K Number
K022983

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-09-25

(16 days)

Product Code
Regulation Number
880.5965
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BardPort® and SlimPort™ Implanted Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products and for the withdrawal of blood samples.

Device Description

The X-Port isp port relies on the same basic, fundamental scientific technology as the predicate Hickman port. Access to the port is made percutaneously with a non-coring needle that enters the port reservoir via the silicone rubber septum. The access path to the vascular system is provided through a catheter attached to the port. The port system serves as a conduit for fluids into, and out of, the central venous system.

The port body consists of an oval shaped plastic base and top. The silicone septum is compressed between the port base and top. The port body profile incorporates smooth transitions, a tapered nose and is intermediate sized. There are only two suture holes, which are adjacent to the stem and are filled with silicone plugs.

The catheters, stems and catheter locks used with the X-Port isp port are unchanged, previously qualified legally marketed devices covered by the predicate device and come in multiple French sizes.

AI/ML Overview

1. Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
Port Tests (based on FDA Guidance)To demonstrate substantial equivalence to the predicate Hickman port or meet predetermined criteria."The X-Port isp port met all predetermined acceptance criteria." These tests included: Dimensions, Septum puncture life, Intermittent port leak, Continuous port leak, Septum blowout, Fluid dynamics clearance. The document states these were performed according to protocols based on "Guidance on 510(k) Submission for Implanted Infusion Ports, dated October 1990" and "in-house protocols."
Additional Verification TestsTo further demonstrate safety and efficacy for intended use."The X-Port isp port met all predetermined acceptance criteria." These tests included: Flow rate, Obturation, Priming volume, Septum needle insertion force, Septum needle retention force, Stem strength, Suture plug retention, Top/base assembly strength. These were performed "in accordance with in-house protocols."
BiocompatibilityMeet ISO-10993 and FDA-Modified ISO 10993 Test Profile."Biocompatibility requirements of ISO-10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing, and the FDA-Modified ISO 10993 Test Profile for a long term implanted device that exhibits tissue contact, indirect blood contact and direct blood contact, were met." (Note: No specific new testing was required as all materials were previously cleared.)
Design ValidationMeet recommendations of "Design Control Guidance for Medical Device Manufacturers.""Design validation was performed to meet the recommendations of the FDA guidance document, Design Control Guidance for Medical Device Manufacturers, dated March 11, 1997."

2. Sample Size Used for the Test Set and Data Provenance:

The provided document does not specify exact sample sizes for each test mentioned (e.g., number of ports tested for septum puncture life). It only states that "Verification and validation testing was performed according to protocols..."

The data provenance is from non-clinical performance testing, meaning internal laboratory and design validation tests. It is not patient or human data; therefore, data provenance regarding country of origin or retrospective/prospective is not applicable in the human study sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable. This study describes non-clinical performance testing of a medical device (an implanted port), not a diagnostic algorithm or image analysis with human interpretation. The "ground truth" for these tests would be established by engineering specifications, material properties, and established test methodologies, not by expert human consensus on clinical data.

4. Adjudication Method for the Test Set:

Not applicable. As detailed above, this is non-clinical device testing, not a study involving human interpretation or subjective assessments that would require an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

No, an MRMC comparative effectiveness study was not done. This document describes the 510(k) submission for an implanted medical device, not an AI or imaging diagnostic algorithm.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

No, a standalone algorithm performance study was not done. This document is for an implanted medical device, not an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The ground truth for the non-clinical performance tests of the X-Port isp Implanted Port would be based on:

  • Engineering Specifications: Defined dimensions, material properties, and performance limits.
  • Physical Measurements and Observations: Direct measurement of dimensions, observation of leaks, tracking of puncture cycles, etc.
  • Material Science Principles: Ensuring materials meet biocompatibility and mechanical strength requirements.
  • Predicate Device Performance: Benchmarking against the previously cleared Hickman port for substantial equivalence.
  • Established Test Standards: Adherence to FDA guidance documents (e.g., "Guidance on 510(k) Submission for Implanted Infusion Ports, dated October 1990") and ISO standards (e.g., ISO-10993).

8. The Sample Size for the Training Set:

Not applicable. This is not a machine learning or AI study that involves a training set. The "training" for the device's design and manufacturing is derived from established engineering practices, past product development, and regulatory standards.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no "training set" in the context of an AI or machine learning model for this device submission. The design and validation relied on established medical device design control procedures, existing scientific and engineering principles, and regulatory guidelines for implanted ports.

{0}------------------------------------------------

Section 6

510(k) Summary X-Port isp Implanted Port

510(k) Summary of Safety and Effectiveness Information 21CFR 807.92

6.1 Submitter Information

Submitter Name:Bard Access Systems, Inc. (BAS)[Subsidiary of C. R. Bard, Inc.]
Address:5425 W. Amelia Earhart DriveSalt Lake City, UT 84116
Telephone Number:(801) 595-0700, Ext. 5439
Fax Number:(801) 595-5425
Contact Person:John Knorpp
Date of Preparation:6 September 2002

6.2 Device Name

Device Name:BardPort® Implanted Port
Trade Name:X-Port isp™
Common/Usual Name:Plastic Subcutaneous Port & Catheter
Classification Name:80LJT - Port & Catheter, Implanted, Subcutaneous, Intravascular21 CFR 880.5965 - Subcutaneous, Implanted, Intravascular InfusionPort and Catheter, Class II

6.3 Predicate Device Name

Device Name:BardPort Implanted Port
Trade Name:Hickman® Plastic Subcutaneous Port
Common/Usual Name:Plastic Subcutaneous Port & Catheter
Classification Name:80LJT - Port & Catheter, Implanted, Subcutaneous, Intravascular21 CFR 880.5965 – Subcutaneous, Implanted, Intravascular InfusionPort and Catheter, Class II
Premarket Notification:K873213, concurrence date – October 27, 1987.

6.4 Device Description

Principles of Operation

There are no new operating principles. The X-Port isp port relies on the same basic, fundamental scientific technology as the predicate Hickman port. Access to the port is made percutaneously with a non-coring needle that enters the port reservoir via the silicone rubber septum. The access path to the vascular system is provided through a catheter attached to the port. The port system serves as a conduit for fluids into, and out of, the central venous system.

Port Body

  • The port body consists of an oval shaped plastic base and top. ●
  • . The silicone septum is compressed between the port base and top.
  • . The port body profile incorporates smooth transitions, a tapered nose and is intermediate sized.

{1}------------------------------------------------

  • There are only two suture holes, which are adjacent to the stem and are filled with silicone . plugs.

Catheters, Stems and Catheter Locks

  • The catheters, stems and catheter locks used with the X-Port isp port are unchanged, previously ● qualified legally marketed devices covered by the predicate device and come in multiple French sizes.

6.5 Intended Use

The BardPort X-Port isp Implanted Port is a totally implantable vascular access device designed to provide long term repeated access to the vascular system.

This is the identical intended use for the predicate Hickman port.

6.6 Summary of Technological Characteristics in Relation to the Predicate Device

  • 6.6.1 Does the new device have the same indication statement?
    Yes.

Does the new device have the same technological characteristics, e.g. design, material, 6.6.2 etc.?

Not in all regards. The X-Port isp port design has some minor differences from the predicate Hickman port, however the basic fundamental scientific technology of the port has not changed. The differences include the following: a larger septum, a lower profile intermediate size port body, two silicone filled suture holes located adjacent to the stem and a tapered nose.

6.6.3 Could the new characteristics affect safety or effectiveness?

Yes. The above features could affect safety or effectiveness of the device.

6.6.4 Do the new characteristics raise new types of safety and effectiveness questions?

No. There are no new issues of safety and effectiveness.

6.6.5 Do accepted scientific methods exist for assessing effects of the new characteristics?

Yes. The FDA's Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990 was used to evaluate the device's performance.

Design validation was performed to meet the recommendations of the FDA guidance document, Design Control Guidance for Medical Device Manufacturers, dated March 11, 1997.

Biocompatibility requirements of ISO-10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing, and the FDA-Modified ISO 10993 Test Profile for a long term implanted device that exhibits tissue contact, indirect blood contact and direct blood contact, were met. No materials are used in the manufacture of the subject device that have not already been cleared for similar applications by Bard Access Systems.

6.6.6 Are performance data available to assess effects of new characteristics?

{2}------------------------------------------------

Yes. Verification and validation testing was performed according to protocols based on the above-referenced guidance document recommendations and standards, as well as in accordance with in-house protocols.

6.6.7 Do performance data demonstrate equivalence?

Yes. Performance data gathered in design verification and validation testing demonstrated that the X-Port isp port is substantially equivalent to the predicate Hickman port and/or met predetermined acceptance criteria, and the risks associated with use of the new device were found acceptable when evaluated by FMEA.

6.7 Nonclinical Performance Testing

Port tests performed in conformance with Guidance on 510(k) Submission for Implanted Infusion Ports, dated October 1990 included assessments of:

  • . Dimensions
  • . Septum puncture life
  • Intermittent port leak .
  • Continuous port leak .
  • . Septum blowout
  • . Fluid dynamics clearance

The following guidance recommended tests were not performed for the following reasons:

  • All catheter to port connection tests. All catheters and catheter/port connection systems are part . of previously qualified legally marketed devices covered by the predicate device and have been incorporated into the X-Port isp port with no change. The only modifications are to the main port body.
  • . No biocompatibility testing was required. All materials used in the manufacture of the subject device have been previously cleared for similar devices.

In addition, the following verification tests were performed to further demonstrate the safety and efficacy of the subject device for its intended use:

  • . Flow rate
  • Obturation .
  • . Priming volume
  • . Septum needle insertion force
  • . Septum needle retention force
  • . Stem strength
  • . Suture plug retention
  • Top/base assembly strength .

6.8 Conclusion

The X-Port isp port met all predetermined acceptance criteria.

Based on FDA's decision tree, the X-Port isp port is substantially equivalent to the predicate device Hickman Port, K873213, cleared October 27, 1987.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized human figures or birds in flight, arranged in a stacked formation.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John C. Knorpp Senior Regulatory Affairs Manager C. R. Bard, Incorporated Bard Access Systems Division 5425W. Amelia Earhart Drive Salt Lake City, Utah 84116

Re: K022983

Trade/Device Name: The BardPort® and SlimPort™ Implanted Ports Regulation Number: 880.5965 Regulation Name: Subcutaneous. Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: September 6, 2002 Received: September 9, 2002

Dear Mr. Knorpp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 1.2

INDICATION(S) FOR USE STATEMENT*

The BardPort® and SlimPort™ Implanted Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products and for the withdrawal of blood samples.

Signature of 510(k) Submitter:

Printed Name of Submitter:

John C. Thoyt

John C. Knorpp Senior Regulatory Affairs Specialist

Date:

9-6-02

*Suggested language and format to meet the requirements of sections 513(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and sections 807.92(a)(5) and 801.4 of the Code of Federal Regulations, Title 21.

Concurrence of Office of Device Evaluation

510(k) Number

K022983

OR

2

Division Sign-Off Office of Device Evaluation

Prescription Use

Over-The-Counter Use

Patux Cicardi

vision Sian-Off Division of Anesthesiology, General Hospital. Infection Control, Dental Devices

510(k) Number: K022983

§ 880.5965 Subcutaneous, implanted, intravascular infusion port and catheter.

(a)
Identification. A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.