K Number
K022873

Validate with FDA (Live)

Date Cleared
2003-03-21

(204 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free surgeon's glove is a device made of synthetic latex that may bear a trace amount of glove powder and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. The radiation attenuating surgical glove is intended to be used during medical procedures where hands are necessarily exposed to radiation in order to offer some degree of protection to the hand from radiation. This includes surgical procedures that require the use of fluoroscopy or radiography.

Device Description

The Powder-Free Radiation Attenuating Surgical Gloves are formulated using neoprene, synthetic latex. The gloves are supplied powder-free and sterile.

AI/ML Overview

This document describes the acceptance criteria and performance of the Powder-Free Radiation Attenuating Surgical Glove by International Biomedical, Inc. It does not pertain to an AI/ML medical device, but rather a physical surgical glove. Therefore, most of the requested information regarding AI/ML studies is not applicable.

Here's the information derived from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Test / Acceptance CriteriaReported Device Performance
Primary Skin Irritation: No irritation potentialGlove does not display irritation potential.
Dermal Sensitization: No sensitization potentialGlove does not display sensitization potential.
Elongation and Tensile Strength: Meets or exceeds ASTM D 3577-01a requirements for synthetic surgical glovesGlove meets or exceeds requirements for synthetic surgical gloves per ASTM D 3577-01a.
Barrier Defects: Meets or exceeds ASTM D3577-01a requirementsGlove meets or exceeds requirements per ASTM D3577-01a.
Powder Level: Meets "Powder Free" designation per ASTM D 6124-01Glove meets powder level requirements for "Powder Free" designation per ASTM D 6124-01.

2. Sample size used for the test set and the data provenance

The document indicates "Summary of Test Results" but does not provide specific sample sizes for each test mentioned (e.g., how many gloves were tested for elongation and tensile strength, or how many subjects were involved in irritation/sensitization tests).

The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective/prospective. However, given that this is a 510(k) submission to the FDA, it is implied that the testing was conducted to meet US regulatory standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI/ML device, and the evaluation involves physical and chemical property testing rather than expert-based ground truth establishment for diagnostic purposes.

4. Adjudication method for the test set

Not applicable. This is not an AI/ML device; the tests are objective measurements of physical and chemical properties.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical surgical glove, not an AI/ML system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical surgical glove, not an AI/ML system.

7. The type of ground truth used

The "ground truth" for this device is based on established industry standards and regulatory requirements for surgical gloves. These include:

  • ASTM D 3577-01a: Standard Specification for Rubber Examination Gloves (likely referencing requirements for synthetic surgical gloves).
  • ASTM D 6124-01: Standard Test Method for Residual Powder on Medical Gloves.
  • Protocols for evaluating primary skin irritation and dermal sensitization.

8. The sample size for the training set

Not applicable. This device is a physical surgical glove, not an AI/ML system that requires a "training set."

9. How the ground truth for the training set was established

Not applicable. This device is a physical surgical glove, not an AI/ML system.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for International Biomedical, Inc. The logo features a stylized letter "i" on the left, followed by the company name in bold, sans-serif font. Below the company name is the address, 8508 Cross Park Dr.

MAR 21 2003

8508 Cross Park Dr. Austin, Texas 78754 512-873-0033 Fax: 512-873-9090 www.int-bio.com

5404) STATEMENT OF SAFETY AND EFFECTIVENESS

510k SUMMARY OF SAFETY AND EFFECTIVENESS RADIATION ATTENUATING SURGICAL GLOVE

Manufacturer:International Biomedical, Inc.8508 Cross Park DriveAustin, TX 78754U.S.A.
Regulatory Affairs Contact:Amy Pieper8508 Cross Park Dr.Austin, TX 78754
Telephone:(512) 873-0033
Date Summary Prepare:August 27, 2002
Common Name:Powder-Free Radiation Attenuating Surgical Glove
Classification:Sugeon's Glove, 79KGO
Predicate Devices:Radiation Attenuating Surgical Glove by International Biomedical,Inc. (82k1720) and the ESP Radiation Recustion ExaminationGloves by Boston Scientific (K891968) and BarrierPlus SyntheticPowder-Free Surgical Gloves y BarrierMed Glove Company(K990710).
Description:The Powder-Free Radiation Attenuating Surgical Gloves areformulated using neoprene, synthetic latex. The gloves aresupplied powder-free and sterile.
Intended Use:A powder-free surgeon's glove is intended to be worn on thehands, usually in a surgical setting, to provide a barrier againstpotentially infectious materials and other contaminants.Additionally, the radiation attenuating surgical glove is intended to

{1}------------------------------------------------

be used during medical procedures where hands are necessarilyexposed to radiation in order to offer some degree of protection tothe hand from radiation. This includes surgical procedures thatrequire the use of fluoroscopy or radiography.
Substantial Equivalence:Powder Free Radiation Attenuating Surgical gloves aresubstantially equivalent to the Radiation Attenuating SurgicalGloves in that they have the same intended use, they are both madeof synthetic latex, and they are both manufactured in the same way.The Powder Free Radiation Attenuating Surgical gloves aresubstantially equivalent to the ESP Radiation RecustionExamination Gloves by Boston Scientific in that they have thesame intended use and polymers.The Powder Free Radiation Attenuating Surgical gloves aresubstantially equivalent to the BarrierPlus Synthetic Powder-FreeSurgical Gloves in intended use, design and product features, andphysical characteristics.
Summary of Test Results:TestResult
Primary Skin IrritationGlove does not display irritation potential.
Dermal SensitizationGlove does not display sensitization potential.
Elongation and TensileStrengthGlove meets or exceeds requirements for synthetic surgical glovesper ASTM D 3577-01a.
Barrier DefectsGlove meets or exceeds requirements per ASTM D3577-01a.
Powder LevelGlove meets powder level requirements for "Powder Free"designation per ASTM D 6124-01.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 1 2003

Ms. Amy Pieper Regulatory Affairs International Biomedical, Incorporated 8508 Cross Park Drive Austin, Texas 78754

Re: K022873

Trade/Device Name: Radiation Attenuating Neoprene Surgical Gloves, Powder Free (Black) Regulation Number: 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: November 25, 2002 Received: January 21, 2003

Dear Ms. Pieper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Pieper

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runner, DDS, MA

Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510k SUBMISSION FOR THE RADIATION ATTENUATING SURGICAL GLOVE (BLACK)

INDICATIONS FOR USE

Applicant's Name: International Biomedical, Inc.

510(k) Number: K022873

Neoprene

Device Name: Radiation Attenuating Surgical Gloves , C

Power Free (BLACK)

Indications for Use:

A powder-free surgeon's glove is a device made of synthetic latex that may bear a trace amount of glove powder and is intended to be worn on the hands, usually in surgical settings, to provide a barrier against potentially infectious materials and other contaminants. The radiation attenuating surgical glove is intended to be used during medical procedures where hands are necessarily exposed to radiation in order to offer some degree of protection to the hand from radiation. This includes surgical procedures that require the use of fluoroscopy or radiography.

Clin S. him

(Division Sign-Off) sion of Anesthesiology, General Hospital Infection Control, Dental De

510(k) Number: K022823

Image /page/4/Picture/12 description: The image shows the logo for International Biomedical, Inc. The logo consists of a stylized lowercase "i" on the left, followed by the words "International Biomedical, Inc." on the right. The "i" is a bold, black color, and the text is also black.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).