Manager GE Medical Systems General Electric Company P.O. Box 414 MILWAUKEE WI 53201
Re: K023442
Trade/Device Name: Extra Large Body Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II
Product Code: 90 MOS Dated: October 8, 2002 Received: October 15, 2002
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|---|---|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): KO234 + 2
Device Name: Extra Large Body Coil
Indications For Use:
The Signa Ovation (formerly known as MFO/I) system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa Ovation system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa Ovation system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Due to the 'open' design of the system, the Signa Ovation may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display, and MR safe biopsy needles.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use OR
David G. Ingram
ological Devic
N/A